Holmium Medical Laser

K231718 · Allengers Global Healthcare Private Limited · GEX · Aug 18, 2023 · General, Plastic Surgery

Device Facts

Record IDK231718
Device NameHolmium Medical Laser
ApplicantAllengers Global Healthcare Private Limited
Product CodeGEX · General, Plastic Surgery
Decision DateAug 18, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Holmium Medical Laser and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynecology, ENT, Pulmonary Surgery and General Surgery.

Device Story

Holmium Medical Laser (CTH:YAG) uses pulsed 2100nm laser energy delivered via optical fiber for soft tissue surgery; includes laser source, power electronics, control electronics, and cooling system. Operated by physicians in OR/clinic settings for incision, ablation, and lithotripsy. Triggered by footswitch; software manages parameters and user selections. Provides precise tissue interaction; benefits include minimally invasive access, reduced bleeding via coagulation, and effective stone fragmentation. Output power ranges up to 100W depending on model (BLAZE-prime, BLAZE).

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 standards for electrical, laser safety, and electromagnetic compatibility. Software verification and validation performed per FDA guidance.

Technological Characteristics

Pulsed CTH:YAG laser; 2100nm wavelength; 3-80Hz frequency; 0.1J-5J pulse energy; 35W-100W power. Aiming beam: 532nm DPSS, <5mW, Class 3R. Optical fiber delivery. Software-controlled functions. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 14971.

Indications for Use

Indicated for surgical incision, resection, ablation, vaporization, coagulation, and haemostasis of soft tissue in urology (including BPH, prostate surgery, bladder/urethral tumors), lithotripsy (urinary/kidney calculi), gastroenterology (polyps, ulcers, GI surgery), arthroscopy (soft/cartilaginous tissue, spinal decompression), gynecology, ENT (sinus surgery), pulmonary, and general surgery.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in a clear, sans-serif font. The FDA logo is a recognizable symbol of the agency's mission to protect and promote public health. August 18, 2023 Allengers Global Healthcare Private Limited Harpreet Singh Director-Technical Room No.5, Khasra no. 79, Bhankarpur, Mubarakpur Road Derabassi, District- Mohali, Punjab 140507 India Re: K231718 Trade/Device Name: Holmium Medical Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 5, 2023 Received: June 13, 2023 Dear Harpreet Singh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Digitally signed by Tanish Tanisha L. 0ate: 2023.08.18 15:22:10 Hithe -S Tanisha Hithe, MS, MHS Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K231718 Device Name Holmium Medical Laser The Holmium Medical Laser and its fiber optic delivery system are in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectory, Gynecology, ENT, Pulmonary Surgery and General Surgery. Urology Open and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) including: - · Urethral Strictures - · Bladder Neck Incisions (BNI) - · Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors, - · Ablation of Benign Prostatic Hypertrophy (BPH), - · Transurethral incision of the prostate (TUIP) - · Holmium Laser Resection of the Prostrate (HoLRP) - · Holmium Laser Enucleation of the Prostate (HoLEP) - · Holmium laser Ablation of the Prostate (HoLAP) - · Condylomas - · Lesions of external genitalia Lithotripsy and Percutaneous Urinary Lithotripsy · Endoscopic fragmentation of ureteral, bladder and renal calculi including cystine, calcium oxalate, - monohydrate and calcium oxalate - · dehydrate stones. - · Endoscopic fragmentation of kidney calculi. - · Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed. #### Gastroenterology Open and endoscopic Gastroenterology surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) including: - · Appendectomy - Polyps - · Biopsy - · Gall Bladder calculi - · Biliary/Bile duct calculi - · Ulcers - · Gastric ulcers - · Duodenal ulcers - · Non Bleeding Ulcers - · Pancreatitas - · Haemorrhoids - · Cholecystectomy - · Benign and Malignant Neoplasm - · Angiodysplasia - · Colorectal cancer {3}------------------------------------------------ · Telangiectasias - · Telangiectasias of the Osler-Weber-Renu disease - Vascular Malformation - · Gastritis - · Esophagitis - · Esophageal ulcers - Varices - · Colitis - · Mallory-Weiss tear - · Gastric Erosions #### Arthroscopy Arthroscopy/Orthopaedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including: - · Ligament and tendon Release - · Contouring and sculpting of articular surfaces - · Capsulectomy in the Knee - · Chondreplasty in the Knee - · Debridement of inflamed synovial tissue - Chondromalacia Ablation - · Chondromalacia and tears - · Plica Removal - · Meniscectomy - · Loose Body Debridement - · Lateral retinecular release Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including · Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-S1 lumbar discs, including Foraminoplasty Percutaneous Cervical Disc Decompression/Discectomy Percutaneous Thoracic Disc Decompression/Discectomy Gynaecology Open and laparoscopic gynaecological surgery (incision, excision, ablation, vaporization, coagulation and haemostasis) of soft tissue #### ENT Endoscopic endonasal surgery (incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and cartilage) including: - · Endonasal/sinus Surgery - · Partial turbinectomy - · Polypectomy - · Dacryocystorhinostomy - · Frontal Sinusotomy - · Ethmoidectomy - · Maxillary antrostomy - · Functional endoscopic sinus surgery #### Pulmonary Surgery Open and endoscopic pulmonary surgery (cutting, vaporization, incision and coagulation of soft tissue) General Surgery Open, laparoscopic and endoscopic surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) including: - · Appendectomy {4}------------------------------------------------ · Skin incision - · Excision of external and internal lesions - · Complete or partial resection of internal organs, turnors and lesions - · Biopsy Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # SECTION 5 # 510(k) Summary K231718 {6}------------------------------------------------ ## 510(k) SUMMARY ## (K231718) This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92 | Applicant /Manufacturer<br>Name and Address | Allengers Global Healthcare Private Limited<br>Room No.5, Khasra no. 79, Bhankarpur, Mubarakpur Road, Derabassi,<br>District- Mohali, 140507, Punjab (India) | | | |---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | | +911762-272600 | | | | 510(K) Contact person | Harpreet Singh<br>Address: Room No.5, Khasra no. 79, Bhankarpur, Mubarakpur Road,<br>Derabassi, District- Mohali, 140507, Punjab (India)<br>Contact Number: +919876615337<br>Email ID: harpreet.singh@allengers.net | | | | Date Prepared | 5th May, 2023 | | | | Device/Trade Name | Holmium Medical Laser | | | | Models Names | BLAZE-prime, BLAZE | | | | Regulation Class | Class II | | | | Regulation Name | Laser Surgical Instrument For Use In General And Plastic Surgery And<br>In Dermatology | | | | Regulation Number | 21 CFR 878.4810 | | | | Product Code | GEX | | | | Panel | General and Plastic Surgery | | | | Predicate Devices | Cyber Ho 100 (K192600)-Quanta system spa | | | | Reference Devices | Empower H65(K180423), Litho(K163009)- Quanta system spa<br>Lumenis Pulse 120H (K170121)-Lumenis Ltd. | | | {7}------------------------------------------------ ## Device Description The devices belonging to Holmium laser family are laser devices based on a Holmium laser source. The main parts (subsystems) of the device are the Holmium laser source, the power electronics, the optical delivery system, the control electronics and the cooling system. Specific software controls the device functions and allows the user selections. Laser emission is triggered by a footswitch. | Feature | Subject Device | | Predicate Device | Justification<br>for<br>differences | |--------------------------|----------------------------------------------|----------------|-----------------------------------|-------------------------------------| | Manufacturer | Allengers Global Healthcare Pvt.<br>Ltd. | | Quanta System SPA | | | Models | BLAZE-prime<br>(100W,65W) | BLAZE<br>(35W) | Cyber Ho 100 | | | 510(k) | K231718 | | K192600 | | | Manufacturer | Allengers Global Healthcare Pvt.<br>Ltd. | | Quanta System SPA | | | Device/Trade<br>Name | Holmium Medical Laser | | Cyber Ho 100 | | | Laser Medium | Pulsed Holmium laser (CTH:YAG) | | Pulsed Holmium laser<br>(CTH:YAG) | Same | | Wavelength | 2100nm | | 2100nm | Same | | Emission | Pulsed | | Pulsed | Same | | Pulse Duration | 95- 1500μs | | 50 - 1100 μs | Similar | | Pulse Frequency | 3-80HZ (100W)<br>3-60HZ (65W) | 3-30HZ | 3-80 Hz | Same | | Pulse Energy | 0.1J to 5J | | Upto 5J | Same | | Maximum<br>Average Power | Up to 100 W<br>(100W)<br>Up to 65 W<br>(65W) | Up to 35 W | Up to 105W | Similar | | Delivery System | Optical Fiber | | Optical Fiber | Same | | Aiming Beam | | | | | | Type | DPSS | | DPSS | Same | | Power | Power <5mW | | Power <5mW | Same | | Wavelength | Wavelength 532nm | | Wavelength 532nm | Same | | Laser Class | Class 3R | | Class 3R | Same | #### Summary of Technical Characteristic ### Intended Use/ Indications for Use: The Holmium Medical Laser and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynecology, ENT, Pulmonary Surgery and General Surgery {8}------------------------------------------------ #### Urology Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including: - Urethral Strictures - Bladder Neck Incisions (BNI) ● - Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors, ● - Ablation of Benign Prostatic Hypertrophy (BPH), - Transurethral incision of the prostate (TUIP) - Holmium Laser Resection of the Prostrate (HoLRP) - Holmium Laser Enucleation of the Prostate (HoLEP) - Holmium laser Ablation of the Prostate (HoLAP) . - Condylomas ● - . Lesions of external genitalia #### Lithotripsy and Percutaneous Urinary Lithotripsy - Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate - dehydrate stones. - Endoscopic fragmentation of kidney calculi. ● - Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed. . #### Gastroenterology Open and endoscopic Gastroenterology surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) including: - Appendectomy ● - Polyps ● - Biopsy ● - Gall Bladder calculi ● - Biliary/Bile duct calculi - Ulcers - Gastric ulcers ● - Duodenal ulcers ● - Non Bleeding Ulcers - Pancreatitas - . Haemorrhoids - . Cholecystectomy - Benign and Malignant Neoplasm - Angiodysplasia - Colorectal cancer - Telangiectasias - Telangiectasias of the Osler-Weber-Renu disease - Vascular Malformation - Gastritis ● - Esophagitis - Esophageal ulcers - Varices ● - Colitis ● - Mallory-Weiss tear ● - Gastric Erosions ● {9}------------------------------------------------ #### Arthroscopy Arthroscopy/Orthopaedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including: - Ligament and tendon Release - . Contouring and sculpting of articular surfaces - Capsulectomy in the Knee ● - Chondreplasty in the Knee - Debridement of inflamed synovial tissue - Chondromalacia Ablation - Chondromalacia and tears - . Plica Removal - Meniscectomy ● - Loose Body Debridement ● - Lateral retinecular release ● Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including - Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-SI lumbar discs, ● including Foraminoplasty Percutaneous Cervical Disc Decompression/Discectomy Percutaneous Thoracic Disc Decompression/Discectomy #### Gynaecology Open and laparoscopic gynaecological surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) of soft tissue #### ENT Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and cartilage) including: - Endonasal/sinus Surgery ● - Partial turbinectomy - Polypectomy ● - Dacryocystorhinostomy ● - Frontal Sinusotomy ● - Ethmoidectomy ● - Maxillary antrostomy ● - . Functional endoscopic sinus surgery #### Pulmonary Surgery Open and endoscopic pulmonary surgery (cutting, vaporization, excision and coagulation of soft tissue) #### General Surgery Open, laparoscopic and endoscopic surgery (incision, excision, ablation, vaporization, coagulation and haemostasis) including: - Appendectomy ● - . Skin incision - Excision of external and internal lesions ● - Complete or partial resection of internal organs, tumors and lesions ● - Biopsy {10}------------------------------------------------ #### Accessories The device is intended to be used together with delivery optical fiber that separately received a FDA clearance for an intended use. #### Performance Testing The Subject device performs testing in accordance with the following recognized consensus standards - Risk analysis activities, in compliance with the requirements of ISO 14971: 2019 Medical devices - Application of Risk Management to Medical Devices. - Electrical and laser safety and electromagnetic compatibility tests, in compliance with: ● IEC 60601-1:2005+ CORR.1:2006+CORR. 2:2007 +AM1:2012 Medical Electrical O Equipment - Part 1: General Requirements for Basic Safety and Essential Performance. - O IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic Compatibility - Requirements and Tests. - IEC 60601-2-22: 2007+A1:2012 Medical Electrical Equipment -Part 2-22: Particular O Requirements For Basic Safety And Essential Performance of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment. - IEC 60825-1:2014 Safety of laser products Part 1: Equipment classification and O requirements. - . Software verification and validations, in compliance with FDA Guidance for the Premarket Submissions for Software Contained in Medical Devices (issue on 2005). The tests verified that the subject Holmium Medical Laser performs according to its specifications. #### Comparison with predicate device The subject device technological characteristics and indication for use are same to the predicate device and its output power ranges are similar to those of the predicate devices. Other difference between the subject device and predicate device is not raise new types of question regarding the subject device's safety and efficacy. ### Conclusion The Holmium Medical Laser do not introduces any new indications for use, nor does the use of the systems result in any new potential hazard. The Holmium Medical Laser, the subject device is substantially equivalent to the predicate and reference device Cyber Ho100 (K192600), Empower65 (K180423), Litho (K163009), Lumenis pulse120H (K170121). The intended use, the design principle and the applicable standards for the subject device to those of the predicate and reference device. Some characteristics, for example, their appearance, user interface and the physical dimension are different. However performance test result demonstrates that these differences do not raise any new question of safety and effectives. Therefore, it is Allengers Global Healthcare Pvt. Ltd. option that the subject device appears to be as safe and effective as the predicate and reference device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...