K231749 · Shenzhenshi Sincoheren S&T Development Co., Ltd. · GEI · Oct 19, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231749
Device Name
Radio Frequency System
Applicant
Shenzhenshi Sincoheren S&T Development Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 19, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Radio Frequency System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Device Story
Radio Frequency System is a bipolar electrosurgical device for dermatologic and general surgical procedures. System comprises a host unit with LCD touchscreen, moving base, treatment handle, and pedal/finger switch. Operator (clinician) selects RF power, duration, and depth via touchscreen. Device delivers RF energy through sterilized tips (microneedle or RF electrodes) to target tissue. Microneedle tips penetrate epidermis; RF electrodes do not. Energy generates electro-thermal reaction for tissue coagulation and hemostasis. Device is used in clinical settings. Output allows clinicians to perform precise coagulation; benefits include controlled thermal effect on target tissue.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including biocompatibility (ISO 10993 series), sterilization validation (ISO 11135), shelf-life validation (ASTM F1980-16), and electrical safety/EMC testing (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6).
Technological Characteristics
Bipolar RF system; 1 MHz frequency; 50 W max output. Materials: Gold-plated stainless steel electrodes. Tips: 49-pin, 25-pin microneedle, and 49-RF electrodes. Control: LCD touchscreen, foot/finger switch. Sterilization: Ethylene Oxide (ISO 11135). Software: Moderate level of concern.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electrocoagulation and hemostasis. No specific patient population age or gender restrictions are provided.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
October 19, 2023
Shenzhenshi Sincoheren S&T Development Co., Ltd. Cai Jialong MD Registered Engineer Floor 4, No 2 plant, No. 14, Zhongxing Road, Xiuxin Community, Kengzi Street, Pingshan District Shenzhen, Guangdong 518000 China
# Re: K231749
Trade/Device Name: Radio Frequency System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: May 11, 2023 Received: June 15, 2023
# Dear Cai Jialong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Trumbore -S Trumbore -S Trumbore -S
Mark Trumbore -S Trumbore -S Trumbore -S
- Date: 2023.10.19 15:32:33 -04'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Radio Frequency System
Indications for Use (Describe) The Radio Frequency System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summarv
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
#### 1. General Information
| Submitter: | Shenzhenshi Sincoheren S&T Development Co., Ltd.<br>Floor 4, No 2 plant, No. 14, Zhongxing Road, Xiuxin<br>Community, Kengzi Street, Pingshan District, Shenzhen<br>City, Guangdong Province, China |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Cai Jialong<br>MD Registered Engineer<br>Tel: +86-13424006986<br>Email: caijl@medsmt.com |
| Summary Preparation Date: | August 21, 2023 |
#### 2. Device Name and Code
Trade/Common Name: Radio Frequency System Classification Name: Electrosurgical, Cutting & Coagulation & Accessories Classification: Class II Product Code: GEI Regulation Number: 21 CFR 878.4400 Review Panel: General & Plastic Surgery
#### 3. Predicate Device
The INFINI Radiofrequency System (K121481). This predicate device has not been subject to a design-related recall. No reference devices were used in this submission.
# 4. Device Description
The Radio Frequency System consists of a host with an LCD touch screen control panel, moving base, treatment handle and pedal switch. The treatment handle includes a handpiece and a sterilized tip. And the sterilized tip is classified into two type, microneedle electrodes and RF electrodes. The treatment parameters are entered via the touchscreen control panel that also displays system output information during treatment. Microneedle electrodes come in light contact with the epidermis of
{4}------------------------------------------------
the patient and minimally penetrate the epidermis while RF electrodes not penetrate. RF POWER, ON TIME (RF POWER×ON TIME = RF Energy), DEPTH, are user-selectable via the control panel. The RF Power output is controlled to insure for a given time that a determinate RF Energy is delivered to the bipolar electrodes. The handpiece is held at right angles to the target tissue. As the RF energy passes into the skin, it generates an electro-thermal reaction capable of coagulating the tissue surrounding the uninsulated portion of the electrodes.
### 5. Indications for Use
The Radio Frequency System is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
#### 6. Technological Characteristics in Comparison to the Predicate Device
The subject device has the same intended use and indications for use and the same fundamental scientific technology as the Predicate device, the INFINI Radiofrequency System cleared in K121481. The subject device design, technology, and the principles of operation are the same as the predicate device. The Radio Frequency System and the Predicate device are bipolar radio frequency systems, with delivery methods through microneedles inserted from the handpiece and the tip. The Radio Frequency System and the predicate device are minimally invasive radio frequency devices employing bipolar microneedle electrode system. The subject device has the same power, same treatment duration, needle diameter, and needle depth as the predicate device. Therefore, the minor differences do not raise any new safety and effectiveness questions because the parameters of the Subject device are similar to those of the predicate device.
| Description | Radio Frequency System | INFINI<br>Radiofrequency<br>System(K121481) | Comments |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Manufacturer | Shenzhen Sincoheren S&T<br>Development Co., Ltd. | Lutronic Corporation | |
| Indication for Use | The Radio Frequency<br>System is intended for use<br>in dermatologic and<br>general surgical<br>procedures for<br>electrocoagulation and<br>hemostasis. | The INFINI<br>Radiofrequency System<br>is intended for use in<br>dermatologic and<br>general surgical<br>procedures for<br>electrocoagulation and | Same as predicate |
| Description | Radio Frequency System | INFINI<br>Radiofrequency<br>System(K121481) | Comments |
| | | hemostasis, and the<br>percutaneous treatment<br>of facial wrinkles. | |
| Prescription/Over-t<br>he Counter Use | Prescription Use | Prescription Use | Same as predicate |
| System Type | Bipolar RF (Radio<br>frequency) | Bipolar RF<br>(Radiofrequency) | Same as predicate |
| Frequency | 1 MHz | 1 MHz | Same as predicate |
| Max Output Power | 50 W | 50 W | Same as predicate |
| Operation Mode | Fractional RF | Fractional RF | Same as predicate |
| Treatment<br>Duration(Time) | 10~1000 ms | 10~1000 ms | Same as predicate |
| Tips | 49-pin: 7×7 microneedle<br>electrodes, 10.5 mm×10.5<br>mm spot size, microneedle<br>electrodes depth<br>adjustment 0.5~3.5 mm<br>25-pin: 5×5 microneedles<br>electrodes, 10.5 mm×10.5<br>mm spot size, microneedle<br>electrodes depth<br>adjustment 0.5~3.5 mm<br>49-RF: 7×7 RF electrodes,<br>10.5 mm×10.5 mm spot<br>size, no electrodes depth<br>adjustment | 49-pin:7×7<br>microneedles, 10×10<br>mm microneedle depth<br>adjustment 0.5~3.5 mm<br>spot size, microneedle<br>depth adjustment<br>0.5~3.5 mm | Different. Additions<br>of tip type in the<br>subject device. |
| Electrode Diameter | Microneedle<br>electrodes:2µm,<br>RF electrodes:600 µm | 200 μm | Different. Additions<br>of tip type in the<br>subject device. |
| Output Control | Foot/ Finger switch | Foot switch | Different. Additions<br>of output control in<br>the subject device. |
| Electrode Material | Gold plated stainless steel | Gold plated stainless<br>steel | Same as predicate<br> |
### 7. Substantial Equivalence Table
{5}------------------------------------------------
Shenzhen Sincoheren S&T Development Co., Ltd. has removed from the subject device the percutaneous treatment of facial wrinkles as an Indication for use. The subject device and the
{6}------------------------------------------------
predicate device have the same Indication for Use.
The following differences exist between the Subject Device and Predicate Device:
· The Subject Device use an additional type of tip(49-RF) and two models of microneedle electrodes(49-pin and 25-pin) but the Predicate Device only use one model of microneedle electrodes(25-pin).
· The finger switch is an addition to the Subject Device.
# 8. Performance Date
The following performance data were provided in support of the substantial equivalence determination:
# Biocompatibility testing
The patient contacting component of the subject device is the biocompatibility tests were conducted in accordance with ISO 10993-1.
- · In Vitro Cytotoxicity Test: ISO 10993-5: 2009
- · Skin Sensitization Test: ISO 10993-10:2021
- · Intracutaneous Reactivity Test: ISO 10993-23:2021
- · Acute Systemic Toxicity Test: ISO 10993-11:2017
- Pyrogen Test: ISO 10993-11:2017
#### Sterilization Validation
Sterilization validation of the tip has been conducted per the standard ISO 10993-7:2008 for Ethylene Oxide(standard ISO 11135:2014).
#### Shelf-life Validation
The shelf-life validation of the device has been conducted using the accelerated aging method in accordance to ASTM F1980-16.
#### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety testing and EMC testing were conducted on the Radio frequency System complying with the IEC 60601-1 standard, the IEC 60601-1-6 standard for safety and the IEC 60601-1-2 standard for EMC.
#### Software Verification and Validation Testing
{7}------------------------------------------------
Software Verification and Validation Testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could indirectly result in minor injury to the patient or operator.
No preclinical and clinical data was provided in this submission.
#### Conclusion
The subject device shares a similar design and intended use to its predicate device(K121481). Additionally, principles of operation, performance characteristics, technological characteristics are similar between the subject device and its predicate devices. The results of verification and validation activities, i.e., testing to standards and performance testing of the devices, have demonstrated substantial equivalence of the subject device to its predicate devices. The minor differences in the subject device do not raise new types of questions regarding safety and efficacy, and the data presented in this 510(k) Premarket Notification supports the contention that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.