K231661 · Medinux (Tianjin) Technologies Co., Ltd. · GEH · Nov 27, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231661
Device Name
CryoThin™ Surgical System and Accessories
Applicant
Medinux (Tianjin) Technologies Co., Ltd.
Product Code
GEH · General, Plastic Surgery
Decision Date
Nov 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CryoThin Surgical System is intended for cryosurgery in the fields of general surgery, urology, gynecology, oncology, neurology, dermatology, ENT, proctology, pulmonary surgery and thoracic surgery. The system is designed to freeze/ablate tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts. The system has the following specific indications: General Surgery - Destruction of warts or lesions - Palliation of tumors of the oral cavity, rectum and skin - Ablation of breast fibroadenomas - Ablation of leukoplakia of the mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemangiomas, recurrent cancerous lesions Urology - Ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia Gynecology - Ablation of malignant neoplasia or benign dysplasia of the female genitalia Oncology - Ablation of cancerous or malignant tissue - Ablation of benign tumors - Palliative intervention Neurology - Freezing of nerve tissue in pain management/cryoanalgesia Dermatology - Ablation or freezing of skin cancers and other cutaneous disorders Proctology - Ablation of benign or malignant growths of the anus or rectum - Ablation of hemorrhoids Thoracic Surgery - Ablation of cancerous lesions
Device Story
CryoThin Surgical System uses Joule-Thomson effect for cryoablation; high-pressure argon gas undergoes throttling expansion within closed-tip cryo probes to freeze target tissue; low-pressure argon is electrically heated and circulated to thaw tissue. System comprises console, cart, cryo probes, and temperature probes. Used by clinicians in surgical settings. Output includes real-time temperature monitoring and controlled freeze/thaw cycles. Enables precise tissue destruction for oncology, urology, and dermatology applications; benefits patients via minimally invasive tumor/lesion removal.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), human factors (IEC 62366), biocompatibility (ISO 10993), sterilization validation (ISO 11137), shelf-life/service-life testing, and iceball/isotherm performance testing to confirm therapeutic cooling capabilities.
Technological Characteristics
Cryosurgical unit utilizing Joule-Thomson effect; Argon gas source; electrical heating for thaw; touch screen interface. Biocompatibility per ISO 10993; sterilization per ISO 11137; shelf-life per ASTM F 1980-16; packaging per ASTM D 4169-16. Class II device.
Indications for Use
Indicated for cryosurgical ablation of benign and malignant tissues in general surgery, urology, gynecology, oncology, neurology, dermatology, ENT, proctology, pulmonary, and thoracic surgery. Includes prostate/kidney tissue, liver metastases, tumors, skin lesions, warts, and nerve tissue for pain management.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
November 27, 2023
Medinux (Tianjin) Technologies Co., Ltd. Ms. Diana Zhang, B.A. Manager, Regulatory Affairs Room 401, Unit 3, Building K1 6 Haitai Fazhan 6th Road, Huayuan Industrial Park Tianjin 300384 China
Re: K231661
Trade/Device Name: CryoThin™ Surgical System and Accessories Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: November 3, 2023 Received: November 3, 2023
Dear Ms. Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.11.27 10:22:03 -05'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical
{2}------------------------------------------------
and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231661
Device Name CryoThin™ Surgical System and Accessories
#### Indications for Use (Describe)
The CryoThin Surgical System is intended for cryosurgery in the fields of general surgery, urology, oncology, neurology, dermatology, ENT, proctology, pulmonary surgery. The system is designed to freeze/ablate tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastase, tumors, skin lesions, and warts. The system has the following specific indications:
General Surgery
- · Destruction of warts or lesions
- · Palliation of tumors of the oral cavity, rectum and skin
- · Ablation of breast fibroadenomas
· Ablation of leukoplakia of the mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemangiomas, recurrent cancerous lesions
#### Urology
· Ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia
Gynecology
· Ablation of malignant neoplasia or benign dysplasia of the female genitalia
#### Oncology
- · Ablation of cancerous or malignant tissue
- · Ablation of benign tumors
- · Palliative intervention
#### Neurology
- · Freezing of nerve tissue in pain management/cryoanalgesia
#### Dermatology
· Ablation or freezing of skin cancers and other cutaneous disorders
#### Proctology
- Ablation of benign or malignant growths of the anus or rectum
- · Ablation of hemorrhoids
Thoracic Surgery
· Ablation of cancerous lesions
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# K231661 510(k) SUMMARY
# CryoThin™ Surgical System and Accessories
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Medinux (Tianjin) Technologies Co., Ltd.<br>Room 401, Unit 3, Building K1<br>6 Haitai Fazhan 6th Road<br>Huayuan Industrial Park<br>Tianjin 300384<br>China |
|---------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact<br>Person: | Ms. Diana Zhang, B.A.<br>Manager, Regulatory Affairs<br>Phone: (86) 22 86 392 536<br>Fax: (86) 22 87 930 633<br>dianazhang@medinux.com |
| Trade/Device Name:<br>Manufacturer:<br>Regulation Number:<br>Regulation Name:<br>Regulation Class:<br>Product Code: | CryoThin™ Surgical System and Accessories<br>Medinux (Tianjin) Technologies Co., Ltd.<br>21 CFR 878.4350<br>Cryosurgical unit and accessories<br>Class II<br>GEH |
| Predicate Devices /<br>Reference 510(k): | CRYOCARE TOUCH System and Accessories (K201588)<br>Visual-ICE® Cryoablation System (K143564) |
| Date of Preparation: | November 23, 2023 |
## Device Description:
Cryosurgery is the use of ultra-low temperatures for selectively freezing and destroying diseased tissues in situ. The CryoThin™ Surgical System is designed to ablate benign and malignant tissues using cryosurgery.
The system utilizes the Joule-Thomson effect for freezing. It uses high-pressure argon gas that undergoes throttling expansion to induce freezing inside the closed tip of the Cryo probe to freeze the target tissue around the probe tip. Low-pressure argon is electrically heated and blown into the probe tip to thaw the tissue around it.
{6}------------------------------------------------
The CryoThin Surgical System consists of a console, a cart, and accessories. The system must be used in conjunction with the Cryo probes for freeze and thaw functions, and the Temp probes for monitoring the temperature in target or adjacent tissues.
## Indications for Use:
The CryoThin Surgical System is intended for cryosurgery in the fields of general surgery, urology, gynecology, oncology, neurology, dermatology, ENT, proctology, pulmonary surgery and thoracic surgery. The system is designed to freeze/ablate tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts. The system has the following specific indications:
## General Surgery
- Destruction of warts or lesions
- Palliation of tumors of the oral cavity, rectum and skin
- Ablation of breast fibroadenomas ●
- Ablation of leukoplakia of the mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemangiomas, recurrent cancerous lesions
## Urology
- Ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia
## Gynecology
- . Ablation of malignant neoplasia or beniqn dysplasia of the female genitalia
Oncology
- Ablation of cancerous or malignant tissue ●
- Ablation of benign tumors ●
- Palliative intervention
## Neurology
- Freezing of nerve tissue in pain management/cryoanalgesia ●
## Dermatology
- . Ablation or freezing of skin cancers and other cutaneous disorders Proctology
{7}------------------------------------------------
- Ablation of benign or malignant growths of the anus or rectum .
- Ablation of hemorrhoids .
Thoracic Surgery
- Ablation of cancerous lesions
## Technological Comparison:
Table 1 below provides a comparison of the technological characteristics of the subject device with the predicate devices.
{8}------------------------------------------------
# Table 1 Subject and Predicate Device Comparison
| | Subject Device | Predicate Device 1 | Predicate Device 2 | |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Feature | CryoThin Surgical System and<br>Accessories | CRYOCARE TOUCH System and<br>Accessories | Visual-ICE Cryoablation System | Comparison |
| 510(k) Number | - | K201588 | K143564 | - |
| Trade/Device<br>Name | CryoThin Surgical System and<br>Accessories | CRYOCARE TOUCH System and<br>Accessories | Visual-ICE Cryoablation System | - |
| Manufacturer | Medinux (Tianjin) Technologies<br>Co., Ltd. | Varian Medical Systems, Inc. | Galil Medical Inc. | Same |
| Regulation<br>Number | 21 CFR 878.4350 | 21 CFR 878.4350 | 21 CFR 878.4350 | Same |
| Regulation<br>Name | Cryosurgical unit and<br>accessories | Cryosurgical unit and<br>accessories | Cryosurgical unit and accessories | Same |
| Regulation<br>Class | Class II | Class II | Class II | Same |
| Product Code | GEH | GEH | GEH | Same |
| Indications for<br>Use | The CryoThin Surgical System is<br>intended for cryosurgery in the<br>fields of general surgery,<br>urology, gynecology, oncology,<br>neurology, dermatology, ENT,<br>proctology, pulmonary surgery<br>and thoracic surgery. The<br>system is designed to<br>freeze/ablate tissue by the<br>application of extreme cold<br>temperatures including | The CRYOCARE TOUCH System<br>is intended for use in open,<br>minimally invasive, or<br>endoscopic surgical procedures<br>in the areas in general surgery,<br>urology, gynecology, oncology,<br>neurology, dermatology, ENT,<br>proctology, pulmonary surgery<br>and thoracic surgery. The<br>system is designed to<br>freeze/ablate tissue by the | The Visual-ICE Cryoablation<br>System is intended for<br>cryoablative destruction of tissue<br>during surgical procedures. The<br>Visual-ICE System is indicated for<br>use as a cryosurgical tool in the<br>fields of general surgery,<br>dermatology, neurology<br>(including cryoanalgesia),<br>thoracic surgery, ENT,<br>gynecology, oncology, | Same |
| Feature | Subject Device<br>CryoThin Surgical System and<br>Accessories | Predicate Device 1<br>CRYOCARE TOUCH System and<br>Accessories | Predicate Device 2<br>Visual-ICE Cryoablation System | Comparison |
| | prostate and kidney tissue, liver<br>metastases, tumors, skin<br>lesions, and warts. | application of extreme cold<br>temperatures including prostate<br>and kidney tissue, liver<br>metastases, tumors, skin<br>lesions, and warts. | proctology, and urology. This<br>system is designed to destroy<br>tissue by the application of<br>extreme cold temperatures<br>including prostate and kidney<br>tissue, liver metastases, tumors,<br>skin lesions, and warts. | |
| | The system has the following<br>specific indications: | In addition, the system is<br>intended for use in the<br>following indications: | The Visual-ICE Cryoablation<br>System has the following specific<br>indications: | |
| | <b>General Surgery</b><br>• Destruction of warts or<br>lesions<br>• Palliation of tumors of the<br>oral cavity, rectum and skin<br>• Ablation of breast<br>fibroadenomas<br>• Ablation of leukoplakia of the<br>mouth, angiomas, sebaceous<br>hyperplasia, basal cell tumors of<br>the eyelid or canthus area,<br>ulcerated basal cell tumors,<br>dermatofibromas, small<br>hemangiomas, mucocele cysts,<br>multiple warts, plantar warts,<br>hemorrhoids, anal fissures,<br>perianal condylomata, pilonidal | <b>General Surgery</b><br>• Destruction of warts or lesions<br>• Palliation of tumors of the oral<br>cavity, rectum and skin<br>• Ablation of breast<br>fibroadenomas<br>• Ablation of leukoplakia of the<br>mouth, angiomas, sebaceous<br>hyperplasia, basal cell tumors of<br>the eyelid or canthus area,<br>ulcerated basal cell tumors,<br>dermatofibromas, small<br>hemangiomas, mucocele cysts,<br>multiple warts, plantar warts,<br>hemorrhoids, anal fissures,<br>perianal condylomata, pilonidal | • <b>Urology</b> : Ablation of prostate<br>tissue in cases of prostate cancer<br>and Benign Prostate Hyperplasia<br>(BPH)<br>• <b>Oncology</b> : Ablation of<br>cancerous or malignant tissue<br>and benign tumors, and<br>palliative intervention<br>• <b>Dermatology</b> : Ablation or<br>freezing of skin cancers and<br>other cutaneous disorders<br>Destruction of warts or lesions,<br>angiomas, sebaceous | |
| Feature | Subject Device<br>CryoThin Surgical System and<br>Accessories | Predicate Device 1<br>CRYOCARE TOUCH System and<br>Accessories | Predicate Device 2<br>Visual-ICE Cryoablation System | Comparison |
| | cysts, actinic and seborrheic<br>keratoses, cavernous<br>hemangiomas, recurrent<br>cancerous lesions | cysts, actinic and seborrheic<br>keratoses, cavernous<br>hemangiomas, recurrent<br>cancerous lesions | hyperplasia, basal cell tumors of<br>the eyelid or canthus area,<br>ulcerated basal cell tumors,<br>dermatofibromas, small<br>hemangiomas, mucocele cysts,<br>multiple warts, plantar warts,<br>actinic and seborrheic keratosis,<br>cavernous hemangiomas, peri-<br>anal condylomata, and palliation<br>of tumors of the skin. | |
| | Urology<br>• Ablation of prostate tissue in<br>cases of prostate cancer and<br>benign prostatic hyperplasia | Urology<br>• Ablation of prostate tissue in<br>cases of prostate cancer and<br>benign prostatic hyperplasia | | |
| | Gynecology<br>• Ablation of malignant<br>neoplasia or benign dysplasia of<br>the female genitalia | Gynecology<br>• Ablation of malignant<br>neoplasia or benign dysplasia of<br>the female genitalia | • Gynecology: Ablation of<br>malignant neoplasia or benign<br>dysplasia of the female genitalia | |
| | Oncology<br>• Ablation of cancerous or<br>malignant tissue<br>• Ablation of benign tumors<br>• Palliative intervention | Oncology<br>• Ablation of cancerous or<br>malignant tissue<br>• Ablation of benign tumors<br>• Palliative intervention | • General surgery: Palliation of<br>tumors of the rectum,<br>hemorrhoids, anal fissures,<br>pilonidal cysts, and recurrent<br>cancerous lesions, ablation of<br>breast fibroadenomas | |
| | Neurology<br>• Freezing of nerve tissue in<br>pain<br>management/cryoanalgesia | Neurology<br>• Freezing of nerve tissue in<br>pain<br>management/cryoanalgesia | • ENT: Palliation of tumors of the<br>oral cavity and ablation of<br>leukoplakia of the mouth | |
| | Dermatology<br>• Ablation or freezing of skin<br>cancers and other cutaneous | Dermatology<br>• Ablation or freezing of skin<br>cancers and other cutaneous | • Thoracic surgery: Ablation of<br>arrhythmic cardiac tissue | |
| | Subject Device | Predicate Device 1 | Predicate Device 2 | |
| Feature | CryoThin Surgical System and<br>Accessories | CRYOCARE TOUCH System and<br>Accessories | Visual-ICE Cryoablation System | Comparison |
| | disorders<br><b>Proctology</b><br>• Ablation of benign or<br>malignant growths of the anus<br>or rectum<br>• Ablation of hemorrhoids<br><b>Thoracic Surgery</b><br>• Ablation of cancerous lesions | disorders<br><b>Proctology</b><br>• Ablation of benign or<br>malignant<br>growths of the anus or rectum<br>• Ablation of hemorrhoids<br><b>Thoracic Surgery</b><br>• Ablation of cancerous lesions | cancerous lesions<br>• Proctology: Ablation of benign<br>or malignant growths of the anus<br>or rectum, and hemorrhoids | |
| Mechanism of<br>action | Joule-Thomson effect | Joule-Thomson effect | Joule-Thomson effect | Same |
| Freeze gas | Argon gas | Argon gas | Argon gas | Same |
| Thaw gas | Argon gas | Helium gas | Helium gas<br>Argon gas | Same as Predicate<br>Device 2 |
| Electrical Thaw | Yes (electrical heated argon<br>gas) | N/A | Yes (electrical heated argon gas) | Same as Predicate<br>Device 2 |
| Human<br>interface<br>device /<br>connections | • 1×touch screen monitor | • 1×touch screen monitor<br>• 1×Remote Keypad | • 1×touch screen monitor | Same as Predicate<br>Device 2 |
{9}------------------------------------------------
# K231661
{10}------------------------------------------------
# K231661
{11}------------------------------------------------
# K231661
{12}------------------------------------------------
## Summary of Performance Data (Non-clinical testing):
The CryoThin Surgical System and Accessories have the same principle of operation, intended use, and very similar technological characteristics as the predicate devices.
Comprehensive tests and analysis have been performed on the subject device which demonstrated the safety and performance characteristics of the device. The non-clinical testing done on the subject device include the following:
- Electrical Safety Testing in accordance with IEC 60601-1. a.
- b. Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2.
- Human Factors and Usability Testing in accordance with IEC 60601-1-6 and C. IEC 62366.
- Software development in accordance with IEC 62304 and the FDA's Guidance, d. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
- Biocompatibility Testing in accordance with ISO 10993 series of standards. e.
- Sterilization Validation in accordance with ISO 11137-1. ISO 11137-2. 11137-3. f. ISO 11737-1, and ISO 11737-2.
- Shelf-life testing in accordance with ASTM F 1980-16 Standard Guide for ರು. Accelerated Aging of Sterile Barrier Systems for Medical Devices.
- Service life testing in accordance with IEC 62506:2013 Methods for Product h. Accelerated Testing.
- Package Verification Testing in accordance with ASTM F 1980-16 Standard i. Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- i. Packaging Testing in accordance with ASTM D 4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems DC13 Level II.
- Cleaning and Disinfection of Reusable Probes testing to support the stated use k. life.
- Iceball and isotherm testing to demonstrate the subject device ability to cool 1. target tissue to desirable therapeutic temperature.
- m. Compatibility testing to demonstrate subject device compatibility with the most commonly used imaging modalities.
## Conclusion:
The subject device has passed all the above testing. The testing results showed its conformance to applicable requirements specifications and assured that the safety measures taken are adequate and effective. The information and data provided in this Traditional 510(k) application demonstrated that the subject device met the safety and performance specifications and did not raise any new safety or effectiveness issues. Thus, the subject device is considered to be substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.