The FemaSeed Intratubal Insemination is intended to introduce washed sperm or in vitro fertilized (IVF) embryos into the uterine ostium via ultrasound guidance.
Device Story
Device is a polymer-based transfer catheter system for assisted reproduction; includes curved transfer catheter with inflatable balloon, guide catheter, and inflation syringe with locking piston. Operated by trained healthcare provider in clinical setting. Procedure: guide catheter inserted transvaginally to fundus under ultrasound guidance; transfer catheter advanced 2 cm into uterine ostium; balloon inflated to secure position; sperm or embryos instilled; balloon deflated; device discarded. Ultrasound confirms balloon placement. System facilitates targeted delivery of reproductive samples to fallopian tube; minimizes specimen loss; ensures safe, controlled placement.
Clinical Evidence
Bench testing only. Studies included sterilization validation (ISO 11135-1, ISO 10993-7), package integrity (ASTM F2096, ASTM F88), transportation simulation (ASTM D4169), and biocompatibility (ISO 10993-5, -10). Performance testing confirmed balloon integrity, burst pressure, insertion depth, and specimen delivery. Biological safety validated via Human Sperm Survival Assay (≥80% motility) and Mouse Embryo Assay (≥80% blastocyst development).
Technological Characteristics
Materials: Pebax/Pellethane polymers. Components: curved transfer catheter with inflatable balloon, guide catheter, inflation syringe. Dimensions: Guide catheter 3.8 mm, transfer catheter 1.5 mm; balloon min 34F. Energy: Manual. Connectivity: None. Sterilization: Ethylene Oxide. Shelf life: 6 months.
Indications for Use
Indicated for patients undergoing assisted reproduction procedures requiring the introduction of washed sperm or IVF embryos into the uterine ostium.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
Predicate Devices
Echosight Jansen-Anderson Intratubal Transfer Set (K983591)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 22, 2023
Femasys, Inc. Christine Thomas SVP. Regulatory & Clinical Affairs 3950 Johns Creek Court. Suite 100 Suwanee, GA 30024
Re: K231730
> Trade/Device Name: FemaSeed Intratubal Insemination Regulation Number: 21 CFR§ 884.6110 Regulation Name: Assisted Reproduction Catheters Regulatory Class: II Product Code: MQF Dated: August 18, 2023 Received: August 22, 2023
Dear Christine Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231730
Device Name FemaSeed Intratubal Insemination
Indications for Use (Describe)
The FemaSeed Intratubal Insemination is intended to introduce washed sperm or in vitro fertilized (IVF) embryos into the uterine ostium via ultrasound guidance
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K231730 Page 1 of 5
## 510(k) Summary - K231730
#### I. SUBMITTER
| Applicant: | Femasys Inc. |
|--------------------|-----------------------------------------------------------|
| Applicant Address: | 3950 Johns Creek Court, Suite 100<br>Suwanee, GA 30024 |
| Phone: | +770-500-3910 x137 |
| Email: | CThomas@femasys.com |
| Contact Person: | Christine Thomas<br>Sr. VP, Regulatory & Clinical Affairs |
| Date Prepared: | September 21, 2023 |
#### II. DEVICE
| Trade Name: | FemaSeed Intratubal Insemination |
|--------------------|---------------------------------------|
| Common Name: | Transfer Catheter |
| Regulation Name: | Assisted Reproduction Catheters |
| Regulation Number: | 21 CFR 884.6110 |
| Regulatory Class: | II |
| Product Code: | MQF (Catheter, Assisted Reproduction) |
#### III. PREDICATE DEVICE
K983591 Intratubal Transfer Sets, currently legally marketed as Echosight Jansen-Anderson Intratubal Transfer Set.
The predicate device has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The FemaSeed Intratubal Insemination consists of polymer curved transfer catheter with balloon, polymer guide catheter and separate balloon inflation syringe. The balloon inflation syringe is a 3 mL syringe fitted with clips that allow for the piston to lock in place to maintain balloon inflation. When the clips are pinched, the piston releases, deflating the balloon. The device is intended to deliver washed sperm or in vitro fertilized embryos into the uterine ostium via ultrasound guidance by a licensed healthcare provider specifically trained in women's health.
Prior to insertion, the slider is moved back until the curved transfer catheter is fully retracted and contained within guide catheter. The preassembled guide catheter is then passed transvaginally through the cervix to the fundus, under ultrasound guidance.
{4}------------------------------------------------
Once in position, the preloaded transfer catheter is advanced approximately 2 cm into the uterine ostium and the balloon is inflated with the provided syringe. The balloon location in the uterine ostium is confirmed using ultrasound. The samples are then instilled into the uterine ostium of the fallopian tube, followed by deflating the balloon and discarding the device. The recommended volume for loading washed sperm should not exceed 1 mL and 50-100 µL when loading embryos.
#### V. INDICATIONS FOR USE
The FemaSeed Intratubal Insemination is intended to introduce washed sperm or in vitro fertilized (IVF) embryos into the uterine ostium via ultrasound guidance.
### VI. COMPARISON OF THE INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICES
The following table compares the FemaSeed Intratubal Insemination Transfer Catheters to the predicate device.
| Comparison Item | Subject Device –<br>K231730 | Predicate Device<br>– K983591 | Comparison |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Classification | 21 CFR 884.6110 | 21 CFR 884.6110 | Same |
| Product code | MQF | MQF | Same |
| Indications for Use | The FemaSeed<br>Intratubal<br>Insemination is<br>intended to introduce<br>washed sperm or in<br>vitro fertilized (IVF)<br>embryos into the<br>uterine ostium via<br>ultrasound guidance. | The intratubal<br>transfer sets are<br>used to inject<br>either sperm,<br>gametes or<br>embryos into the<br>uterine ostium of<br>the fallopian tube<br>via ultrasound<br>guidance. The<br>device is sterile<br>and intended for<br>one time use. | Similar |
| Design Features | The FemaSeed<br>Intratubal<br>Insemination consists<br>of a polymer curved<br>transfer catheter,<br>polymer guide<br>catheter and a<br>separate balloon | The Intratubal<br>Transfer Set<br>consists of a<br>polymer curved<br>guide catheter,<br>stainless steel,<br>malleable<br>obturator, stainless | Different |
| Comparison Item | Subject Device –<br>K231730 | Predicate Device –<br>K983591 | Comparison |
| | inflation syringe. The<br>transfer catheter<br>includes an inflatable<br>balloon. | steel wire guide<br>and a polymer<br>transfer catheter. | |
| Catheter Assembly | Preassembled<br>catheters, separate<br>syringe | Separate catheters,<br>separate stainless-<br>steel obturator | Different |
| Outer and Inner<br>Diameters | Guide catheter 3.8 mm<br>Transfer catheter 1.5 mm<br>There are two (2) lumens of the transfer<br>catheter:<br>Specimen lumen ID 0.036" +0.000 / -0.003"Balloon inflation lumen ID 0.010" +0.000 / -0.002"The dimension of the<br>balloon when inflated is a minimum of 34F | Guide catheter 1.9 mm<br>Transfer catheter 1 mm | Different |
| Materials | Polymer<br>(Pebax/Pellethane) | Polymer<br>Stainless steel | Different |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Single Use | Yes | Yes | Same |
| Shelf Life | 6-months | 36-months | Different |
| Mouse Embryo Assay | 2-Cell MEA: ≥ 80%<br>embryos developed to<br>expanded blastocyst<br>at 72 hours. | Not available<br>publicly | Not<br>applicable |
| Endotoxin | < 20 EU/device | Not available<br>publicly | Not<br>applicable |
| HSSA | ≥80% of control<br>motility at 24 hours<br>following 30-minute<br>exposure. | Not available<br>publicly | Not<br>applicable |
Table 1. Intended use and technological characteristics comparison
{5}------------------------------------------------
{6}------------------------------------------------
The subject and predicate devices have similar indications for use statements and have the same intended use. As shown in the table above, the subject and predicate device have some technological characteristics that are the same or similar. However, there are differences in design, materials and shelf-life. These differences do not raise different questions of safety and effectiveness.
#### SUMMARY OF NON-CLINICAL PERFORMANCE DATA VII.
The following studies have been performed to support substantial equivalence to the predicate device:
- . Ethylene Oxide Sterilization Validation testing per:
- A ISO 11135-1:2014
- A AAMI TIR 28:2016
- A ISO 10993-7: 2008
- . Package Integrity testing:
- A Visual inspection
- A Bubble Leak test per ASTM F2096-11
- A Seal Strength testing per ASTM F88/ F88M-15
- Transportation Simulation testing per ASTM D4169-14
- Biocompatibility studies conducted in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process. " Testing included the following assessments:
- Cytotoxicity per ISO 10993-5: 2009 A
- A Sensitization ISO 10993-10: 2010
- A Irritation per ISO 10993-10: 2010
Testing showed the device material to be non-cytotoxic, non-sensitizing, and nonirritating.
- Endotoxin testing per USP <85> and AAMI/ANSI ST72:2019 ● Specification: < 20 EU/device A
- Bench performance studies before and after accelerated aging to the equivalent of 6-months of real-time aging in accordance with ASTM F1980-16 demonstrated that all predetermined acceptance criteria were met in the following tests:
- A Appearance
- A Dimensional analysis
- A Mechanical performance
- A Device interface compatibility
- Balloon integrity, maximum burst pressure analysis, maximum cycles for stop >
{7}------------------------------------------------
lock clips to maintain balloon inflation
- Maximum extension force for transfer and guide catheters A
- > Depth of insertion and transfer catheter advancement testing
- Aspiration and delivery testing >
- > Minimize lost specimen during aspiration and delivery
- > Indicate when transfer catheter is fully advanced or retracted
- > Guide catheter insertion without kinking and minimize patient discomfort or risk of injury
- > Human Sperm Survival Assay (HSSA) demonstrating ≥80% of control motility at 24 hours following 30-minute exposure.
- > Mouse Embryo Assay (MEA) per the 2021 FDA guidance Mouse Embryo Assay for Assisted Reproduction Technology Devices: Specification - 2-Cell MEA: ≥ 80% embryos developed to expanded blastocyst at 72 hours.
## VIII. CONCLUSIONS
The results of the testing described above demonstrate that the FemaSeed Intratubal Insemination transfer catheters are as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.