The Neteera 130H-Plus device is intended for spot and continuous measurement of heart rate and respiration rate in adult patients (in healthcare facilities and home monitoring) and inform on bed exit. The indications provided are to be used by health care professionals and are intended to be reviewed by clinicians to inform patient care. The Neteera device is not intended to be used as an alarm system for potentially acute life-threatening situations in which medical intervention is necessary (e.g., ICU).
Device Story
Neteera 130H-Plus is a contact-free vital signs monitor using high-frequency (122.25-123 GHz) micro-radar technology. It detects ballistocardiograph (BCG) micro-movements of skin through nonmetallic materials (clothing, furniture). The device measures heart rate (HR) and respiration rate (RR) during rest or mild movement and identifies bed exit events. Used in healthcare facilities and home environments, it is operated by healthcare professionals to inform patient care. Data is processed and transmitted to a cloud dashboard for clinician review. The device provides real-time, non-invasive monitoring, benefiting patients by enabling remote observation without physical contact or leads.
Clinical Evidence
Confirmatory clinical validation study conducted on 46 subjects under GCP compliance. Compared subject device against FDA-cleared ECG and Capnograph (EPM-10M) for spot and continuous HR and RR measurement. All outcomes met pre-specified performance criteria, demonstrating that hardware modifications and new features did not negatively affect clinical performance.
Technological Characteristics
Contact-free micro-radar sensor (122.25-123 GHz). Measures ballistocardiograph (BCG) micro-movements. Components include sensor, USB cable, power supply. Connectivity via cloud dashboard. Complies with IEC 60601-1, IEC 60601-1-2 (EMC), IEC 60601-1-11 (home healthcare), IEC 62304 (software lifecycle), and FCC 47 CFR Part 15 (116-123 GHz band).
Indications for Use
Indicated for adult patients in healthcare facilities and home monitoring settings for spot and continuous heart rate and respiration rate measurement, and bed exit notification. Not for use as an alarm system for acute life-threatening situations requiring medical intervention (e.g., ICU).
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
February 9, 2024
Neteera Technologies Ltd. Rakefet Shohat VP QA/RA/CA Prof Rakah 3 St. Jerusalem, Jerusalem 9139002 Israel
Re: K231733
Trade/Device Name: Neteera 130H-Plus Vital Sign Monitoring Sensor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: DRT, BZQ Dated: January 11, 2024 Received: January 11, 2024
Dear Rakefet Shohat:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
#### Hetal B. Odobasic -S for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
510(k) Number (if known)
K231733
Device Name
Neteera 130H-Plus Indications for Use (Describe)
The Neteera 130H-Plus device is intended for spot and continuous measurement of heart rate and respiration rate in adult patients (in healthcare facilities and home monitoring) and inform on bed exit.
The indications provided are to be used by health care professionals and are intended to be reviewed by clinicians to inform patient care.
The Neteera device is not intended to be used as an alarm system for potentially acute life-threatening situations in which medical intervention is necessary (e.g., ICU).
Type of Use (Select one or both, as applicable)
□ Over-The-Counter Use (21 CFR 801 Subpart C) 2 Prescription Use (Part 21 CFR 801 Subpart D)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number
{3}------------------------------------------------
# 510(k) Summary
## Device Name: Neteera 130H-Plus Vital Signs Monitoring Sensor
### 1. 510K Submission Sponsor
Neteera Technologies Ltd. Prof. Rakah 3 St. PO Box 39088 Jerusalem 9139002, ISRAEL Company Phone No.: +972-2-5808733
Contacts:
- a. Janice M. Hogan, Consultant, Hogan Lovells US LLP Email: mjanice.hogan@hoganlovells.com
- b. Rakefet Shohat, VP QA/RA/CA Email: rakefet.shohat(@neteera.com
# 2. Date Prepared
9-Jan 2024
## 3. Device Identification
Trade/Proprietary Name: Neteera 130H-Plus Common/Usual Name: Vital Signs Monitoring Sensor Classification Name: Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm) & Monitor, Breathing Frequency Regulation Number: 21 CFR 870.2300 (Product Code DRT) & 21 CFR 868.2375 (Product Code BZQ) Product Code: DRT & BZQ Device Class: 2 Classification Panel: Cardiovascular
- 4. Legally Marketed Predicate Device(s): Neteera's Vital Sign Monitoring Sensor Model: 130H (K212143)
- 5. Indication for Use Statement: The Neteera 130H-Plus device is intended for spot and continuous measurement of heart rate and respiration rate in adult patients (in facilities and home monitoring) and inform healthcare on bed exit.
Neteera Technologies Ltd
Page 1 of 5
{4}------------------------------------------------
The indications provided are to be used by health care professionals and are intended to be reviewed by clinicians to inform patient care.
The Neteera device is not intended to be used as an alarm system for potentially acute life-threatening situations in which medical intervention is necessary (e.g., ICU).
- 6. Device Description: Neteera 130H-Plus device is a modification of the predicate device Neteera 130H (K212143). It is a contact-free vital-signs monitor based on a high frequency (122.25-123 GHz) micro-radar on-chip and algorithm, capable of detecting Respiration Rate (RR) and Heart Rate (HR) during rest or subject's mild body movement, with the additional capability of identifying and notifying on bed exit.
Neteera's micro- radar-based solution enables remote measurement, in real-time, in a non-invasive and contact-free manner. The system works by measuring only the ballistocardiograph micro-movements of of the skin (BCG) through nonmetallic materials such as furniture and clothing at a high resolution, and it has several different mounting options and measuring ranges.
## 7. Performance Testing (Non-Clinical):
As part of demonstrating Substantial Equivalence for safety (electrical, EMC, etc.) and effectiveness of the Neteera 130H-Plus to the predicate device, Neteera Technologies Ltd. repeated several non-clinical performance (SW) and electrical safety tests that were performed for the predicate device. Testing results confirmed that the design output meets the design inputs and specifications of the subject device.
Testing and/or compliance with the following standards has been achieved for the Neteera 130H-Plus Device:
- IEC 60601-1-2: 2014 + A1: 2020 Edition 4.1Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [tested as Class B, Group 2].
- EN 60601-1-2:2015 + A1: 2021 Edition 4.1 - Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-6:2010/AMD1:2013 for use in conjunction with IEC 62366:2007, ● AMD1:2014 and IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012 Medical electrical equipment - Part 1-6: General requirements for safety -Collateral standard: Usability.
- EN 60601-1-6:2010/AMD1:2015 for use in conjunction with EN 60601-1:2006 + A11:2011 + A1:2013.
- EN 60601-1-11:2015+A1:2021 Medical electrical equipment. ●
- IEC 60601-1-11:2015+A1:2020 Medical electrical equipment Part 1-11 for use ● in conjunction with IEC 60601-1:2005, AMD1:2012: General requirements for basic safety and essential performance - Collateral Standard: Requirements for
Neteera Technologies Ltd
Page 2 of 5
{5}------------------------------------------------
medical electrical equipment and medical electrical systems used in the home healthcare environment.
- EN 60601-1:2006 +A11:2011 + A1: 2013 Medical electrical equipment General requirements for basic safety and essential performance.
- IEC 60601-1:2005 AMD1: 2012 Medical electrical equipment General ● requirements for basic safety and essential performance.
- IEC 62366-1 Edition 1.0 2015-02 Medical devices - Part 1: Application of usability engineering to medical devices [Including CORRIGENDUM 1 (2016)]
- IEC 62304:2006/A1:2015 Medical device software - Software life cycle processes.
- AIM 7351731 Rev. 3.00 (2021-06-04) Medical Electrical Equipment and ● System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers.
- IEEE/ANSI C63.27-2017– American National Standard for Evaluation of ● Wireless Coexistence.
- AAMI TIR69 2017 - Risk management of radio-frequency wireless coexistence for medical devices and systems.
- EN 305 550-1 V1.2.1 sections 7.1 thru 7.5, sec 8.1. Electromagnetic compatibility and Radio spectrum Matters (ERM); Short Range Devices (SRD); Radio equipment to be used in the 40 GHz to 246 GHz frequency range; Part 1: Technical characteristics and test methods.
- FCC 47 CFR part 15 (Operation in the bands 116-123 GHz) section 15.258(b), transmitter power test; section 15.215(c), occupied bandwidth; section 15.258(c) (2), Out of band radiated emissions below 40 GHz; section 15.258(c) (3), Out of band radiated emissions below 40 GHz up to 370 GHz; section 15.258(d), Frequency stability test; section 15.203, Antenna requirement.
- EN/IEC 60601-1-8 General requirements for safety - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
In addition to the testing above, software validation was also conducted. This validation demonstrated that the Neteera 130H-Plus performed and functioned as intended and according to the design specifications and as such is equivalent to the predicate.
Neteera Technologies Ltd
Page 3 of 5
{6}------------------------------------------------
## 8. Clinical Testing
As part of the validation of the effectiveness of the device, Neteera Technologies Ltd. completed a GCP-compliant confirmatory clinical validation study under Independent Ethical Committee approval to demonstrate that minor changes to the device did not affect clinical performance. The study compared the subject device to FDA-cleared devices in the same way and using the same FDA-cleared comparators as done for the clearance of the predicate device. Specifically, the comparators consisted of FDAcleared devices ECG and Capnograph, EPM-10M S/N: AD1-03 for measuring Spot HR, Spot RR, Continuous HR and Continuous RR. Testing was performed on 46 subjects and all outcomes met the same pre-specified performance criteria.
### 9. Statement of Substantial Equivalence
The Neteera 130H-Plus device has the same intended use and same indications for use and technological characteristics as the predicate device, Neteera 130H, except it has the additional ability to detect and notify on Bed Exit of the monitored person, and minor hardware modifications were introduced.
The changes introduced for the 130H-Plus device and warranted this submission are hardware modification that required performance validation using human subjects and notification on Bed Exit which entails a change in the Intended Use (labeling).
The minor difference of having a feature that identifies and notifies on Bed Exit of the monitored subjects, and the hardware modifications do not raise different questions regarding the safety and effectiveness of the device. A Substantial Equivalence Table is presented below.
Therefore, based on the information provided in this 510(k) premarket notification, Neteera Technologies Ltd. concludes that the Neteera 130H-Plus device is substantially equivalent to the Neteera 130H predicate device.
| | Subject Device | Predicate Device |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | K231733 | K212143 |
| Device Name | Vital Sign Monitoring Sensor<br>Neteera 130H-Plus | Vital Sign Monitoring Sensor<br>Neteera 130H |
| Common name | Heart Rate and Respiratory Rate<br>Monitor | Heart Rate and Respiratory Rate<br>Monitor |
| Manufacturer | Neteera Technologies Ltd. | Neteera Technologies Ltd. |
| Intended Use | The Neteera System, Neteera<br>130H-Plus device, is intended<br>for spot and continuous<br>measurement of heart rate and<br>respiration rate in adult patients<br>(in healthcare facilities and<br>home monitoring) and inform<br>on bed exit.<br>The indications provided are to<br>be used by health care<br>professionals and are intended | The Neteera System, Neteera<br>130/131H device, is intended for<br>spot and continuous<br>measurement of heart rate and<br>respiration rate in adult patients<br>(in healthcare facilities and<br>home monitoring).<br>The indications provided are to<br>be used by health care<br>professionals and are intended to |
| | to be reviewed by clinicians to<br>inform patient care.<br>The Neteera device is not<br>intended to be used as an alarm<br>system for potentially acute<br>life-threatening situations in<br>which medical intervention is<br>necessary (e.g., ICU). | be reviewed by clinicians to<br>inform patient care.<br>The Neteera device is not<br>intended to be used as an alarm<br>system for potentially acute life-<br>threatening situations in which<br>medical intervention is necessary<br>(e.g., ICU). |
| Components | Sensor; USB Cable, Power<br>Supply | Sensor; USB Cable, Power<br>Supply |
| Technology | Contactless, Radar-based<br>measurement of micro-motions | Contactless, Radar-based<br>measurement of micro-motions |
| Medical<br>Parameters<br>Monitored and<br>Displayer | Heart Rate; Respiratory Rate,<br>bed exit. | Heart Rate; Respiratory Rate |
| Patient Type | Adult | Adult |
| Use Environment | Healthcare facilities and home<br>monitoring.<br>The indications provided are to<br>be used by health care<br>professionals and are intended to<br>be reviewed by clinicians to<br>inform patient care.<br>Not intended for potentially<br>acute life-threatening situations. | Healthcare facilities and home<br>monitoring.<br>The indications provided are to<br>be used by health care<br>professionals and are intended to<br>be reviewed by clinicians to<br>inform patient care.<br>Not intended for potentially acute<br>life-threatening situations. |
| HR Measurement<br>Range | 40-160 Beats Per Minute | 40-160 Beats per Minute |
| RR Measurement<br>Range | 5-40 Breaths Per Minute | 5-40 Breaths per Minute |
| EMC<br>Compliance | IEC 60601-1-2 | IEC 60601-1-2 |
| Product<br>Electrical Safety<br>and coexistence | IEC 60601-1;<br>IEC 60601-1-6;<br>IEC 60601-1-11<br>AIM 7351731 Rev. 3.00<br>IEEE/ANSI C63.27-2017<br>AAMI TIR69 2017<br>EN 305 550-1 V1.2.1<br>FCC 47 CFR part 15<br>EN/IEC 60601-1-8 | IEC 60601-1;<br>IEC 60601-1-6;<br>IEC 60601-1-11<br>AIM 7351731 Rev. 3.00<br>IEEE/ANSI C63.27-2017<br>AAMI TIR69 2017<br>EN 305 550-1 V1.2.1<br>FCC 47 CFR part 15<br>EN/IEC 60601-1-8 |
| Data Display | Cloud Dashboard | Cloud Dashboard |
Table 5A - Substantial Equivalence Chart for the Neteera 130H-Plus
Neteera Technologies Ltd
Page 4 of 5
{7}------------------------------------------------
Neteera Technologies Ltd
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.