The Maxtec MaxBlend 2+p is intended to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen and pressure being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e., hospital, subacute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life-supporting device or life sustaining device.
Device Story
Maxtec MaxBlend 2+p integrates air/oxygen blender, battery-powered oxygen/pressure monitor, and adjustable flowmeter into single assembly. Device receives medical-grade air and oxygen inputs; blender provides precise mixing; flowmeter controls delivery rate. Oxygen monitor uses galvanic fuel cell (MAX-550E) to measure concentration; solid-state transducer monitors airway pressure. Used in professional healthcare settings by trained personnel. Output displayed on LCD; includes audible/visual alarms for oxygen concentration and pressure limits. Assists clinicians in managing gas delivery; benefits patients by ensuring precise oxygen concentration and pressure monitoring.
Clinical Evidence
No clinical data. Bench testing only, including AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 80601-2-55, and ISO 11195. Biocompatibility of patient-contacting materials (disposable pressure tubing) leveraged from secondary predicate (K221734) per ISO 10993 and ISO 18562 standards.
Technological Characteristics
Microprocessor-controlled device. Components: air/oxygen blender, flowmeter, galvanic fuel cell oxygen sensor (MAX-550E), solid-state pressure transducer. Power: 4x AA alkaline batteries. Connectivity: standalone. Standards: ANSI/AAMI/ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 80601-2-55, ISO 11195. Materials: identical to predicates K161718/K221734.
Indications for Use
Indicated for infant, pediatric, and adult patients requiring continuous air/oxygen gas mixture delivery and monitoring of oxygen concentration and airway pressure. Intended for use by qualified, trained personnel under physician direction in professional healthcare settings (hospitals, sub-acute, nursing-care). Not for life-support/life-sustaining use.
Regulatory Classification
Identification
A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is a blue square with the letters 'FDA' in white. Next to the blue square, the words 'U.S. FOOD & DRUG' are written in blue, followed by 'ADMINISTRATION' in a smaller font size.
December 4, 2023
Maxtec, LLC % Paul Dryden Consultant ProMedic Consulting LLC 131 Bav Point Dr NE Saint Petersburg, Florida 33704
Re: K231895
Trade/Device Name: Maxtec MaxBlend2+p Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: Class II Product Code: BZR, CCL, CAP Dated: October 31, 2023 Received: November 1, 2023
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231895
Device Name
Maxtec MaxBlend 2+p
Indications for Use (Describe)
The Maxtec MaxBlend 2+p is intended to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen and pressure being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healtheare settings, i.e., hospital, subacute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required.
This is not intended as a life-supporting device or life sustaining device.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
1-Dec-23
#### Date Prepared:
Maxtec, Inc. 2305 South 1070 West Salt Lake City, UT 84119 Tel - 1-732-244-0010 x1151
| Official Contact: | Rebecca Hudson - QARA |
|----------------------------|---------------------------------------------------------------------------------|
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC<br>131 Bay Point Dr NE<br>St. Petersburg, FL 33704 |
| Proprietary or Trade Name: | Maxtec MaxBlend2+p |
| Classification CFR: | 21 CFR 868.5330 |
| Classification Code: | BZR |
| Classification Name: | Mixer, Breathing Gases, Anesthesia Inhalation |
| Classification CFR: | 21 CFR 868.1720 |
| Classification Code: | CCL |
| Classification Name: | Analyzer, Gas, Oxygen, Gaseous-Phase |
| Classification CFR: | 21 CFR 868.2600 |
| Classification Code: | CAP |
| Classification Name: | Monitor, Airway Pressure |
| Predicate Device: | Maxtec MaxBlend 2 - K161718 |
| Secondary Predicate: | MaxO2ME+p -K221734 |
#### Device Description:
The Maxtec MaxBlend 2+p is an oxygen delivery device which incorporates an air/oxygen blender, battery powered oxygen and pressure monitor, and an adjustable flowmeter, all in a single assembly. The integral air/oxygen blender provides precise mixing of medical grade air and oxygen. The flowmeter provides control of the flow rate delivered. The oxygen monitor measures the oxygen concentration from the blender's gas flow, displays these measured concentrations, and provides user selectable high and low oxygen alarms. It also allows the user to monitor pressure simultaneously using adjustable high and low alarm limits.
#### Indications for Use:
The Maxtec MaxBlend 2+p is designed to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen and pressure being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e.. hospital, sub-acute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required.
This is not intended as a life-supporting device or life sustaining device.
#### Patient Population:
Infant, pediatric, and adult patients for MaxBlend 2+p.
{5}------------------------------------------------
# Environments of use:
Hospital, sub-acute, and nursing-care facilities
We present the proposed device vs. the predicates in the Tables below.
{6}------------------------------------------------
# 510(k) Summary Page 3 of 11 Table 1 Comparison of MaxBlend 2+p to Primary Predicate
| Attributes | Subject Device | Primary Predicate |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Maxtec MaxBlend 2+p | MaxBlend 2 K161718 |
| Product Classification | Primary - BZR Secondary - CCL<br>New Secondary - CAP | Primary - BZR<br>Secondary - CCL |
| Indications for Use | The Maxtec MaxBlend 2+p is designed to provide a<br>continuous air/oxygen gas mixture and to continuously<br>monitor the concentration of oxygen and pressure being<br>delivered to infant, pediatric, and adult patients. It is a<br>restricted medical device intended for use by qualified, trained<br>personnel, under the direction of a physician, in professional<br>healthcare settings, i.e., hospital, sub-acute, and nursing-care<br>facilities where the delivery and monitoring of air/oxygen<br>mixtures is required.<br><br>This is not intended as a life-supporting device or life<br>sustaining device. | The MaxBlend 2 and MaxBlend Lite are designed to provide<br>a continuous air/oxygen gas mixture and to continuously<br>monitor the concentration of oxygen being delivered to infant,<br>pediatric, and adult patients. It is a restricted medical device<br>intended for use by qualified, trained personnel, under the<br>direction of a physician, in professional healthcare settings,<br>i.e., hospital, sub-acute, and nursing-care facilities where the<br>delivery and monitoring of air/oxygen mixtures is required.<br><br>This is not intended as a life supporting device. |
| New indication | Pressure monitoring | N/A See Table 2 for this feature |
| Environments of Use | Professional healthcare settings, i.e., hospital, sub-acute, and<br>nursing-care facilities where the delivery and monitoring of<br>air/oxygen mixtures is required. | Professional healthcare settings, i.e., hospital, sub-acute, and<br>nursing-care facilities where the delivery and monitoring of<br>air/oxygen mixtures is required. |
| | Not for use in MRI environments | Not for use in MRI environments |
| Patient Population | Infant, pediatric, and adult patients | Infant, pediatric, and adult patients |
| Components | Oxygen monitoring<br>Air/Oxygen Blender<br>Flowmeter<br>Bleed<br>Adding pressure feature | Oxygen monitoring<br>Air/Oxygen Blender<br>No Flowmeter<br>Bleed |
| Air/Oxygen Mixer Features | | |
| Blender | Blender is<br>CareFusion / Bird - K883038 or<br>BioMed Devices - K925982 | Blender is<br>CareFusion / Bird - K883038 or<br>BioMed Devices - K925982 |
| Gas Supply Type | Air / Oxygen | Air / Oxygen |
| Pressure | 30 to 75 psi | 30 to 75 psi |
| Attributes | Subject Device<br>Maxtec MaxBlend 2+p | Primary Predicate<br>MaxBlend 2 K161718 |
| Mixed gas stability | + 1% oxygen | + 1% oxygen |
| Flow range of Blenders | Low flow model – 0-30 Lpm<br>High flow model – 0-70 Lpm | Low flow model – 0-30 Lpm<br>High flow model – 0-100 Lpm |
| Pressure supply<br>differential alarm | Air / oxygen pressure must be < 20 psi an alarm sounds | Air / oxygen pressure must be < 20 psi an alarm sounds |
| Pressure Drop | < 6 psi @ 50 psi | < 6 psi @ 50 psi |
| Bleed flow | 3-13 Lpm at 50 psi depending upon model | 3-13 Lpm at 50 psi depending upon model |
| Oxygen Monitor Features | | |
| Oxygen measurement<br>range | 0-100% | 0-100% |
| Total Accuracy | ±3% Actual oxygen level over full operating temperature<br>range | ±3% Actual oxygen level over full operating temperature<br>range |
| Response Time | 90% of final value in approx. 15 seconds at 23°C | 90% of final value in approx. 15 seconds at 23°C |
| Warm-up Time | None required | None required |
| High alarm range | 16 – 100% | 16 – 100% |
| Operating Temperature | 15°C – 40°C (59°F – 104°F) | 15°C – 40°C (59°F – 104°F) |
| Storage Temperature | -15°C – 50°C (5°F – 122°F) | -15°C – 50°C (5°F – 122°F) |
| Humidity | 0-95% (non-condensing) | 0-95% (non-condensing) |
| Expected use-life of sensor | 1,500,000 O2% hours (approx. 2 years) | 1,500,000 O2% hours (approx. 2 years) |
| Flowmeter Features | | |
| Accuracy | + 10% of indicated value when inlet pressure is 50 psi | + 10% of indicated value when inlet pressure is 50 psi |
| Flow meter ranges | 0 – 3 Lpm<br>0 – 15 Lpm<br>0 – 30 Lpm<br>0 – 70 Lpm | 0 – 3 Lpm<br>0 – 30 Lpm<br>0 – 70 Lpm<br>0 – 100 Lpm |
| Attributes | Subject Device | Primary Predicate |
| | Maxtec MaxBlend 2+p | MaxBlend 2 K161718 |
| Pressure Monitor Features (adding this feature) | | |
| Technology | Microprocessor controlled device | See Table 2 |
| Pressure sensor type | Solid-state pressure transducer | See Table 2 |
| Displayed information | Low pressure alarm setting | See Table 2 |
| | High pressure alarm setting | |
| | Alarm silence indicator | |
| | Power supply indicator | |
| | Average (mean) pressure | |
| | Audible and visual alarm | |
| Pressure Measurement | -15 to +60 cmH2O | See Table 2 |
| Range | | |
| Pressure Resolution | 1 cmH2O | See Table 2 |
| Display resolution | 0.5 cm H2O | See Table 2 |
| Total Accuracy | + 1 cmH2O | See Table 2 |
| Low alarm range | 1 - 30 cmH2O | See Table 2 |
| High alarm range | 1 - 60 cmH2O | See Table 2 |
| Alarm delay | 3 seconds (pressure only) | See Table 2 |
| Zero calibration | Yes | See Table 2 |
| System interface | Disposable tubing with filter and with or without Nafion | See Table 2 |
| | connects to tee or face mask | |
| Operating Temperature | 15°C - 40°C, 0-95% RH…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.