K231778 · Nihon Kohden Orangemed, LLC · CBK · Mar 1, 2024 · Anesthesiology
Device Facts
Record ID
K231778
Device Name
Nihon Kohden NKV-550 Series Ventilator System
Applicant
Nihon Kohden Orangemed, LLC
Product Code
CBK · Anesthesiology
Decision Date
Mar 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-550 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-550 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation.
Device Story
NKV-550 Series Ventilator System provides continuous respiratory support via invasive or non-invasive modes. System comprises graphic user interface (GUI) and breath delivery unit (BDU). GUI allows clinician parameter setting (tidal volume, inspiratory pressure), alarm configuration, and waveform/loop monitoring. BDU utilizes microprocessor-controlled pneumatic system with proportional valves to deliver oxygen (21-100%) and medical air. Device supports A/CMV, SIMV, and SPONT modes; includes new Adaptive Ventilation Mode (AVM), Adaptive Trigger (ATRIG), and Adaptive Cycle (Acycle) software features. Used in hospitals and for in-hospital transport by healthcare professionals. Output provides mechanical ventilation and real-time respiratory monitoring to assist patient breathing; aids clinical decision-making through automated breath delivery and alarm management.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, software verification, device functionality testing, human factors/usability testing, and risk management/regression analysis. Compliance with ISO 80601-2-12 requirements for critical care ventilators confirmed.
Technological Characteristics
Continuous ventilator; AC/DC power; proportional valve pneumatic system; microprocessor-controlled. Features AVM, ATRIG, and Acycle software. Meets ISO 80601-2-12. No material changes to gas path; biocompatibility unchanged.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring invasive or noninvasive respiratory support for continuous ventilation.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
Nihon Kohden NKV-550 Series Ventilator System (K192307)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nihon Kohden OrangeMed, Inc. Sheryl Higgins V.P. of Regulatory Affairs & Quality Assurance 1800 E. Wilshire Avenue Santa Ana, California 92705
#### Re: K231778
Trade/Device Name: Nihon Kohden NKV-550 Series Ventilator System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: February 6, 2024 Received: February 6, 2024
#### Dear Sheryl Higgins:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by John S. John S. Bender -S Date: 2024.03.01 10:47:10 Bender -S -05'00' for Ethan Nyberg Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K231778
Device Name
Nihon Kohden NKV-550 Series Ventilator System
Indications for Use (Describe)
The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-550 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-550 is intended for use in hospitals and hospital-type facilities, as well as, for in-hospital transportation.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|--|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |
| <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for NIHON KOHDEN. The logo consists of a blue abstract graphic on the left, followed by the company name in blue capital letters. The graphic appears to be a stylized representation of a wave or a rising sun.
# Section 5 - 510(k) Summary
#### Submission Date
February 29, 2024
### Submitter / Manufacturing Location
Nihon Kohden OrangeMed, LLC 1800 E. Wilshire Avenue Santa Ana, CA 92705 USA
Phone: (949) 502-6448
### Company Contact
Sheryl Higgins Vice President of Regulatory Affairs and Quality Assurance
shiggins@orange-med.com Email:
(949) 502-6448 x7012 Phone:
#### Establishment Registration Number
3014631252 - Nihon Kohden OrangeMed, LLC
#### Common Name of Device
Critical Care Ventilator
#### Trade Name
Nihon Kohden NKV-550 Series Ventilator System
#### Classification Name
Product Code: CBK - Ventilator, Continuous, Facility Use Regulation Number: 21 CFR 868.5895 - Continuous ventilator Device Class: II Review Panel: Anesthesiology
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for NIHON KOHDEN. The logo consists of a blue square with a white design inside, followed by the text "NIHON KOHDEN" in blue. The text is in a sans-serif font and is aligned to the right of the square.
### Primary Predicate Device
| Trade Name: | Nihon Kohden NKV-550 Series Ventilator System |
|-----------------------|-----------------------------------------------|
| 510k #: | K192307 |
| Product Code: | CBK |
| Manufacturer: | Nihon Kohden OrangeMed, LLC |
| Establishment Reg. #: | 3014631252 |
#### Secondary Predicate Device
| Trade Name: | Hamilton-G5 |
|-----------------------|---------------------|
| 510k #: | K193228 |
| Product Code: | CBK |
| Manufacturer: | Hamilton Medical AG |
| Establishment Reg. #: | 3001421318 |
### Device Description
The Nihon Kohden NKV-550 Series Ventilator System consists of a graphic user interface (GUI) and a breath delivery unit (BDU). The GUI allows clinicians to set ventilator control parameters such as tidal volume and inspiratory pressure, to set alarm limits such as high inspiratory pressure alarm, to view monitored numeric values, to view waveform and loops, and to operate various features through the apps.
The BDU contains a microprocessor that receives inputs from the electronic system and controls the pneumatic system for breath delivery to the patient. It also provides various alarms, a safety valve, and other design features to maximize patient safety.
### Product Intended Function
The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation using medical oxygen and external sources of compressed medical air to deliver oxygen concentrations of 21% to 100%. Ventilatory support is intended to be delivered invasively or non-invasively to patients who require Assisted/Control Mandatory Ventilation (A/CMV), Synchronized Intermittent Mandatory Ventilation (SIMV) or Spontaneous Ventilation (SPONT).
## Indication For Use (Intended Medical Indication)
The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support. The NKV-550 offers mandatory and spontaneous ventilation modes as well as respiratory monitoring. The NKV-550 is intended for use in hospitals and hospitaltype facilities, as well as, for in-hospital transportation.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for NIHON KOHDEN. The logo consists of a blue square with a white design inside, followed by the text "NIHON KOHDEN" in blue. The design inside the square features a white curved line at the bottom and a white diagonal line extending from the bottom left corner to the top right corner.
### Summary of Technical Characteristics
The Nihon Kohden NKV-550 Series Ventilator System technological characteristics with the addition of Adaptive Ventilation Mode (AVM) and Adaptive Trigger (ATRIG) /Adaptive Cycle (Acycle) are substantially equivalent as compared to the predicate device and the reference device as summarized in the table below.
| Table 7-1<br>Characteristic | Nihon Kohden<br>NKV-550 Series<br>Ventilator System<br>(Modified Device) | Nihon Kohden<br>NKV-550 Series<br>Ventilator System<br>(Primary<br>Predicate Device -<br>k192307) | Hamilton G5<br>(Secondary<br>Predicated Device<br>- k193228) | Comparison |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indication for Use | The Nihon Kohden<br>NKV-550 Series<br>Ventilator System is<br>intended to provide<br>continuous ventilation<br>for adult, pediatric and<br>neonatal patients who<br>require invasive or<br>noninvasive<br>respiratory support.<br>The NKV-550 offers<br>mandatory and<br>spontaneous<br>ventilation modes as<br>well as respiratory<br>monitoring. The NKV-<br>550 is intended for use<br>in hospitals and<br>hospital-type facilities,<br>as well as for in-<br>hospital transportation. | The Nihon Kohden<br>NKV-550 Series<br>Ventilator System is<br>intended to provide<br>continuous<br>ventilation for adult,<br>pediatric and<br>neonatal patients who<br>require invasive or<br>noninvasive<br>respiratory support.<br>The NKV-550 offers<br>mandatory and<br>spontaneous<br>ventilation modes as<br>well as respiratory<br>monitoring. The<br>NKV-550 is intended<br>for use in hospitals<br>and hospital-type<br>facilities, as well as<br>for in-hospital<br>transportation. | The Hamilton-G5<br>ventilator is designed<br>for intensive care<br>ventilation of adult<br>and pediatric<br>patients, and<br>optionally infant and<br>neonatal patients.<br>The device is<br>intended for use in<br>the hospital and<br>institutional<br>environment where<br>health care<br>professionals provide<br>patient care | Same |
| Environment of<br>Use | Hospitals, hospital-<br>type facilities and in-<br>hospital transportation<br>for patients who need<br>ventilation therapy | Hospitals, hospital-<br>type facilities and in-<br>hospital<br>transportation for<br>patients who need<br>ventilation therapy | Hospital and<br>institutional<br>environment where<br>health care<br>professionals provide<br>patient care | Same |
| Anatomical Site | Patient airways | Patient airways | Patient airways | Same |
| Target Population | Adult, pediatric and<br>neonatal patients | Adult, pediatric and<br>neonatal patients | Adult, pediatric, and<br>neonatal patients | Same |
| Performance | Met ISO 80601-2-12<br>requirements on<br>essential performance<br>of critical care<br>ventilator | Met ISO 80601-2-12<br>requirements on<br>essential performance<br>of critical care<br>ventilator | Met ISO 80601-2-12<br>requirements on<br>essential performance<br>of critical care<br>ventilator | Same |
| Table 7-1<br>Characteristic | Nihon Kohden<br>NKV-550 Series<br>Ventilator System<br>(Modified Device) | Nihon Kohden<br>NKV-550 Series<br>Ventilator System<br>(Primary<br>Predicate Device -<br>k192307) | Hamilton G5<br>(Secondary<br>Predicated Device<br>- k193228) | Comparison |
| Design | Consists of a graphic<br>user interface to set<br>and monitor<br>ventilation, breath<br>delivery unit,<br>breathing circuit;<br>Controls air and<br>oxygen deliveries by<br>proportional valves<br>through<br>microprocessors | Consists of a graphic<br>user interface to set<br>and monitor<br>ventilation, breath<br>delivery unit,<br>breathing circuit;<br>Controls air and<br>oxygen deliveries by<br>proportional valves<br>through<br>microprocessors | Consists of a graphic<br>user interface to set<br>and monitor<br>ventilation, breath<br>delivery unit,<br>breathing circuit;<br>Controls air and<br>oxygen deliveries by<br>proportional valves<br>through<br>microprocessors | Same |
| Chemicals<br>Delivered to<br>Patient | Medical Air and<br>Oxygen | Medical Air and<br>Oxygen | Medical Air and<br>Oxygen | Same |
| Delivery method to<br>Patient | Positive pressure | Positive pressure | Positive pressure | Same |
| Energy Used for<br>Device | AC Power and DC<br>Power (battery) | AC Power and DC<br>Power (battery) | AC Power and DC<br>Power (battery) | Same |
| Control principle | time-cycled, volume-<br>constant, pressure-<br>controlled | time-cycled, volume-<br>constant, pressure-<br>controlled | time-cycled, volume-<br>constant, pressure-<br>controlled | Same |
| Therapy Types | Invasive, Non-<br>invasive, O2 Therapy | Invasive, Non-<br>invasive, O2 Therapy | Invasive, Non-<br>invasive, O2 Therapy | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image features the logo of NIHON KOHDEN. The logo consists of a stylized blue graphic on the left, resembling a stylized mountain range or waveform. To the right of the graphic, the text "NIHON KOHDEN" is displayed in a bold, sans-serif blue font. The overall design is clean and corporate, suggesting a professional and established organization.
The differences between the modified Nihon Kohden NKV-550 Ventilator System and the predicate device (k192307) are as follows:
- The subject device adds a breath mode, Adaptive Ventilation Mode (AVM), as well 1. as Adaptive Trigger (ATRIG)/Adaptive Cycle (AcycLE). These are software-only features that do not require any new hardware.
The differences between the modified Nihon Kohden NKV-550 Ventilator System and the secondary predicate device (k193228) are described in section 12 – Substantial Equivalent Discussion.
In conjunction with AVM and Adaptive Trigger (ATRIG)/Adaptive Cycle (AcYCLE), this 510(k) brings FDA up-to-date on non-significant changes implemented since the last 510(k) clearance. Therefore, the following modifications are included:
- 1. Mechanical Hardware changes
- 2. Software Revision History changes
- 3. Labeling changes
Device modifications were made in compliance with Design Control procedures.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Nihon Kohden, a Japanese medical equipment manufacturer. The logo consists of a blue abstract graphic to the left of the company name, also in blue. The graphic appears to be a stylized representation of a wave or a wing. The text is in a sans-serif font and is easily readable.
Traditional 510(k): NKV-550 Series Ventilator System 510(k) #: K231778 510(k) Summary
OrangeMed
### Summary of Non-Clinical Performance Data
Performance of the Nihon Kohden NKV-550 Series Ventilator System was demonstrated by the following testing performed in compliance with Design Controls:
- Software Verification
- Device Functionality Testing ●
Human Factors/Usability Testing
- Bench Test Comparison with Reference Device
- Risk Management
- . Regression
Additional biocompatibility testing was not required since no material changes were made to the gas path of the ventilator.
### Summary of Clinical Performance Data
Not Applicable – Clinical performance data was not required to demonstrate substantial equivalence.
### Conclusion
●
The evaluation and testing performed in compliance with Design Controls demonstrates that the Nihon Kohden NKV-550 Series Ventilator System is as safe and as effective as its primary and secondary predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.