The XO Cath Microcatheter is intended for peripheral vascular use. The microcatheter can be used for controlled and selective infusion of diagnostic, embolic or therapeutic materials into the vessel.
Device Story
XO Cath Microcatheter is a single-lumen, metal alloy shaft device with micro-cuts for flexibility; designed to support guidewire during peripheral vascular access and provide conduit for delivery of embolic/diagnostic/therapeutic materials. Available in 2.0Fr and 2.6Fr lumen sizes with straight or bern tip shapes. Features hydrophilic coating on distal outer surface and radiopaque markers at distal tip for fluoroscopic visualization. Proximal end includes standard luer for syringe connection. Used by physicians in clinical settings (e.g., interventional radiology/cardiology) to navigate peripheral vasculature. Device facilitates targeted delivery of embolics (spheres, particles, coils, plugs, liquid embolics) to treat vascular conditions. Benefits include improved access to challenging anatomy and controlled material delivery.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-1), dimensional verification, functional/simulated use, static burst pressure, maximum infusion pressure, flow rate, kink radius, tensile/torque testing, coating lubricity/durability, fluid/air leak testing, embolic compatibility, and radiopacity. All testing met pre-determined acceptance criteria.
Technological Characteristics
Single-lumen metal alloy shaft with micro-cuts; 2.0Fr and 2.6Fr sizes; hydrophilic coating; radiopaque marker bands; standard luer hub. Sterilization method: not specified. Connectivity: none. Energy source: none (mechanical).
Indications for Use
Indicated for peripheral vascular use in patients requiring controlled and selective infusion of diagnostic, embolic, or therapeutic materials into the vessel.
Regulatory Classification
Identification
A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.
Predicate Devices
Direxion and Direxion Hi-Flo Torqueable Microcatheter (K163701)
Submission Summary (Full Text)
{0}------------------------------------------------
November 29, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Transit Scientific, LLC % Spencer Walker Director of Regulatory Affairs University of Utah, Center for Medical Innovation 10 North 1900 East, EHSL Rm. 22 Salt Lake City, Utah 84112
## Re: K231821
Trade/Device Name: XO Cath Microcatheter (E20-090-S, E20-110-S, E20-130-S, E20-175-S, E20-220-S, E20-090-B, E20-110-B, E20-130-B, E20-150-B, E20-175-B, E20-220-B, E26-090-S, E26-110-S, E26-130-S, E26-150-S, E26-175-S, E26-220-S, E26-090-B, E26-110-B, E26-130-B, E26-150-B, E26-175-B, E26-220-B,) Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: October 27, 2023 Received: October 30, 2023
Dear Spencer Walker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Samuel G. Raben -S
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices
{2}------------------------------------------------
OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231821
#### Device Name
XO Cath Microcatheter (E20-090-S, E20-130-S, E20-175-S, E20-175-S, E20-220-S, E20-090-B, E20-110-B, E20-130-B, E20-150-B, E20-175-B, E20-220-B, E26-110-S, E26-130-S, E26-175-S, E26-175-S, E26-220-S, E26-110-B, E26-130-B, E26-150-B, E26-175-B, E26-220-B,)
Indications for Use (Describe)
The XO Cath Microcatheter is intended for peripheral vascular use. The microcatheter can be used for controlled and selective infusion of diagnostic, embolic or therapeutic materials into the vessel.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> |
|----------------------------------------------|-------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | □ |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Transit Scientific. The logo is in black and white and features the words "TRANSIT" stacked on top of "SCIENTIFIC." The word "TRANSIT" is in a bold, sans-serif font, while the word "SCIENTIFIC" is in a thinner, sans-serif font. There is a trademark symbol to the right of the word "SCIENTIFIC."
## 510(K) SUMMARY
#### (21 CFR 807.92)
### GENERAL INFORMATION
| Submitter: | Transit Scientific, LLC |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Spencer Walker, MSc – Director of Regulatory Affairs<br>University of Utah/ Center for Medical Innovation<br>10 North 1900 East, Rm. 22B<br>Salt Lake City, UT 84112<br>(801) 708-2238<br>Spencer.walker@hsc.utah.edu |
| Date Prepared: | June 20, 2023 |
| Trade Name: | XO Cath Microcatheter |
| Classification Name: | Continuous Flush Catheter<br>21 CFR §870.1210, Product Code KRA |
| Device Class: | Class II |
| Predicate Device: | 510(k) No.: K163701<br>Model: Direxion and Direxion Hi-Flo Torqueable Microcathete<br>Manufacture: Boston Scientific Corporation<br>Classification: KRA |
#### Device Description:
The subject XO Cath Microcatheter is a single-lumen, metal alloy shaft with micro-cuts for flexibility, designed to support a guidewire during access of the peripheral vasculature, and to provide a conduit for the delivery of embolic applications. The XO Cath Microcatheter is available in two lumen sizes, 2.0Fr and 2.6Fr.
The XO Cath 2.0Fr and 2.6Fr Microcatheters are available with a straight or bern tip shape to aid with accessing challenging anatomy. The distal outer surface of the microcatheter is coated with a hydrophilic coating. Radiopaque markers are located at the distal tip to facilitate fluoroscopic visualization. The proximal end incorporates a standard luer for ease of use and to connect with a syringe.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Transit Scientific. The logo is in black and white and features the word "TRANSIT" in large, bold letters on the top line. Below that, the word "SCIENTIFIC" is written in smaller letters. The logo has a clean and modern design.
| Table 1: XO Cath Microcatheter Model Numbers | |
|----------------------------------------------|-------------------------------------------------------------------------|
| Model No. | Description |
| E20-090-S | XO Cath Microcatheter – 2.0 Fr diameter, and 90cm length, Straight Tip |
| E20-110-S | XO Cath Microcatheter - 2.0 Fr diameter, and 110cm length, Straight Tip |
| E20-130-S | XO Cath Microcatheter – 2.0 Fr diameter, and 130cm length, Straight Tip |
| E20-150-S | XO Cath Microcatheter – 2.0 Fr diameter, and 150cm length, Straight Tip |
| E20-175-S | XO Cath Microcatheter - 2.0 Fr diameter, and 175cm length, Straight Tip |
| E20-220-S | XO Cath Microcatheter – 2.0 Fr diameter, and 220cm length, Straight Tip |
| E20-090-B | XO Cath Microcatheter - 2.0 Fr diameter, and 90cm length, Bern Tip |
| E20-110-B | XO Cath Microcatheter – 2.0 Fr diameter, and 110cm length, Bern Tip |
| E20-130-B | XO Cath Microcatheter – 2.0 Fr diameter, and 130cm length, Bern Tip |
| E20-150-B | XO Cath Microcatheter – 2.0 Fr diameter, and 150cm length, Bern Tip |
| E20-175-B | XO Cath Microcatheter – 2.0 Fr diameter, and 175cm length, Bern Tip |
| E20-220-B | XO Cath Microcatheter – 2.0 Fr diameter, and 220cm length, Bern Tip |
| E26-090-S | XO Cath Microcatheter – 2.6 Fr diameter, and 90cm length, Straight Tip |
| E26-110-S | XO Cath Microcatheter – 2.6 Fr diameter, and 110cm length, Straight Tip |
| E26-130-S | XO Cath Microcatheter – 2.6 Fr diameter, and 130cm length, Straight Tip |
| E26-150-S | XO Cath Microcatheter – 2.6 Fr diameter, and 150cm length, Straight Tip |
| E26-175-S | XO Cath Microcatheter - 2.6 Fr diameter, and 175cm length, Straight Tip |
| E26-220-S | XO Cath Microcatheter – 2.6 Fr diameter, and 220cm length, Straight Tip |
| E26-090-B | XO Cath Microcatheter - 2.6 Fr diameter, and 90cm length, Bern Tip |
| E26-110-B | XO Cath Microcatheter - 2.6 Fr diameter, and 110cm length, Bern Tip |
| E26-130-B | XO Cath Microcatheter – 2.6 Fr diameter, and 130cm length, Bern Tip |
| E26-150-B | XO Cath Microcatheter – 2.6 Fr diameter, and 150cm length, Bern Tip |
| E26-175-B | XO Cath Microcatheter - 2.6 Fr diameter, and 175cm length, Bern Tip |
| E26-220-B | XO Cath Microcatheter - 2.6 Fr diameter, and 220cm length, Bern Tip |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Transit Scientific. The word "TRANSIT" is in all caps and bolded on the top line. The word "SCIENTIFIC" is in all caps on the second line, but it is not bolded.
### Indications for Use:
The XO Cath Microcatheter is intended for peripheral vascular use. The microcatheter can be used for controlled and selective infusion of diagnostic, embolic or therapeutic materials into the vessel.
#### Comparative Analysis:
It has been demonstrated that the XO Cath Microcatheter is comparable to the predicate device in intended use, fundamental scientific technology, design, principles of operation and functional performance evaluations. The XO Cath Microcatheter has been fully assessed within the Transit Scientific Risk Management and Design Controls systems. All necessary verification steps met pre-determined acceptance criteria to determine performance as compared to the predicate.
It has been demonstrated that the XO Cath Microcatheter is comparable to the predicate device in the following manner:
- Same intended use ●
- Same indications for use
- Same fundamental scientific technology ●
- Same or similar material properties ●
- Same or similar operating principle
- . Similar performance specifications
- Similar patient-user interface
| | Subject Device –<br>(XO Cath 2.0 Fr) | Predicate –<br>K163701 (Direxion<br>Microcatheter) | Subject Device –<br>(XO Cath 2.6 Fr) | Predicate –<br>K163701 (Direxion<br>Hi-Flo<br>Microcatheter) | |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Ind. for Use | The XO Cath<br>Microcatheter is intended<br>for peripheral vascular<br>use. The microcatheter<br>can be used for<br>controlled and selective<br>infusion of diagnostic,<br>embolic, or therapeutic<br>materials into the vessel | The Direxion<br>Microcatheters is<br>intended for peripheral<br>vascular use. The<br>microcatheter can be<br>used for controlled and<br>selective infusion of<br>diagnostic, embolic, or<br>therapeutic materials into<br>the vessel. | The XO Cath<br>Microcatheter is<br>intended for peripheral<br>vascular use. The<br>microcatheter can be<br>used for controlled and<br>selective infusion of<br>diagnostic, embolic, or<br>therapeutic materials<br>into the vessel. | The Direxion HI-FLO<br>Torqueable<br>Microcatheters is<br>intended for peripheral<br>vascular use. The<br>microcatheter can be<br>used for controlled and<br>selective infusion of<br>diagnostic, embolic, or<br>therapeutic materials<br>into the vessel. | |
| Classification<br>Name | Cardiovascular Device - Continuous Flush Catheter<br>21 CFR §870.1210<br>Product Code: KRA<br>Class II | | | | |
| Single Use | Yes | | | | |
| Prescription (Rx<br>Only) | Yes | | | | |
| Table 2: Substantial Equivalence Comparison Chart | | | | | |
| | Subject Device -<br>(XO Cath 2.0 Fr) | Predicate -<br>K163701 (Direxion<br>Microcatheter) | Subject Device -<br>(XO Cath 2.6 Fr) | Predicate -<br>K163701 (Direxion<br>Hi-Flo<br>Microcatheter) | |
| Anatomical<br>Access | Peripheral vasculature | | | | |
| Effective<br>Length (cm) | 90cm, 110cm,<br>130cm, 150cm,<br>175cm, and 220cm | 105cm, 130cm,<br>155cm | 90cm, 110cm,<br>130cm, 150cm,<br>175cm, and 220cm | 105cm, 130cm,<br>155cm | |
| Shaft Outer<br>Diameter | Distal: 2.0 Fr<br>Proximal: 2.6 Fr | Distal: 2.4 Fr<br>Proximal: 2.7 Fr | Distal: 2.6 Fr<br>Proximal: 2.75Fr | Distal: 2.8 Fr<br>Proximal: 3.0 Fr | |
| Inner Lumen | 0.021" | 0.021" | 0.027" | 0.027" | |
| Max Pressure<br>(PSI) | 1200 | | | | |
| Max Flow Rate<br>(mL/s) | 2.0Fr x 130cm:<br>3.0 mL/s | 2.4Fr x 130cm:<br>3.1 mL/s | 2.6Fr x 130cm:<br>5.8 mL/s | 2.8Fr x 130cm:<br>5.6 mL/s | |
| Dead Space<br>Volume (mL) | 2.0Fr x 130cm:<br>0.47 mL | 2.4Fr x 130cm:<br>0.46 mL | 2.6Fr x 130cm:<br>0.66 mL | 2.8Fr x 130cm:<br>0.64 mL | |
| Hydrophilic<br>Coating | Yes | | | | |
| Visibility/<br>Radiopacity | Radiopaque Marker Bands on catheter body | | | | |
| Guidewire<br>Compatibility | Max 0.018" | Max 0.018" | Max 0.021" | Max 0.021" | |
| Guide<br>Catheter<br>Compatibility | Min. 0.038 in | Min. 0.035 in | Min. 0.038 in | Min. 0.038 in | |
| Embolic<br>Compatibility | ≤ 700 Micron spheres<br>≤ 500 Micron particles<br>Coils & Plugs (max<br>0.021" recommended<br>microcatheter<br>compatible)<br>Liquid Embolics<br>(DMSO.<br>Lipiodol/Ethiodol,<br>Cyanoacrylate<br>adhesives) | ≤ 700 Micron spheres<br>≤ 500 Micron particles<br>Coils (max 0.018"<br>recommended<br>microcatheter<br>compatible)<br>Liquid Embolics<br>(DMSO.<br>Lipiodol/Ethiodol,<br>Cyanoacrylate<br>adhesives, Ethanol) | ≤ 900 Micron spheres<br>≤ 700 Micron particles<br>Coils & Plugs (max<br>0.027" recommended<br>microcatheter<br>compatible)<br>Liquid Embolics<br>(DMSO,<br>Lipiodol/Ethiodol,<br>Cyanoacrylate<br>adhesives) | ≤ 900 Micron spheres<br>≤ 710 Micron particles<br>Liquid Embolics<br>(DMSO,<br>Lipiodol/Ethiodol,<br>Cyanoacrylate<br>adhesives, Ethanol) | |
| Tip Shape | Straight, Bern | Straight, Bern, J,<br>Swan | Straight, Bern | Straight, Bern, J,<br>Swan | |
| Shapable | No | Some SKUs (via<br>steam) | No | Some SKUs (via<br>steam) | |
{7}------------------------------------------------
# Functional Testing:
The following testing was conducted to validate and verify that the subject device was substantially equivalent to the predicate devices. All data met pre-determined acceptance criteria.
{8}------------------------------------------------
TRANSIT
SCIENTIFIC™
- Biocompatibility Biocompatibility of the complete and finished XO Cath Microcatheter ● has been verified according to the requirements and testing prescribed in ISO 10993-1 and in accordance with FDA guidance document "Use of International Standard ISO 10993-1" for an external communicating device with limited exposure (<24hrs) to circulating blood.
- Cytotoxicity O
- Sensitization O
- Irritation/ Intracutaneous Toxicity O
- Acute Systemic Toxicity O
- Material Mediated Pyrogenicity O
- Hemolysis Assay O
- Complement Activation Assay O
- Partial Thromboplastin Time (PTT) O
- Blood Platelet and Leukocyte Count Testing O
- LAL Pyrogenicity O
- Design Verification Performance bench testing was conducted to ensure that the XO ● Cath Microcatheter met the applicable design and performance requirements throughout its shelf life, verify conformity to applicable standards, and demonstrate substantial equivalence to the predicate system. The following performance testing was performed or fulfilled with the XO Cath Microcatheter.
- o Material Verification
- Dimensional Verification o
- Visual Verification O
- Functional/ Simulated Use Testing O
- Static Burst Pressure o
- Maximum Infusion Pressure O
- O Maximum Flow Rate
- Kink Radius O
- Tensile Testing O
- Torque Testing o
- Coating Lubricity O
- Coating Durability O
- Fluid Leak Testing O
- Hub Assembly Air Leak O
- Physical Embolic Testing O
- Liquid Embolic Exposure Testing O
- Radiopacity o
- Packaging O
- Sterilization O
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Transit Scientific. The word "TRANSIT" is in large, bold, sans-serif font on the top line. Below that, in a smaller, thinner font, is the word "SCIENTIFIC" with a trademark symbol to the right. The logo is simple and modern.
## Conclusion:
Based on the similarities in design between the subject and predicate devices, and the performance testing performed, the subject XO Cath Microcatheter is substantially equivalent to the cited predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.