The Intraosseous Fusion Device System is indicated for use in skeletally mature individuals for fracture repair and arthrodesis, osteotomy, joint fusion and fragment fixation appropriate with the size of the implant including: Minimally invasive reconstruction of fractures and joints; Fractures of the foot and ankle; Osteochrondritis dissecans, Ostero-Chondral Fractures, Other small fragment, cancellous bone fractures, Small joint fusion. Metatarsal and phalangeal osteotomies; Fractures of the tarsals and other fractures of the foot: Avulsion fractures and fractures of metatarsal V; Tarso-metatarsal and metatarso-phalangeal arthrodesis; Tarsal Fusions; Calcaneal and talar fractures, Subtalar arthrodesis. Ankle arthrodesis. Fractures of small joints, such as: Ankle fractures, Fractures of the fibula, malleolus, and calcaneus. The Intraosseous Fusion Device System is not intended for spinal use.
Device Story
Intraosseous Fusion Device (IFD) is a threaded, hollow, fenestrated titanium implant designed for bone stabilization and arthrodesis. Input: bone graft material (packed into hollow center). Operation: implant provides mechanical fixation via interrupted external threads; fenestrations allow bone ingrowth. Output: stabilized bone segments. Used in clinical settings by orthopedic surgeons for foot/ankle/small joint reconstruction. Healthcare providers use the device to facilitate fusion and fracture healing; benefits include minimally invasive reconstruction and enhanced bone integration. System includes implants (6, 9, 12mm diameters; 10, 15, 20mm lengths), surgical instruments, and storage/sterilization trays.
Clinical Evidence
Bench testing only. Mechanical testing included torsional strength, driving torque, and axial pullout (ASTM F543), and static/cyclic radial compression (ASTM F2077). Biocompatibility (cytotoxicity), steam sterilization validation (ISO 17665-1, ISO 14937), manual cleaning validation (AAMI TIR12, AAMI TIR30), and shipping simulation (ISTA 3A) were performed.
Indicated for skeletally mature individuals requiring fracture repair, arthrodesis, osteotomy, joint fusion, or fragment fixation in the foot, ankle, and small joints. Not for spinal use.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
September 29, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Auxano Medical LLC James Price Senior Design Engineer 8006 Katherine Blvd. Brecksville, Ohio 44141
Re: K231649
Trade/Device Name: Intraosseous Fusion Device System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: August 21, 2023 Received: August 21, 2023
Dear Mr. Price:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali-S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231649
Device Name Intraosseous Fusion Device System
#### Indications for Use (Describe)
The Intraosseous Fusion Device System is indicated for use in skeletally mature individuals for fracture repair and arthrodesis, osteotomy, joint fusion and fragment fixation appropriate with the size of the implant including:
Minimally invasive reconstruction of fractures and joints; Fractures of the foot and ankle; Osteochrondritis dissecans, Ostero-Chondral Fractures, Other small fragment, cancellous bone fractures, Small joint fusion. Metatarsal and phalangeal osteotomies: Fractures of the tarsals and other fractures of the foot: Avulsion fractures and fractures of metatarsal V; Tarso-metatarsal and metatarso-phalangeal arthrodesis; Tarsal Fusions; Calcaneal and talar fractures, Subtalar arthrodesis. Ankle arthrodesis. Fractures of small joints, such as: Ankle fractures, Fractures of the fibula, malleolus, and calcaneus.
The Intraosseous Fusion Device System is not intended for spinal use.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------|
| <div>区 Prescription Use (Part 21 CFR 801 Subpart D)</div> |
| <div> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K231649 Page 1 of 6
## 510(k) Summary
Intraosseous Fusion Device System
### SUBMITTER
Auxano Medical LLC 8006 Katherine Boulevard Brecksville, OH 44141 Phone: 440-262-2000
Contact Person: James Price - Senior Design Engineer Email: James.Price@auxanomedical.net Date Prepared: August 28, 2023
### DEVICE INFORMATION
Trade/Device Name: Intraosseous Fusion Device System Common Name: Intraosseous Fusion Device System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC
### PREDICATE INFORMATION
Predicate Device: K191995 (S4 Screw System; Subchondral Solutions, Inc.) Additional Predicate Device: K163489 (OrthoSolutions 'System26' Bone Screws; Ortho Solutions UK Limited)
### INDICATIONS FOR USE:
The Intraosseous Fusion Device System is indicated for use in skeletally mature individuals for fracture repair and arthrodesis, osteotomy, joint fusion and fragment fixation appropriate with the size of the implant including:
Minimally invasive reconstruction of fractures and joints; Fractures of the foot and ankle; Osteochrondritis dissecans, Ostero-Chondral Fractures, Other small fragment, cancellous bone fractures, Small joint fusion. Metatarsal and phalangeal osteotomies; Fractures of the tarsals, metatarsals and other fractures of the foot; Avulsion fractures of metatarsal V; Tarsometatarsal and metatarso-phalangeal arthrodesis; Tarsal Fusions; Calcaneal and talar fractures, Subtalar arthrodesis. Ankle arthrodesis. Fractures of small joints, such as: Ankle fractures, Navicular fractures, Fractures of the fibula, malleolus, and calcaneus.
The Intraosseous Fusion Device System is not intended for spinal use.
# DEVICE DESCRIPTION
The Intraosseous Fusion Device (IFD) is a threaded titanium implant that is designed to provide stabilization of bones, bone segments or bone fragments to facilitate arthrodesis. The implant has an interrupted external thread to facilitate placement and fixation. The implant is hollow, which allows packing of bone graft material, and fenestrated, which allows bone growth and formation throughout the implant. The implants come in a variety of diameters (6, 9, 12mm) and lengths (10, 15 & 20mm) to address different anatomy locations. The implants are made from Ti6-6Al-4V alloy. The Intraosseous Fusion Device System includes implants, the instruments necessary to implant them, and the tray system for transport, storage and sterilization. The system is provided non-sterile and
{4}------------------------------------------------
requires steam sterilization at point of use. Single-use metallic implants are time stable. Reusable instruments are inspected in reprocessing for conditions compromising function. All instruments (except for k-wires) are reusable and can be reprocessed.
Image /page/4/Figure/2 description: The image shows a cylindrical object with a complex, repeating pattern on its outer surface. The object appears to be hollow, with a smooth inner surface and several circular holes distributed along its length. The outer surface is covered in a grid-like arrangement of raised, rectangular blocks, creating a textured and segmented appearance. The object is rendered in a neutral gray color, giving it a metallic or industrial look.
Figure 1. Intraosseous Fusion Device
{5}------------------------------------------------
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATES:
# Table 1.
| | Subject Device:<br>Auxano Medical- Intraosseous<br>Fusion Device System | Predicate (K191995):<br>Subchondral Solutions- S4 Screw | Additional Predicate (K163489):<br>OrthoSolutions- System26 Bone Screws |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class<br>and Product<br>Code: | Class II; HWC | Class II; HWC | Class II; HWC & HTN |
| | Substantial Equivalence: | Same | Similar |
| Sterilization: | Implants & Instruments- provided<br>NON-STERILE; Steam sterilize on-site<br>to SAL of 10-6 | Implant & Instruments- provided STERILE | Implants & Instruments- provided NON-STERILE |
| | Substantial Equivalence: | Different | Similar |
| Prescription: | Prescription Only | Prescription Only | Prescription Only |
| | Substantial Equivalence: | Same | Same |
| Indications for<br>Use: | The Intraosseous Fusion Device System is<br>indicated for use in skeletally mature<br>individuals for fracture repair and<br>arthrodesis, osteotomy, joint fusion and<br>fragment fixation appropriate with the size<br>of the implant including:<br>Minimally invasive reconstruction of<br>fractures and joints; Fractures of the foot<br>and ankle; Osteochrondritis dissecans,<br>Ostero-Chondral Fractures, Other small<br>fragment, cancellous bone fractures, Small<br>joint fusion. Metatarsal and phalangeal<br>osteotomies; Fractures of the tarsals,<br>metatarsals and other fractures of the foot;<br>Avulsion fractures and fractures of<br>metatarsal V; Tarso-metatarsal and<br>metatarso-phalangeal arthrodesis; Tarsal | Minimally invasive reconstruction of fractures and joints;<br>Adjuvant for osteosynthesis in complex joint fractures;<br>Multifragment joint fractures; Simple metaphyseal fractures,<br>Fractures of the wrist, ankle, elbow, and shoulder, Condylar<br>fractures: Osteochrondritis dissecans, Osteochondral<br>Fractures, Ligament avulsion injuries, Ligament fixation,<br>Other small fragment, cancellous bone fractures, Small joint<br>fusion. Areas where accurate screw placement is vital.<br>Metatarsal and phalangeal osteotomies; Fractures of the<br>tarsals, metatarsals and other fractures of the foot; Avulsion<br>fractures and fractures of metatarsal V; Tarso-metatarsal and<br>metatarsophalangeal arthrodesis; Tarsal Fusions; Calcaneal<br>and talar fractures, Subtalar arthrodesis, Ankle arthrodesis.<br>Fractures of small joints, such as: Ankle fractures, Navicular<br>fractures, Fractures of the fibula, malleolus, and calcaneus,<br>Distal tibia and pilon fractures, Acetabular fractures. Other<br>fractures of the pelvic ring; Fractures of the femoral head and | The OrthoSolutions 'System26' cannulated screws (headed and<br>headless compression) and washers are indicated for use over a<br>guide pin or wire for aligned bone fracture repair and arthrodesis,<br>osteotomy, joint fusion and bone fragment fixation appropriate with<br>the size of the screw. Washers of matching size to the headed<br>cannulated bone screw may be used in certain applications for<br>deficient osteopenic bone. The non-cannulated 2.0 (headed, headless<br>compression and Twist-Off) bone screws are applicable, as well, for<br>bone fracture repair and arthrodesis, osteotomy, joint fusion and<br>bone fragment fixation appropriate with the size of the screw.<br>'System26' non-cannulated 2.0mm (headed, headless compression<br>and Twist-Off) bone screws and cannulated (headed and headless<br>compression) bone screw sizes of 2.0mm, 2.5mm 3.0mm are<br>indicated for treating small bone fractures as well as performing<br>osteotomies, arthrodesis and joining cancellous bone fragments in<br>the upper and lower limb and extremities. |
| | Subject Device:<br>Auxano Medical- Intraosseous<br>Fusion Device System | Predicate (K191995):<br>Subchondral Solutions- S4 Screw | Additional Predicate (K163489):<br>OrthoSolutions- System26 Bone Screws |
| | Fusions; Calcaneal and talar fractures,<br>Subtalar arthrodesis. Ankle arthrodesis.<br>Fractures of small joints, such as: Ankle<br>fractures, Navicular fractures, Fractures of<br>the fibula, malleolus, and calcaneus.<br>The Intraosseous Fusion Device System is<br>not intended for spinal use. | neck, Supracondylar femoral fractures, Slipped capital<br>femoral epiphyses, An adjunct to DHIS in basilar neck<br>fractures, Intercondylar femur fractures, Intracapsular<br>fractures of the hip, Fractures of the distal femur and proximal<br>tibia, Patellar fractures, Tibial plateau fractures. Small<br>fragments of the hand and wrist, Fractures of the carpals and<br>metacarpals, Carnal and metacarpal arthrodesis, Scaphoid<br>fracture and other fractures of the hand, Phalangeal and<br>interphalangeal fractures, Fractures of the ulna and radius.<br>Radial head fractures, Fractures of the olecranon and distal<br>humerus, Humeral head fractures, Ligament fixation at the<br>proximal humerus, Glenoid fractures. | 'System26' cannulated bone screw sizes (headless compression and<br>headed with optional matching washer) of 4.0mm, 5.0mm, 6.5mm,<br>and 8.0mm are indicated to be used with large and long bones.<br>Specific indications, which are dependent in part on the diameter of<br>the screw include: Minimally invasive bone fracture/joint<br>reconstructions; Additive osteosynthesis for complex joint fractures;<br>Multiple-fragment joint fractures; Femoral neck and femoral head<br>fractures; Femoral supracondylar fractures; Tibial plateau fractures;<br>Fractures of the head of the humerus; Fractures of the tibia; Cooper<br>fractures of the tibia; Bone fractures of the radius, wrist, ankle,<br>elbow, and shoulder; Ligament fixation of the proximal humerus;<br>Bone fractures of the acetabulum and dorsal pelvic ring; Condylar<br>fractures; Ligament avulsion injuries; Malleolar and navicular<br>fractures; Bone fractures of the calcaneus and talus; Arthrodesis of<br>the ankle joint; Arthrodesis of foot joints and; Avulsion fractures.<br>The OrthoSolutions 'System26' Bone Screws are not intended for<br>spinal use. |
| | Substantial Equivalence: | Similar | Similar |
| Intended Use: | Bone fixation of fractures, fusions<br>(arthrodesis), osteotomies and<br>fragments. | Bone fixation of fractures, fusions (arthrodesis) and<br>osteotomies. | Bone fixation of fractures, fusions (arthrodesis), osteotomies and<br>fragments. |
| | Substantial Equivalence: | Similar | Same |
| Components: | K-wires, sizers, drills, taps, drivers,<br>driver handles | K-wires, countersink, depth gauge, reamer, driver | K-wires, drill guides, drills, countersinks, depth gauges, drivers,<br>driver handles, forceps, clamps, easy-outs |
| | Substantial Equivalence: | Similar | Similar |
{6}------------------------------------------------
{7}------------------------------------------------
| | Subject Device:<br>Auxano Medical- Intraosseous<br>Fusion Device System | Predicate (K191995):<br>Subchondral Solutions- S4 Screw | Additional Predicate (K163489):<br>OrthoSolutions- System26 Bone Screws |
|-------------------------------|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Technical<br>Characteristics: | Threaded<br>Cannulated<br>Headless<br>Fenestrated<br>Interrupted thread | Threaded<br>Cannulated<br>Headless<br>Fenestrated<br>Interrupted thread<br>HA-coated | Threaded<br>Cannulated and Non-cannulated<br>Headed and Headless (Smooth/Threaded) |
| | Substantial Equivalence: | Similar | Similar |
| Materials: | Ti-6Al-4V (ASTM F136 & F3001) | Ti-6Al-4V (ASTM F136)<br>(w/HA coating) | Ti-6Al-4V (ASTM F136) |
| | Substantial Equivalence: | Similar | Same |
{8}------------------------------------------------
## PERFORMANCE DATA:
- . Mechanical testing (torsional strength, driving torque, axial pullout strength) was carried out per ASTM F543 on the subject device and additional predicate, OrthoSolutions System26 bone screws.
- Mechanical testing was carried out per ASTM F2077 on subject device for static and cyclic radial compression.
- . Implant cleanliness validation was carried out at implant vendor (Marle Tangible) and internally following subsequent shipping, inspection and warehouse tasks.
- . GLP cytotoxicity testing was conducted on the IFD implants and instruments.
- . Steam sterilization validation testing was conducted on the IFD system according to ISO 17665-1 and ISO 14937.
- . Manual cleaning validation testing was conducted on the worst-case instrument following simulated use cycles according to AAMI TIR12 and AAMI TIR30.
- Shipping simulation was conducted on the IFD system and packaging according to ISTA 3A.
# CONCLUSION
The subject Intraosseous Fusion Device System has same intended use and similar indications for use, the similar technology, comparable principles of operation, and similar materials as the identified predicate systems: Subchondral Solutions- S4 Screw (K191995) and OrthoSolutions-System26 Bone Screws (K163489). The IFD System does not raise any new questions of safety or effectiveness as compared to the predicate systems. Performance testing according to ASTM F543 demonstrates substantially equivalent mechanical performance of the subject device as compared to the System26 Bone screws currently on the market. Therefore, the Intraosseous Fusion Device System is substantially equivalent to the identified predicate systems.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.