K231691 · TeDan Surgical Innovations, Inc. · GXZ · Jun 29, 2023 · Neurology
Device Facts
Record ID
K231691
Device Name
Phantom XL Insulated Dilators
Applicant
TeDan Surgical Innovations, Inc.
Product Code
GXZ · Neurology
Decision Date
Jun 29, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1350
Device Class
Class 2
Indications for Use
Phantom XL Insulated Dilators are indicated for use during surgery of the spine to deliver an electrical stimulus to the tissues and nerves at the operative site, to assist in locating those nerves at risk during the surgical procedure.
Device Story
Phantom XL Insulated Dilators are single-use, sterile, monopolar instruments used during spinal surgery. The device delivers electrical stimulation to tissues and nerves at the operative site to assist surgeons in identifying nerves at risk during procedures. The device is operated by a surgeon in an intraoperative setting. The output is the delivery of electrical current, which the surgeon uses to monitor neurological responses and guide surgical navigation. The device is provided in four diameter sizes and is designed to facilitate safe surgical access by helping avoid nerve injury.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including packaging integrity, seal strength, shipping container performance, and accelerated aging studies to validate the 5-year shelf life. Sterilization validation was performed per ISO 11135 and related standards.
Technological Characteristics
Monopolar needle electrode; sterile; single-use; EtO sterilized. Compliant with IEC 60601-1 (electrical safety) and ISO 10993 (biocompatibility). Packaging compliant with ASTM F1886, F2096, F88, D4169, and ISO 11607. Shelf life: 5 years.
Indications for Use
Indicated for use during spinal surgery to deliver electrical stimulation to tissues and nerves at the operative site to assist in nerve localization.
Regulatory Classification
Identification
A needle electrode is a device which is placed subcutaneously to stimulate or to record electrical signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 29, 2023
TeDan Surgical Innovations, Inc. Lynne Davies Vice President, Regulatory Affairs and Quality Assurance 12320 Cardinal Meadow Dr. Ste. 150 Sugar Land, TX 77478
Re: K231691
Trade/Device Name: Phantom XL Insulated Dilators Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: II Product Code: PDQ Dated: June 7, 2023 Received: June 9, 2023
Dear Lynne Davies:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Patrick Antkowiak -S
for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) not assigned
Device Name Phantom XL Insulated Dilators
#### Indications for Use (Describe)
Phantom XL Insulated Dilators are indicated for use during surgery of the spine to deliver an electrical stimulus to the tissues and nerves at the operative site, to assist in locating those nerves at risk during the surgical procedure.
Type of Use (Select one or both, as applicable)
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-----------------------------------------------------------------------------|
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K231691
| | 510(k) Summary |
|----------------------|-------------------------------------------------------------------------------------------------------------|
| Submitter: | TeDan Surgical Innovations, Inc.<br>12320 Cardinal Meadow Drive<br>Suite 150<br>Sugar Land, Texas 77478 USA |
| Contact Person | Lynne Davies<br>Vice President, Regulatory Affairs and Quality Assurance |
| Telephone: | 713-726-0886 |
| Submission Date: | June 7, 2023 |
| Device Name: | Phantom XL Insulated Dilators |
| Common Name: | Needle electrode |
| Regulatory Class: | II |
| Classification Name: | Neurosurgical nerve locator, 21 CFR 874.1820 (PDQ) |
# A. Predicate Device
The predicate device is the Phantom XL Insulated Dilators (K140088) also manufactured by TeDan Surgical Innovations, Inc.
## B. Device Description
TSI's Phantom XL Insulated Dilators are used as instruments to deliver electrical stimulation to tissue during intraoperative neurological monitoring. Phantom XL Insulated Dilators are available in monopolar configuration and four diameter sizes. They are supplied sterile, are non-pyrogenic, and are intended for single use only.
## C. Indications for Use
TSI's Phantom XL Insulated Dilators are indicated for use during surgery of the spine to deliver an electrical stimulus to the tissues and nerves at the operative site, to assist in locating those nerves at risk during the surgical procedure.
{4}------------------------------------------------
## D. Comparison of Technological Characteristics with the Predicate Device
The following technological differences exist between the subject and predicate devices:
- . Change in packaging from backer card/Tyvek pouch/Shelf Box configuration to preformed tray sealed in Tyvek Lid placed in Shelf Box
- Extended shelf life from 2 years to 5 years .
| Device Name | Predicate Device<br>Phantom XL Insulated<br>Dilators<br>K140088 | Subject Device<br>Phantom XL Insulated Dilators<br>K231619 |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | TeDan Surgical<br>Innovations | TeDan Surgical Innovations |
| 510(k) number | K140088 | K231619 |
| Regulation | 21CFR 882.1350,<br>(GZX)<br>and 21 CFR 874.1820<br>(PDQ) | 21 CFR 874.1820 |
| Review panel | Neurology | Neurology |
| Product code | Initially filed as GXZ,<br>subsequently changed to<br>PDQ | PDQ |
| Regulation<br>description | Needle electrode | Surgical nerve stimulator/locator. |
| Classification | Class II | Class II |
| Indication for use<br>and Intended use | Phantom XL Insulated<br>Dilators are indicated<br>for use during surgery of<br>the spine to deliver an<br>electrical stimulus to the<br>tissues and nerves at the<br>operative site, to assist<br>in locating those nerves<br>at risk during the<br>surgical procedure. | Phantom XL Insulated Dilators are<br>indicated for use during surgery of the<br>spine to deliver an electrical stimulus to<br>the tissues and nerves at the operative<br>site, to assist in locating those nerves at<br>risk during the surgical procedure. |
| Biocompatibility | Compliant with ISO<br>10993 for contact of<br>limited duration | Compliant with ISO 10993 for<br>contact of limited duration |
| Sterile | Yes; EtO | Yes; EtO |
| Single Use | Yes | Yes |
| Shelf Life | 2 years | 5 years |
| Device Name | Predicate Device<br>Phantom XL Insulated<br>Dilators<br>K140088 | Subject Device<br>Phantom XL Insulated Dilators<br>K231619 |
| Principle of Operation | Used as instrument to<br>deliver electrical<br>stimulation to tissue<br>during neurological<br>monitoring | Used as instrument to deliver electrical<br>stimulation to tissue during neurological<br>monitoring |
| Polarity of<br>stimulation | Monopolar | Monopolar |
| Exposed<br>area of<br>stimulation | Tissues and nerves at the<br>operative site selected by<br>the user | Tissues and nerves at the operative site<br>selected by the user |
| Electrical<br>Safety | Compliant with IEC<br>60601-1 | Compliant with IEC 60601-1 |
| Packaging | Product is loaded into<br>backer cards and sealed<br>in a Tyvek Pouch; Pouch<br>is placed in Shelf Box;<br>compliant with applicable<br>ASTM standards | Product is loaded into preformed tray<br>and sealed in a Tyvek Lid; Tray is<br>placed in Shelf Box; compliant with<br>applicable ASTM standards |
{5}------------------------------------------------
{6}------------------------------------------------
# E. Performance Data
The following performance data were provided in support of the substantial equivalence determination:
| Type of change | Description of Tests |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Packaging | ASTM F1886:2016, Standard Test Method for Determining Integrity of<br>Seals for Medical Packaging by Visual Inspection |
| | ASTM F2096-11, Standard test method for detecting Gross Leaks in<br>Packaging by internal Pressurization |
| | ASTM F88/F88M-21,2021, Standard Test Method for Seal Strength of<br>Flexible Barrier Materials |
| | (Dupont Medical Packaging, Description of Documents Supporting the<br>Compliance of Tyvek to ISO 11607 Standard) |
| | ASTM D4169:2022, Standard Practices for Performance Testing of<br>Shipping Containers and Systems |
| | ISTA 3A: 2018, Packaged-Products for Parcel Delivery System<br>Shipments 70kg (150lb) or Less |
| | ISO 11607-2:2019, Packaging for terminally sterilized medical devices-<br>Part 2: Validation requirements for sealing and assembly processes |
| | ISO 16269-6:2014, Statistical interpretation of data – Part 6:<br>Determination of statistical tolerance intervals |
| Expiration Date | ASTM F1980; 2021, Guide to Accelerated Aging of Sterile Medical<br>Device Packages |
| Type of change | Description of Tests |
| Sterilization | ISO 11607-1:2019, Packaging for terminally sterilized medical devices<br>- Part 1: Requirements for materials, sterile barrier systems and<br>packaging systems |
| | AAMI/ISO TIR16775: 2014, Technical Information Report, Packaging<br>for terminally sterilized medical devices-Guidance on the application of<br>ISO 11607-1 and ISO 11607-2 |
| | AAMI TIR28:2016, Product adoption and process equivalency for<br>ethylene oxide sterilization |
| | ANSI/AAMI ST67:2019 – Sterilization of health care products -<br>Requirements and guidance for selecting a sterility assurance level<br>(SAL) for products labeled sterile |
| | ANSI/AAMI/ISO 11135:2014, Sterilization of health care products -<br>Ethylene oxide - Requirements for development, validation and routine<br>control of a sterilization process for medical devices- Part 1,<br>Sterilization of health care products, Ethylene oxide, Part 1 |
| | ANSI/AAMI ST72; 2019, Bacterial Endotoxins-Test methods, routine<br>monitoring, and alternatives to batch testing |
{7}------------------------------------------------
{8}------------------------------------------------
Additional verification and validation testing were completed in accordance with the company's Design Control process in compliance with 21 CFR Part 820.30, including the following:
- 1. Bench testing for sterility in compliance with the standards cited above
- 2. Bench testing for new expiration dating in compliance with the standards cited above
- 3. Bench testing for packaging integrity in compliance with the standards cited above
## F. Conclusion
Potential risks were identified according to the ISO 14971 Standard. The risks were analyzed with regard to risk/benefit category and mitigations were implemented and tested as part of the performance testing described above for the changes submitted. No new risks were identified for the modifications to the devices. All risk mitigations were satisfactorily verified and validated in compliance with the company's Design Control Process. Where there were differences from the predicate, these were shown not to result in any new issues of safety or effectiveness according to the performance data submitted. Therefore, the non-clinical performance data provided demonstrates that the TSI Phantom XL Dilators are safe for use perform comparably to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.