K231896 · Beijing Adss Development Co., Ltd. · GEX · Feb 20, 2024 · General, Plastic Surgery
Device Facts
Record ID
K231896
Device Name
Diode Laser Therapy System
Applicant
Beijing Adss Development Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diode Laser Therapy System (Model: FG2000-B/FG2000-B Pro) is indicated for temporary hair reduction. The Diode Laser Therapy System(Model: FG2000-D+Pro/FG2000-D) is indicated for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen. It is suitable for all skin types (Fitzpatrick skin type I-VI), including tanned skin.
Device Story
Diode Laser Therapy System consists of main unit, handpiece, and accessories; utilizes selective photothermal interaction principle where specific laser wavelengths are absorbed by target chromophores. Models FG2000-B/B Pro are desktop units combining 755/808/1064 nm wavelengths for temporary hair reduction. Models FG2000-D+/D+Pro/E are vertical units using 808 nm wavelength for permanent hair reduction. Operated by clinicians in professional settings; user interface via LCD color touchscreen. Device emits laser energy to target hair follicles; clinical benefit is reduction of unwanted hair. No clinical data provided; safety established via bench testing and compliance with electrical/laser safety standards.
Clinical Evidence
No clinical data.
Technological Characteristics
Class IV diode laser; wavelengths: 755/808/1064 nm (multi) or 808 nm (single). Handpiece tip material: Quartz crystal. Energy fluence: 2-100 J/cm2; pulse frequency: 1-10 Hz; pulse width: up to 200ms. User interface: LCD color touchscreen. Biocompatibility: ISO 10993-5, 10993-10, 10993-23. Electrical safety: IEC 60601-1, IEC 60601-2-22, IEC 60825-1.
Indications for Use
Indicated for temporary hair reduction (FG2000-B/B Pro) or permanent hair regrowth reduction (FG2000-D+/D+Pro/E) in all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
February 20, 2024
Beijing ADSS Development Co., Ltd. Song Ying International Registration Manager Room 609, F6, Building 13, Yard 5 Tianhua Street, Daxing District, Beijing, 102600 China
Re: K231896
Trade/Device Name: Diode Laser Therapy System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 17, 2024 Received: January 18, 2024
Dear Song Ying:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shlomit Halachmi
Digitally signed by Shlomit Shlomit Halachmi -S Halachmi -S Date: 2024.02.20 18:39:10-05'00'
For Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231896
Device Name Diode Laser Therapy System
#### Indications for Use (Describe)
The Diode Laser Therapy System (Model: FG2000-B/FG2000-B Pro) is indicated for temporary hair reduction.
The Diode Laser Therapy System(Model: FG2000-D+Pro/FG2000-D) is indicated for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen. It is suitable for all skin types (Fitzpatrick skin type I-VI), including tanned skin.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="display:inline-block;"> <span style="font-size:16px;"> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span style="display:inline-block;"> <span style="font-size:16px;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
Submission Number: K231896
### 1. Submitter's Information
#### Establishment Registration Information
#### Beijing ADSS Development Co., Ltd.
Room 609, F6, Building 13, Yard 5 Tianhua Street, Daxing District, Beijing, 102600, P. R. China
## Contact Person of applicant Name: Song Ying International Registration Manager TEL: +86 13051615111 Email: irm@adss.com.cn Summary Preparation date: February 18, 2024
#### 2. Device Information
| Trade Name: | Diode Laser Therapy System |
|--------------------|--------------------------------------------------------|
| Model: | FG2000-B/FG2000-B Pro/FG2000-D+/FG2000-D+Pro/ FG2000-E |
| Common name: | Powered Laser Surgical Instrument |
| Regulation Number: | 21 CFR 878.4810 |
| Product code: | GEX |
| Regulation Class: | Class II |
| Review panel: | General & Plastic Surgery |
### 3. Predicate Device Information
Predicate Device of Diode Laser Therapy System (Model: FG2000-B/FG2000-B Pro) 510(K) Number: K230371 Trade Name: Alma Soprano Titanium Manufacturer: Alma Lasers Inc.
Predicate Device of Diode Laser Therapy System (Model: FG2000-D+Pro/ FG2000-D+Pro/ FG2000-E) 510(K) Number: K220381 Trade Name: Diode Laser Therapy Systems Beijing LaserTell Medical Co., Ltd Manufacturer:
{5}------------------------------------------------
## 4. Device description
The Diode Laser Therapy System consists of main unit, handpiece, and its accessories. Diode Laser Therapy System is intended for hair removal, mainly based on the principle of selective photothermal interaction, which means that lasers of specific wavelengths can only be selectively absorbed by the target color base.
The Diode Laser Therapy System(Model: FG2000-B/FG2000-B Pro) are desktop devices, which combines 3 wavelengths (755+808+1064 nm) into a single handpiece to achieve purpose for temporary hair reduction.
The Diode Laser Therapy System (Model: FG2000-D+Fro/ FG2000-E) are vertical device, which is a single wavelength (808 nm only) device.
### 5. Indications for Use
The Diode Laser Therapy System (Model: FG2000-B/FG2000-B Pro) is indicated for temporary hair reduction.
The Diode Laser Therapy System(Model: FG2000-D+Fro/FG2000-E) is indicated for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs regrowing when measured at 6,9 and 12 months after the completion of a treatment regimen. It is suitable for all skin types (Fitzpatrick skin type I-VI), including tanned skin.
# 6. Summary of technological characteristics of device compared to the predicate device
| Item | Subject device | Predicate device<br>(K230371) | Comparison |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Device Name | Diode Laser Therapy<br>System | Alma Soprano Titanium | / |
| Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 | SE |
| Class | Class II | Class II | SE |
| Product Code | GEX | GEX | SE |
| Indication for<br>use | The Diode Laser Therapy<br>System is indicated for<br>temporary hair reduction. | The Soprano Titanium diode laser module is<br>intended for use in dermatology procedures<br>requiring coagulation.<br><b>The indications for use for the Soprano<br/>Trio diode laser module include:</b> | SE<br>The indication of<br>subject device is<br>covered by the<br>predicate device.<br>The subject |
| | The Super Hair Removal (SHR) Mode is intended for temporary hair reduction. | device is only intended to use for temporary hair reduction. | |
| | The Soprano Trio diode laser module HR mode is intended for use in dermatology procedures requiring coagulation. The indications for use for the Soprano Trio diode laser HR module include: Benign vascular and vascular dependent lesions. | | |
| | 810nm Applicator<br>Soprano Titanium 810 nm applicator intended use and indications for use:<br>The indications for use for the 810nm Modified Diode Laser Module 2 cm2 include:<br>The Hair Removal (HR) and Super Hair Removal (SHR) Mode are intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen.<br>Use on all skin types (Fitzpatrick I-VI), including tanned skin.(HR, and SHR Modes) | | |
| | 755nm applicator<br>Soprano Titanium 755 nm applicator intended use and indications for use:<br>The indications for use for the 755nm Diode Laser Module include:<br>• The Hair Removal (HR) and Super Hair Removal (SHR) Mode are intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen.<br>• Use on all skin types (Fitzpatrick I-VI), including tanned skin.(HR, SHR Modes) | | |
| | NIR Applicator<br>NIR Applicator intended use and indications for use<br>The Alma Lasers NIR Modules intended use is to emit energy in the near infrared (NIR) | | |
| | | | |
| | | spectrum to provide topical heating.<br>The indications for use for NIR Modules are:<br>• Elevating the tissue temperature for the<br>temporary relief of minor muscle pain and joint<br>pain and stiffness,<br>• The temporary relief of minor joint pain<br>associated with arthritis,<br>• The temporary increase in local circulation<br>where applied, and<br>• The relaxation of muscles; may also help<br>muscle spasms, minor sprains and strains, and<br>minor muscular back pain. | |
| Laser Type | Diode laser | Diode laser | SE |
| Laser<br>Classification | Class IV | Class IV | SE |
| Laser<br>Wavelength | combination of<br>755nm/808nm/1064nm | combination of 755nm/810nm/1064nm | SE |
| Spot size | 12×16 mm*mm(1.92cm²) | 20*10 mm*mm (2cm²); | Different 1. |
| Pulse width | up to 200ms | Up to 200ms | SE |
| Energy<br>fluence | 2~8J/cm2 | 2 to 8J/cm2 | SE |
| Pulse<br>frequency | 1~10Hz | up to 10 Hz | SE |
| User<br>interface | LCD Color Touchscreen | LCD Color Touchscreen | SE |
| Electrical<br>Safety | Comply with ANSI/AAMA<br>ES 60601-1,<br>IEC 60601-2-22 | Comply with ANSI/AAMA ES 60601-1,<br>IEC 60601-2-22 | SE |
| EMC | Comply with IEC 60601-1-<br>2 | Comply with IEC 60601-1-2 | SE |
| Laser Safety | Comply with IEC 60601-2-<br>22,<br>IEC 60825-1 | Comply with IEC 60601-2-22,<br>IEC 60825-1 | SE |
| Handpiece tip<br>material | Quartz crystal | Sapphire | SE |
| Cytotoxicity | comply with ISO 10993-<br>5:2009 | comply with ISO 10993-5:2009 | SE |
| Sensitization | comply with ISO 10993-<br>10:2021 | comply with ISO 10993-10:2010 | SE |
| Irritation | comply with ISO 10993-23:<br>2021 | comply with ISO 10993-10:2010 | SE |
| Item | Subject device | Predicate device (K220381) | Comparison |
| Device<br>Name | Diode Laser Therapy System | Diode Laser Therapy Systems | / |
| Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 | SE |
| Class | Class II | Class II | SE |
| Product<br>Code | GEX | GEX | SE |
| Indication for<br>use | The Diode Laser Therapy System<br>is indicated for permanent<br>reduction in hair regrowth defined<br>as a long term, stable reduction in<br>the number of hairs re-growing<br>when measured at 6,9 and 12<br>months after the completion of a<br>treatment regimen. It is suitable for<br>all skin types (Fitzpatrick skin type<br>I-VI), including tanned skin. | The Diode Laser Therapy Systems<br>is intended for hair removal,<br>permanent hair reduction on all skin<br>types (Fitzpatrick skin type I-VI),<br>including tanned skin.<br>Permanent hair reduction is defined<br>as the long-term, stable reduction in<br>the number of hairs regrowing when<br>measured at 6, 9, and 12 months<br>after the completion of a treatment<br>regime. | SE |
| Laser Type | Diode laser | Diode laser | SE |
| Laser<br>Classification | Class IV | Class IV | SE |
| Laser<br>Wavelength | 808nm | 808nm | SE |
| Spot size | 10×10mm*mm | 15x15mm*mm; | Different 1 |
| Pulse width | 2~100ms | 1-300ms | Different 2 |
| Energy<br>fluence | 2~100J/cm2 | 1~100J/cm2 | Different 3 |
| Pulse<br>frequency | 1~10Hz | 1~10Hz | SE |
| User<br>interface | LCD Color Touchscreen | LCD Color Touchscreen | SE |
| Electrical<br>Safety | Comply with<br>ANSI/AAMA ES 60601-1,<br>IEC 60601-2-22 | Comply with<br>ANSI/AAMA ES 60601-1,<br>IEC 60601-2-22 | SE |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE |
| Laser Safety | Comply with IEC 60601-2-22,<br>IEC 60825-1 | Comply with IEC 60601-2-22,<br>IEC 60825-1 | SE |
| Handpiece<br>tip materia | Quartz crystal | Sapphire | SE |
| Cytotoxicity | comply with ISO 10993-5:2009 | comply with ISO 10993-5:2009 | SE |
| Sensitization | comply with ISO 10993-10:2021 | comply with ISO 10993-10:2010 | SE |
| Irritation | comply with ISO 10993-23: 2021 | comply with ISO 10993-10:2010 | SE |
## 6.1 For Model: FG2000-B/FG2000-B Pro
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# 6.2 For Model: FG2000-D+/FG2000-D+Pro/FG2000-E
{9}------------------------------------------------
# 7. Non-Clinical Tests Performed for Safety and effectiveness
Non clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
IEC60601-1:2005, AMD1:2012, AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
AAMI ES60601-1:2005, ES60601-1:2005/AMD1 1:2012, ES60601-1:2005/AMD2:2021 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC /TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems IEC 60601-2-22:2007 + A1:2012 +A2:2019 Medical Electrical Equipment - Part 2: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment. IEC 60825-1:2014 (Third Edition) Safety of laser products - Part 1: Equipment classification, and requirements
ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity ISO 10993-10: 2021 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation.
# 8. Clinical Testing Performed
There was no clinical testing performed.
# 9. Conclusions
The subject device Diode Laser Therapy System (Model: FG2000-B/FG2000-B Pro) have the similar intended use and similar characteristics as the cleared predicate device. And the bench testing contained in this submission supplied demonstrate that the differences existed do not raise any new questions of safety or effectiveness. Thus, subject device Diode Laser Therapy System is Substantially Equivalent (SE) to the predicate device(Alma Soprano Titanium, K230371).
The subject device Diode Laser Therapy System (Model: FG2000-D+FG2000-D+Pro/FG2000-E) have the same intended use and similar characteristics as the cleared predicate device. And the bench testing contained in this submission supplied demonstrate that the differences existed do not raise any
{10}------------------------------------------------
new questions of safety or effectiveness. Thus, subject device Diode Laser Therapy System is Substantially Equivalent (SE) to the predicate device (K220381).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.