K231877 · Lineage Biomedical, Inc. · FMF · Nov 30, 2023 · General Hospital
Device Facts
Record ID
K231877
Device Name
NavaClick Injection System
Applicant
Lineage Biomedical, Inc.
Product Code
FMF · General Hospital
Decision Date
Nov 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NavaClick Injection System is intended for injection of fluids into the body. It is specifically indicated for injection of botulinum neurotoxin, consistent with the neurotoxin's approved labeling.
Device Story
NavaClick Injection System is a manually-operated 1 mL piston syringe kit; includes needles and vial adapter. Features calibrated barrel, luer-lock connector, and ratcheted plunger. Provides audible and tactile feedback upon delivery of 0.02 mL fluid increments; feedback mechanism engaged by 90-degree twist of thumb rest. Used by clinicians for precise delivery of botulinum neurotoxin. Device enhances injection efficiency and precision. Sterile, single-use, non-pyrogenic; sterilized via electron beam radiation. Kit components facilitate convenient neurotoxin administration.
Clinical Evidence
No clinical data. Bench testing only. Performance testing conducted per ISO 7886-1, ISO 80369-7, and ISO 7864. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute systemic toxicity, hemolysis, material-mediated pyrogenicity). Sterility validated per ISO 11137. Shelf-life and package integrity validated per ASTM F1980, ASTM D4332, and ASTM D4169.
Technological Characteristics
Piston syringe; 1 mL volume. Materials: polypropylene barrel/plunger, polyisoprene seal. Features ratcheted plunger for 0.02 mL tactile/audible feedback. Luer-lock connector. Sterilized via electron beam radiation (min 25kGy). Compliant with ISO 7886-1 (syringes), ISO 80369-7 (connectors), and ISO 7864 (needles).
Indications for Use
Indicated for injection of fluids into the body, specifically botulinum neurotoxin, consistent with the neurotoxin's approved labeling.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
November 30, 2023
Bradley Pliskow Director of R&D and Operations 620 NW Dixie Hwy Stuart, Florida 34994
Re: K231877
Trade/Device Name: NavaClick Injection System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, FMF, FMI, LHI Dated: September 22, 2023 Received: November 3, 2023
Dear Bradley Pliskow:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
#### Juliane C. Lessard -S Sincerely,
Juliane C. Lessard, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K231877
Device Name
NavaClick Injection System
Indications for Use (Describe)
The NavaClick Injection System is intended for injection of fluids into the body. It is specifically indicated for injection of botulinum neurotoxin, consistent with the neurotoxin's approved labeling.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K231877 510(k) Summary
#### SUBMITTED BY:
| Company Name: | Lineage Biomedical. |
|---------------|--------------------------------------|
| Address: | 620 NW Dixie Hwy<br>Stuart, FL 34994 |
| Telephone: | 1-772-539-9749 |
| Fax: | N/A |
| CONTACT PERSON: | Bradley Pliskow |
|---------------------------------|----------------------------------------------|
| DATE PREPARED: | Nov 27, 2023 |
| TRADE NAME: | NavaClick Injection System |
| COMMON and CLASSIFICATION NAME: | Piston Syringe |
| CLASSIFICATION REG/PANEL: | 21 CFR 880.5860/ General and Plastic Surgery |
| PROCODE: | FMF |
#### PREDICATE DEVICE:
Merit Syringe K171362
#### DEVICE DESCRIPTION:
The NavaClick Injector is a manually-operated, 1 mL piston syringe consisting of a calibrated, hollow barrel with a standard luer-lock connector into which is inserted a closely-fitted, ratcheted, moveable plunger. The syringe is printed with a graduated scale (1 mL volume), and has audible and tactile feedback upon delivery of each 0.02 mL fluid. The syringe has been designed to enhance the efficiency and precision of neurotoxin injections.
The syringe has been packaged with other components for convenience. The NavaClick Injection System is a kit containing the NavaClick Injectors (syringes), needles, and a vial adapter for delivery of neurotoxin per the labeled indication.
The product is sterile, single use only, non-pyrogenic and sterilized by electron beam radiation.
#### INDICATIONS FOR USE/INTENDED USE:
The NavaClick Injection System is intended for injection of fluids into the body and is specifically indicated for injection of botulinum neurotoxin, consistent with the neurotoxin's approved labeling.
The general intended use for injection of fluids into the body is identical to that of the predicate device; however, the subject device is not intended for withdrawal of fluids from the body. I In addition, the subject device is specifically used to inject botulinum neurotoxin per the neurotoxin product's intended use. This difference does not alter the general intended use of the subject
{4}------------------------------------------------
device for injection of fluids which is identical to a section of the predicate device's intended use statement and covered in 21 CFR 880.5860.
## COMPARISON of TECHNOLOGICAL CHARACTERISTICS:
Results from testing based on "Guidance on the Content of Premarket Notification [510(K)] Submissions for Piston Syringes" April 1993, support substantial equivalence of the NavaClick Injector to the predicate Merit Syringe cleared under K171362.
| Attribute | NavaClick Injection<br>System (Syringe) | Predicate Device:<br>Merit Syringe K171362 | Analysis |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Lineage Biomedical | Merit Medical Systems | |
| Design | Piston Syringe | Piston Syringe | Same |
| Device Class | II | II | Same |
| Procode | FMF | FMF | Same |
| Syringe Type | Piston Syringe | Piston Syringe | Same |
| Intended Use | Intended to inject fluids into<br>the body and is specifically<br>indicated for injection of<br>botulinum neurotoxin<br>consistent with the<br>neurotoxin's approved<br>labeling | Used to inject fluids into, or<br>withdraw fluids from, the<br>body | Different<br>The general intended use<br>for injection of fluids is the<br>same for both the subject<br>and predicate device and is<br>covered in 21 CFR<br>880.5860 (piston syringe).<br>No new questions of safety<br>or effectiveness. |
| Principle of<br>Operations | Manually operated by<br>advancing and withdrawing<br>the plunger within the barrel | Manually operated by<br>advancing and withdrawing<br>the plunger within the barrel | Same |
| Specific Drug<br>Use | Neurotoxin | Not specified | Different<br>Injection of botulinum toxin<br>(neurotoxin) is typically<br>achieved using any<br>conventional piston syringe;<br>a specific delivery device is<br>not required or identified in<br>published prescribing<br>information for common<br>botulinum toxin drugs.<br>Therefore, no new<br>questions of safety or<br>effectiveness for use of the<br>NavaClick piston syringe for<br>delivery of neurotoxin. |
| Design | Standard piston syringe<br>construction. The device<br>uses a two-piece molded<br>ratcheted plunger. Fitting<br>offered with male luer lock<br>connector. | Standard three-piece piston<br>syringe. The device uses a<br>single-piece molded<br>ratcheted plunger. Fitting<br>offered with male luer lock<br>connector | Different<br>The subject device has a<br>free mode (plunger<br>operates freely with no<br>feedback), and a feedback<br>mode (ratchet provides |
| Attribute | NavaClick Injection<br>System (Syringe) | Predicate Device:<br>Merit Syringe K171362 | Analysis |
| Materials | The barrel is constructed<br>using polypropylene, the<br>plunger from polypropylene,<br>and the seal from<br>polyisoprene | The barrel is constructed<br>using polypropylene, the<br>plunger from polypropylene,<br>and the seal from silicone. | Different<br>Both polyisoprene and<br>silicone are commonly used<br>plunger seal materials.<br>Testing confirms<br>performance and<br>biocompatibility of subject<br>device. No new questions<br>of safety or effectiveness. |
| Operational<br>Volume | Operational Volume of 1 mL | Operational volume of 10<br>mL | Different<br>Use of a plunger within the<br>syringe barrel to move fluid<br>into and out of the syringe<br>for operation of the syringe<br>is the same for both<br>devices. No new questions<br>of safety or effectiveness. |
| Graduation | Printed with accurate<br>graduation lines that are<br>compliant with ISO 7886-1 | Printed with accurate<br>graduation lines that are<br>compliant with ISO 7886-1 | Different<br>The subject device has a<br>graduation scale division of<br>0.1 mL and sub-division of<br>0.01 mL. Although volume<br>increments for the<br>graduation scale are<br>different, graduation for<br>both the subject and<br>predicate is in compliant<br>with ISO 7886-1. No new<br>questions of safety or<br>effectiveness. |
| Barrel<br>Transparency | Clear as required by ISO<br>7886-1 | Clear as required by ISO<br>7886-1 | Same |
| Biocompatibility | Per ISO 10993-1 | Per ISO 10993-1 | Same |
| Performance<br>Testing | ISO 7886-1<br>ISO 80369-7 | ISO 7886-1<br>ISO 594-2 (withdrawn) | Same |
| Re-use | Single use only | Single use only | Same |
| Sterilization | Electron Beam | Ethylene Oxide | Different |
| Attribute | NavaClick Injection<br>System (Syringe) | Predicate Device:<br>Merit Syringe K171362 | Analysis |
| | | | Both the subject and<br>predicate devices are<br>sterilized using well-<br>established methods.<br>Electron Beam sterilization<br>of the subject device is<br>compliant with ISO 11137-<br>1, ISO 11137-2 and ISO<br>11137-3. No new questions<br>of safety or effectiveness. |
| Accessories | Packaged as a kit | Not a kit configuration | Different<br>The NavaClick syringe is<br>packaged with needles and<br>a vial adapter for<br>convenience of the user.<br>The kit components are<br>previously 510k cleared<br>devices to be utilized per<br>the cleared indications. No<br>new questions of safety of<br>effectiveness. |
{5}------------------------------------------------
{6}------------------------------------------------
The NavaClick Injector has the same intended use, materials of construction, and principle of operations as the predicate device. There are minor differences between the subject and predicate devices. Although they are both 3-part piston syringes consisting of a barrel, plunger and tip, the Nava device also has grips to enhance control. Both devices have ratcheted plungers; however, the mechanism to produce an audible clicking and tactile feedback is different. The Nava device allows the feedback mechanism to engage by twisting the thumb rest 90 degrees. These differences are not significant with regard to the function of the syringe to inject fluids; these minor differences do not raise different questions of safety and effectiveness.
### Performance Testing
The NavaClick Injection System components described in this summary have been tested and demonstrated to be in conformance with the following FDA recognized standards:
- . ISO 7886-1:2017 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
- ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare applications • - Part 7: Connectors for intravascular or hypodermic applications
- . ISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods
{7}------------------------------------------------
## Biocompatibility Testing
Biocompatibility testing of the predicate device was conducted per ISO 10993-1:2018 and the risk-based approach outlined in FDA quidance document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" Sep 4, 2020. Based on the standard, the kit components were defined as externally communicating, limited exposure and either blood path indirect (syringe and adapter) or circulating blood (needle). The following testing was conducted on all kit components and no deviations were noted.
- . Cytotoxicity (ISO 10993-5:2009 "Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity")
- Sensitization (ISO 10993-10:2021 "Biological evaluation of medical device Part 10: . Tests for skin sensitization")
- Irritation (ISO 10993-23:2021 "Biological evaluation of medical device – Part 23: Test for irritation")
- . Acute Systemic Toxicity (ISO 10993-11:2017 "Biological evaluation of medical device -Part 11: Tests for systemic toxicity")
- Hemolysis (ISO 10993-4:2017 "Biological evaluation of medical device – Part 4: Selection of tests for interactions with blood")
- Material Mediated Pyrogenicity (ISO 10993-11:2017 "Biological evaluation of medical . device - Part 11: Tests for systemic toxicity")
Evaluation of particulate in the fluid path per USP <788> has been conducted on the NavaClick Injection System. Although there are no USP specifications for particulate matter found in or on medical devices, by comparison, the kit components met the specification for small volume parenteral fluids.
Results from the biocompatibility testing support the biocompatibility and safety of the subject device.
## Sterility and Shelf Life
The sterilization method for the NavaClick Injection System is electron beam irradiation with a sterilization dose minimum of 25kGy. Acceptable test results from the sterilization validation demonstrated the ability of the electron beam process to consistently vield a sterility assurance level (SAL) of 10° for the NavaClick Injection System. Sterilization and endotoxin testing was completed in accordance with the following standards. No deviations were noted.
- ISO 11137-1: 2006/Amd 2:2019 Sterilization of health care products Radiation Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 11137-2: 2013 Sterilization of health care products Radiation Part 2: Establishing the sterilization dose
{8}------------------------------------------------
- ISO 11137-3:2017 Sterilization of health care products Radiation Part 3: Guidance on dosimetric aspects of development, validation and routine control
- AAMI TIR29:2012 (R2017) Guide For Process Characterization And Control In Radiation Sterilization Of Medical Devices
- USP <85> Bacterial Endotoxin Test ●
Accelerated stability testing demonstrates the sterility and performance of the device to support the shelf-life of 1 year.
Product stability testing of the NavaClick Injection System was conducted after accelerated aging per ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices. Product performance testing of each component was conducted according to the relevant ISO standard as well as functional performance testing as a kit.
Package (sterile barrier) testing was performed on the subject device after accelerated aging per ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices. Package testing was also performed after simulated environmental conditioning (ASTM D4332-22) and distribution (ASTM D4169-22). Sterile barrier integrity testing was completed in accordance with the following standards. No deviations were noted.
- Visual inspection (ASTM F1886)
- Bubble Leak Test (ASTM 2096-11) ●
- Seal Strength (ASTM F88/F88M-21)
## Conclusion
Based on the intended use, technological characteristics, performance testing, and biocompatibility studies, the NavaClick Injection System meets the requirements that are considered essential for its intended use. The technological differences between the predicate (K171362) Merit Syringe and the subject device do not raise any new questions of safety or effectiveness. The NavaClick Injection System is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.