Lucitone Digital Print™ 3D Denture Base is a light-cured resin indicated for the fabrication of denture bases including full and partial dentures and implant overdentures, and implant overdentures and other dental appliances. Lucitone Digital Value™ 3D Economy Tooth & Trial Placement is used as a try-in material for evaluation prior to fabrication of the final restoration or a temporary denture. Lucitone Digital IPN™ 3D Premium Tooth is a light-cured resin intended for printing denture teeth. Lucitone Digital Value™ 3D Economy Tooth & Trial Placement is intended for printing full arch and tooth segments. Lucitone Digital Fuse™ Step 1 - 3D Tooth Conditioning Agent is indicated for use in enhancing the bond of denture teeth to denture base and denture base to denture base. Lucitone Digital Fuse™ Step 2 - 3D Denture Bonding Resin is utilized as an aid in bonding denture teeth to denture base as well as repair using traditional techniques. Lucitone Digital Fuse™ Step 3 – Total 3D Sealer is a light-cured sealant that produces a smooth, glossy surface finish on the denture.
Device Story
System comprises methacrylate-based, light-cured resins and ancillary agents for additive manufacturing of dental prostheses; utilizes CAD/CAM workflow; includes denture base resin, tooth resin, try-in/temporary material, conditioning agent, bonding resin, and surface sealer. Operated by dental clinicians/technicians in dental labs or clinical settings. Workflow involves digital design, 3D printing, and light-curing. Output is a finished dental prosthesis (denture base, teeth, or try-in). System enhances bond strength between components and provides smooth surface finish. Benefits include efficient fabrication of custom dental appliances.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including biocompatibility and mechanical property testing per ISO 20795-1:2013.
Technological Characteristics
Methacrylate-based, light-cured resins. Additive manufacturing (3D printing) workflow. Conforms to ISO 20795-1:2013 Type 4. Components: denture base resin, tooth resin, conditioning agent, bonding resin, sealant. Packaging: cartridge. Biocompatibility confirmed via extractables testing.
Indications for Use
Indicated for fabrication of denture bases (full, partial, implant overdentures), denture teeth, and try-in/temporary dentures by dental professionals in laboratory or clinical settings. Includes conditioning, bonding, and sealing agents for denture assembly and repair.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
October 20, 2023
Dentsply Sirona Rebecca Sporer Principle Regulatory Affairs Specialist 221 West Philadelphia Street Suite 60W York, Pennsylvania 17401
Re: K231578
Trade/Device Name: Lucitone Digital Print Denture System Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI, KLE, PZY, EBD Dated: Mav 31, 2023 Received: May 31, 2023
Dear Rebecca Sporer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bobak Shirmohammadi -2
Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and ENT Devices
{2}------------------------------------------------
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K231578
Device Name
Lucitone Digital Print Denture System
Indications for Use (Describe)
· Lucitone Digital Print™ 3D Denture Base is a light-cured resin indicated for the fabrication of denture bases including full and partial dentures and implant overdentures, and implant overdentures and other dental appliances.
· Lucitone Digital Value™ 3D Economy Tooth & Trial Placement is used as a try-in material for evaluation prior to fabrication of the final restoration or a temporary denture.
· Lucitone Digital IPN™ 3D Premium Tooth is a light-cured resin intended for printing denture teeth.
· Lucitone Digital Value™ 3D Economy Tooth & Trial Placement is intended for printing full arch and tooth segments.
· Lucitone Digital Fuse™ Step 1 - 3D Tooth Conditioning Agent is indicated for use in enhancing the bond of denture teeth to denture base and denture base to denture base.
· Lucitone Digital Fuse™ Step 2 - 3D Denture Bonding Resin is utilized as an aid in bonding denture teeth to denture base as well as repair using traditional techniques.
· Lucitone Digital Fuse™ Step 3 – Total 3D Sealer is a light-cured sealant that produces a smooth, glossy surface finish on the denture.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Image /page/4/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized, abstract shape on the left, resembling a leaf or a curved form with a white line running through it. To the right of the shape are the words "Dentsply" on the top line and "Sirona" on the bottom line, both in a simple, sans-serif font.
#### 510(k) SUMMARY
for
Lucitone Digital Print Denture System (K231578)
- 1. Submitter Information: Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17404
Contact Person: Ms. Rebecca Sporer Telephone Number: 717-353-1150 Email address: Corporate-RA@dentsplysirona.com
19 October 2023 Date Prepared:
- 2. Device Name:
| Proprietary Name: | Lucitone Digital Print Denture System |
|----------------------|-----------------------------------------------|
| Common Name: | Denture relining, repairing or rebasing resin |
| Classification Name: | Resin, Denture, Relining, Repairing, Rebasing |
| CFR Number: | 21 CFR 872.3760 |
| Device Class: | II |
| Product Code: | EBI, KLE, EBD, EBG |
# 3. Predicate Device:
| Table FS.1: Predicate and Reference device information | | | | | | |
|--------------------------------------------------------|------------------------------------|---------|-----------------|--|--|--|
| Device Type | Predicate Device Name | 510(k) | Company Name | | | |
| Primary Predicate Device | Halley Resin System | K191427 | Dentsply Sirona | | | |
| Predicate Device | Eclipse Bonding Agent | K051707 | Dentsply Sirona | | | |
| Reference Device | Halley Resin | K190043 | Dentsply Sirona | | | |
| Reference Device | TempFX Esthetic Provisional System | K061264 | Dentsply Sirona | | | |
{5}------------------------------------------------
# 4. Description of Device:
Lucitone Digital Print Denture System is a methacrylate-based, light-cured denture base resin which is intended for use by dental clinician to fabricate dental prostheses using an additive printer and CAD/CAM technologies or as a try-in denture. The workflow utilizes several support components that aid in conditioning, bonding or sealing to provide a smooth finish to the device. The Lucitone Digital Print Denture System conforms to the physical and mechanical property requirements for ISO 20795-1:2013.
## 5. Indications for Use:
Indications for Use for the proposed device are:
- Lucitone Digital Print™ 3D Denture Base is a light-cured resin indicated for the fabrication of ● denture bases including full and partial dentures and implant overdentures, and implant overdentures and other dental appliances.
- Lucitone Digital Value™ 3D Economy Tooth & Trial Placement is used as a try-in material for . evaluation prior to fabrication of the final restoration or a temporary denture.
- . Lucitone Digital IPN™ 3D Premium Tooth is a light-cured resin intended for printing denture teeth.
- . Lucitone Digital Value™ 3D Economy Tooth & Trial Placement is intended for printing full arch and tooth segments.
- Lucitone Digital Fuse™ Step 1 3D Tooth Conditioning Agent is indicated for use in enhancing . the bond of denture teeth to denture base and denture base to denture base.
- Lucitone Digital Fuse™ Step 2 3D Denture Bonding Resin is utilized as an aid in bonding . denture teeth to denture base as well as repair using traditional techniques.
- Lucitone Digital Fuse™ Step 3 Total 3D Sealer is a light-cured sealant that produces a smooth, . glossy surface finish on the denture.
Table FS.2 compares the proposed Indications for Use with those of the predicate and reference devices.
{6}------------------------------------------------
| Table FS.2 - Indications for Use Side-by-Side Comparison | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proposed Device<br>Lucitone Digital Print<br>Denture System<br>(K231578) | Primary<br>Predicate<br>Device<br>Halley Resin<br>System<br>(K191427) | Reference<br>Device<br>Halley Resin<br>(K190043) | Predicate<br>Device Eclipse<br>Bonding Agent<br>(K051707) | Reference Device<br>TempFX<br>Esthetic<br>Provisional<br>System<br>(K061264) | Differences |
| • Lucitone Digital<br>Print 3D Denture<br>Base is a light-cured<br>resin indicated for<br>the fabrication of<br>denture bases<br>including full and<br>partial dentures and<br>implant overdentures<br>and other dental<br>appliances. | Halley resin is a light-cured resin<br>indicated for the fabrication of<br>dentures bases in dental<br>laboratories, including full and<br>partial dentures as well as implant<br>overdentures, and other dental<br>appliances. | | N/A | N/A | Similar.<br>-The proposed device<br>indication does not state "in<br>dental laboratories" as the<br>proposed device is targeted<br>for use by dental<br>professionals in a dental<br>laboratory or a clinical<br>setting.<br>-Product name "Lucitone<br>Digital Print 3D Denture<br>Base" instead of "Halley<br>resin" |
| • Lucitone Digital<br>IPNTM 3D Premium<br>Tooth is a light-cured<br>resin intended for<br>printing denture<br>teeth. | | | N/A | N/A | Different. Additional<br>component in indication.<br>-The additional indication<br>for use falls under the scope<br>of product code PZY<br>(additively manufactured,<br>preformed, resin denture<br>tooth) which is 510(k)<br>exempt device. |
| • Lucitone Digital<br>Value 3D Economy<br>Tooth & Trial<br>Placement is used as<br>a try-in material for<br>evaluation prior to | Halley resin can be used as a try-in<br>material for evaluation prior to<br>fabrication of the final restoration. | | N/A | N/A | Similar.<br>-The proposed Try-in<br>denture can also be used as a<br>temporary denture. The<br>extension of use is supported<br>by physical and mechanical |
| Table FS.2 - Indications for Use Side-by-Side Comparison | | | | | |
| Proposed Device<br>Lucitone Digital Print<br>Denture System<br>(K231578) | Primary<br>Predicate<br>Device<br>Halley Resin<br>System<br>(K191427) | Reference<br>Device<br>Halley Resin<br>(K190043) | Predicate<br>Device Eclipse<br>Bonding Agent<br>(K051707) | Reference Device<br>TempFX<br>Esthetic<br>Provisional<br>System<br>(K061264) | Differences |
| fabrication of the<br>final restoration, or a<br>temporary denture | | | | | properties per ISO 20795-<br>1:2013 Dentistry – Base<br>polymers - Part 1: Denture<br>base polymers;<br><br>-Product name "Lucitone<br>Digital Value 3D Premium<br>Tooth & Trial Placement"<br>instead of "Halley resin" |
| • Lucitone Digital<br>Value™ 3D<br>Economy Tooth &<br>Trial Placement is<br>intended for printing<br>full arch and tooth<br>segments. | | | N/A | N/A | Different. Additional<br>component in indication.<br>-The additional indication<br>for use falls under the scope<br>of product code PZY<br>(additively manufactured,<br>preformed, resin denture<br>tooth) which is 510(k)<br>exempt device. |
| • Lucitone Digital<br>Fuse Step 1- 3D<br>Tooth Conditioning<br>Agent is indicated<br>for use in enhancing<br>the bond of denture<br>teeth to denture base<br>and denture base to<br>denture base. | Halley Denture Tooth Bonding<br>Agent is indicated for the use in<br>enhancing the bond of denture base<br>to denture base. | | Eclipse®<br>Bonding Agent<br>is indicated for<br>use in enhancing<br>the bond of<br>acrylic teeth to<br>acrylic<br>removable<br>denture base. | N/A | Similar.<br>-Additional indication for<br>use for bonding denture<br>teeth to the denture base<br>cleared in the predicate<br>device (K051707) (identical<br>formulation). |
| Table FS.2 - Indications for Use Side-by-Side Comparison | | | | | |
| Proposed Device<br>Lucitone Digital Print<br>Denture System<br>(K231578) | Primary<br>Predicate<br>Device<br>Halley Resin<br>System<br>(K191427) | Reference<br>Device<br>Halley Resin<br>(K190043) | Predicate<br>Device Eclipse<br>Bonding Agent<br>(K051707) | Reference Device<br>TempFX<br>Esthetic<br>Provisional<br>System<br>(K061264) | Differences |
| | | | | | -Product name "Lucitone<br>Digital Fuse Step 1 – 3D<br>Tooth Conditioning Agent"<br>instead of "Halley Denture<br>Tooth Bonding Agent" |
| • Lucitone Digital<br>Fuse Step 2- 3D<br>Denture Bonding<br>Resin in utilized as<br>an aid in bonding<br>denture teeth to<br>denture base as well<br>as repair using<br>traditional<br>techniques. | Halley resin can be utilized as an<br>aid in bonding denture base to<br>denture teeth as well as repair using<br>traditional techniques. | | N/A | N/A | Similar.<br>-Reworded for clarification.<br>-Product name "Lucitone<br>Digital Fuse Step 2 - 3D<br>Denture Bonding Resin"<br>instead of "Halley resin" |
| • Lucitone Digital<br>Fuse Step 3 - Total<br>3D Sealer is a light-<br>cured sealant that<br>produces a smooth,<br>glossy surface finish<br>on the denture. | Halley Sealer is light-cured sealant<br>that produces a smooth, glossy<br>surface finish on the denture. | | N/A | TEMPFX<br>ESTHETIC<br>PROVISIONAL<br>SYSTEM is<br>indicated for<br>fabrication of<br>provisional dental<br>restorations. | Same<br>-Product name "Lucitone<br>Digital Fuse Step 3 – Total<br>3D Sealer" instead of<br>"Halley Sealer" |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# 6. Comparison of Technological Characteristics:
The proposed Lucitone Digital Print Denture System includes the addition of the Primeprint 3D printer workflow. The workflow has been modified to include new packaging (cartridge), the use of a new washing agent, and minor modifications when compared to the primary predicate device workflow (K191427). The overall sequential workflow between the dentist (or dental professional) and the dental laboratory of the proposed Lucitone Digital Print Denture System is identical to the workflow of the primary predicate device Halley Resin System (K1 91427). Although the packaging of the resin was modified, the formulation of the denture base resin and try-in denture resin remain the same as the predicate device (K191427). Both the proposed and predicate device have comparable percent extractables, which confirms the biological safety of the proposed device.
The proposed device was also modified to include cumulative changes made to the predicate device (K191427). There were minor formulation updates to ancillary components of the system in trace percentages. Biological safety testing was performed to confirm the biological safety of the proposed device in comparison with the predicate device.
## 7. Non-Clinical Performance Data
The proposed device was tested and conforms to ISO 20795-1:2013 Type 4 materials and follows FDA guidance documents:
- 1) Denture Base Resins - Performance Criteria for Safety and Performance Based Pathway: Guidance for Industry and Food and Drug Administration Staff (April 2022)
- 2) FDA Guidance Technical Considerations for Additive Manufactured Medical Devices (December 2017). The proposed device meets the same technological requirements as the predicate device (Kl 91427).
# 8. Clinical Performance Data
Not applicable, no data from human clinical studies has been included to support the substantial equivalence of the proposed Lucitone Digital Print Denture System.
# 9. Conclusion
Minor differences in the technological characteristics between the proposed and predicate (K191427) device were evaluated through appropriate performance and biocompatibility testing which demonstrates that the proposed device, when compared to the predicate device,
does not raise new questions regarding safety and effectiveness. Therefore, nonclinical testing data supports the conclusion that the proposed device is substantially equivalent to the predicate device (K191427).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.