K231887 · Jeil Medical Corporation · HRS · Dec 14, 2023 · Orthopedic
Device Facts
Record ID
K231887
Device Name
ARIX Ankle Distal Tibia System
Applicant
Jeil Medical Corporation
Product Code
HRS · Orthopedic
Decision Date
Dec 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ARIX Ankle Distal Tibia System(Ankle fusion) is indicated for facilitating arthrodesis of the ankle including tibiotalocalcaneal and tibiotalar joints.
Device Story
The ARIX Ankle Distal Tibia System is a rigid orthopedic fixation system consisting of metallic plates and screws. It is used by surgeons in an operating room setting to facilitate arthrodesis (joint fusion) of the ankle, including the tibiotalocalcaneal and tibiotalar joints. The device provides mechanical stabilization to the bone segments to promote fusion. It is a purely mechanical implant system; it does not involve electronic inputs, software, or automated processing.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including compliance with ASTM F382 (metallic bone plates) and ASTM F543 (metallic bone screws).
Technological Characteristics
Constructed from Titanium Alloy (Ti-6Al-4V) meeting ASTM F136. System consists of plates and screws in various configurations, shapes, and sizes. Mechanical fixation device; no energy source, connectivity, or software. Sterilization method not specified.
Indications for Use
Indicated for patients requiring arthrodesis of the ankle, specifically involving the tibiotalocalcaneal and tibiotalar joints.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
December 14, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the full name. The full name of the agency is "U.S. Food & Drug Administration".
Jeil Medical Corporation Dajung Lee RA Specialist 702-703-704-705-706-804-805-807-812-815-ho. 55 Digital-ro 34-gil, Guro-gu Seoul, 08378 Korea. South
Re: K231887
Trade/Device Name: ARIX Ankle Distal Tibia System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: November 7, 2023 Received: November 13, 2023
Dear Dajung Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
### K231887 - Dajung Lee
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumaya Ali-S
Shumaya Ali, M.P.H.
Shumaya Ali, M.P.H Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231887
Device Name ARIX Ankle Distal Tibia System
Indications for Use (Describe)
The ARIX Ankle Distal Tibia System(Ankle fusion) is indicated for facilitating arthrodesis of the ankle including tibiotalocalcaneal and tibiotalar joints.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Digital-ro 34-gil. Guro-gu. Se Tel: +82 2 850 3591 / Fax: +82 2 850 3536
510(k) Summary
[As required by 21 CFR 807.92]
## 1. Date Prepared [21 CFR 807.92(a)(a)]
7th December 2023
### 2. Submitter's Information [21 CFR 807.92(a)(1)]
| • Name of Sponsor: | Jeil Medical Corporation |
|--------------------|-------------------------------------------------------------------------------------------------|
| • Address: | 702,703,704,705,706,804,805,807,812,815-ho, 55, Digital-ro 34-gil, Guro-gu, Seoul, 08378, Korea |
| Contact Name: | Dajung Lee / RA Specialist |
|----------------|----------------------------|
| Telephone No.: | +82 2 850 3591 |
| Fax No.: | +82 2 850 3536 |
| Email Address: | dajunglee@jeilmed.co.kr |
#### . Registration Number: 3004049923
| • Name of Manufacturer: | Same as Sponsor |
|-------------------------|-----------------|
| - Address: | Same as Sponsor |
### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| • | Trade Name; | ARIX Ankle Distal Tibia System |
|---|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| • | Common Name; | - Plate, Fixation, Bone (Primary)<br>- Screw, Fixation, Bone |
| • | Classification Name; | - Single/multiple component metallic bone fixation appliances<br>and Accessories (Primary)<br>- Smooth or threaded metallic bone fixation fastener |
| • | Classification Panel; | Orthopedic |
| • | Classification regulation; | 21 CFR 888.3030 (Primary), 21 CFR 888.3040 |
| • | Product code; | HRS (Primary), HWC |
| • | Device Class; | II |
{4}------------------------------------------------
Tel: +82.2.850.3591 / Fax:
### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[
The legally marketed device(s) to which substantial equivalence is claimed is/are:
| Primary Predicate Device | K181067 – AlignX Ankle Fusion System,<br>Extremity Medical, LLC. |
|----------------------------|-----------------------------------------------------------------------|
| Secondary Predicate Device | K141735 – Arthrex Ankle Fusion Plating System,<br>ARTHREX, INC. |
| Reference Device | K202912 – ARIX Ankle Distal Tibia System,<br>Jeil Medical Corporation |
There are no significant differences between the subject device and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to these devices in intended use and technological characteristics as internal fixation components.
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
The ARIX Ankle Distal Tibia System is rigid fixation consisting of plates and screws in various configurations, shapes, and sizes.
The ARIX Ankle Distal Tibia System is made of Titanium Alloy (Ti-6AI-4V), which meet ASTM F136, Standard Specification for Wrought Titanium-6 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well-known biocompatibility.
#### 6. Indications for use [21 CFR 807.92(a)(5)]
The ARIX Ankle Distal Tibia System(Ankle fusion) is indicated for facilitating arthrodesis of the ankle including tibiotalocalcaneal and tibiotalar joints.
# 7. Technological Characteristics [21 CFR 807.92(a)(6)]
### ARIX Ankle Distal Tibia System, Bone Plate, Bone Screw:
Based on the technological feature comparison, the subject device was found that there are no significant differences between the subject device and predicate devices that would adversely affect the use of the product and it is substantially equivalent to predicate device in technological characteristics.
### Non-Clinical Test Summary:
Bench tests were conducted to ensure the safety and effectiveness of the device as well as to demonstrate substantial equivalence to the predicate device. The test results demonstrated that the subject device complies with the following standards:
- ASTM F382, Standard Specification and Test Method for Metallic Bone Plates
- . ASTM F543, Standard Specification and Test Methods for Metallic Medical Bone Screws
The results of this testing indicate that the ARIX Ankle Distal Tibia System is equivalent to predicate device.
### Clinical Test Summary
No clinical studies were considered necessary and performed.
{5}------------------------------------------------
### 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
Based on the submitted information in this premarket notification, the subject device is substantially equivalent to the predicate devices in terms of:
- Intended use
- . Technological characteristics (Design features, Material, Surface Treatment, Sterilization methods, Biocompatibility and Performance)
The subject device has met the performance, safety, and effectiveness of the device for its intended use.
#### 9. Conclusion [21 CFR 807.92(b)(3)]
In all respects, the ARIX Ankle Distal Tibia System is substantially equivalent to the legally marketed device. Above all, the subject device has equivalent intended use and technological characteristics. Further, nonclinical verification and validation to determine substantial equivalence provide additional evidence that subject device is substantially equivalent to the predicate device in terms of safety, efficacy, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.