← Product Code [GCJ](/productcode/GCJ) · K231725

# Laparoscope (4KA0, 4KA0R, 4KA30, 4KA30R, 4KA45, 4KA45R) (K231725)

_Hunan Endoso Life Technology Co., Ltd. · GCJ · Oct 16, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K231725

## Device Facts

- **Applicant:** Hunan Endoso Life Technology Co., Ltd.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Oct 16, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

It is used in medical institutions for observation and imaging in abdominal surgery by entering the human body through the wound.

## Device Story

Laparoscope is a tubular optical instrument for visualization/imaging of anatomy during minimally invasive abdominal surgery. Device transmits visible and near-infrared light to illuminate surgical site; relays optical image to external camera system for processing and display. Used in hospital environments by trained surgeons. Provides visual feedback to assist in diagnostic and therapeutic procedures; enables minimally invasive access via small skin punctures. Benefits include reduced surgical trauma compared to open procedures.

## Clinical Evidence

Bench testing only. Performance verified via optical and mechanical testing per ISO 8600 series. Biocompatibility testing per ISO 10993 (cytotoxicity, irritation, sensitization, pyrogenicity, systemic toxicity). Electrical safety/EMC per IEC 60601-1/2-18. Reprocessing validated per AAMI TIR 12/30 and ISO 17664/17665.

## Technological Characteristics

Reusable tubular optical instrument. Materials: Optical glass, sapphire, epoxy glue. Complies with ISO 10993-1. Sterilization: Autoclave, STERRAD 100NX. Standards: ISO 8600-1/3/4/7, IEC 60601-1/2-18, ISO 17664/17665. No internal software or electronic processing; passive optical relay device.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- O-Mec Laparoscopes 690 Series ([K201151](/device/K201151.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 16, 2023

Hunan Life Technology Co., Ltd. Fu Wang Official Correspondent 4th floor, Building 10, Innovation and Entrepreneurship Center, 31 Dongfeng Road, Heping Street, Xiangtan Economic D HuNan, 411199 China

Re: K231725

Trade/Device Name: Laparoscope (4KA0, 4KA0R, 4KA30R, 4KA30R, 4KA45, 4KA45R) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: September 11, 2023 Received: September 11, 2023

Dear Fu Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark
Digitally signed by
Mark Trumbore -S
Trumbore - Date: 2023.10.16
08:26:23 -04'00'

Mark Trumbore, Ph.D. Assistant Director

{2}------------------------------------------------

DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K231725

Device Name

Laparoscope (4KA04KA0R4KA304KA30R4KA454KA45R)

Indications for Use (Describe)

It is used in medical institutions for observation and imaging in abdominal surgery by entering the human body through the wound.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

### 510(k) Summary

| Date Prepared:                | April. 11st, 2023                                                                                                                                                                                            |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:                 | Hunan Endoso Life Technology Co., Ltd.<br>4th floor, Building 10, Innovation and Entrepreneurship Center, 31<br>Dongfeng Road, Heping Street, Xiangtan Economic Development<br>District, HuNan, China 411199 |
| Contact person:               | Wang Fu<br>Title: Director<br>Tel: (86) 0731-52331999<br>Email: services@endoso.com.cn<br>Or wangfu@endoso.com.cn                                                                                            |
| Identification of the Device: |                                                                                                                                                                                                              |
| Proprietary/Trade Name:       | Laparoscope                                                                                                                                                                                                  |
| Model:                        | 4KA0、4KA0R、4KA30、4KA30R、4KA45、4KA45R                                                                                                                                                                         |
| Common name:                  | Laparoscope                                                                                                                                                                                                  |
| Classification Name:          | Laparoscope, General & Plastic Surgery                                                                                                                                                                       |
| Regulatory Number:            | 21 CFR Part 876.1500                                                                                                                                                                                         |
| Product Code:                 | GCJ                                                                                                                                                                                                          |
| Device Class:                 | Class II                                                                                                                                                                                                     |
| Review Panel:                 | General & Plastic Surgery                                                                                                                                                                                    |

### Identification of the Legally Marketed Predicate Device:

| Trade Name:              | O-Mec Laparoscopes 690 Series              |
|--------------------------|--------------------------------------------|
| Classification Name:     | Laparoscope, General & Plastic Surgery     |
| Regulatory Number:       | 21 CFR Part 876.1500                       |
| Product Code:            | GCJ                                        |
| Device Class:            | Class II                                   |
| Review Panel:            | General & Plastic Surgery                  |
| Submitter/510(k) Holder: | Qingdao O-Mec Medical Technology Co., Ltd. |
| Clearance:               | K201151                                    |

This predicate has not been subject to a design-related recall.

### Device Description:

The Laparoscope (Model: 4KA0、4KA0R、4KA30、4KA30R、4KA45、4KA45R) is used in medical institutions. Laparoscopes are tubular optical instruments used to visualize or image a patient's anatomy during minimally invasive, endoscopic procedures for examination, diagnosis or therapy. The Laparoscope transmits light in both the visible and near infrared spectrum to illuminate and image the anatomy, then

{5}------------------------------------------------

forms and relays the image of the surgical site to a camera system for processing and display.

### Indications for Use:

It is used in medical institutions for observation and imaging in abdominal surgery by entering the human body through the wound.

## Substantial Equivalence:

The Laparoscope employs the same fundamental scientific technology as its predicate device, as below table:

|                                                          | Subject Device                                                                                                                                                                                                                                                                                                                   | Predicate Device                                                                                                                                                                                                                                                                                                                                                        | Comparison |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| K Number                                                 | --                                                                                                                                                                                                                                                                                                                               | K201151                                                                                                                                                                                                                                                                                                                                                                 |            |
| Name                                                     | Laparoscope                                                                                                                                                                                                                                                                                                                      | O-Mec 690 Series Laparoscopes                                                                                                                                                                                                                                                                                                                                           | /          |
| Model                                                    | 4KA0、4KA0R、4KA30、4KA30R、4KA45、4KA45R                                                                                                                                                                                                                                                                                             | 690-331000H, 690-331030H, 690-300500H, 690-300530H                                                                                                                                                                                                                                                                                                                      | /          |
| Manufacturer                                             | Hunan Endoso Life Technology Co., Ltd.                                                                                                                                                                                                                                                                                           | QINGDAO O-MEC MEDICAL TECHNOLOGY CO., LTD.                                                                                                                                                                                                                                                                                                                              | /          |
| Classification Name                                      | Laparoscope, General & Plastic Surgery                                                                                                                                                                                                                                                                                           | Laparoscope, General & Plastic Surgery                                                                                                                                                                                                                                                                                                                                  | Same       |
| Product Code                                             | GCJ                                                                                                                                                                                                                                                                                                                              | GCJ                                                                                                                                                                                                                                                                                                                                                                     | Same       |
| Indications For Use                                      |                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                         |            |
| Intended Use                                             | It is used in medical institutions for observation and imaging in abdominal surgery by entering the human body through the wound.                                                                                                                                                                                                | The O-Mec Laparoscopes 690 Series (Models, 90-331000H, 690-331030H, 690-300500H, 690-300530H) are intended to be used by surgeons in diagnostic and therapeutic procedures. Laparoscopic minimally invasive procedures are performed in the abdominal cavity by means of small skin punctures that allow the insertion of the laparoscope and laparoscopic instruments. | Similar    |
| Working place/User                                       | Use in a hospital environment by trained surgical physicians                                                                                                                                                                                                                                                                     | Use in a hospital environment by trained                                                                                                                                                                                                                                                                                                                                | Same       |
|                                                          | Subject Device                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>K201151                                                                                                                                                                                                                                                                                                                                             | Comparison |
| K Number                                                 | who are familiar with<br>endoscopic procedures.                                                                                                                                                                                                                                                                                  | surgical physicians who are<br>familiar with endoscopic<br>procedures.                                                                                                                                                                                                                                                                                                  |            |
| Population                                               | Adult                                                                                                                                                                                                                                                                                                                            | Adult                                                                                                                                                                                                                                                                                                                                                                   | Same       |
| Physical Characteristics                                 |                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                         |            |
| Maximum width<br>of insertion section                    | $Φ$ 10mm                                                                                                                                                                                                                                                                                                                         | $Φ$ 10.1mm                                                                                                                                                                                                                                                                                                                                                              | Similar    |
| Working length                                           | 345mm±3%.                                                                                                                                                                                                                                                                                                                        | 334mm                                                                                                                                                                                                                                                                                                                                                                   | Similar    |
| Optical Characteristics                                  |                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                         |            |
| Field of view                                            | 80°±15%                                                                                                                                                                                                                                                                                                                          | 75°                                                                                                                                                                                                                                                                                                                                                                     | Similar    |
| Direction of view                                        | 30                                                                                                                                                                                                                                                                                                                               | 30°                                                                                                                                                                                                                                                                                                                                                                     | Same       |
| Clear Observation Range                                  | 3mm~200mm                                                                                                                                                                                                                                                                                                                        | 25-150mm                                                                                                                                                                                                                                                                                                                                                                | Similar    |
| Resolution power                                         | 7lp/mm                                                                                                                                                                                                                                                                                                                           | 7lp/mm                                                                                                                                                                                                                                                                                                                                                                  | Same       |
| Lighting Edge Uniformity                                 | Inspection within the effective<br>depth of field, the<br>illumination spot should be<br>filled with all direction of<br>view, and the illuminance at<br>90% of the maximum<br>direction of view should be<br>uniform. Select four<br>orthogonal azimuth in this<br>direction of view for testing.<br>Uniformity should be ≤25%. | Inspection within the<br>effective depth of field, the<br>illumination spot should be<br>filled with all direction of<br>view, and the illuminance at<br>90% of the maximum<br>direction of view should be<br>uniform. Select four<br>orthogonal azimuth in this<br>direction of view for testing.<br>Uniformity should be ≤25%.                                        | Same       |
| Patient Contacting Materials                             |                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                         |            |
| General material<br>type of main<br>patient-contact part | Compliance with ISO10993-1                                                                                                                                                                                                                                                                                                       | Compliance with<br>ISO10993-1                                                                                                                                                                                                                                                                                                                                           | Same       |
| Duration and<br>type of contact                          | "Surface -Mucosal<br>Membrane" with a contact<br>duration of "Limited (< 24<br>hours)"                                                                                                                                                                                                                                           | "Surface -Mucosal<br>Membrane" with a contact<br>duration of "Limited (< 24<br>hours)"                                                                                                                                                                                                                                                                                  | Same       |
|                                                          | Subject Device                                                                                                                                                                                                                                                                                                                   | Predicate Device                                                                                                                                                                                                                                                                                                                                                        | Comparison |
| K Number                                                 | --                                                                                                                                                                                                                                                                                                                               | K201151                                                                                                                                                                                                                                                                                                                                                                 |            |
| ng Materials                                             | Optical Glass<br>Sapphire<br>Epoxy Glue                                                                                                                                                                                                                                                                                          | Stainless Steel 316L<br>Optical Glass<br>Sapphire<br>Epoxy Glue                                                                                                                                                                                                                                                                                                         |            |
| Sterilization Methods                                    |                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                         |            |
| Number of Users                                          | Reusable                                                                                                                                                                                                                                                                                                                         | Reusable                                                                                                                                                                                                                                                                                                                                                                | Same       |
| Cleaning                                                 | Manual and Automated                                                                                                                                                                                                                                                                                                             | Manual and Automated                                                                                                                                                                                                                                                                                                                                                    | Same       |
| Disinfection                                             | Manual and Automated                                                                                                                                                                                                                                                                                                             | Manual and Automated                                                                                                                                                                                                                                                                                                                                                    | Same       |
| Sterilization                                            | Autoclave,<br>STERRAD® 100NX                                                                                                                                                                                                                                                                                                     | Autoclave,<br>STERRAD® 100NX                                                                                                                                                                                                                                                                                                                                            | Same       |
| Sterility Assurance Level                                | 10-6                                                                                                                                                                                                                                                                                                                             | 10-6                                                                                                                                                                                                                                                                                                                                                                    | Same       |
| Biocompatibility                                         | No Cytotoxicity                                                                                                                                                                                                                                                                                                                  | No Cytotoxicity                                                                                                                                                                                                                                                                                                                                                         | Same       |
|                                                          | No Irritation to Skin                                                                                                                                                                                                                                                                                                            | No Irritation to Skin                                                                                                                                                                                                                                                                                                                                                   | Same       |
|                                                          | No significant evidence of<br>sensitization                                                                                                                                                                                                                                                                                      | No significant evidence of<br>sensitization                                                                                                                                                                                                                                                                                                                             | Same       |
|                                                          | No pyrogen                                                                                                                                                                                                                                                                                                                       | No pyrogen                                                                                                                                                                                                                                                                                                                                                              | Same       |
|                                                          | No acute Systemic Toxicity                                                                                                                                                                                                                                                                                                       | No acute Systemic Toxicity                                                                                                                                                                                                                                                                                                                                              | Same       |

{6}------------------------------------------------

#### 510(k) Premarket Notification Submission Laparoscope

{7}------------------------------------------------

### 510(k) Premarket Notification Submission Laparoscope

# Summary of Testing:

| Test                         | Method                                                                                                                                                | Result |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Electrical Safety and<br>EMC | • ANSI/AAMI ES:60601-1:2005/A2:2010<br>• IEC 60601-2-18:2009                                                                                          | PASS   |
| Packaging                    | • ASTM D4169:2016                                                                                                                                     | PASS   |
| Biocompatibility             | • ISO 10993- 1:2018<br>• ISO 10993-5:2009<br>• ISO 10993- 10:2010<br>• ISO 10993- 11:2017<br>• FDA Guidance: Use of International Standard ISO10993-1 | PASS   |

{8}------------------------------------------------

#### 510(k) Premarket Notification Submission Laparoscope

| Cleaning,<br>Disinfection and<br>Sterilization | • AAMI TIR 12:2020<br>• AAMI TIR 30: 2011/(R)2016<br>• ISO 17664- 1:2021<br>• ISO 17665-1: 2006<br>FDA Guidance: Reprocessing Medical Devices in Health<br>Care Settings: Validation Methods and Labeling          | PASS |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Performance -<br>Bench                         | • Optical and mechanical verification in accordance with device<br>input specifications and comparison to predicate device<br>• ISO 8600- 1:2015<br>• ISO 8600-3: 2019<br>• ISO 8600-4: 2014<br>• ISO 8600-7: 2012 | PASS |
| Design Validation                              | • IEC 62366-1:2020<br>• In accordance with device user needs                                                                                                                                                       | PASS |

## Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the subject device, the Laparoscope is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K231725](https://fda-staging.innolitics.com/device/K231725)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
