K231786 · Cosm Medical · HHW · Mar 6, 2024 · Obstetrics/Gynecology
Device Facts
Record ID
K231786
Device Name
Gynethotics™ Pessary
Applicant
Cosm Medical
Product Code
HHW · Obstetrics/Gynecology
Decision Date
Mar 6, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.3575
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Gynethotics pessaries are removable devices placed in the vaginal cavity to provide a nonsurgical alternative for the treatment of pelvic organ prolapse and to relieve the symptoms of pelvic relaxation with or without urinary incontinence in adults.
Device Story
Gynethotics pessaries are silicone vaginal devices used to provide structural support to the vagina, uterus, bladder, and rectum. They serve as a nonsurgical treatment for pelvic organ prolapse and pelvic floor dysfunction symptoms, including stress urinary incontinence. The device is inserted into the vaginal cavity to provide mechanical support. Clinicians select from various shapes (Ring, Gellhorn, Marland, Cube) and personalized configurations to match patient anatomy. The device functions by physically supporting prolapsed organs, thereby relieving symptoms. It is intended for use in a clinical setting or by the patient as directed by a physician. Benefits include a nonsurgical management option for pelvic floor dysfunction.
Clinical Evidence
Bench testing only. Testing included mechanical performance (folding, 3-point bend, cord detachment, end of life, squeeze, knob displacement), biocompatibility (ISO 10993 series), TSST-1 risk assessment, cleaning validation (AAMI ST98:2022, AAMI TIR12), and packaging validation (ASTM D4169-22).
Technological Characteristics
Silicone and colorant construction. Shapes: Ring, Gellhorn, Marland, Cube. Mechanical support principle. Biocompatibility per ISO 10993-1:2018. Cleaning validation per AAMI ST98:2022 and AAMI TIR12. Packaging per ASTM D4169-22.
Indications for Use
Indicated for adult patients with pelvic organ prolapse (Stage I-III) or stress urinary incontinence, depending on specific pessary model (Ring, Ring with Support, Ring with Knob, Ring with Support and Knob, Incontinence Ring, Marland, Marland with Support, Gellhorn, or Cube).
Regulatory Classification
Identification
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse (falling down of uterus), uterine retroposition (backward displacement), or gynecologic hernia.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
March 6, 2024
Cosm Medical Linda Wu Head of Clinical, Regulatory and Quality 101 College Street Toronto, ON M5G1L7 CANADA
Re: K231786 Trade/Device Name: Gynethotics™ Pessary Regulation Number: 21 CFR 884.3575 Regulation Name: Vaginal Pessary Regulatory Class: II Product Code: HHW Dated: February 3, 2024 Received: February 5, 2024
Dear Linda Wu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231786
Device Name Gynethotics™ Pessary
#### Indications for Use (Describe)
Gynethotics Ring Pessary is indicated for use in cases of mild (Stage I) pelvic organ prolapse.
Gynethotics Ring with Support Pessary is indicated for use in cases of mild (Stage I) pelvic organ prolapse.
Gynethotics Ring with Knob Pessary is indicated for use in cases of mild (Stage I) pelvic organ prolapse, and stress urinary incontinence.
Gynethotics Ring with Support and Knob is indicated for use in cases of mild (Stage I) pelvic organ prolapse, and stress urinary incontinence.
Gynethotics Incontinence Ring Pessary is indicated for use in cases of stress urinary incontinence.
Gynethotics Marland Pessary is indicated for use in cases of mild (Stage I) and moderate (Stage II) pelvic organ prolapse, and stress urinary incontinence.
Gynethotics Marland with Support Pessary is indicated for use in cases of mild (Stage I) and moderate (Stage II) pelvic organ prolapse, and stress urinary incontinence.
Gynethotics Gellhorn Pessary is indicated for use in cases of mild (Stage I) to severe (Stage III or procidentia) pelvic organ prolapse.
Gynethotics Cube Pessary is indicated for use in cases of moderate (Stage II) to severe (Stage III or procidentia) pelvic organ prolapse.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
## 16.2.1. Submitter Information
| Company Name | Cosm Medical Corp. |
|-----------------|--------------------------------------------------|
| Company Address | 101 College Street, Toronto, ON M5G 1L7 |
| Company Phone | 1-888-589-2676 |
| Contact Person | Linda Wu (Head of Clinical, Regulatory, Quality) |
| Date Prepared | March 5, 2024 |
## 16.2.2. Device Information
| Device Trade Name | Gynethotics™ Pessary |
|-----------------------|---------------------------|
| Device Common Name | Vaginal Pessary |
| Product Code Name | Pessary, Vaginal |
| Product Code | HHW |
| Classification Panel | Obstetrical/Gynecological |
| Regulatory Class | 2 |
| Classification Number | 21 CFR 884.3575 |
| Classification Name | Vaginal pessary |
### 16.2.3. Predicate Devices
| Name | Manufacturer | 510(k) |
|---------------------------------------|----------------------------|---------|
| Bioteque Vaginal Pessary<br>(primary) | Bioteque America, Inc. | K013289 |
| Panpac vaginal pessary | Panpac Medical Corporation | K092981 |
| Flexi-Shelf | Panpac Medical Corporation | K173351 |
| PelvX Incontinence Ring | DesChutes Medical Products | K974116 |
The predicate devices have not been subject to a design-related recall.
## 16.2.4. Intended Use
Gynethotics pessaries are removable devices placed in the vaginal cavity to provide a nonsurgical alternative for the treatment of pelvic organ prolapse and to relieve the symptoms of pelvic relaxation with or without urinary incontinence in adults.
#### 16.2.5. Indications for use
Gynethotics Ring Pessary is indicated for use in cases of mild (Stage I) pelvic organ prolapse.
Gynethotics Ring with Support Pessary is indicated for use in cases of mild (Stage I) pelvic organ prolapse.
Gynethotics Ring with Knob Pessary is indicated for use in cases of mild (Stage I) pelvic organ prolapse, and stress urinary incontinence.
{5}------------------------------------------------
K231786 Page 2 of 4
Gvnethotics Ring with Support and Knob is indicated for use in cases of mild (Stage I) pelvic organ prolapse, and stress urinary incontinence.
Gynethotics Incontinence Ring Pessary is indicated for use in cases of stress urinary incontinence.
Gynethotics Marland Pessary is indicated for use in cases of mild (Stage I) and moderate (Stage II) pelvic organ prolapse, and stress urinary incontinence.
Gynethotics Marland with Support Pessary is indicated for use in cases of mild (Stage I) and moderate (Stage II) pelvic orqan prolapse, and stress urinary incontinence.
Gynethotics Gellhorn Pessary is indicated for use in cases of mild (Stage I) to severe (Stage III or procidentia) pelvic organ prolapse.
Gynethotics Cube Pessary is indicated for use in cases of moderate (Stage II) to severe (Stage III or procidentia) pelvic organ prolapse.
# 16.2.6. Device Description
Gynethotics pessaries are vaginal pessaries made from silicone that are inserted into the vagina to function as a supportive structure of the vagina, uterus, bladder and/or rectum to manage and treat the symptoms of pelvic floor dysfunction, such as pelvic organ prolapse and incontinence. Gynethotics pessaries come in four general shapes, including: Ring, Gellhorn, Marland, and Cube. Each shape comes in a range of configurable features and dimensions.
## 16.2.7. Comparison to Predicate Devices
Gynethotics pessaries are vaqinal pessaries used to provide support to prolapse organs and treat symptoms of pelvic floor dysfunction. Similar to predicates. Gynethotics pessaries are available in multiple shapes and sizes. The kev difference between the subject and predicate devices is that Gynethotics pessaries allow for personalized configurations of certain shape features, as determined by the clinician. The differences do not raise different questions of safety or effectiveness, and can be evaluated through performance testing.
| Shape Description | Indications for Use<br>(subject device) | Indications for Use<br>(predicate devices) |
|-------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Ring | Indicated for use in cases of mild<br>(Stage I) pelvic organ prolapse. | K013289: Indicated for use<br>in cases of first or mild<br>second-degree prolapse. |
| Ring with Support | Indicated for use in cases of mild<br>(Stage I) pelvic organ prolapse. | K013289: Indicated for use<br>in cases of first or mild<br>second-degree prolapse. |
| Ring with Knob | Indicated for use in cases of mild<br>(Stage I) pelvic organ prolapse, and<br>stress urinary incontinence. | K092981: Indicated for first<br>or second-degree prolapse<br>and stress urinary<br>incontinence. |
{6}------------------------------------------------
| Page 3 of 4 | | |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Shape Description | Indications for Use<br>(subject device) | Indications for Use<br>(predicate devices) |
| Ring with Support<br>and Knob | Indicated for use in cases of mild<br>(Stage I) pelvic organ prolapse, and<br>stress urinary incontinence. | K092981: Indicated for first<br>or second-degree prolapse<br>and stress urinary<br>incontinence. |
| Incontinence Ring | Indicated for use in cases of stress<br>urinary incontinence. | K974116: for treatment of<br>urinary incontinence. |
| Marland | Indicated for use in cases of mild<br>(Stage I) and moderate (Stage II)<br>pelvic organ prolapse, and stress<br>urinary incontinence. | K013289: Indicated for use<br>for the control of stress<br>urinary incontinence and first<br>or mild second-degree<br>prolapse. |
| Marland with<br>Support | Indicated for use in cases of mild<br>(Stage I) and moderate (Stage II)<br>pelvic organ prolapse, and stress<br>urinary incontinence. | K013289: Indicated for use<br>for the control of stress<br>urinary incontinence and first<br>or mild second-degree<br>prolapse. |
| Gellhorn | Indicated for use in cases of mild<br>(Stage I) to severe (Stage III or<br>procidentia) pelvic organ prolapse. | K173351: Indicated for the<br>support of a second to third<br>degree prolapse or<br>procidentia. |
| Cube | Indicated for use in cases of<br>moderate (Stage II) to severe<br>(Stage III or procidentia) pelvic<br>organ prolapse. | K013289: Indicated for use<br>in the support of a third<br>degree prolapse, including<br>procidentia, as well as<br>cystocele and rectocele. |
Any differences in wording of the Indications for use statements between subject and predicate devices, do not impact the intended use of the device. Indications for use language of the subject device is reflective of currently recognized medical terminology and reflects the same understanding of Indications for use in predicate devices.
| Technological characteristic | Differences |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimensions and sizing increments | The min and max feature sizes are within range of predicate pessaries. Subject device sizes increment with greater degrees of freedom within the cleared ranges, thereby offering greater sizing options. |
| Materials | Subject device is made of silicone and colorant. Predicate devices are also made of silicone and colorant, but formulation differs. |
# 16.2.8. Performance Data
The following testing was conducted to evaluate that the device meets its specifications:
- Mechanical testing was performed to confirm that the device meets key performance ● specifications related to the functionality and safety of the device. Testing included: folding, 3-point bend, cord detachment, end of life, squeeze, knob displacement.
{7}------------------------------------------------
- Biocompatibility according to ISO 10993-1:2018 and applicable FDA quidance ● documents. Testing included:
| Test name | Results | Standards |
|----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Cytotoxicity | Non-cytotoxic | ISO 10993-5:2009 |
| Sensitization | Non-sensitizer | ISO 10993-10:2021 |
| Irritation | Non-irritant | ISO 10993-23:2021 |
| Acute systemic toxicity | No mortality or evidence of systemic toxicity | ISO 10993-11:2017 |
| Implantation | Compared to negative control:<br>Macroscopic - difference not significant; Microscopic – minimal or no reaction | ISO 10993-6:2016 |
| Pyrogenicity | Non-pyrogenic | ISO 10993-11:2017 |
| Chemical characterization and<br>Materials and<br>Toxicological Risk assessments | Acceptable risk of daily dose-exposure to compounds from device use | ISO 10993-17:2002<br>ISO 10993-17:2023<br>ISO 10993-18:2020 |
- TSST-1 Risk laboratory testing to evaluate the impact of the device on Staphylococcus ● aureus growth and TSST-1 production, following FDA guidance for Menstrual Tampons and Pads: Information for Premarket Notification Submissions, which demonstrated a low risk profile for the device.
- Cleaning validation conducted as per AAMI ST98:2022, AAMI TIR12 and FDA guidance on Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labelling, demonstrated that the cleaning removes soil and contamination effectively.
- Packaging validation conducted as per ASTM D4169-22 confirmed that the packaging ● materials are capable of withstanding the distribution environment during the rigors of shipping and handling.
# 16.2.9. Conclusions
Subject and predicate devices share the same intended use and fundamental scientific technology, including principles of operation and mechanism of action. Differences in design and technological characteristics between the subject and predicate devices do not raise any new types of questions of safety and effectiveness. The results of performance testing demonstrate that the subject device is as safe and effective as the predicate devices to support a substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.