K231680 · L&K BIOMED Co., Ltd. · MAX · Aug 24, 2023 · Orthopedic
Device Facts
Record ID
K231680
Device Name
AccelFix Lumbar Expandable Cage System
Applicant
L&K BIOMED Co., Ltd.
Product Code
MAX · Orthopedic
Decision Date
Aug 24, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AccelFix Lumbar Expandable Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft and/or allogenous bone graft composed of cancellous and/or corticocancellous bone. AccelFix Lumbar Expandable Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
Device Story
AccelFix Lumbar Expandable Cage System consists of hollow, rectangular titanium alloy (Ti-6Al-4V ELI) implants designed for intervertebral body fusion. Implants feature serrated superior/inferior surfaces for fixation and are available in various dimensions and lordotic angles. Implants are inserted in an unexpanded state and expanded in situ using system instrumentation. Used in spinal surgery via transforaminal, posterior, anterior-to-psoas, or lateral approaches. Operated by surgeons in clinical settings. Output is a stabilized intervertebral space filled with bone graft to facilitate fusion. Benefits include restoration of disc height and spinal stability in patients with degenerative disc disease.
Clinical Evidence
Bench testing only. Mechanical performance demonstrated via static/dynamic compression, shear, torsion, expulsion, and subsidence testing per ASTM F2077 and ASTM F2267. No clinical data provided.
Technological Characteristics
Material: Ti-6Al-4V ELI (ASTM F136). Instruments: Medical-grade stainless steel (ASTM F899-20: STS 630, STS 304, STS 455) and silicone. Expandable cage mechanism. Sterilization: Gamma radiation (sterile) or steam autoclave (non-sterile).
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, including those with up to Grade 1 spondylolisthesis or retrolisthesis, who have failed at least 6 months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract image of a human figure, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" in a larger font size than the rest of the text. The logo is clean and professional, and it is easily recognizable as the symbol of the FDA.
August 24, 2023
L&K BioMed Co., Ltd. Katherine Kim RA #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil, Giheung-gu Yongin-si, Gyeonggi-do 17015 Korea, South
Re: K231680
Trade/Device Name: AccelFix Lumbar Expandable Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: July 31, 2023 Received: August 2, 2023
Dear Katherine Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K231680
#### Device Name
AccelFix Lumbar Expandable Cage System
#### Indications for Use (Describe)
AccelFix Lumbar Expandable Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft and/or allogenous bone graft composed of cancellous and/or corticocancellous bone. AccelFix Lumbar Expandable Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):
#### 1. SUBMITTER
| Submitter's Name: | L&K BIOMED Co., Ltd. |
|------------------------|----------------------------------------------------------------------------------------------------------|
| Submitter's Address: | #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil<br>Giheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea |
| Submitter's Telephone: | +82-2-6717-1983 |
| Submission date: | July 6, 2023 |
| Contact Person: | Katherine Kim<br>khkim@lnkbiomed.com/khkim3747@gmail.com |
#### 2. DEVICE NAME
| Trade or Proprietary Name | AccelFix Lumbar Expandable Cage System |
|---------------------------|----------------------------------------------------------------|
| Common or Usual Name | Intervertebral Body Fusion Device, Intervertebral cage, Spacer |
| Regulation class / Number | ClassII, CFR 888.3080 |
| Regulation Name | Intervertebral Body Fusion Device |
| Product Code | MAX |
| Classification Panel | Orthopedics |
#### 3. PREDICATE DEVICE
The changes to this product, the additional approach, do not affect the product's design, dimensional sizes, indication for use, material, scientific technology and manufacturing process.
The design feature and indications for use and manufacturing process for the subject device 'AccelFix Lumbar Expandable Cage System' is substantially equivalent to the predicate device.
Primary Predicate Device: AccelFix Lumbar Expandable Cage System (K190708)
#### 4. DESCRIPTION OF THE DEVICE
The AccelFix Lumbar Expandable Cage System's implants are interbody fusion devices intended for use as an aid in spinal fixation. They are made of Titanium 6AL-4V Alloy (ASTM F136). These hollow, rectangular implants are offered in a variety of widths, lengths, heights and lordotic angles designed to adapt to a variety of patient anatomies. The implants can be expanded in height after insertion in the unexpanded state using the system instrumentation. The implants have serrations on the superior and inferior surfaces designed for fixation.
- · AccelFix-XT Expandable Cage is to be implanted via transforaminal and posterior approach.
- . AccelFix-XTP Expandable Cage is to be implanted via Anterior to Psoas approach and lateral Approach.
- AccelFix-XL Expandable Cage is to be implanted via lateral approach.
#### 5. INDICATION FOR USE
AccelFix Lumbar Expandable Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft and/or allogenous bone graft composed of cancellous and/or corticocancellous bone. AccelFix Lumbar Expandable Cage System is to be used with supplemental fixation systems. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
{4}------------------------------------------------
#### 6. PERFORMANCE DATA
Changes made by this submission are not subject to additional mechanical testing.
The predicate AccelFix Lumbar Expandable Cage System was tested according to the ASTM F 2077. ASTM F 2267 and ASTM draft as described below.
Static compression, dynamic compression, static and dynamic shear testing according to ASTM F2077, was presented to demonstrate the substantial equivalency of the AccelFix Lumbar Expandable Cage System to the predicate devices.
- Static Axial Compression Test ASTM F 2077 -17
- Static Compression-Shear Test ASTM F 2077 -17
- Static Torsion Test ASTM F 2077 -17
- · Static Expulsion Test
- Static Subsidence Test ASTM F 2267 04 (Reapproved 2018)/F 2077-17
- Dynamic Axial Compression Test- ASTM F 2077 -17
- Dynamic Compression-Shear Test ASTM F 2077 -17
Bench testing to evaluate the mechanical properties of the AccelFix Lumbar Expandable Cage System showed a higher or similar mechanical value than predicate marketed devices.
#### 7. MATERIAL
The AccelFix Lumbar Expandable Cage is manufactured from Ti-6Al-4V ELI titanium alloy (ASTM F136). This this is the same material used in the predicate devices. And additional instruments and changed instrument pieces that are patient anatomy contacting are manufactured of medical grade materials (medical-grade stainless steel, medical grade silicone) and are tissue contacting in limitedexposure duration of less than 24 hours. Only the materials that contact with tissue are considered in this regard. The additional instruments that contact tissue area during surgery are made of Stainless-Steel ASTM F 899-20(STS455, STS630, STS 304). These materials have been widely recognized to be safe materials and known to be biocompatible.
#### 8. COMPARISON OF TECHNICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject and predicate devices are identical technological characteristics and there is not any new issues of safety and effectiveness. The characteristics are similar or identical between the subject and the predicate, but the data for the characteristics are as compared to the data below.
- . Instruction for use (added lateral approach to AccelFix-XTP Cage)
- · Design and sizes
- Expanding Mechanism
- · Material (material change due to design change of Expandable Cage Inserter)
- " Approach (added lateral approach to AccelFix-XTP Cage)
- Sterilization & Method
- Manufacturing process (for the subject device of Expandable Cage Inserter, the handle manufacturing process and the integrating process(assembling) of indicator module, cage inserter body and handle are excluded from the predicate device manufacturing process).
| INDICATION FOR USE | |
|------------------------------|-----------------------------|
| Similarities | |
| Subject Devices | |
| Predicate Devices<br>K190708 | Indication for use are same |
{5}------------------------------------------------
#### DESIGN AND SIZES OF IMPLANTS
| Characteristics | Subject Devices | Predicate Devices K190708 | Similarities/Differences |
|-----------------------|----------------------------------------------------------------------------|------------------------------------------|--------------------------|
| Design | Subject implant has the same design as predicate implant | | Similar |
| Sizes | Subject implant has the same sizes as predicate implant | | Similar |
| Material | Ti-6Al-4V ELI titanium alloy (ASTM F136) | Ti-6Al-4V ELI titanium alloy (ASTM F136) | Similar |
| Manufacturing Process | Subject implant's manufacturing process is the same as predicate implant's | | Similar |
#### INSTRUMENTS (EXPANDABLE CAGE INSERTER XTP01-0001)
| Characteristics | Subject Devices | Predicate Devices K190708 | Similarities/Differences |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------|
| Design(changed) | The Height Indicator<br>Module and handle are<br>separated and excluded<br>from the expandable cage<br>inserter. | The indicator module, cage<br>inserter body and handle<br>are integrated | Difference |
| Material | STS 630, STS 304-tissue<br>contact | STS 630(ASTM F 899-20)<br>STS 304(ASTM F 899-20)<br>STS 455(ASTM F 899-20)<br>Ti6AL4V (ASTM F 136) Silicone | Difference |
| Manufacturing<br>process | The handle and indicator module are separated and<br>removed from the cage insert of the subject device.<br>Therefore, the manufacturing process of the handle and<br>indicator module and the process of integrating (assembly)<br>the cage insert body are excluded, and the rest of the<br>processes are the same as the predicate device<br>manufacturing process. | | Difference |
#### ADDITIONAL INSTRUMENTS (CLASS II)
| Characteristics | Material | Similarities/Differences |
|-----------------------------------------------|-----------------------------------------------------------------------|--------------------------|
| Subject Devices<br>tissue contact<br>material | ASTM F 899-20 (STS 630, STS 304, STS455)<br>Tissue contacts materials | Similar |
| Predicate Devices<br>K190708 | ASTM F 899-20 (STS 630, STS 304, STS455)<br>Tissue contacts materials | |
#### APPROACH
| Characteristics | Subject Devices | Predicate Devices K190708 | Similarities/Differences |
|----------------------------------|-------------------------------------------------|---------------------------|--------------------------------------|
| Approach<br>AccelFix-XTP<br>Cage | Anterior to Psoas Approach<br>Lateral Approach. | Lateral Approach. | Difference<br>added Lateral Approach |
#### STERILIZATION
| Subject Devices | Predicate Devices<br>K190708 | Similarities/Differences |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Sterile device: Gamma radiation Non-sterile device: recommended<br>steam sterilization (autoclave<br>sterilization) | Sterile device: Gamma radiation Non-sterile device: recommended<br>steam sterilization (autoclave<br>sterilization) | Similar |
| Subject Devices | Predicate Devices K190708 | Similarities<br>/Differences |
| (1) Soak the instrument in the tap water<br>for 1 minute.<br>(2) Prepare cleaning solution at the<br>(0.8%~1.6<br>concentration<br>%)<br>by the<br>suggested<br>detergent<br>manufacturer and soak the instrument<br>in the prepared cleaning solution for 5<br>minutes, and then remove the<br>contamination from the instrument<br>surface with a soft brush. However,<br>the operating part should be operated<br>to clean the contamination.<br>(3) Rinse the instrument for at least 3<br>minutes with tap water. However, the<br>small diameter hole is rinsed using a<br>syringe or the like.<br>(4) Rinse the instrument by shaking with<br>purified water for at least 5 minutes.<br>(5) Wipe off water in the instruments with<br>a clean cloth. However, compressed<br>air can be used for the device with<br>complex structures.<br>(6) Prepare for storage and sterilization. | - Use the neutral pH enzyme soaking<br>solution that has been prepared.<br>- Completely submerge the instrument in<br>enzyme solution and allow it to soak for<br>20 minutes (water temperature: follow<br>the manufacturer's instruction). Scrub<br>instrument using a soft-bristled brush to<br>gently clean the device (particular<br>attention shall be given to crevices,<br>lumens, mated surfaces, and other hard-<br>to-clean areas) until all visible soil has<br>been removed. Lumens should be<br>cleaned with a long, narrow, soft-<br>bristled brush (i.e., pipe cleaner brush).<br>Note: Any assembled instruments, please<br>disassemble the parts before submerging.<br>Note: The enzyme solution should be<br>changed on a regular basis in order to<br>ensure its effectiveness.<br>- Remove the device from the enzyme<br>solution and rinse in purified water<br>(from one or any combination of the<br>following processes: ultra-filter, RO,<br>DI and/or distilled) for a minimum of 3<br>minutes. Thoroughly flush lumens,<br>holes and other difficult to reach areas.<br>Prepare the neutral pH cleaning<br>(detergent) solution according to the<br>manufacturer's instructions, dilution<br>recommendations, and temperatures<br>and place in a sonication unit.<br>- Completely submerge device<br>in<br>cleaning solution and sonicate for 10<br>minutes, preferably at 45-50 kHz.<br>- Rinse instrument in purified water<br>(from one or any combination of the<br>following processes: ultra-filter, RO,<br>DI and/or distilled) thoroughly for at<br>least 3 minutes.<br>- Repeat Steps 5 and 6 with freshly<br>prepared cleaning solution.<br>- Dry the instrument with a clean,<br>disposable, absorbent, non-shedding<br>wipe.<br>- Visually inspect the devices under<br>normal room lighting condition to<br>verify all foreign debris has been<br>removed thoroughly clean.<br>- Verify that the devices are visually<br>clean. | changed |
| Semi-auto Cleaning Procedure: Added Semi-auto cleaning procedure on subject devices | | Similarities<br>/Differences |
| Subject Devices | Predicate Devices K190708 | |
| (1) Prepare cleaning solution at the<br>concentration (0.8%~1.6 %) suggested<br>by the detergent manufacturer and soak<br>the instrument in the prepared cleaning<br>solution for 20 minutes. Clean the<br>instrument with a soft brush. (Especially,<br>carefully wipe the gaps, interiors, and<br>surfaces that are difficult to clean.) The<br>inside of the instrument should be<br>cleaned with a long, narrow, and soft<br>brush.<br>(2) Rinse with purified water for 3 minutes<br>to remove the cleaning solution. Use<br>rinsing water temperature between 35°C<br>and 45°C for part and holes that are<br>difficult to clean.<br>(3) Immerse the product in the cleaning<br>solution and clean using ultrasonic<br>cleaner at 45 kHz to 50 kHz for 10<br>minutes.<br>(4) Rinse with purified water for at least 3<br>minutes until there is no blood or<br>contaminants. If contaminants remain,<br>ultrasonic cleaning (step 3) and rinsing<br>(step 4) are performed once more.<br>(5) Wipe the instrument cleaned with a<br>clean, absorbent disposable wipe to dry<br>instrument. If additional drying is<br>needed after drying step, use medical<br>compressed air (psi 20~30) to dry it.<br>Check that the instrument is visually<br>cleaned. | None | Added |
{6}------------------------------------------------
#### CLEANING METHOD
Manual Cleaning Procedure: The manual cleaning procedure for subject devices changed the cleaning time from the manual cleaning procedure for predicate devices and deleted the sonicate part.
{7}------------------------------------------------
- removed thoroughly clean. removed thoroughly clean. Verify that the instruments are in Verify that the instruments are in " visually clean. visually clean.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the text 'K231680' and 'Page 6 of 6' at the top. Below that is the logo for L&K BIOMED, which includes the company's name in bold letters and a graphic of blue circles to the right. The text and logo are positioned against a white background.
#### 9. SUBSTANTIAL EQUIVALENCE AND CONCLUSION
The subject AccelFix Lumbar Expandable Cage System have been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, material, intended use and manufacturing process. The information provided within this premarket notification supports substantial equivalence of the subject device to the predicate devices. The overall data lead to the conclusion that the AccelFix Lumbar Expandable Cage System is substantially equivalent to the predicate devices (K190708).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.