Virtue Male Sling System with Alexis Wound Retractor Convenience Kit
K231891 · Coloplast · OTM · Sep 25, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231891
Device Name
Virtue Male Sling System with Alexis Wound Retractor Convenience Kit
Applicant
Coloplast
Product Code
OTM · General, Plastic Surgery
Decision Date
Sep 25, 2023
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K231891 · Sep 25, 2023
Virtue Male Sling System with Alexis Wound Retractor Convenience Kit
Coloplast
Post-market surveillance data; Clinical literature
The sponsor utilized post-market surveillance data and clinical literature analysis to update the device labeling, specifically regarding adverse events and procedural information, to align with current clinical practice.
Post-market surveillance; Literature analysis; Labeling update
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Post-market surveillance and literature analysis; Retrospective analysis of post-market data and clinical literature
Patients treated with the Virtue Male Sling System
Not applicable for this study
Adverse event profile and procedural updates
Indications for Use
The Virtue Male Sling System is an implantable, suburethral support sling indicated for the treasment of male stress urinary incontinence (SUI). The Applied Alexis Wound Retractor is indicated for use to: Access the abdominal cavity during surgery through an atraumatically retracted incision. Deliver maximum exposure of the abdominal cavity with minimum incision. Protect against wound contamination during laparoscopic and open surgery. The smaller two sizes of Alexis are also intended to be used to: Seal off the incision opening to permit insufflating the peritoneum. Convert the incision wound to an additional trocar port site. Access the thoracic cavity or other soft tissue retraction during cardiac and general surgical procedures through an atraumatically-retracted incision.
Device Story
Virtue Male Sling System is an implantable, suburethral support sling for male stress urinary incontinence (SUI). Device features four-arm design providing dual mechanism: compression of bulbous urethra and proximal urethral elevation. Constructed of non-absorbable, medical-grade knitted monofilament polypropylene. System includes Virtue Introducer (polycarbonate/stainless steel) for transobturator and pre-pubic passage and Alexis Wound Retractor for surgical site access. Used by surgeons in clinical settings. Device provides mechanical support to urethra to treat incontinence. No changes to device design or materials in this submission; updates limited to labeling, contraindications, and procedural guidance.
Clinical Evidence
No clinical data provided. Substantial equivalence determined via comparison of design, materials, and intended use to predicate device. Labeling updates based on post-market surveillance and literature analysis.
Technological Characteristics
Sling: non-absorbable medical grade knitted monofilament polypropylene. Introducer: polycarbonate thermoplastic elastomer handle, medical grade stainless steel needle. Four-arm design. Permanent implant. Mechanical support principle.
Indications for Use
Indicated for male patients with stress urinary incontinence (SUI) requiring suburethral support. Alexis Wound Retractor indicated for surgical access, wound protection, and retraction in abdominal, thoracic, and general surgical procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Virtue Male Sling System with Alexis Wound Retractor (K113496)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 25, 2023
Coloplast A/S Jennifer Mrkvicka Principal Regulatory Affairs Specialist 1601 West River Road North Minneapolis, Minnesota 55411
Re: K231891
Trade/Device Name: Virtue Male Sling System with Alexis Wound Retractor Convenience Kit Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTM, GAD Dated: June 27, 2023 Received: June 27, 2023
Dear Jennifer Mrkvicka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Angel A. Soler-garcia -S
for Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231891
Device Name
Virtue® Male Sling System with Alexis® Wound Retractor
Indications for Use (Describe)
The Virtue Male Sling System is an implantable, suburethral support sling indicated for the treasment of male stress urinary incontinence (SUI).
The Applied Alexis Wound Retractor is indicated for use to:
Access the abdominal cavity during surgery through an atraumatically retracted incision. Deliver maximum exposure of the abdominal cavity with minimum incision. Protect against wound contamination during laparoscopic and open surgery.
The smaller two sizes of Alexis are also intended to be used to:
Seal off the incision opening to permit insufflating the peritoneum.
Convert the incision wound to an additional trocar port site.
Access the thoracic cavity or other soft tissue retraction during cardiac and general surgical procedures through an atraumatically-retracted incision.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Virtue Male Sling System
# 510(k) SUMMARY
#### I. SUBMITTER
| 510(K) Owner's Name: | Coloplast A/S |
|----------------------------|-----------------------------------------------------------------------------------|
| Address: | Holtedam 1<br>3050 Humlebaek, Denmark |
| Phone/Email: | Office: 612-707-5062<br>Email: usjmrk@coloplast.com |
| Name of Contact Person: | Jennifer Mrkvicka<br>Principal Regulatory Affairs Specialist |
| Address/Contact: | 1601 West River Road North<br>Minneapolis, MN 55411 |
| Date Prepared: | September 18, 2023 |
| II.<br>DEVICE | |
| Trade or Proprietary Name: | Virtue® Male Sling System with Alexis® Wound Retractor |
| Common or Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical, For Stress Urinary Incontinence, Male<br>(OTM)<br>Retractor (GAD) |
| Review Panel: | Gastroenterology/Urology |
| Classification Number: | 21 CFR 878.3300 (OTM)<br>21 CFR 878.4800 (GAD) |
| Product Code: | OTM (Virtue)<br>GAD (Alexis) |
| Product Classification: | Class II (Virtue)<br>Class I (Alexis) |
| III.<br>PREDICATE DEVICE | |
### K113496 510(k) Number: Trade Name: Virtue Male Sling System with Alexis Wound Retractor
The predicate has not been subject to a design-related recall. No reference devices were used in this submission.
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The Virtue Male Sling is an implantable, suburethral support sling intended for the surgical treatment of male stress urinary incontinence (SUI). The four-arm design of the Virtue Male Sling provides a dual mechanism of action by providing both compression of the bulbous urethra and proximal urethral elevation. The Virtue Male Sling is constructed of non-absorbable medical grade knitted, monofilament polypropylene and is intended to be permanently implanted.
The Virtue Introducer is used to facilitate transobturator and pre-pubic passages during the surgical placement of the Virtue Male Sling. The introducer is manufactured from polycarbonate thermoplastic elastomer (handle) and medical grade stainless steel (needle). The Virtue Introducer is only to be used with the Virtue Male Sling System in accordance with the Instructions for Use.
The Virtue Male Sling System with Alexis Wound Retractor includes one (1) Virtue Male Sling, one (1) single-use Virtue Introducer and one (1) Alexis Wound Retractor intended to facilitate the placement of the Virtue Male Sling.
#### V. INDICATIONS FOR USE
The Virtue Male Sling System is an implantable, suburethral support sling indicated for the treatment of male stress urinary incontinence (SUI).
The Applied Alexis Wound Retractor is indicated for use to:
- | Access the abdominal cavity during surgery through an atraumatically retracted incision.
- Deliver maximum exposure of the abdominal cavity with minimum incision. |
- Protect against wound contamination during laparoscopic and open surgery. -
The smaller two sizes of Alexis are also intended to be used to:
- | Seal off the incision opening to permit insufflating the peritoneum
- Convert the incision wound to an additional trocar port site. |
- -Access the thoracic cavity or other soft tissue retraction during cardiac and general surgical procedures through an atraumatically-retracted incision.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Virtue Male Sling System has identical indications for use, design, materials, and target population as the predicate device. In this 510(k) submission, Coloplast A/S only provided revised labeling; there are no changes to the device itself. The labeling revisions include, but are not limited to, the addition of a contraindication to align with other Coloplast mesh Instructions for Use (IFU) manuals, updates to adverse events based on post market surveillance and literature analysis, and procedure updates aligned with current literature. The differences in device labeling do not raise questions of safety or effectiveness, and the subject device was found to be substantially equivalent with the predicate.
{5}------------------------------------------------
#### VII. PERFORMANCE DATA
Performance data was not necessary for the substantial equivalence determination as the subject device and the predicate device have identical intended use, target population, sterilization technique, biocompatibility features, overall device design features, and duration of use.
# VIII. LABELING
In this 510(k), Coloplast A/S revised the IFU to add a contraindication to align with other Coloplast mesh IFUs, updated adverse events based on post market surveillance and literature analysis, and aligned the surgical procedure with current literature.
#### IX. CONCLUSIONS
Based on the information presented in this submission, it can be concluded that the Virtue Male Sling System with Alexis Wound Retractor is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.