← Product Code [GEI](/productcode/GEI) · K231732

# CleanSeal Advanced Bipolar Vessel Sealer Maryland (K231732)

_Conmed Corporation · GEI · Oct 23, 2023 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K231732

## Device Facts

- **Applicant:** Conmed Corporation
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Oct 23, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

CONMED CleanSeal Vessel Sealers are intended to be used in open and minimally invasive procedures for the ligation and division of tissue bundles, lymphatics, and vessels up to and including 6 mm vessels (arteries, veins, pulmonary arteries, pulmonary veins). The handpiece is indicated for use in gynecological and general surgical procedures, including urologic, thoracic, and vascular. These procedures include, but are not limited to, hysterectomies, colectomies, Nissen fundoplication, and adhesiolysis. CleanSeal vessel sealer has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use CleanSeal vessel sealer for these procedures.

## Device Story

CleanSeal Advanced Bipolar Vessel Sealer Maryland is a sterile, single-use bipolar electrosurgical handpiece; used with compatible MEP-1 generator. Device features double-action non-stick coated jaws, three shaft lengths (23cm, 37cm, 44cm), and a one-hand controlled handle with latchable lever. User operates device in open or minimally invasive surgical settings. Mechanism: RF energy applied to tissue/vessels between jaws; integrated stainless-steel knife for transection. Rotator knob allows 360° shaft rotation. Smart chip in ESU plug communicates parameters to generator. Healthcare provider uses device to seal and divide vascular structures; benefits include effective hemostasis and tissue management during procedures like hysterectomies and colectomies.

## Clinical Evidence

Benchtop, ex-vivo, and in-vivo simulated use testing performed. Chronic animal studies (porcine and canine models) evaluated performance against predicates. Results demonstrated effective sealing of arteries, veins, and AV bundles after 21 days. Biocompatibility testing met ISO 10993-1 requirements. Mechanical and electrical verification confirmed compliance with AAMI/ANSI ES60601-1, IEC 60601-2-2, and 60601-2-18.

## Technological Characteristics

Bipolar electrosurgical vessel sealer; 5mm jaw size; 23/37/44cm shaft lengths. Non-stick coated jaws; stainless-steel knife. Energy source: MEP-1 electrosurgical generator. Connectivity: Three-prong ESU plug with smart chip. Sterilization: Ethylene oxide (SAL 10^-6). Single-use. Complies with AAMI/ANSI ES60601-1, IEC 60601-2-2, and 60601-2-18.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- LigaSure Maryland Jaw Sealer/Divider One-step Sealing, Nano-coated ([K170869](/device/K170869.md))
- LigaSure Maryland Jaw Thoracic Sealer/Divider, One-step sealing Nano-coated ([K181085](/device/K181085.md))

## Reference Devices

- CleanSeal Vessel Sealer 5mm ([K213354](/device/K213354.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 23, 2023

Comed Corporation Ali Abusaleh Senior Regulatory Affairs Specialist 525 French Road Utica, New York 13502

Re: K231732

Trade/Device Name: CleanSeal Advanced Bipolar Vessel Sealer Maryland Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: September 18, 2023 Received: September 18, 2023

Dear Ali Abusaleh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.10.23
11:57:16-04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### Device Name

CleanSeal Advanced Bipolar Vessel Sealer Maryland

#### Indications for Use (Describe)

CONMED CleanSeal Vessel Sealers are intended to be used in open and mimimally invasive procedures for the ligation and division of tissue bundles, lymphatics, and vessels up to and including 6 mm vessels (arteries, veins, pulmonary arteries, pulmonary veins). The handpiece is indicated for use in gynecological and general surgical procedures, including urologic, thoracic, and vascular. These procedures include, but are not limited to, hysterectomies, colectomies, Nissen fundoplication, and adhesiolysis. CleanSeal vessel sealer has not been shown to be effective for ubal sterilization or tubal coagulation for sterilization procedures. Do not use CleanSeal vessel sealer for these procedures.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## Section 6 - 510(k) Summary of Safety and Effectiveness

## CONMED CleanSeal Advanced Bipolar Vessel Sealer Maryland

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21CFR 807.92, ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness

## A. Submitter

ConMed Corporation 525 French Road Utica, NY 13502

## B. Company Contact

Ali AbuSaleh Senior Regulatory Affairs Specialist (708) 407-4324 Aliabusaleh@conmed.com Date prepared: 06/13/2023

## C. Device Name

| Proprietary Name:    | CleanSeal Advanced Bipolar Vessel Sealer Maryland            |
|----------------------|--------------------------------------------------------------|
| Catalog Numbers:     | CS-M5-23, CS-M5-37, CS-M5-44                                 |
| Common Name:         | Bipolar Vessel Sealing Device                                |
| Classification Name: | Electrosurgical cutting & coagulation device and accessories |
| Regulation Number:   | 878.4400                                                     |
| Product Code:        | GEI                                                          |
| Regulatory Class:    | II                                                           |
| Panel:               | General and Plastic Surgery                                  |

## D. Predicate Device

| Primary Predicate |                                                                    |
|-------------------|--------------------------------------------------------------------|
| Device Name:      | LigaSure Maryland Jaw Sealer/Divider One-step Sealing, Nano-coated |
| Company Name:     | Covidien LLC                                                       |
| 510(k):           | K170869                                                            |

#### Secondary Predicate

LigaSure Maryland Jaw Thoracic Sealer/Divider, One-step sealing Nano-coated Device Name: Covidien LLC Company Name: 510(k): K181085

## Reference Device

| Device Name:  | CleanSeal Vessel Sealer 5mm |
|---------------|-----------------------------|
| Company Name: | CONMED                      |
| 510(k):       | K213354                     |

{5}------------------------------------------------

# E. Device Description

CleanSeal Advanced Bipolar Vessel Sealer Maryland is designed to be used, with a compatible MEP-1 electrosurqical generator, in open and minimally invasive procedures for the ligation and division of tissue bundles, lymphatics, and vessels up to and including 6 mm arteries and veins. They are sterile, single use bipolar vessel sealing hand pieces featuring double action non-stick coated jaws, three shaft length options, a one-hand controlled handle with latchable lever. The instrument creates a seal by application of radio frequency (RF) electrosurgical energy to vascular structures or tissue bundles interposed between the jaws.

## F. Intended Use/ Indications for Use

CONMED CleanSeal Vessel Sealers are intended to be used in open and minimally invasive procedures for the ligation and division of tissue bundles, and vessels up to and including 6 mm vessels (arteries, veins, pulmonary arteries, pulmonary veins). The handpiece is indicated for use in gynecological and general surqical procedures, including urologic, thoracic, and vascular. These procedures include, but are not limited to, hysterectomies, colectomies, Nissen fundoplication, and adhesiolysis. CleanSeal vessel sealer has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use CleanSeal vessel sealer for these procedures.

## G. Technological Characteristics

CleanSeal Advanced Bipolar Vessel Sealers Maryland (23cm, 37cm and 44cm) connect to the CleanSeal/Bipolar Resection port on MEP-1 via a three-prong ESU plug with a smart chip containing unique parameters specific to the handpiece model. A hand actuated lever allows the user to open or close the instrument jaws during vessel sealing and cutting. An activation button is incorporated into the body of the handle. When connected to MEP-1 and the jaws are shut, the activation button can be pressed to allow the ESU to deliver energy to the CleanSeal Maryland vessel sealer. The cut trigger allows for the transection of tissue, when the jaws are shut, through the deployment of a stainless-steel knife. A rotator knob on the handle allows for 360° rotation of the shaft to position the jaw accurately on the targeted tissue or vessel. The CleanSeal Maryland comes in three shaft lengths (23 cm, 37 cm, and 44 cm). The device is single use.

## H. Performance Testing

CleanSeal Advanced Bipolar Vessel Sealers Maryland have been tested according to the applicable requirements listed in FDA guidance "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" and "Premarket Notification (510(k)) Submissions for Bipolar Electrosurgical Vessel Sealers for General Surgery". Benchtop and ex-vivo comparison testing, in-vivo simulated use testing, and chronic animal studies support that the CleanSeal Maryland is substantially equivalent to the Ligasure Maryland Jaw nano-coated (primary predicate), Ligasure Maryland Jaw thoracic nano-coated (secondary predicate) and the CleanSeal Vessel Sealer 5mm (reference device) in intended use, technology, and system performance.

Mechanical and electrical verification activities demonstrate the CleanSeal Advanced Bipolar Vessel Sealers Maryland complies with the applicable sections of AAMI/ANSI ES60601-1, IEC 60601-2-2 and 60601-2-18 . Risk management activities in accordance with ISO 14971 demonstrate the risks associated with the use of CleanSeal Maryland is mitigated to an acceptable level. Analyses of these activities conclude the benefits associated with the use of the CleanSeal Maryland outweigh the residual risks.

{6}------------------------------------------------

#### Biocompatibility Test -

The CleanSeal Advanced Bipolar Vessel Sealer Marylance with 2020 FDA Guidance of Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biological test results support that the CleanSeal Maryland met acceptance criteria for all biocompatibility test endpoints.

# J. Chronic Animal Study

A chronic animal study was conducted according to FDA guidance, "Premarket Notification (510(k)) Submissions for Bipolar Electrosurgical Vessel Sealers for General Surgery" Guidance for Industry and Food and Drug Administration Staff, issued on August 15, 2016; and "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" Guidance for Industry and Food and Drug Administration Staff, issued on March 9, 2020. The first test evaluated, in a porcine model, the performance of the CleanSeal Maryland with MEP-1 (Test system) in comparison to the LigaSure Maryland with Nanocoating and Valleylab FT10 ESU. (Primary predicate). After 21 days in the porcine model, the treatment of vascular structures (i.e., arteries, veins, and arterio-venous [AV] bundles) with the test system was determined to result in the effective sealing of treated vessels. The second test evaluated, in a canine model, the performance of the CleanSeal Maryland with MEP-1 (Test system) in comparison to the LigaSure Maryland thoracic with Nanocoating and Valleylab FT10 ESU, (Secondary predicate). After 21 days in the canine model, the treatment of vascular structures in the thoracic cavity (i.e., Pulmonary arteries, veins, and arterio-venous [A/V] bundles) with the test system was determined to result in the effective sealing of treated vessels.

# K. Shelf Life

CleanSeal Advanced Bipolar Vessel Sealer Maryland has a one-year shelf-life based on results of an accelerated aging study. The shelf-life study evaluated the device specifications and package integrity/sterile barrier. The device met all acceptance criteria to support a shelf-life of one year.

## L. Substantial Equivalence

The differences between the predicate devices and the proposed device do not raise any new risks of safety or efficacy. Supporting information per this premarket submission confirms that the CleanSeal Advanced Bipolar Vessel Sealer Maryland is safe and effective for its intended use and is substantially equivalent in design, intended use, principals of operation, and technical characteristics to the LigaSure Maryland Jaw Sealer/Divider One-step Sealing, Nano-coated, LigaSure Maryland Jaw Thoracic Sealer/Divider, One-step sealing Nano-coated and CleanSeal 5mm Vessel Sealer.

{7}------------------------------------------------

| Features                                                                 | Subject Device: CleanSeal<br>Advanced Bipolar Vessel Sealer<br>Maryland                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Reference Device: CleanSeal<br>Vessel Sealer 5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate Device:<br>LigaSure<br>Maryland Jaw Open<br>sealer/Divider, One-step sealing<br>Nano-coated<br>LF1923, LF1937 and LF1944                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Secondary Predicate:<br>LigaSure<br>Maryland Jaw Thoracic<br>Sealer/Divider, One-step sealing<br>Nano-coated<br>LF1930T                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Comparison/<br>Discussion                                                                                    |                           |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------|
| Model<br>Numbers<br>510(k)<br>Intended<br>Use/<br>Indications<br>for Use | CS-M5-23, CS-M5-37 and CS-<br>M5-44<br><br>K231732<br><br>CONMED CleanSeal Vessel<br>Sealers are intended to be used<br>in open and minimally invasive<br>procedures for the ligation and<br>division of tissue bundles,<br>lymphatics, and vessels up to and<br>including 6 mm vessels (arteries,<br>veins, pulmonary arteries,<br>pulmonary veins). The handpiece<br>is indicated for use in<br>gynecological and general<br>surgical procedures, including<br>urologic, thoracic, and vascular.<br>These procedures include, but<br>are not limited to, hysterectomies,<br>oophorectomies, colectomies,<br>Nissen fundoplication, and<br>adhesiolysis. CleanSeal vessel<br>sealer has not been shown to be<br>effective for tubal sterilization or<br>tubal coagulation for sterilization<br>procedures. Do not use<br>CleanSeal vessel sealer for these<br>procedures. | CS-M5-23, CS-M5-37 and CS-<br>M5-44<br><br>K213354<br><br>CONMED CleanSeal Vessel<br>Sealers are intended to be used in<br>open and minimally invasive<br>procedures for the ligation and<br>division of tissue bundles,<br>lymphatics, and vessels up to and<br>including 7 mm vessels (arteries<br>and veins). The handpiece is<br>indicated for use in gynecological<br>and general surgical procedures,<br>including urologic, thoracic, and<br>vascular. These procedures<br>include, but are not limited to,<br>hysterectomies, oophorectomies,<br>colectomies, Nissen<br>fundoplication, and adhesiolysis.<br>CleanSeal VS has not been<br>shown to be effective for tubal<br>sterilization or tubal coagulation<br>for sterilization procedures. Do not<br>use CleanSeal VS for these<br>procedures. | K170869<br><br>The LigaSure Sealer/Divider is a<br>bipolar electrosurgical instrument<br>intended for use in minimally<br>invasive or open surgical<br>procedures where ligation and<br>division of vessels, tissue<br>bundles, and lymphatics is<br>desired. The LigaSure Sealer/<br>Divider can be used on vessels<br>(arteries and veins) up to and<br>including 7 mm. It is indicated for<br>use in general surgery and such<br>surgical specialties as urologic,<br>vascular, thoracic, and<br>gynecologic. Procedures may<br>include, but are not limited to,<br>Nissen fundoplication,<br>colectomy, cholecystectomy,<br>adhesiolysis, hysterectomy,<br>oophorectomy, etc.<br><br>The LigaSure system has not<br>been shown to be effective for<br>tubal sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the<br>LigaSure system for these<br>procedures. | K181085<br><br>The LigaSure Sealer/Divider is a<br>bipolar electrosurgical instrument<br>intended for use in minimally invasive<br>or open surgical procedures where<br>ligation and division of vessels, tissue<br>bundles, and lymphatics is desired.<br>The LigaSure Sealer/ Divider can be<br>used on vessels (arteries, veins,<br>pulmonary arteries, pulmonary veins)<br>up to and including 7 mm. It is<br>indicated for use in general surgery<br>and such surgical specialties as<br>urologic, vascular, thoracic, and<br>gynecologic. Procedures may<br>include, but are not limited to, Nissen<br>fundoplication, colectomy,<br>cholecystectomy, adhesiolysis,<br>hysterectomy, oophorectomy, etc.<br><br>The LigaSure system has not been<br>shown to be effective for tubal<br>sterilization or tubal coagulation for<br>sterilization procedures. Do not use<br>the LigaSure system for these<br>procedures. | Similar                                                                                                      |                           |
|                                                                          | Polarity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Bipolar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Bipolar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Bipolar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Bipolar                                                                                                      |                           |
|                                                                          | Features                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Subject Device: CleanSeal<br>Advanced Bipolar Vessel Sealer<br>Maryland                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Reference Device: CleanSeal<br>Vessel Sealer 5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device:<br>LigaSure<br>Maryland Jaw Open<br>sealer/Divider, One-step sealing<br>Nano-coated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Secondary Predicate:<br>LigaSure<br>Maryland Jaw Thoracic<br>Sealer/Divider, One-step sealing<br>Nano-coated | Comparison/<br>Discussion |
|                                                                          | Model<br>Numbers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | CS-M5-23, CS-M5-37 and CS-<br>M5-44                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | CS-M5-23, CS-M5-37 and CS-<br>M5-44                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | LF1923, LF1937 and LF1944                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | LF1930T                                                                                                      |                           |
| 510(k)                                                                   | K231732                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | K213354                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K170869                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K181085                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                              |                           |
| Dimensional<br>Design                                                    | Jaw size: 5mm<br>Length of shaft: 23cm, 37cm and<br>44cm<br>Tip Shape: Curve similar to<br>Maryland curve;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Jaw size: 5mm<br>Length of shaft: 23cm, 37cm and<br>44cm<br>Tip Shape: Curve similar to<br>Maryland curve;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Jaw size: 5mm<br>Length of shaft: 23cm, 37cm and<br>44cm<br>Tip Shape: Curve similar to<br>Maryland curve;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Jaw size: 5mm<br>Length of shaft: 37 cm<br>Tip Shape: Curve similar to Maryland<br>curve;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Similar                                                                                                      |                           |
|                                                                          | Rated<br>Voltage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 350Vpk                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 350Vpk                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 288Vpk                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 288Vpk                                                                                                       | Similar                   |
| Shelf-life                                                               | 1 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 1 year                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similar                                                                                                      |                           |
| Sterilization<br>Method                                                  | Ethylene oxide, SAL of 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Ethylene oxide, SAL of 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Ethylene oxide, SAL of 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Ethylene oxide, SAL of 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                    |                           |
| Disposable<br>device                                                     | Single patient use device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Single patient use device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Single patient use device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Single patient use device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                    |                           |

{8}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K231732](https://fda-staging.innolitics.com/device/K231732)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com