Rotium Bioresorbable Wick

K231641 · Nanofiber Solutions, LLC · MAI · Jul 5, 2023 · Orthopedic

Device Facts

Record IDK231641
Device NameRotium Bioresorbable Wick
ApplicantNanofiber Solutions, LLC
Product CodeMAI · Orthopedic
Decision DateJul 5, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Rotium™ Bioresorbable Wick is intended to be used in conjunction with suture anchors for the reattachment of tendon to bone in rotator cuff repairs.

Device Story

Rotium Bioresorbable Wick is an electrospun, non-woven, microporous, microfiber matrix accessory for rotator cuff repair. Composed of PLCL and PGA polymer fibers. Used in conjunction with suture anchors; placed over the tendon to facilitate tendon-bone reattachment. Intended for single use; supplied gamma-sterilized. Device is identical to predicate; labeling update allows placement on top of the rotator cuff tendon. No changes to technological characteristics or performance profile.

Clinical Evidence

No clinical data. Bench testing only (relying on previous predicate data).

Technological Characteristics

Electrospun, non-woven, microporous, microfiber matrix. Materials: Poly(lactide-co-caprolactone) (PLCL) and Polyglycolic acid (PGA). Supplied gamma-sterilized. Single-use. No changes to technological characteristics from predicate.

Indications for Use

Indicated for patients undergoing rotator cuff repair requiring reattachment of tendon to bone, used in conjunction with suture anchors.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 5, 2023 Nanofiber Solutions, LLC Jason Chakroff QA/RA Manager 5164 Blazer Parkway Dublin, Ohio 43017 Re: K231641 Trade/Device Name: Rotium Bioresorbable Wick Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI Dated: May 31, 2023 Received: June 5, 2023 Dear Jason Chakroff: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Jesse Muir -S Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K231641 Device Name Rotium™ Bioresorbable Wick Indications for Use (Describe) The Rotium™ Bioresorbable Wick is intended to be used in conjunction with suture anchors for the reattachment of tendon to bone in rotator cuff repairs. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| |× Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Nanofiber Solutions. The logo consists of a square graphic to the left of the company name. The graphic is a purple square with white lines inside that create a web-like pattern. The company name is written in a large, purple font, with the word "SOLUTIONS" written in a smaller font below. ## SUBMITTER'S INFORMATION | Owner: | Nanofiber Solutions, LLC | |-------------------------|----------------------------------------------------------------------------| | Address: | 5164 Blazer Parkway<br>Dublin, OH 43017 | | Official Correspondent: | Jason T. Chakroff<br>614-565-4161<br>jason.chakroff@nanofibersolutions.com | | Date Summary Prepared: | May 31, 2023 | ### DEVICE INFORMATION | Name of Device: | Rotium™ Bioresorbable Wick | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common/Usual Name: | Bioresorbable Wick Accessory | | Classification Name: | Fastener, fixation, biodegradable, soft tissue fixation<br>appliances and accessories (21 CFR 888.3030) | | Regulatory Class: | Class II | | Product Code: | MAI | | Primary Predicate Device: | Rotium™ Bioresorbable Wick, K183236, K201414 | | Reason for Submission: | Updating labeling to allow for placement on top of the rotator<br>cuff tendon. | | Indication for Use: | The Rotium™ Bioresorbable Wick is intended to be used in<br>conjunction with suture anchors for the reattachment of tendon<br>to bone in rotator cuff repairs. | | Device Description: | The Rotium™ Bioresorbable Wick is an accessory to be used<br>in conjunction with suture anchors for rotator cuff repair. The<br>wick is placed over the tendon and is designed to facilitate<br>tendon-bone reattachment. The wick is an electrospun, non-<br>woven, microporous, microfiber matrix. The wick is made from<br>two types of polymer fibers: Poly(lactide-co-caprolactone)<br>(PLCL) and Polyglycolic acid (PGA). The wick is packaged in a<br>primary foil pouch with a desiccant pouch, sealed within a<br>secondary Tyvek pouch. The device is supplied gamma-<br>sterilized. The device is single use only. | {4}------------------------------------------------ # 510(K) Summary Technological No changes have been made to the technological Characteristics: characteristics of the device. - Performance Data: No new performance data is required because the labeling changes made to the device will have no impact on the Biocompatibility, Bench Testing, or Animal Testing results of the predicate (K183236, K201414). - Substantial Equivalence: The Rotium™ Bioresorbable Wick accessory is identical to the predicate, Rotium™ Bioresorbable Wick (K183236, K201414). The labeling changes raise no new safety or effectiveness questions. - Conclusion: The Rotium™ Bioresorbable Wick accessory is substantially equivalent to the primary predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...