CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System
K231870 · Ge Medical Systems Information Technologies, Inc. · DQK · Dec 13, 2023 · Cardiovascular
Device Facts
Record ID
K231870
Device Name
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System
Applicant
Ge Medical Systems Information Technologies, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Dec 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Pediatric
Indications for Use
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are intended to be used by trained operators under direct supervision of a licensed health care practitioner on adult and pediatric patients. The CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are designed to acquire, process, record, archive, analyze and output (12 and 15 lead) ECG data during a period of physiologic stress or during a resting ECG test, acquire data from ancillary devices (such as Spirometry and Ambulatory Blood Pressure), provide median morphology recordings and record ECG in real-time with and without arrhythmia detection. The arrhythmia detection portion of CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are provided to the user for the convenience of automatic detection of arrhythmias but do not provide alarms. CASE V7.0 Cardiac Testing System and Cardiac Testing System provide the control of external devices (typically a treadmill or bicycle ergometer) and communicate with centralized electronic/digital storage system via data networks. CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System provide a user selectable option for printouts of prognostic scores on selected reports. Vector loops are also available. CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System can be configured in a network environment for multiple CASE and CardioSoft stations allowing the user to create a central database of patient demographics and collected patient physiological data. CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are intended to be used primarily in the hospital but can also be used in clinics, physician offices, outreach centers or wherever exercise, stress testing, ECG, Spirometry or ambulatory blood pressure testing is performed. CASE V7.0 Cardiac Testing System and Cardiac Testing System offer no diagnostic opinion to the user. Instead, it provides interpretive statements of morphology, rhythm, and conduction for which the physician renders his/her own medical opinion. CASE V7.0 Cardiac Testing System and Cardiac Testing System are not intended to be used as a transport device or for home use. CASE V7.0 Cardiac Testing System and Cardiac Testing System are not intended for use as a vital signs physiological monitor. CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for intracardiac use. CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for use as an emergency device. CASE V7.0 Cardiac Testing System and Cardiac Testing System will not cause abnormal operation of a patient's cardiac pacemaker or other electronic stimulators. CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for use with high frequency surgical units. Disconnect the patient from CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System before using the high frequency surgical unit.
Device Story
System acquires, processes, records, archives, and analyzes 12/15-lead ECG data during resting or physiologic stress tests; integrates ancillary data from spirometry and ambulatory blood pressure devices. Operates via medical cart (CASE) or customer-provided PC (CardioSoft) using wired (CC14) or wireless (GEH ECG 1200) acquisition modules. Controls external exercise equipment (treadmills/ergometers). Provides arrhythmia detection (no alarms), prognostic scores, and vector loops. Used in hospitals, clinics, and physician offices by trained operators. Outputs interpretive statements for physician review; does not provide diagnostic opinions. Facilitates network-based central database management of patient demographics and physiological data. Benefits include standardized cardiac testing, automated data acquisition, and streamlined clinical workflow through EMR/MUSE integration.
Clinical Evidence
No clinical studies required. Substantial equivalence supported by bench testing, risk analysis, requirements reviews, verification testing, and EMI-EMC/safety testing per IEC 60601-2-25 and ANSI/AAMI ES60601-1 standards.
Technological Characteristics
System uses wired (CC14) or wireless (GEH ECG 1200) acquisition modules. Connectivity via network (DICOM, TLS 1.2, DCP). Software runs on Windows 10 IoT/Professional/Server. Database migrated to Microsoft SQL Server. Complies with ANSI/AAMI ES60601-1 and IEC 60601-2-25. ECG bandwidth 0.04-150 Hz. Sample rate 500 Hz.
Indications for Use
Indicated for adult and pediatric patients requiring resting ECG, stress test ECG, spirometry, or ambulatory blood pressure monitoring. Used by trained operators under licensed healthcare practitioner supervision. Not for home, transport, emergency, or intracardiac use.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
CASE Cardiac Testing System, CS Cardiac Testing System (K103678)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
December 13, 2023
GE Medical Systems Information Technologies, Inc. Manjunatha Kn Senior Regulatory Affairs Leader 9900 Innovation Drive Wauwatosa, Wisconsin 53226
Re: K231870
Trade/Device Name: CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: November 15, 2023 Received: November 15, 2023
## Dear Manjunatha Kn:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K231870
#### Device Name
CASE V7.0 Cardiac Testing System CardioSoft V7.0 Cardiac Testing System
#### Indications for Use (Describe)
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are intended to be used by trained operators under direct supervision of a licensed health care practitioner on adult and pediatric patients.
The CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are designed to acquire, process, record, archive, analyze and output (12 and 15 lead) ECG data during a period of physiologic stress or during a resting ECG test, acquire data from ancillary devices (such as Spirometry and Ambulatory Blood Pressure), provide median morphology recordings and record ECG in real-time with and without arrhythmia detection.
The arrhythmia detection portion of CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are provided to the user for the convenience of automatic detection of arrhythmias but do not provide alarms.
CASE V7.0 Cardiac Testing System and Cardiac Testing System provide the control of external devices (typically a treadmill or bicycle ergometer) and communicate with centralized electronic/digital storage system via data networks.
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System provide a user selectable option for printouts of prognostic scores on selected reports. Vector loops are also available.
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System can be configured in a network environment for multiple CASE and CardioSoft stations allowing the user to create a central database of patient demographics and collected patient physiological data.
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are intended to be used primarily in the hospital but can also be used in clinics, physician offices, outreach centers or wherever exercise, stress testing, ECG, Spirometry or ambulatory blood pressure testing is performed.
CASE V7.0 Cardiac Testing System and Cardiac Testing System offer no diagnostic opinion to the user. Instead, it provides interpretive statements of morphology, rhythm, and conduction for which the physician renders his/her own medical opinion.
CASE V7.0 Cardiac Testing System and Cardiac Testing System are not intended to be used as a transport device or for home use.
CASE V7.0 Cardiac Testing System and Cardiac Testing System are not intended for use as a vital signs physiological monitor.
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for intracardiac use.
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for use as an emergency device.
CASE V7.0 Cardiac Testing System and Cardiac Testing System will not cause abnormal operation of a patient's cardiac pacemaker or other electronic stimulators.
{3}------------------------------------------------
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for use with high frequency surgical units. Disconnect the patient from CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System before using the high frequency surgical unit.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is a stylized version of the letters "GE" inside a circle. The text "GE HealthCare" is written in a clean, sans-serif font.
# 510(k) Summary
| | In accordance with 21 CFR 807.92 the following summary of information is provided: |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | 01-June-2023 |
| Submitter: | GE Medical Systems Information Technologies, Inc. |
| | 9900 Innovation Drive |
| | Wauwatosa, WI 53226 |
| Primary Contact Person: | Manjunatha K N |
| | Senior Regulatory Affairs Leader |
| | GE Medical Systems Information Technologies, Inc. |
| | Phone: +91 (98453) 58045 |
| Secondary Contact Person: | Shlomi Deler |
| | Regulatory Affairs Director |
| | GE Medical Systems Information Technologies, Inc. |
| | Phone: +(972) 544301056 |
| Device Trade Name: | CASE V7.0 Cardiac Testing System |
| | CardioSoft V7.0 Cardiac Testing System |
| Common/Usual Name: | CASE V 7.0 / Cardiosoft V7.0 |
| Classification Names: | Computer, diagnostic, programmable |
| Product Code: | DQK 21CFR.870.1425 |
| Predicate Device(s): | CASE Cardiac Testing System, CS Cardiac Testing System (K103678) |
| | |
| Device Description: | The CASE V7.0 Cardiac Testing System and the CardioSoft V7.0<br>Cardiac Testing System are designed to be used for resting ECG,<br>stress test ECG, Spirometry, Ambulatory Blood Pressure and for<br>recording ECG in real-time with and without arrhythmia detection.<br>The CardioSoft V7.0 Cardiac Testing System will be offered as a<br>software only package including a front end for data acquisition. The |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a purple circle with the letters "GE" inside, followed by the words "GE HealthCare" in purple. The logo is simple and modern, and it is likely used to represent the company's brand.
CASE V7.0 Cardiac Testing System is a turnkey product utilizing the CardioSoft V7.0 Cardiac Testing software.
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Indications for Use: Testing System are intended to be used by trained operators under direct supervision of a licensed health care practitioner on adult and pediatric patients.
> The CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are designed to acquire, process, record, archive, analyze and output (12 and 15 lead) ECG data during a period of physiologic stress or during a resting ECG test, acquire data from ancillary devices (such as Spirometry and Ambulatory Blood Pressure), provide median morphology recordings and record ECG in real-time with and without arrhythmia detection.
> The arrhythmia detection portion of CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are provided to the user for the convenience of automatic detection of arrhythmias but do not provide alarms.
> CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System provide the control of external devices (typically a treadmill or bicycle ergometer) and communicate with centralized electronic/digital storage system via data networks.
> CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System provide a user selectable option for printouts of prognostic scores on selected reports. Vector loops are also available.
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System can be configured in a network environment for multiple CASE and CardioSoft stations allowing the user to create a central database of patient demographics and collected patient physiological data.
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are intended to be used primarily in the hospital but can also be used in clinics, physician offices, outreach centers or wherever exercise, stress testing, ECG, Spirometry or ambulatory blood pressure testing is performed.
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System offer no diagnostic opinion to the user. Instead, it provides interpretive statements of morphology, rhythm, and conduction for which the physician renders his/her own medical opinion.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circle with a stylized "GE" inside. To the right of the circle is the text "GE HealthCare" in purple. The text is in a sans-serif font.
| | CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended to be used as a transport device or for home use. |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for use as a vital signs physiological monitor. |
| | CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for intracardiac use. |
| | CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for use as an emergency device. |
| | CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System will not cause abnormal operation of a patient's cardiac pacemaker or other electronic stimulators. |
| | CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System are not intended for use with high frequency surgical units. Disconnect the patient from CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System before using the high frequency surgical unit. |
| Technology: | CASE V7.0 Cardiac Testing System and the CardioSoft V7.0 Cardiac Testing System employ the same fundamental scientific technology, basic design, construction, materials, energy source, control mechanism, and operating principles as the predicate device CASE / CS Cardiac Testing System (K103678) to acquire, process, record, archive, analyze, and output (12 and 15 lead) ECG data during a period of physiologic stress or during a resting ECG test, acquire data from ancillary devices (such as Spirometry and Ambulatory Blood Pressure), provide median morphology recordings, and record ECG in real-time with and without arrhythmia detection. |
| Determination of Substantial Equivalence: | CASE V7.0 Cardiac Testing System and the CardioSoft V7.0 Cardiac Testing System is substantially equivalent to predicate device CASE / CS Cardiac Testing System (K103678) as described in the table below |
CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the alphanumeric string "K231870" in a bold, sans-serif font. The string is horizontally oriented and appears to be a code or identifier. The characters are uniformly sized and spaced, creating a clear and legible sequence.
Image /page/7/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem with a stylized "GE" monogram inside, followed by the text "GE HealthCare" in a simple, sans-serif font. The emblem and text are both in a matching shade of purple.
| Feature/Function | Predicate Device<br>CASE V6.6 and CS V6.6<br>(K103678) | Proposed Device<br>CASE V7.0 and CardioSoft V7.0 | Explanation of Differences |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification /<br>Product Code | 21CFR.870.1425<br>/DQK | 21CFR.870.1425<br>/ DQK | Identical |
| Intended Use<br>Statement | CASE Cardiac Testing System and<br>CS Cardiac Testing System are<br>intended to be used by trained<br>operators under direct supervision<br>of a licensed health care<br>practitioner on adult and pediatric<br>patients. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM are intended to be<br>used by trained operators under<br>direct supervision of a licensed health<br>care practitioner on adult and<br>pediatric patients. | Equivalent:<br>The CardioSoft Cardiac<br>Testing System was<br>formerly named CS<br>Cardiac Testing System. |
| | The CASE Cardiac Testing System<br>and CS Cardiac Testing System are<br>designed to acquire, process,<br>record, archive, analyze and output<br>(12 and 15 lead) ECG data during a<br>period of physiologic stress or<br>during a resting ECG test, acquire<br>data from ancillary devices (such as<br>Spirometry and Ambulatory Blood<br>Pressure), provide median<br>morphology recordings and record<br>ECG in real-time with and without<br>arrhythmia detection. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM are designed to<br>acquire, process, record, archive,<br>analyze and output (12 and 15 lead)<br>ECG data during a period of<br>physiologic stress or during a resting<br>ECG test, acquire data from ancillary<br>devices (such as Spirometry and<br>Ambulatory Blood Pressure), provide<br>median morphology recordings and<br>record ECG in real-time with and<br>without arrhythmia detection. | |
| | The arrhythmia detection algorithm<br>of CASE Cardiac Testing System and<br>CS Cardiac Testing System are<br>provided to the user for the<br>convenience of automatic detection<br>of arrhythmias but does not provide<br>alarms. | The arrhythmia detection algorithm<br>of CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM are provided to the<br>user for the convenience of automatic<br>detection of arrhythmias but do not<br>provide alarms. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System provide<br>the control of external devices<br>(typically a treadmill or Ergometer)<br>and communicate with centralized<br>electronic/digital storage system<br>via data network. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM provide the control<br>of external devices (typically a<br>treadmill or Ergometer) and<br>communicate with centralized<br>electronic/digital storage system via<br>data network. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System provide<br>a user selectable option for<br>printouts of prognostic scores on<br>select reports. Vector loops are also<br>available. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM provide a user<br>selectable option for printouts of<br>prognostic scores on select reports.<br>Vector loops are also available. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System can be<br>configured in a network<br>environment for multiple CASE<br>stations and CS stations allowing<br>the user to create a central<br>database of patient demographics<br>and collected patient physiological<br>data. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM can be configured in<br>a network environment for multiple<br>CASE and CardioSoft stations<br>allowing the user to create a central<br>database of patient demographics<br>and collected patient physiological<br>data. | |
| Feature/Function | Predicate Device<br>CASE V6.6 and CS V6.6<br>(K103678) | Proposed Device<br>CASE V7.0 and CardioSoft V7.0 | Explanation of Differences |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System are<br>intended to be used primarily in the<br>hospital but can be used in clinics,<br>physician offices, outreach centers<br>or wherever exercise, stress testing,<br>ECG, Spirometry or ambulatory<br>blood pressure testing is<br>performed. | TESTING SYSTEM intended to be used<br>primarily in the hospital but can be<br>used in clinics, physician offices,<br>outreach centers or wherever<br>exercise, stress testing, ECG,<br>Spirometry or ambulatory blood<br>pressure testing is performed. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System offer no<br>diagnostic opinion to the user.<br>Instead, it provides interpretive<br>statements of morphology, rhythm,<br>and conduction for which the<br>physician renders his/her own<br>medical opinion. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM offer no diagnostic<br>opinion to the user. Instead, it<br>provides interpretive statements of<br>morphology, rhythm, and conduction<br>for which the physician renders<br>his/her own medical opinion. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System are not<br>intended to be used as a transport<br>device or for home use. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM are not intended to<br>be used as a transport device or for<br>home use. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System are not<br>intended for the use as a vital signs<br>physiological monitor. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM are not intended for<br>use as a vital signs physiological<br>monitor. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System are not<br>intended for intracardiac use. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM are not intended for<br>intracardiac use. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System are not<br>intended for the use as an<br>emergency device. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM are not intended for<br>use as an emergency device. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System will not<br>cause abnormal operation of a<br>patient's cardiac pacemaker or<br>other electronic stimulators. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM will not cause<br>abnormal operation of a patient's<br>cardiac pacemaker or other electronic<br>stimulators. | |
| | CASE Cardiac Testing System and<br>CS Cardiac Testing System are not<br>intended for use with high<br>frequency surgical units. Disconnect<br>the patient from CASE Cardiac<br>Testing System and CS Cardiac<br>Testing System before using the<br>high frequency surgical unit. | CASE V7.0 CARDIAC TESTING<br>SYSTEM, CARDIOSOFT V7.0 CARDIAC<br>TESTING SYSTEM are not intended for<br>use with high frequency surgical units.<br>Disconnect the patient from CASE<br>V7.0 CARDIAC TESTING SYSTEM,<br>CARDIOSOFT V7.0 CARDIAC TESTING<br>SYSTEM before using the high<br>frequency surgical unit | |
| Device Type | ECG recording system | ECG recording system | Identical |
| Modalities | Resting ECG<br>Exercise Test<br>Spirometry<br>Ambulatory Blood Pressure | Resting ECG<br>Exercise Test<br>Spirometry<br>Ambulatory Blood Pressure | Equivalent<br>Ergospirometry<br>measurement in the<br>Exercise testing, will not |
| Feature/Function | Predicate Device<br>CASE V6.6 and CS V6.6<br>(K103678) | Proposed Device<br>CASE V7.0 and CardioSoft V7.0 | Explanation of Differences |
| | Exercise Test and<br>ergospirometry/metabolic<br>test) | | be offered for sale with<br>the subject device in US. |
| Device<br>Description | CASE is a medical cart<br>including a thermal writer.<br>The system cart is based on<br>an embedded PC with CAM<br>14 acquisition module. The<br>application software is pre-<br>installed. CASE can be<br>configured to maintain a<br>local database or use a<br>shared database on a server.<br>Additional devices like<br>treadmills, ergometers<br>ambulatory blood pressure<br>devices and spirometer<br>devices can be connected.<br>CS utilizes a customer-<br>provided PC and consists of<br>the application software and<br>the CAM USB acquisition<br>module (CAM14 and CAM-<br>USB Interface), which<br>connects to the PC's USB<br>port. CS can be configured to<br>maintain a local database or<br>use a shared database on a<br>server. There are a number<br>of peripheral devices and<br>accessories that can be<br>connected. | CASE is a medical cart<br>including a thermal writer. The<br>system cart is based on an<br>embedded PC with wired<br>acquisition module CAM<br>Connect 14 (CC14) and GEH<br>ECG 1200 wireless acquisition<br>module. The application<br>software is pre-installed.<br>CASE can be configured to<br>maintain a local database or<br>use a shared database on a<br>server. Additional devices like<br>treadmills, ergometers<br>ambulatory blood pressure<br>devices and spirometer<br>devices can be connected.<br>CardioSoft utilizes a<br>customer-provided PC and<br>consists of the application<br>software and acquisition<br>modules (CAM Connect 14,<br>CAM-USB interface, and GEH<br>ECG 1200), which connects to<br>the PC's USB port. CardioSoft<br>can be configured to maintain<br>a local database or use a<br>shared database on a server.<br>There are a number of<br>peripheral devices and<br>accessories that can be<br>connected. | Equivalent<br>The proposed device<br>support both wired (CC-<br>14) and wireless (GEH<br>ECG 1200) ECG<br>acquisition modules<br>which don't impact safety<br>and effectiveness of the<br>device. Refer to section<br>18 for Performance data |
| Operating<br>System | CASE:<br>Windows XP embedded<br><br>CS:<br>Windows XP,<br>Windows XP Pro,<br>Windows Vista Home<br>Premium, Windows Vista<br>Business, | CASE:<br>Windows 10 IoT Enterprise 64<br>bit<br>CardioSoft:<br>Windows 10 Professional (64<br>bit)<br>Windows 10 Enterprise (64<br>bit).<br>Windows Server 2016 (64 bit).…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.