K231801 · Medhealth Medical Limited · JOW · Dec 14, 2023 · Cardiovascular
Device Facts
Record ID
K231801
Device Name
MedHealth DVT compression Devices MHH900S/MHH900E
Applicant
Medhealth Medical Limited
Product Code
JOW · Cardiovascular
Decision Date
Dec 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MedHealth DVT Compression Devices MHH900S and MHH900E are intended to increase venous blood flow in at risk patients in order to help prevent deep vein thrombosis.
Device Story
External pneumatic compression (EPC) device; consists of air pump and soft pliable compression garment (foot, calf, or thigh). Controller delivers preset inflation/deflation cycles (12s inflation, 48s deflation) at 40 mmHg (leg) or 120 mmHg (foot). Air pressure transfers to extremity to augment venous blood flow, reduce stasis, and stimulate fibrinolysis. Used in clinical settings; operated by healthcare personnel. Reusable device. Benefits patient by reducing risk of DVT and pulmonary embolism.
Class II EPC device. Powered by AC 100-240V or internal Li-ion battery (14.4V). Intermittent pneumatic compression principle. Applied parts: garments and air tubes. Dimensions: 8.6" x 6.2" x 8.98"; Weight: 2.48 kg. Class I Type BF. Software-controlled timing cycles.
Indications for Use
Indicated for at-risk patients requiring prevention of deep vein thrombosis (DVT) following surgeries such as orthopedics. Contraindications include severe congestive cardiac failure, severe arteriosclerosis or ischemic vascular disease, extreme limb deformity, known or suspected DVT, and local conditions where garments interfere (gangrene, dermatitis, untreated/infected wounds, recent skin grafts).
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2023
MedHealth Medical Limited Steven Jiang General Manager 701/101, BLDG 24, Block B, Yuanshan Zone Shangcun, Gongming Shenzhen, Guangdong 518106 China
### Re: K231801
Trade/Device Name: MedHealth DVT compression Devices MHH900S/MHH900E Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: June 1, 2023 Received: June 20, 2023
### Dear Steven Jiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Eric E. Richardson -STDV
for Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K231801
Device Name
MedHealth DVT compression Devices MHH900S/MHH900E
Indications for Use (Describe)
MedHealth DVT Compression Devices MHH900S/MHH900E are intended to increase venous blood flow in at risk patients in order to help prevent deep vein thrombosis.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Submission
MHH900S/MHH900E DVT Compression Devices
# 510(k) Summary [As required by 21 CFR 807.92]
- 1. Date Prepared [21 CFR807.92 (a) (1)] Nov 21,2023
- 2. Submitter's Information [21 CFR807.92 (a) (1)] Name of Sponsor: MedHealth Medical Limited
Address: 701/101, BLDG 24, Block B, Yuanshan Zone, Shangcun, Gongming, Shenzhen, 518106, China Contact Name: Steven Jiang Telephone No.: +86-13502865807 Fax No.: +86-755-88210289 Email Address: steven.jiang@medhealth.com.cn
### 3. Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)]
Trade Name/Model: MedHealth DVT compression Devices MHH900S/MHH900E Common Name, MHH900S/MHH900E DVT Pumps Classification Name: Sleeve, Limb, Compressible Regulation Number: 21 CFR 870.5800 Product code: JOW Classification Panel: Cardiovascular Device Class: Class II
### 4. Identification of predicated Device (S) [21 CFR807.92 (a) (3)]
The predicate device is B&J Manufacturing Ltd. Deep Vein Thrombosis (DVT) compression device MHH800 (K200154)
### 5. Description of the Device [21 CFR807.92 (a) (4)]
The Deep Vein Thrombosis (DVT) Pumps MHH900S and MHH900E are external pneumatic compression (EPC) devices that aid in the prevention of DVT from a potentially life-threatening condition which can lead to pulmonary embolism. MHH900S and MHH900E are non-invasive mechanical prophylactic devices function as secondary
{4}------------------------------------------------
#### 510(k) Submission
pumps to propel venous blood for patients whose deep vein thrombosis must be prevented after surgeries in Orthopedics etc. Additionally, the devices are reusable and can be used for more than one patient within its lifecycle.
The two devices separately consist of an air pump and a soft pliable compression garment(s) (Purchased separately) for the foot, calf, or thigh. For MHH900S and MHH900E, the controller supplies compression on a preset timing cycle (12 seconds inflation and maintenance followed by 48 seconds of deflation) at a suggested pressure setting of 40 mmHg for the Leg and 120mmHg for the Foot. The pressure in the qarments is transferred to the extremity, augmenting venous blood flow when the leg is compressed, reducing stasis. This process also stimulates fibrinolysis; thus, reducing the risk of early clot formation.
The DVT pumps produce automatically timed cycles of compressed air. This compressed air forces blood out of the deep veins, helping to prevent slowed or stopped blood flow. Bursts of air are delivered to the specially designed garments wrapped around extremities. This air helps to move blood out of the deep veins and reduce the risk of developing DVT.
#### 6. Intended Use [21 CFR807.92 (a) (5)]
MedHealth DVT Compression Devices MHH900S and MHH900E are intended to increase venous blood flow in at risk patients in order to help prevent deep vein thrombosis.
{5}------------------------------------------------
510(k) Submission
# 7. Technological Characteristics [21 CFR807.92 (a) (6)]
# MHH900S DVT Pump
| General Specification: | | |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Model No. | MHH900S | |
| Size | 8.6" (L) x 6.2" (W) x 8.98" (H) | |
| Weight | 2.48 kg | |
| Pressure Range | Calf/Thigh Garment: 40 mmHg +10/-5mmHg<br>Foot Garment: 120mmHg +10/-5mmHg | |
| Input Rating | AC 100V-240V 50/60Hz 1A | |
| Fuse Rating | 1A or T1AH 250V | |
| Classification | Class I Type BF<br>Not AP or AGP type | |
| Humidity | Operation: 30% to 80 %<br>Storage & Transportation: 30% to 93% | |
| Air pressure | 75 - 106KPA | |
| Temperature | Operation: 5° C- 40° C<br>Storage & Transportation: 5° C - 55° C | |
| Cycle Time | Inflation 12 seconds +/- 1 second<br>Deflation 48 seconds +/- 2 seconds | |
| Applied Part | Garments (Purchased separately) and Air Tube | |
| Battery | Battery pack: 4 x series Li-ion battery cell<br>Battery pack capacity: 2900mAh (Nominal),<br>2700mAh (minimum)<br>Nominal voltage: 14.4V<br>Temperature: Operation 5°C ~ 40°C Storage 5°C ~ 55°C | |
| Garment for the DVT Pump (Purchased separately) : | | |
| Model Name | Applied Part | Specification |
| 801P | Calf | Small, for calf circumference up to 14" (35 cm) |
| 801M | Calf | Medium, for calf circumference up to 18" (45 cm) |
| 801L | Calf | Large, for calf circumference up to 24" (61 cm) |
| 801B | Calf | Extra Large, for calf circumference up to 32" (81 cm) |
| 820M | Foot | Medium, for foot sizes up to 13" |
| 820L | Foot | Large, for foot sizes up to 16" |
| 830S | Full-Leg | Small, for thigh circumference up to 22" (55 cm) |
| 830M | Full-Leg | Medium, for thigh circumference up to 29" (73 cm) |
| 830L | Full-Leg | Large, for thigh circumference up to 36" (91 cm) |
| 830B | Full-Leg | Extra Large, for thigh circumference up to 42" (106 cm) |
| Air Hose Extension for the DVT Pump: | | |
| 810 | | Air hose extension of 59" (Pair) |
| 810L | | Air hose extension of 118" (Pair) |
{6}------------------------------------------------
510(k) Submission
| General Specification: | |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model No. | MHH900E |
| Size | 8.6" (L) x 6.2" (W) x 8.98" (H) |
| Weight | 2.48 kg |
| Pressure Range | Calf/Thigh Garment: 40mmHg +10/-5mmHg<br>Foot Garment: 120mmHg +10/-5mmHg |
| Input Rating | AC 100V-240V 50/60Hz 1A |
| Fuse Rating | 1A or T1AH 250V |
| Classification | Class I Type BF<br>Not AP or AGP type<br>Image: [Person Symbol] |
| Humidity | Operation: 30% to 80 %<br>Storage & Transportation: 30% to 93% |
| Air pressure | 75 - 106KPA |
| Temperature | Operation: 5° C- 40° C<br>Storage & Transportation: 5° C - 55° C |
| Cycle Time | Inflation 12 seconds +/- 1 second<br>Deflation 48 seconds +/- 2 seconds |
| Applied Part | Garments (Purchased separately) and Air Tub |
| Battery | Battery pack: 4 x series Li-ion battery cell<br>Battery pack capacity: 2900mAh (Nominal),<br>2700mAh (minimum)<br>Nominal voltage: 14.4V<br>Temperature: Operation5°C ~ 40°C Storage 5°C ~ 55°C |
| Garment for the DVT Pump (Purchased separately) : | |
| Model Name | Applied Part Specification |
| 801P | Calf Small, for calf circumference up to 14" (35 cm) |
| 801M | Calf Medium, for calf circumference up to 18" (45 cm) |
| 801L | Calf Large, for calf circumference up to 24" (61 cm) |
| 801B | Calf Extra Large, for calf circumference up to 32" (81 cm) |
| 820M | Foot Medium, for foot sizes up to 13" |
| 820L | Foot Large, for foot sizes up to 16" |
| 830S | Full-Leg Small, for thigh circumference up to 22" (55 cm) |
| 830M | Full-Leg Medium, for thigh circumference up to 29" (73 cm) |
| 830L | Full-Leg Large, for thigh circumference up to 36" (91 cm) |
| 830B | Full-Leg Extra Large, for thigh circumference up to 42" (106 cm) |
| Air Hose Extension for the DVT Pump: | |
| 810 | Air hose extension of 59" (Pair) |
| 810L | Air hose extension of 118" (Pair) |
# MHH900E DVT Pump
{7}------------------------------------------------
510(k) Submission
MHH900S/MHH900E DVT Compression Devices
### 8. Substantial Equivalence [21 CFR807.92 (a) (3) and 807.92]
### 8.1 Intended uses:
### Table 1 Intended Use Comparison
| ID | Comparison<br>Item | Proposed Device<br>MedHealth<br>DVT<br>Compression Devices<br>(MHH900S/MHH900E) | Predicate Device<br>B&J DVT Compression<br>Device<br>(MHH800) |
|----|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | 510(K) No. | To be assigned | K200154 |
| 2 | Intended<br>Use | The MedHealth Medical<br>Limited. MHH900S and<br>MHH900E Deep Vein<br>Thrombosis (DVT)<br>Compression Devices are<br>intended to increase venous<br>blood flow in at risk patients<br>in order to help prevent deep<br>vein thrombosis. | The B&J Manufacturing Ltd.<br>MHH800 Deep Vein<br>Thrombosis (DVT)<br>Compression Devices are<br>intended to increase venous<br>blood flow in at risk patients<br>in order to help prevent deep<br>vein thrombosis. |
| 3 | Type of use | Prescription Use | Prescription Use |
| 4 | Single use | Reusable | Reusable |
It is clear that the intended use of MHH900S and MHH900E is same as the predictive device
MHH800
### 8.2 Comparison table
## Table 2 General Comparison between MHH900S and MHH800
| ID | Comparison<br>Item | Proposed Device<br>MedHealth DVT<br>Compression<br>Device (MHH900S) | Predicate Device<br>B&J DVT<br>Compression<br>Device (MHH800) | Explanation<br>of Difference |
|-----|-----------------------|---------------------------------------------------------------------|---------------------------------------------------------------|-----------------------------------------------------------------------------------|
| 5 | General Specification | | | |
| 5.1 | Size | 8.6"x6.2"x8.98" | 6.5"x7.0"x12.8" | Different but<br>does not raise<br>any new issue<br>of substantial<br>equivalence |
{8}------------------------------------------------
| MHH900S/MHH900E DVT Compression Devices | | | | | 510(k) Submission |
|-----------------------------------------|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| 5.2 | Weight | 2.48kg | 2. 1kg | | Different but<br>does not raise<br>any new issue<br>of substantial<br>equivalence |
| 5.3 | Pressure Range | Calf/Thigh:40mmHg;<br>Foot:120mmHg | Calf/Thigh:40mmHg;<br>Foot:120mmHg | | Same |
| 5.4 | Input Rating | AC 100-240V,<br>50/60Hz | AC 100-240V,<br>50/60Hz | | Same |
| 5.5 | Fuse Rating | 1A or T1AH 250V | 1A or T1AH 250V | | Same |
| 5.6 | Classification | Class I Type BF<br>Not AP or AGP<br>type | Class I Type BF<br>Not AP or AGP<br>type<br>Image: Man in doorway | | Same |
| 5.7 | Operation<br>Humidity | 30-80% | 30-75% | Different but<br>does not raise<br>any new issue<br>of substantial<br>equivalence | |
| 5.8 | Storage &<br>Transportation<br>Humidity | 30-93% | 30-75% | Different but<br>does not raise<br>any new issue<br>of substantial<br>equivalence | |
| 5.9 | Operation<br>Temperature | 5 ° C-40 ° C | 15 ° C-35 ° C | Different but<br>does not raise<br>any new issue<br>of substantial<br>equivalence | |
| 5.10 | Storage &<br>Transportation<br>Temperature | 5 ° C-55 ° C | 5 ° C-60 ° C | Different but<br>does not raise<br>any new issue<br>of substantial<br>equivalence | |
| 5.11 | Applied Part | Garment(Purchased separately) and Air Hose | Garment(Purchased separately) and Air Hose | Same | |
| 5.12 | Applied Mode<br>of Pressure | Intermittent | Intermittent | Same | |
| 5.13 | Number of<br>Chambers in<br>Garment | No | No | Same | |
| 5.14 | Inflation<br>chamber<br>(Calf/Thigh) | 12 seconds | 12 seconds | Same | |
| 5.15 | Deflation time<br>per chamber<br>(Calf/Thigh) | 48 seconds | 48 seconds | Same | |
| 5.16 | Inflation time per<br>chamber (Foot) | 12 seconds | 12 seconds | Same | |
| 5.17 | Deflation time<br>per chamber<br>(Foot) | 48 seconds | 48seconds | Same | |
| 5.18 | Pressure<br>Range<br>Calf/ Thigh | 40mmHg | 40mmHg | Same | |
| 5.19 | Pressure Range<br>Foot | 120mmHg | 120mmHg | Same | |
| 5.20 | Pre- Programme<br>d Controls | Yes | Yes | Same | |
| 5.21 | Battery Pack | Yes | Yes | Same | |
| 5.22 | Garments | Purchased separately,<br>listed<br>in<br>510(K)<br>K201532 & K201547 | Purchased separately,<br>listed<br>in<br>510(K)<br>K201532 & K201547 | Same | |
| 5.23 | Contraindications<br>for Use | • Severe<br>congestive cardiac<br>failure.<br>• Severe<br>arteriosclerosis or<br>other ischemic<br>vascular disease.<br>• Extreme<br>deformity of the<br>limbs.<br>• Known or<br>suspected deep<br>vein thrombosis.<br>• Any local<br>condition in which<br>the garments would<br>interfere:<br>1) Gangrene<br>2) Dermatitis<br>3)Untreated or<br>infected wounds<br>4)Recent skin<br>grafts | - Severe congestive<br>cardiac failure.<br>- Severe<br>arteriosclerosis or<br>other ischemic<br>vascular disease.<br>- Extreme deformity<br>of the limbs.<br>- Known or<br>suspected deep vein<br>thrombosis.<br>- Any local condition<br>in which the<br>garments would<br>interfere:<br>- Gangrene<br>- Dermatitis<br>- Untreated or<br>infected wounds<br>- Recent skin grafts | Same | |
{9}------------------------------------------------
510(k) Submission
{10}------------------------------------------------
510(k) Submission
{11}------------------------------------------------
510(k) Submission
Table3 General Comparison between MHH900E and MHH800
| ID | Compariso<br>n Item | Proposed Device<br>MedHealth<br>DVT<br>Compression<br>Device<br>(MHH900E) | Predicate Device<br>B&J DVT<br>Compression<br>Device<br>(MHH800) | Explanation of<br>Difference |
|------|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| 5 | General Specification | | | |
| 5.1 | Size | 8.6"x6.2"x8.98" | 6.5"x7.0"x12.8" | Different but<br>does not raise any<br>new issue of<br>substantial<br>equivalence |
| 5.2 | Weight | 2.48kg | 2.1kg | Different but<br>does not raise any<br>new issue of<br>substantial<br>equivalence |
| 5.3 | Pressure Range | Calf/Thigh:40mmHg;<br>Foot:120mmHg | Calf/Thigh:40mmHg;<br>Foot:120mmHg | Same |
| 5.4 | Input Rating | AC 100-240V,<br>50/60Hz | AC 100-240V,<br>50/60Hz | Same |
| 5.5 | Fuse Rating | 1A or T1AH 250V | 1A or T1AH 250V | Same |
| 5.6 | Classification | Class I Type BF<br>Not AP or AGP<br>type<br>Image: Man in wheelchair | Class I Type BF<br>Not AP or AGP<br>type<br>Image: Man in wheelchair | Same |
| 5.7 | Operation<br>Humidity | 30-80% | 30-75% | Different but does<br>not raise any new<br>issue of<br>substantial<br>equivalence |
| 5.8 | Storage &<br>Transportation<br>Humidity | 30-93% | 30-75% | Different but does<br>not raise any new<br>issue of<br>substantial<br>equivalence |
| 5.9 | Operation<br>Temperature | 5 °C-40 °C | 15 °C-35 °C | Different but does<br>not raise any new<br>issue of<br>substantial<br>equivalence |
| 5.10 | Storage &<br>Transportation<br>Temperature | 5 °C-55 °C | 5 °C-60 °C | Different but<br>does not raise<br>any new issue of<br>substantial<br>equivalence |
| 5.11 | Applied Part | Garment and Air<br>Hose | Garment and Air<br>Hose | Same |
| 5.12 | Applied<br>Mode<br>of Pressure | Intermittent | Intermittent | Same |
| 5.13 | Number of<br>Chambers in<br>Garment | No | No | Same |
| 5.14 | Inflation<br>time per<br>chamber<br>(Calf/Thigh) | 12 seconds | 12 seconds | Same |
| 5.15 | Deflation time<br>per chamber<br>(Calf/Thigh) | 48 seconds | 48 seconds | Same |
| 5.16 | Inflation time per<br>chamber (Foot) | 12 seconds | 12 seconds | Same |
| 5.17 | Deflation<br>time<br>per chamber<br>(Foot) | 48 seconds | 48seconds | Same |
| 5.18 | Pressure<br>Range<br>Calf/Thig<br>h | 40mmHg | 40mmHg | Same |
| 5.19 | Pressure<br>Range<br>Foot | 120mmHg | 120mmHg | Same |
| 5.20 | Pre- Progr<br>ammed<br>Controls | Yes | Yes | Same |
| 5.21 | Battery Pack | Yes | Yes | Same |
| 5.22 | Garments | Purchased separately,<br>listed in 510(K) K201532<br>& K201547 | Purchased separately,<br>listed in 510(K) K201532<br>& K201547 | Same |
| 5.23 | Contraindications for Use | • Severe congestive<br>cardiac failure.<br>• Severe<br>arteriosclerosis or<br>other ischemic<br>vascular disease.<br>• Extreme deformity<br>of the limbs.<br>• Known or suspected<br>deep vein thrombosis.<br>• Any local condition<br>in which the garments<br>would interfere:<br>• Gangrene<br>• Dermatitis<br>• Untreated or<br>infected wounds<br>• Recent skin grafts | - Severe congestive<br>cardiac failure.<br>- Severe<br>arteriosclerosis or other<br>ischemic vascular<br>disease.<br>- Extreme deformity of<br>the limbs.<br>- Known or suspected<br>deep vein thrombosis.<br>- Any local condition in<br>which the garments<br>would interfere:<br>- Gangrene<br>- Dermatitis<br>- Untreated or infected<br>wounds<br>- Recent skin grafts | Same |
{12}------------------------------------------------
{13}------------------------------------------------
It is cleared that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the MHH900S and MHH900E
{14}------------------------------------------------
### 8.3 Non-clinical testing [21 CFR 807.92(b) (1)]
The following safety and performance test were conducted to assess MHH900S and MHH900E DVT compression Devices.
- A Software validation
- Cleaning and Disinfection validation—Testing was performed to, and passed, the A following standards:
- AAMI TIR12: 2020 Designing, Testing And Labeling Reusable Medical Devices For Reprocessing In Health Care Facilities: A Guide For Medical Device Manufacturers
● ASTM F3208-20 Standard Guide for Selecting Test Soils for Validation of Cleaning Methods for Reusable Medical Devices
● AAMI…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.