Fabrication of temporary crowns, bridges, inlays, onlays and veneers; Fabrication of long-lasting temporary restorations; Rebasing of prefabricated temporary crowns made of composite, polycarbonate or metal.
Device Story
PrimmaArt is a self-curing composite resin for indirect dental procedures, including temporary crowns, bridges, inlays, onlays, and veneers. The device consists of a base paste and a catalyzing paste stored in a cartridge and double-body syringe; these are automatically dispensed in a 1:1 ratio to minimize errors and air bubbles. It is intended for use by dental professionals in clinics or laboratories. The material is applied to the patient's teeth, requiring finishing and polishing after polymerization. The device provides temporary restorations, benefiting patients by protecting prepared teeth during the fabrication of permanent prosthetics. It is a non-reusable, temporary-use dental material.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 4049, ISO 7491, and ISO 10477 standards. Metrics included setting time (2m48s), heat release (0.30 mW/mg), viscosity (Base: 92733 mPa.s; Catalyst: 90133 mPa.s), radiopacity (1.68 mm), flexural strength (65.5 MPa), water sorption (9.389 µg/mm³), solubility (1.542 µg/mm³), and Vickers microhardness (11.81 HV). Accelerated stability studies (225 days) showed no significant physical-chemical degradation.
Technological Characteristics
Self-curing composite resin. Supplied in cartridge/double-body syringe for 1:1 dispensing. Complies with ISO 7491 (color stability), ISO 10477 (polymer-based crown/veneering materials), ISO 4049 (restorative materials), and ISO 10993 (biocompatibility). Non-sterile, single-use. Storage temperature 5° to 25°C.
Indications for Use
Indicated for fabrication of temporary crowns, bridges, inlays, onlays, and veneers; fabrication of long-lasting temporary restorations; and rebasing of prefabricated temporary crowns (composite, polycarbonate, or metal). No specific age or gender restrictions noted.
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 26, 2024
DentsCare LTDA % Rodrigo Abreu Regulatory Specialist United Regulatory LLC 2950 West Cypress Creek Rd. Ste 110 Fort Lauderdale, Florida 33309
Re: K231859
Trade/Device Name: PrimmaArt Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown And Bridge Resin Regulatory Class: Class II Product Code: EBG, POW Dated: September 27, 2024 Received: September 27, 2024
Dear Rodrigo Abreu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231859
Device Name PrimmaArt
Indications for Use (Describe)
- Fabrication of temporary crowns, bridges, inlays, onlays and veneers;
- Fabrication of long-lasting temporary restorations;
- Rebasing of prefabricated temporary crowns made of composite, polycarbonate or metal.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------|
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo and contact information for DentsCare LTDA. The logo features a stylized letter 'D' with wave-like lines inside. The text includes the company name, address (AV. EDGAR NELSON MEISTER, 474, JOINVILLE, SANTA CATARINA 89219-501 BRAZIL), and phone number (Ph: 55 - 47 - 3441-6131).
# K231859
# Section 5
# 510(k) SUMMARY
A) Submitter's Name: DENTSCARE LTDA
Owner / Operator Registration Number: 3007210751
### Manufacture Registration Number: 3007210751
B) Address: AV. EDGAR NELSON MEISTER, 474, JOINVILLE, SANTA CATARINA 89219-501 BRAZIL
#### C) Phone and Fax Numbers
Phone: +55 (47) 34416131
D) Contact Person:
Roberta Uyara
Tel.: +55 (47) 99160-4500
E) Preparation Date: November 25, 2024
F) Classification Name: Crown And Bridge, Temporary, Resin
Common / Usual Name: Temporary crown and bridge resin
Proprietary Name: PrimmaArt
Product Code: EBG
Class: Class II
Regulation: 21 CFR 872.3770
### G) Device Description
Self-curing composite that is highly efficient for the production of temporary pieces for indirect dental procedures such as inlays, onlays, crowns, veneers and bridges. The product consists of a base paste and a catalyzing paste stored in a cartridge and in a double-body syringe, from where they are automatically dispensed in the correct proportion (1:1), providing practical clinical procedures with reduced chances of errors or incorporated bubbles.
### H) Substantial Equivalence (Predicates):
| Predicate<br>Type | Predicate<br>Description | 510(k) Number | Model | Company |
|-------------------|-------------------------------|---------------|------------------|-------------------|
| Predicate | Main Predicate | K120779 | Structur 3 | VOCO<br>GmbH |
| Reference | Biocompatibility<br>Reference | K191306 | Llis, Vittra APS | Dentscare<br>LTDA |
The PrimmaArt is equivalent with the following product:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for DentsCare. The logo features a stylized letter "D" in a gray color, with three horizontal lines inside the upper part of the "D", resembling waves or airflow. Below the symbol is the text "DentsCare" in a sans-serif font, also in gray, with the first letter capitalized.
# DENTSCARE LTDA BENTOGARE ETDA
AV. EDGAR NELSON MEISTER, 474, JOINVILLE, SANTA
CATARINA 89219-501 BRAZIL Ph: 55 - 47 - 3441-6131
| Reference | Biocompatibility<br>Reference | K201707 | Opallis, Opallis Flow | Dentscare<br>LTDA |
|-----------|-------------------------------|---------|-----------------------|-------------------|
| Reference | Biocompatibility<br>Reference | K183465 | Allcem, Allcem Core | Dentscare<br>LTDA |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo and contact information for DentsCare LTDA. The logo features a stylized letter 'D' with three horizontal lines inside it, and the name 'DentsCare' written below. The contact information includes the company name, address (AV. EDGAR NELSON MEISTER, 474, JOINVILLE, SANTA CATARINA 89219-501 BRAZIL), and phone number (Ph: 55 - 47 - 3441-6131).
### I) Indications for Use Statement:
- · Fabrication of temporary crowns, bridges, inlays, onlays and veneers;
- · Fabrication of long-lasting temporary restorations;
- · Rebasing of prefabricated temporary crowns made of composite, polycarbonate or metal.
| Indications for Use Comparison | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| PrimmaArt | Structur 3 |
| - Fabrication of temporary crowns, bridges, inlays, onlays and veneers; | - Fabrication of temporary crowns, bridges, inlays, onlays, partial crowns, veneers and temporary posts. |
| - Fabrication of long-lasting temporary restorations; | - Fabrication of long-term temporary restorations |
| - Rebasing of prefabricated temporary crowns made of composite, polycarbonate or metal. | - Rebasing of prefabricated temporary crowns made of composite, polycarbonate or metal. |
#### Discussion:
The products are similar. Both devices are indicated to produce temporary pieces (crowns, bridges, inlays, onlays and veneers), long-term temporary restorations and rebasing of prefabricated temporary crowns made or metal.
#### J) Technological Characteristics Comparison:
The predicate device used to establish equivalence for the PrimmaArt device is outlined below. This section of this submission will provide a comparison of dechnical specifications of the PrimmaArt to the predicate device stratified by functional modality.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo and contact information for DentsCare LTDA. The logo features a stylized letter 'D' with three horizontal lines inside it, along with the name 'DentsCare' written below. The contact information includes the address 'AV. EDGAR NELSON MEISTER, 474, JOINVILLE, SANTA CATARINA 89219-501 BRAZIL' and the phone number 'Ph: 55 - 47 - 3441-6131'.
| Device<br>Manufacturer<br>and<br>Common Name | PrimmaArt<br>Dentscare Ltda | Structur 3<br>VOCO GmbH |
|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| 510k # | Not assigned yet | K120779 |
| Classification | Class II | Class II |
| Regulation # | 21 CFR 872.3770 | 21 CFR 872.3770 |
| Product Code | EBG | EBG |
| Classification Name | Temporary crown and bridge resin | Temporary crown and bridge resin |
| Patient Population | All the groups | All the groups |
| Prescription Use | RX only | RX only |
| Environment | Dental prosthetics and authorized laboratories<br>and clinics.<br>PrimmaArt must be stored in temperatures<br>between 5° to 25°C. | Dental prosthetics and authorized laboratories<br>and clinics.<br>Structur 3 must be stored in temperatures<br>between 4° to 23°C |
| Applicable Standards | ISO 7491; ISO 10477; ISO 10993; ISO 4049 | ISO 7491; ISO 10477; ISO 10993; ISO 4049 |
| Device Sterilization | Not Applicable | Not Applicable |
| Primary Package Container: | Syringe and cartridge | Syringe and cartridge |
| Shelf life | 2 years | Not disclosed by the manufacturer |
| Use the same materials or<br>substances in contact with<br>the same human tissues<br>or body fluids? | YES | YES |
| Is the product in compliance<br>to EN ISO 10993? | YES | YES |
| Tissues | Enamel and Dentin | Enamel and Dentin |
| Reusable | NO | NO |
| Duration | Temporary | Temporary |
| Part of body | Oral, teeth | Oral, teeth |
| it<br>used<br>for<br>the<br>ls<br>same<br>clinical condition? | yes | yes |
| Is it used at the same site in the body? | yes | yes |
| Is it used in a similar population? | yes | yes |
| Is it used for the same intended purpose? | yes | yes |
| Is not foreseen to deliver significantly different performances? | no | no |
| Is it similar conditions of use? | yes | yes |
| Is it similar specifications and properties | yes | yes |
| Is it similar principles of operation? | yes | yes |
| CLINICAL STEP | PrimmaArt (Dentscare) | Structur 3 (VOCO) |
| Two options: total isolation or<br>relative isolation | YES | YES |
| Application<br>according<br>to<br>adhesive technique | YES | YES |
| Require<br>finishing<br>and<br>polishing | YES | YES |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for DentsCare. The logo features a stylized letter D with horizontal lines inside it, above the word "DentsCare" in a sans-serif font. The text "Dents" is also visible, but it is unclear what it refers to.
| Specification | PrimmaArt | Structur 3 |
|----------------------------------------------------------|--------------------------------------------------------|------------------------------------------------------|
| Shade consistency and color<br>stability ISO 4049 e 7491 | No more than a small color change | Not performed. Not disclosed by the<br>manufacturer. |
| Setting time and heat<br>released<br>ISO 4049 | Setting time: 2m48s<br>Heat released: 0.30 mW/mg | Setting time: 3m02s<br>Heat released: 0.31 mW/mg |
| Viscosity | Base Paste: 92733 mPa.s<br>Catalyst paste: 90133 mPa.s | Not disclosed by the manufacturer |
| Surface Finish ISO 10477 | According to ISO 10477, the polymerized | Not disclosed by the manufacturer |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo and contact information for DentsCare LTDA. The logo features a stylized letter D with three horizontal lines inside. The contact information includes the address: AV. EDGAR NELSON MEISTER, 474, JOINVILLE, SANTA CATARINA 89219-501 BRAZIL, and the phone number: Ph: 55 - 47 - 3441-6131.
| | composite has a shiny surface | |
|-------------------------------------------|------------------------------------------------------------------|-------------------------------------------------------------------|
| Radio-opacity ISO 4049 | 1.68 mm | < 1 |
| Flexing Resistance ISO 10477 | 65.5 MPa | 68.0 MPa |
| Water sorption and solubility<br>ISO 4049 | Sorption: 9.389 ± 1.26 µg/mm³<br>Solubility: 1.542 ± 0.32 µg/mm³ | Sorption: 10.527 ± 1.29 µg/mm³<br>Solubility: 2.016 ± 0.31 µg/mm³ |
| Working time | More than 60 seconds | Not disclosed by the manufacturer |
| Microhardness Vickers | 11,81 HV | 4,55 HV |
#### Discussion:
The subject device is similar to the predicate device in that they are light activated, usar self-curing, restorative composite, permanently cementing restorations of temporary pieces for indirect dental procedures. The subject device and the predicate device have substantially equivalent of indications for use, technological properties and fulfil the ISO 7491 e ISO 10477 ISO 4049 minimum requirements. Despite the minor differences in formulation between the predicate, they can be considered equivalent, since tifferences does not affect the products laboratorial and clinical performance and safety.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for DentsCare. The logo features a stylized letter "D" with three horizontal lines emanating from the top left corner, suggesting movement or airflow. Below the symbol, the word "DentsCare" is written in a clean, sans-serif font, with the "D" in a slightly larger size and a darker shade of gray compared to the rest of the word.
DENTSCARE LTDA AV. EDGAR NELSON MEISTER, 474, JOINVILLE, SANTA CATARINA 89219-501 BRAZIL Ph: 55 - 47 - 3441-6131
## K)Applicable Standards:
In order to reach substantially equivalent to the predicate device PrimmaArt was developed, as well produced in compliance with recognized international regulations and standards for the medical device industry.
ISO 7491:2000 – Dental materials – Determination of colour stability
ISO 10477:2018 – Dentistry – Polymer-based crown and veneering materials
ISO 4049:2009 - Dentistry - Polymer - based, restorative materials
ISO 10993-1 - Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process (EN ISO 10993-1:2009)
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows a logo for a company called DentsCare. The logo features a stylized letter "D" with horizontal lines emanating from it, suggesting movement or flow. Next to the logo is the text "DentsCare" in a clear, sans-serif font.
DENTSCARE LTDA AV. EDGAR NELSON MEISTER, 474, JOINVILLE, SANTA CATARINA 89219-501 BRAZIL Ph: 55 - 47 - 3441-6131
# L) Non-clinical Testing:
In order to study the performance of the product, pre-clinical tests were performed according to the table below.
| Test | Specification | Results |
|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Setting<br>time<br>and<br>heat<br>released | According to EN ISO 4049, for<br>self curing restorative<br>materials, the setting time<br>shall be no more than 5<br>minutes.<br><br>The time considered for the<br>project is 2\'30\" and 3\'10\" and<br>the heat released less than<br>0,60mW/mg. | Product was evaluated using the<br>Differential Scanning Calorimetry<br>Test (DSC) and presented results<br>according to the design input. |
| Viscosity | According to the design input<br>the viscosity of the base and<br>the catalyzer should be<br>between 60000 and 100000<br>mPa.s. | The test was done using the<br>Brooksfield digital viscosimeter and<br>the results were within the defined<br>limits. |
| Surface Finish | According to the ISO 10477<br>standard, the specification for<br>finishing of a surface is that<br>the polymerised composite<br>presents a reflective surface. | It was observed that the surface of<br>the specimen was reflective, so this<br>material is considered to be in<br>accordance with the specifications<br>of the ISO 10477 standard. |
| Shade<br>consistency<br>and<br>color<br>stability | According to EN ISO 4049 and<br>ISO 7491, acceptance must be<br>made as long as there is no<br>more than a small color<br>change. | Only a small color difference<br>(acceptable according to ISO 4049)<br>was attested, when comparing<br>Sample 3 (color A3) with Bisacrilic<br>Technical Profile FGM. In the other<br>analyzes, no changes were<br>identified. |
| Radiopacity | According to ISO 4049, the<br>material<br>is<br>considered<br>radiopaque if the specimen has<br>a value of > 1.0 mm, when<br>compared to the aluminum<br>scale, therefore, the value of<br>"X" obtained through the<br>equation of the line, must be ><br>1. | The FGM Bisacrylic resin, presented<br>results above that specified by ISO<br>4049. |
| Flexing<br>Resistance | According to ISO 10477, the<br>specified<br>flexural strength is<br>≥50MPa. | The FGM Bisacrylic<br>type<br>resin<br>presented all the results within the<br>values specified in the ISO<br>10477 standard,<br>therefore,<br>the<br>material<br>meets<br>the<br>requirements<br>determined<br>in |
| | | terms of 3-point Flexural Strength. |
| Water sorption<br>and solubility | According to ISO 4049, the<br>specified for sorption is a<br>maximum of 40 µg/mm³ and<br>for solubility a maximum of 7.5<br>µg/mm³. | All results below are within the<br>specified limit, therefore the<br>material is considered to conform. |
| Working time | According to EN ISO 4049, the<br>working time for cementation<br>materials must be at least 60<br>seconds. | The results show that the product<br>has a working time of more than<br>60 seconds. |
| Accelerated<br>Stability<br>Studies | Study created to accelerate the<br>possible chemical degradation<br>and/or physical changes of the<br>product in forced conditions of<br>storage. | The FGM Bisacrylic Resin product<br>did not show significant changes in<br>terms of its physical-chemical<br>properties during the accelerated<br>stability process (225 days). |
| Evaluation<br>of<br>Report<br>Long-Term<br>Stability<br>(Shelf) | Study designed to verify the<br>physical and chemical<br>characteristics of the product<br>during the expected shelf life.<br>The results are used to confirm<br>the expiration date and storage<br>conditions. | We don't have any conclusive data<br>about shelf life yet. At the moment<br>we are using the estimated shelf<br>life in 2 years in the storage<br>condition of 25°C, based on no<br>accelerated stability test. |
| Michohardness<br>Vickers | According to the Project Entry<br>Requirements, the Hardness<br>(HV) must be between 5 and<br>15 HV, and in the same way,<br>the results must be similar or<br>superior (justified) when<br>compared to the competitor<br>under study. | All results are within the specified<br>limit, so the material is considered<br>as conform. |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for DentsCare. The logo features a stylized letter "D" in a muted blue-gray color, with three horizontal lines emanating from the left side of the "D", suggesting movement or flow. Below the symbol, the word "DentsCare" is written in a sans-serif font, also in the same muted blue-gray color, completing the branding for the dental care company.
Conclusion: Based on the performance test applied to this PrimmaArt and the predicate comparison (Structur 3), we conclude that the device in question is substantially equivalent with the predicate, as all products meet the same recognized standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.