← Product Code [HGX](/productcode/HGX) · K231785

# Perifit Pump (K231785)

_X6 Innovations · HGX · Jan 26, 2024 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K231785

## Device Facts

- **Applicant:** X6 Innovations
- **Product Code:** [HGX](/productcode/HGX.md)
- **Decision Date:** Jan 26, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Perifit Pump is a wearable electric breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

## Device Story

Wearable electric breast pump for home use; collects milk from lactating women. Hub contains motor, valve, PCB, and rechargeable battery; connects to silicone flange, diaphragm, duckbill valve, and collection bottle. Operates in stimulation and expression modes with adjustable vacuum levels (30-285 mmHg) and cycle speeds (28-120 cpm). Closed-system design uses anti-backflow diaphragm to separate air-pumping from milk-flow systems. User controls pump via hub buttons or optional smartphone app. App provides session history and milk volume estimation. Benefits include portable, hands-free milk expression. Healthcare provider does not operate device; patient self-manages pumping sessions.

## Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601 series), battery safety (IEC 62133, UN 38.3), software verification/validation (IEC 62304), and performance testing (vacuum/cycle speed verification, backflow protection, use life, battery capacity, ingress protection, and packaging/drop testing).

## Technological Characteristics

Materials: silicone, polypropylene, Tritan copolyester, PC/ABS plastic. Energy: 3.7V 1500mAh Li-ion polymer battery. Form factor: wearable hub with integrated collection container. Connectivity: optional smartphone app via wireless. Sterilization: not specified. Software: embedded firmware. Standards: ISO 10993, IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 62133-2, IEC 62304, 21 CFR 177.1520/177.2420/177.2600.

## Regulatory Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

## Predicate Devices

- Elvie Stride ([K210936](/device/K210936.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 26, 2024

X6 Innovations % Lina Kontos Partner Hogan Lovells US LLP 555 13th Street NW Washington, DC 20004

Re: K231785

Trade/Device Name: Perifit Pump Regulation Number: 21 CFR\$ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: December 27, 2023 Received: December 27, 2023

Dear Lina Kontos:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Monica D. Garcia -S

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231785

Device Name Perifit Pump

Indications for Use (Describe)

The Perifit Pump is a wearable electric breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### 510(k) SUMMARY - K231785

#### X6 Innovation's Perifit Pump

#### l. Submitter:

X6 Innovations 128 rue la Boétie, 75008 Paris, France Phone: +33 6 71 26 46 45 Contact Person: Erica Perrier

Date Prepared: January 22, 2024

#### II. Device

Name of Device: Perifit Pump Common or Usual Name: Powered breast pump Regulation Name: Powered breast pump Regulation Number: 21 CFR 884.5160 Regulatory Class: Class II Product Code: HGX (Pump, Breast, Powered)

#### III. Predicate Device:

Elvie Stride (K210936), Chiaro Technology Limited, 2nd Floor 63-66 Hatton Garden, London, GB EC1M 8LE

The predicate device has not been subject to a design-related recall.

#### IV. Device Description

The Perifit Pump is an electrically powered wearable single breast pump consisting of the following key components: a housing ("hub") containing pump electronics (motor, valve, PCB, rechargeable battery and associated components), and detachable milk-contacting components: a silicone flange (breast shield), a plastic (polypropylene) connector, silicone diaphragm, a silicone duckbill valve, a plastic bottle. Accessories provided with the Perifit Pump include a screw-on container lid, a USB charging cable, and bra adjustment buckle. The Perifit pump is a closed-system breast pump.

The hub user interface allows the user to power the breast pump on and off, and to operate the breast pump in two modes (stimulation, expression) and to control the vacuum settings within these modes. Each mode consists of 8 vacuum levels, capable of providing suction from 30 to 285 mmHg, with cycle speeds ranging from 68 to 120 cycles per minute in stimulation mode and 28 to 80 cycles per minute in expression mode.

The Perifit Pump is powered by an internal rechargeable lithium-ion polymer battery (3.7V 1500mAh). The breast pump cannot be operated while it is being charged. The Perifit Pump 

{5}------------------------------------------------

operates on embedded software, and software updates by end-users are supported. The subject device is intended for use in a home healthcare environment.

An optional smartphone app may be used with the Perifit Pump. The app may be used to control the operation of the breast pump (select mode and vacuum level, start and stop pumping), save pumping session history and display the quantity (in ml) of milk in the bottle on the user's smartphone. Use of the app is not mandatory to operate the Perifit Pump.

The Perifit Pump is a closed-system breast pump with an anti-backflow diaphragm which physically separates the air-pumping system from the milk-flow system.

#### V. Intended Use / Indications for Use

The Perifit Pump is a wearable electric breast pump intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

#### VI. Comparison of Intended Use and Technological Characteristics

The table below compares the intended use and technological characteristics of the subject and predicate device.

|                              | Perifit Pump<br>Subject Device                                                                                                                                                                          | Elvie Stride (K210936)                                                                                                                                                                  | Comparison                                                                                                                                                                                                                                        |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name                 | Perifit Pump                                                                                                                                                                                            | Elvie Stride                                                                                                                                                                            | N/A                                                                                                                                                                                                                                               |
| Product Code                 | HGX                                                                                                                                                                                                     | HGX                                                                                                                                                                                     | Identical                                                                                                                                                                                                                                         |
| Regulation Number            | 21 CFR 884.5160                                                                                                                                                                                         | 21 CFR 884.5160                                                                                                                                                                         | Identical                                                                                                                                                                                                                                         |
| Regulatory Class             | Class II                                                                                                                                                                                                | Class II                                                                                                                                                                                | Identical                                                                                                                                                                                                                                         |
| Patient Population           | Lactating Women                                                                                                                                                                                         | Lactating Women                                                                                                                                                                         | Identical                                                                                                                                                                                                                                         |
| Indication for Use           | The Perifit Pump is a wearable<br>electric breast pump intended<br>to express milk from lactating<br>women in order to collect milk<br>from their breasts. The device<br>is intended for a single user. | The Elvie Stride is a powered<br>breast pump to be used by<br>lactating women to express<br>and collect milk from their<br>breasts.  The Elvie Stride is<br>intended for a single user. | Identical                                                                                                                                                                                                                                         |
| Single/double pump           | Single (or Double by using two<br>units simultaneously)                                                                                                                                                 | Single or Double                                                                                                                                                                        | Identical                                                                                                                                                                                                                                         |
| Cycling control<br>mechanism | Microcontroller                                                                                                                                                                                         | Microcontroller                                                                                                                                                                         | Identical                                                                                                                                                                                                                                         |
| Backflow Protection          | Yes                                                                                                                                                                                                     | Yes                                                                                                                                                                                     | Identical                                                                                                                                                                                                                                         |
| Suction Modes                | Stimulation Mode and<br>Expression Mode                                                                                                                                                                 | Stimulation mode and<br>Expression mode                                                                                                                                                 | Identical                                                                                                                                                                                                                                         |
| Suction levels               | 8                                                                                                                                                                                                       | 10                                                                                                                                                                                      | Similar: the difference in the<br>number of suction levels does<br>not raise different questions of                                                                                                                                               |
|                              | Perifit Pump<br>Subject Device                                                                                                                                                                          | Elvie Stride (K210936)                                                                                                                                                                  | Comparison                                                                                                                                                                                                                                        |
| Adjustable suction<br>levels | Yes                                                                                                                                                                                                     | Yes                                                                                                                                                                                     | Identical<br>safety or effectiveness as the<br>vacuum range is comparable.                                                                                                                                                                        |
| Flange Size                  | 18mm, 21 mm, 24 mm, and 27<br>mm                                                                                                                                                                        | 15, 17, and 19mm (silicone<br>inserts), 21, 24, 28mm (plastic<br>flanges)                                                                                                               | Similar: the difference in<br>flange sizes does not raise<br>different questions of safety or<br>effectiveness as the size<br>range they cover is<br>comparable. The user will<br>select the specific size that<br>works best and is comfortable. |
| Vacuum range                 | -30 to -300 mmHg                                                                                                                                                                                        | -55 to -300mmHg in single<br>pumping mode                                                                                                                                               | Similar: the difference in<br>specification does not raise<br>different questions of safety or<br>effectiveness.                                                                                                                                  |
|                              |                                                                                                                                                                                                         | (-35 to -260mmHg in double<br>pumping mode)                                                                                                                                             |                                                                                                                                                                                                                                                   |
| Cycle speed:<br>Stimulation  | 68 to 120 cycles per minute                                                                                                                                                                             | 31 to 83 cycles per minute                                                                                                                                                              | Similar: the difference in<br>specification does not raise<br>different questions of safety or<br>effectiveness.                                                                                                                                  |
| Cycle speed:<br>Expression   | 28 to 80 cycles per minute                                                                                                                                                                              | 18 to 51 cycles per minute                                                                                                                                                              | Similar: the difference in<br>specification does not raise<br>different questions of safety or<br>effectiveness.                                                                                                                                  |
| Controls                     | On/Off button (which also<br>controls Play/Pause pump<br>function); Increase/decrease<br>vacuum buttons; Mode<br>selection button                                                                       | On/Off button;<br>Play/pause button;<br>Increase/decrease vacuum<br>buttons, Mode selection button                                                                                      | Identical                                                                                                                                                                                                                                         |
| Power Supply                 | Li-Ion Battery                                                                                                                                                                                          | Li-ion Battery                                                                                                                                                                          | Identical                                                                                                                                                                                                                                         |
| Indicators                   | Yes, LED                                                                                                                                                                                                | Yes, LED                                                                                                                                                                                | Identical                                                                                                                                                                                                                                         |
| Materials                    | Milk-contacting parts are made<br>of silicone, polypropylene, and<br>Tritan copolyester (bottle only)<br>Pump Outer Housing: PC/ABS<br>plastic                                                          | Milk-contacting parts are made<br>of silicone and polypropylene<br>Pump Outer Housing: ABS<br>plastic                                                                                   | Similar: the difference in<br>bottle/container material does<br>not raise different questions of<br>safety or effectiveness.                                                                                                                      |
| Software                     | Pump operates on embedded<br>software.                                                                                                                                                                  | Pump operates on embedded<br>software.                                                                                                                                                  | Identical                                                                                                                                                                                                                                         |
|                              | A smartphone app (optional)<br>may also be used to control<br>the pump                                                                                                                                  | A smartphone app (optional)<br>may also be used to control<br>the pump.                                                                                                                 | Similar: the smartphone apps<br>for both devices provide<br>similar functionalities and any<br>differences do not raise<br>different questions of safety or<br>effectiveness.                                                                     |

# Table 1: Substantial Equivalence Table

{6}------------------------------------------------

{7}------------------------------------------------

The subject device and predicate device have the same intended use. Both the subject and predicate devices work according to the same technological principle. Suction is applied, then released, to the user's nipple in order to stimulate milk expression from a lactating woman's breasts. The user can control the vacuum intensity and select between two suction modes offering faster and slower suction cycles. The milk expressed from the breasts is collected in a collection container that is self-contained within the user's bra. Both systems are closed-system breast pumps with an anti back-flow diaphraqm physically separating the air-pumping system from the milk-collecting system. The subject device (hub containing the pump electronics plus milk-contacting elements) is fully contained within the user's bra, while the predicate's milkcontacting elements are fully contained within the nursing bra and connected to the pump using external tubing. Despite the difference in configuration, both pumps operate in the same manner, using an anti-backflow diaphragm to physically separate the pumping system from the milk collection system.

Both the subject and predicate device have the option of being used in conjunction with a smartphone app. Use of this app is not mandatory for either device, and both the subject and predicate breast pumps can be operated using only the user control interface on the hub.

The Perifit Pump app also includes some additional features that are not available with the predicate (e.g., estimate of milk volume during pumping session, selecting between several suction patterns).

The technological differences between the Perifit Pump and the predicate device do not raise different questions of safety or effectiveness.

# VII. Summary of Non-Clinical Performance Testing

# Biocompatibility

- Cytotoxicity (ISO 10993-5:2009) .
- Skin Sensitization (ISO 10993-10:2021) ●
- Skin Irritation (ISO 10993-23:2021). ●

Results of this testing support the biocompatibility of the device. The user-contacting materials were found to be non-cytotoxic, non-sensitizing, and non-irritating.

# Electrical Safety and Electromagnetic Compatibility

Electrical safety and electromagnetic compatibility testing were conducted in accordance with the following standards:

- . IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601- 1:2005/AMD2:2020 plus US National Differences: National standard AAMI ES60601-1:2005, ES60601-1:2005/AMD1 1:2012, ES60601-1:2005/AMD2:2021
- IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020 for use in conjunction with IEC 60601-1:2005, IEC 60601- 1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 plus US NATIONAL DIFFERENCES - Differences according to US National standard ANSI/AAMI HA60601-1-11:2015+A1: 2021
- . IEC 60601- 1-2:2014 + A1:2020, EN 60601- 1-2:2015 + A1:2021, IEC 60601-1 -11:2015 + A1:2020 Clause 12, EN 60601-1 -11:2015 + A1:2021 Clause 12
- IEC 62133-2:2017/AMD1:2021
- Altitude testing of the device battery per UN 38.3

{8}------------------------------------------------

Software verification and validation testing was conducted in accordance with FDA guidance ("Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", 2005 and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", 2023) and IEC 62304:2006.

All milk-contacting components were tested for compliance with material-specific requirements as outlined in 21 CFR 177.1520, 21 CFR 177.2420 and 21 CFR 177.2600.

Performance testing to substantiate the design and performance of the Perifit Pump included:

- Vacuum level and cycle speed verification testing to demonstrate that the device met ● cycle speed and suction specifications at all intensity levels in both modes of operation.
- Backflow protection testing to ensure that liquid was not able to come into contact with ● the air-pumping system.
- Device use life testing to demonstrate that the device maintained its vacuum level and cycle speed specifications over its proposed use life.
- . Battery use life testing to demonstrate that the battery maintained its capacity over the number of charging cycles associated with the device use life.
- . Battery charge and discharge time to demonstrate that the operating time and charge time of the breast pump device fall within specifications communicated to the user.
- . Battery status indicator testing
- Milk level validation testing
- Ingress protection (IP22)
- Packaging drop and vibration testing according to ISTA-3A ●

#### Conclusion

The results of the performance testing described above demonstrate that the Perifit Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K231785](https://fda-staging.innolitics.com/device/K231785)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
