Disposable Endoscopic Bipolar is a sterile, single use electrosurgical instrument intended to conduct electrosurgical current for coagulation of tissue during general surgical procedures.
Device Story
Disposable Endoscopic Bipolar is a sterile, single-use electrosurgical instrument for tissue coagulation during general surgery. Device consists of handle, pushrod, inner rod, and electrode tips. Operated by professional medical staff; connects to high-frequency surgical generator (e.g., OBS-350A). Principle of operation: bipolar electrocoagulation; electrosurgical current passes between two electrode tips; thermal effect coagulates blood proteins to achieve hemostasis. Handle mechanism uses two-finger pinch to actuate electrode clamping; spring-loaded rebound opens electrodes. Protective insulation and stainless steel armor prevent pushrod friction. Output is localized thermal effect on tissue between electrodes; tissue outside tips remains largely unaffected. Benefits include precise coagulation of small blood vessels and structures.
Clinical Evidence
Bench testing only. Included performance bench tests, mechanical strength verification, and ex-vivo experimental study on thermal effects conducted per FDA guidance for electrosurgical devices. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, and pyrogenicity.
Technological Characteristics
Materials: Stainless steel, plastic. Sensing/Actuation: Mechanical two-finger pinch handle with spring rebound. Energy: Bipolar electrosurgical current (Rated 500Vpk). Dimensions: 3.2mm pushrod diameter, 175mm working length, 0.5-1.2mm tip width. Connectivity: Wired connection to high-frequency surgical generator. Sterilization: Ethylene oxide.
Indications for Use
Indicated for coagulation of tissue during general surgical procedures in patients requiring electrosurgical hemostasis.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
December 5, 2023
Jiangsu Hope Biomedical Science & Technology Co., Ltd. % Kitty Zhang Regulation Affair Staff Shanghai Jiushun Enterprise Management Technology Service Co 15 floor, 25 floor, Bao An Tower, No.800 Dongfang Road Shanghai, Shanghai 200122 China
Re: K231615
Trade/Device Name: Disposable Endoscopic Bipolar Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: November 6, 2023 Received: November 6, 2023
Dear Kitty Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Digitally signed by Mark Trumbore -S Trumbore -Date: 2023.12.05 റ 13:25:49 -05'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231615
Device Name Disposable Endoscopic Bipolar
Indications for Use (Describe)
Disposable Endoscopic Bipolar is a sterile, single use electrosurgical instrument intended to conduct electrosurgical current for coagulation of tissue during general surgical procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### Submitter Information 1.
Company Name: Jiangsu Hope Biomedical Science&Technology Co.,Ltd.
Address: No.3 building, High-tech Innovation Service Center, Jiangdu District,
Yangzhou City, Jiangsu Province, People's Republic of China.
Contact Person(including title): Shaote Geng (Registered Engineer)
Phone Number: +86-514-87185186
Fax Number: +86-514-86445567-8808
E-mail: hoboatfda@hoboat.com
Correspondent Name: Shanghai Jiushun Enterprise Management Technology
Service Co., Ltd.
Correspondent Address: 15 floor,25 floor,Bao An Tower,No.800,Dongfang
Road,200122 Shanghai,China
Correspondent Contact Telephone: +86-21-50931939
Correspondent Contact: Mrs. Kitty Zhang
Correspondent Contact Email: kitty-zhang@isosh.com
#### 2. Subject Device Information
Common Name: Electrosurgical cutting and coagulation device and accessories
Trade Name: Disposable Endoscopic Bipolar
Model: WZX-8XZWZ、WZX-8CZWW
Classification Name: Electrosurgical Cutting and Coagulation Device and
Accessories
Review Panel: General and Plastic Surgery
{5}------------------------------------------------
Product Code: GEI
Regulation Number: 21 CFR 878.4400
Regulation Class: II
#### 3. Predicate Device Information
Sponsor: Günter Bissinger Medizintechnik GmbH
Common Name: Electrosurgical Instrumentation
Trade Name: Bipolar Micro-Coagulation Forceps
510(k) number: K172368
Classification Name: Electrosurgical Cutting and Coagulation Device and
Accessories
Review Panel: General and Plastic Surgery
Product Code: GEI
Regulation Number: 21 CFR 878.4400
Regulation Class: II
### 4. Device Description Summay
1). the mechanism of action The mechanism of action of disposable endoscopic bipolar is to hemostasis of tissue by bipolar electrocoagulation. The disposable endoscopic bipolar can carry out more precise electrocoagulation and treatment of small blood vessels and other structures. The principle of disposable endoscopic bipolar is that the electrosurgical current is released through the two electrode tips, so that the human tissue between the electrode tips of the disposable endoscopic bipolar produces the thermal effect to coagulate proteins in the blood to achieve hemostasis. Due to the mutual insulation of the two electrodes of the disposable endoscopic bipolar, only the electrode tips conduct current. During electrocoagulation, the current passes from one electrode tip to the other, and the tissue between the two electrode tips is affected by the thermal effect, while the
2 / 7
{6}------------------------------------------------
tissue outside the electrode tips is little or not affected, which has a protective effect on the tissue.
2). mechanical structure Disposable endoscopic bipolar consists of electrode, power line, socket, handle group, inner rod, pushrod. Handle group consists of block, set screw, handle, fixed block, rivet pin, rivet cap, connecting rod, sliding block, spring, front insert, upper cover, lower cover, tail sleeve. The inner rod is set in the pushrod and handle group mechanism, and is fixed with set screws and fixed block, and the pushrod is connected to the sliding block. The back end of the push rod is shaped like a semi-circle for easy bending and use. The product is divided in straight and curved based on poshrod shape type. The electrodes are equipped with the insulating sleeve, and the outer side of the sleeve is provided with a stainless steel armor plate to prevent the push rod from rubbing the sleeve. The electrodes are fixed in the inner rod. The electrodes are connected to the socket through power line, and the socket can be connected to the high frequency surgical equipment. Disposable endoscopic bipolar materials: Stainless Steel, Plastic Material. 3). principle of operation and technological characteristics The handle group mechanism of the disposable endoscopic bipolar is a two-finger hand pinch connecting rod structure, and the rebound action is realized by the spring rebound force. Principle of operation: When kneading the handle, the two connecting rods push the sliding block and the pushrod forward, the pushrod pushes the electrode to achieve the clamping action, and then the hand is loosened, and the sliding block is pushed back by the spring, and the electrode is opened by its own elasticity, so that the closure and opening of the electrode are realized. Technological characteristics: Rated Accessory Voltage is 500Vpk. The external diameter of the pushrod is 3.2mm, working length is 175mm. Electrode tips width are 0.5-1.2mm. 4). operating procedures The disposable endoscopic bipolar specifications should be selected according to the clinical needs. Take the sterile pack, tear up and take out the disposable endoscopic bipolar, connect the product to high frequency device
3 / 7
{7}------------------------------------------------
with special connection, the tissue attachments on the electrode tips could be removed with gauze during surgery. Disposable endoscopic bipolar has to be connected to bipolar receptacle of OBS Electrosurgical Generator (Model: OBS-350A) of Baisheng Medical Co.,Ltd. Aseptic operation and related procedures should be strictly followed, also kindly observe the patient signs during surgery. This product should be used by professional medical staff.
| Elements of<br>Comparison | Subject<br>Device | Predicate<br>Device | Result |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Company | Jiangsu Hope<br>Biomedical<br>Science&Techno<br>logy Co.,Ltd | Günter Bissinger<br>Medizintechnik<br>GmbH | / |
| Device Name | Disposable<br>Endoscopic<br>Bipolar | Bipolar<br>Micro-Coagulation<br>Forceps (K172368) | / |
| Model | WZX-8XZWZ,<br>WZX-8CZWW | POWERGRIP 3.0,<br>MITHRAS | / |
| Prescription/OTC | Prescription | Prescription | Same |
| Regulation Number | 21 CFR<br>878.4400 | 21 CFR 878.4400 | Same |
| Code | GEI | GEI | Same |
| Class | II | II | Same |
| Indication for Use | Disposable<br>Endoscopic<br>Bipolar is a<br>sterile, single<br>use<br>electrosurgical<br>instrument<br>intended to<br>conduct<br>electrosurgical<br>current for<br>coagulation of<br>tissue during<br>general surgical<br>procedures | The Bipolar<br>Micro-Coagulation<br>Forceps are<br>intended to remove<br>tissue and control<br>bleeding. | Similar<br>Note1 |
| Single Use | Yes | No | Different<br>Note 2. |
| Device design | Disposable<br>endoscopic<br>bipolars are<br>electrosurgical<br>tools available<br>with different<br>handles and<br>different tip<br>designs. | Bipolar<br>Micro-Coagulation<br>Forceps are<br>electrosurgical<br>tools available with<br>different handles<br>and different tip<br>designs. | same |
| Length | WZX-8CZWW,<br>WZX-8XZWZ:<br>$320\pm50mm$ ; | 30cm | Similar<br>Note 3 |
| Tip Size | $1.2\pm0.3mm$ | 0.35 – 2.8 mm | Similar<br>Note 4 |
| Color | Black | Black | Same |
| Tips material | Medical<br>Stainless Steel | Stainless Steel | Same |
| Maximum peak<br>Voltage(Vp) | $500\pm10%$ | 300 | Different<br>Note 5 |
| Electrical Safety<br>Testing | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | Same |
| Biocompatibility | Cytotoxicity<br>Sensitization<br>Intracutaneous<br>Reactivity<br>Acute systemic<br>toxicity test<br>Pyrogenicity test | Cytotoxicity<br>Sensitization<br>Irritation or<br>Intracutaneous<br>Reactivity | Different<br>Note 6 |
| Sterility | Yes | Non-sterile,<br>reusable | Different<br>Note 7 |
| Sterilization<br>Method | Ethylene oxide | NA | Different<br>Note 7 |
| Shelf Life | 3 years | NA | Different<br>Note 8 |
## 5. General Comparison
{8}------------------------------------------------
{9}------------------------------------------------
|--|--|--|--|
Based on available 510(k) information provided herein, Disposable Endoscopic Bipolar is considered substantially equivalent to the predicate devices in terms of indications for use, material, technology, design and performance specifications. The performed tests show that the subjected device is in compliance with the applicable recognized standards. Performed comparison testing with the predicate device shows substantial equivalence to the FDA-cleared predicate device Sponsor: Jiangsu Hope Biomedical Science&Technology Co.,Ltd Subject Device: Disposable Endoscopic Bipolar Bissinger Medizintechnik GmbH Bipolar Micro-Coagulation Forceps (K172368). There are no differences between the devices which would raise new issues of safety or effectiveness. The Bipolar Micro-Coagulation Forceps is substantially equivalent to predicate device Bissinger Medizintechnik GmbH Bipolar Micro-Coagulation Forceps (K172368).
### 6. Non-Clinical and/or Clinical Tests Summary & Conclusions
To demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified, we follow the Premarket Notification (510(k) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff, and Factory Inspection Instruction for Disposable Endoscopic Bipolar, to conduct the performances bench tests, details please refer to Performance Bench Test Report; Ex-vivo experimental study on Thermal Effects ,was conducted follow Premarket Notification (510(k) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff which refer to Ex-vivo experimental study on Thermal Effects of Disposable Endoscopic Bipolar; Mechanical Strength (refer to Mechanical Strength Verification and Report,included in the EMC, Wireless, & EMT Documentation attachments).
### Conclusion:
{10}------------------------------------------------
The subject device Disposable Endoscopic Bipolar has all features of the predicate device based on its intended use, The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.