Correlation (R) = 0.98 – 0.99 compared to predicate; Test-retest variability (TRT) = 1.2% – 1.8% depending on tracer and reference region.
—
—
Performance evaluation using 2,275 subjects including ADNI data.
>1 (readers)
PET brain amyloid load quantification
—
—
Agreement with histopathology SoT = 94.4% (PET-only) and 96.3% (PET-MR); Agreement with visual read SoT = 94.8% (PET-only) and 92.8% (PET-MR); Centiloid correlation (R2) = 0.85 – 0.99.
—
—
Performance evaluation using 2,275 subjects including ADNI data and Florbetaben Phase III data.
5 (readers)
Indications for Use
cNeuro cPET aids physicians in the evaluation of patient pathologies via assessment and quantification of PET brain scans. The software aids in the assessment of human brain PET scans enabling automated analysis through quantification of tracer uptake and comparison with the corresponding tracer uptake in normal subjects. The resulting quantification is presented using volumes of interest and voxel-based maps of the brain. cNeuro cPET allows the user to generate information relative changes in PET-FDG glucose metabolism. cNeuro cPET additionally allows the user to generate information relative changes in PET brain amyloid load between a subject's images and a normal database, which may be the result of brain neurodegeneration. PET co-registration and fusion display capabilities with MRI allow PET findings to be related to brain anatomy. cNeuro cPET aids physicians in the image interpretation of PET studies conducted on patients being evaluated for cognitive impairment, or other causes of cognitive decline.
Device Story
cNeuro cPET is a software tool for clinicians to analyze PET brain scans. It accepts DICOM PET images (mandatory) and MRI images (optional) from PACS or local folders. The device performs automated registration of a volume-of-interest (VOI) atlas to the PET image, comparing tracer uptake against a database of healthy, amyloid-negative subjects. It calculates Standard Uptake Value Ratios (SUVR), Z-scores, and Centiloids. Outputs include VOI-based metrics, voxel-based maps, and 3D stereotactic surface projections, presented in a PDF report for PACS. The device supports PET-only and PET-MR pipelines. By quantifying glucose metabolism and amyloid load, it assists physicians in interpreting studies for cognitive impairment and neurodegeneration, potentially aiding earlier or more accurate diagnosis.
Clinical Evidence
Performance evaluated using 2,275 PET images (Flutemetamol, Florbetaben, FDG). Comparison with predicate showed correlations (R) of 0.98–0.99. Test-retest variability was low (1.2%–1.8%). Categorization agreement with histopathology-based Standard of Truth was 94.4% (PET-only) and 96.3% (PET-MR); agreement with visual read consensus was 94.8% (PET-only) and 92.8% (PET-MR). Centiloid correlation with Centiloid Project values ranged R2 = 0.85–0.99.
Technological Characteristics
Software-based image processing system. Complies with NEMA PS 3.1-3.20 (DICOM). Features automated registration, VOI-template fitting, and statistical comparison to tracer-specific reference data. Supports PET-only and PET-MR fusion. Outputs include SUVR, Z-scores, and Centiloids. Moderate level of concern.
Indications for Use
Indicated for patients being evaluated for cognitive impairment or other causes of cognitive decline. Aids physicians in assessment and quantification of PET brain scans (FDG, Flutemetamol, Florbetaben, Florbetapir) to evaluate pathologies and neurodegeneration.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
September 14, 2023
Combinostics Oy % Erin Gontang Senior Consultant Ram+ 2251 San Diego Avenue, B-257 San Diego, California 92110
Re: K231576
Trade/Device Name: cNeuro cPET Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: Class II Product Code: KPS, LLZ Dated: May 31, 2023 Received: May 31, 2023
Dear Erin Gontang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Signature
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
k231576
Device Name
cNeuro cPET
Indications for Use (Describe)
cNeuro cPET aids physicians in the evaluation of patient pathologies via assessment and quantification of PET brain scans.
The software aids in the assessment of human brain PET scans enabling automated analysis through quantification of tracer uptake and comparison with the corresponding tracer uptake in normal subjects. The resulting quantification is presented using volumes of interest and voxel-based maps of the brain. cNeuro cPET allows the user to generate information relative changes in PET-FDG glucose metabolism.
cNeuro cPET additionally allows the user to generate information relative changes in PET brain amyloid load between a subject's images and a normal database, which may be the result of brain neurodegeneration.
PET co-registration and fusion display capabilities with MRI allow PET findings to be related to brain anatomy.
cNeuro cPET aids physicians in the image interpretation of PET studies conducted on patients being evaluated for cognitive impairment, or other causes of cognitive decline.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped design made up of blue and red squares on the left, followed by the word "Combinostics" in a dark gray, sans-serif font. The brain design is abstract, with the squares arranged to suggest the shape of a brain.
# 510(k) Summary (K231576)
#### DATE PREPARED
August 17, 2023
#### MANUFACTURER AND 510(k) OWNER
Combinostics OY Hatanpään valtatie 24 FI 33100 Tampere, Finland Telephone: Official Contact: Lennart Thurfjell, CEO
## REPRESENTATIVE/CONSULTANT
Erin A. Gontang, Ph.D. Allison C. Komiyama, Ph.D., RAC RQM+ +1 (412) 816-8253 Telephone: Email: egontang@rqmplus.com akomiyama@rqmplus.com
#### DEVICE INFORMATION
| Proprietary Name/Trade Name: | cNeuro cPET |
|------------------------------|-------------------------------------------------------------------------------------------------|
| Common Name: | System, Image Processing, Radiological<br>System, Tomography, Computed, Emission |
| Regulation Number: | 892.2050<br>892.1200 |
| Class: | Class II |
| Product Code: | LLZ<br>KPS |
| Premarket Review: | Radiology |
| Review Panel: | Office of Radiological Health (OHT8)<br>Div. of Imaging Devices and Electronic Products (DHT8B) |
#### PREDICATE DEVICE IDENTIFICATION
The cNeuro cPET is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|------------------------------------------|-------------------|
| K141074 | CortexID Suite / GE Medical Systems, LLC | ✓ |
The predicate device have not been subject to a design related recall.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped design made up of blue and red squares on the left, followed by the company name "Combinostics" in a dark gray, sans-serif font. The logo is simple and modern, with a focus on the company name.
## DEVICE DESCRIPTION
cNeuro cPET has been developed to aid clinicians in the assessment and quantification of pathologies derived from PET scans. The software enables the display, co-registration, and fusion of PET images with those from MRI. Additionally, cPET enables automated quantitative and statistical analysis of tracer uptake by registration of a volume-of-interest atlas to the PET images and by comparing voxel and region-based uptake with corresponding uptake in healthy, amyloid-negative subjects. There are two quantification pipelines, one when the patient's MRI is not available (PET-only) and one when the patient's MRI is available (PET-MR). Quantification results are presented using volumes of interest, voxel-based or 3D stereotactic surface projection maps of the brain.
#### INDICATIONS FOR USE
cNeuro cPET aids physicians in the evaluation of patient pathologies via assessment and quantification of PET brain scans.
The software aids in the assessment of human brain PET scans enabling automated analysis through quantification of tracer uptake and comparison with the corresponding tracer uptake in normal subjects. The resulting quantification is presented using volumes of interest and voxel-based maps of the brain. cNeuro cPET allows the user to generate information regarding relative changes in PET-FDG glucose metabolism.
cNeuro cPET additionally allows the user to generate information regarding relative changes in PET brain amyloid load between a subject's images and a normal database, which may be the result of brain neurodegeneration.
PET co-registration and fusion display capabilities with MRI allow PET findings to be related to brain anatomy.
cNeuro cPET aids physicians in the image interpretation of PET studies conducted on patients being evaluated for cognitive impairment, or other causes of cognitive decline.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Combinostics believes that the cNeuro cPET is substantially equivalent to the predicate device based on the information summarized here:
The subject device and predicate device are software for the automated quantification, visualization, and reporting of PET FDG and amyloid brain scans. Both devices compute standard uptake value ratios (SUVR) where the value in a target region or voxel is divided with the value in a reference region. In addition, both devices provide comparison to tracer specific
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped icon made up of small squares in red and blue, followed by the word "Combinostics" in a dark gray, sans-serif font. The logo is simple and modern, and the colors are eye-catching.
reference data with results presented as region and voxel-based Z-scores and with results summarized in a report.
The subject device and predicate device achieve their intended use based on a similar principle, which is the quantification system relies on fitting a volume-of-interest atlas to the data, and through statistical comparison to reference data. The main difference being that the subject device performs the analysis in PET native space whereas the predicate device performs the analysis in a standardized template space.
| | Subject Device | Predicate Device |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Combinostics OY | GE Healthcare |
| | cNeuro cPET | CortexID Suite |
| | K231576 | K141074 |
| Indications for Use | cNeuro cPET aids physicians in the<br>evaluation of patient pathologies via<br>assessment and quantification of PET<br>brain scans. | CortexID software has been developed<br>to aid physicians in the evaluation of<br>patient pathologies via assessment and<br>quantification of PET brain scans. |
| | The software aids in the assessment of<br>human brain PET scans enabling<br>automated analysis through<br>quantification of tracer uptake and<br>comparison with the corresponding<br>tracer uptake in normal subjects. The<br>resulting quantification is presented<br>using volumes of interest and voxel-<br>based maps of the brain. cNeuro cPET<br>allows the user to generate information<br>regarding relative changes in PET-FDG<br>glucose metabolism. | The software aids in the assessment of<br>human brain PET scans enabling<br>automated analysis through<br>quantification of tracer uptake and<br>comparison with the corresponding<br>tracer uptake in normal subjects. The<br>resulting quantification is presented<br>using volumes of interest, voxel-based<br>or 3D stereotactic surface projection<br>maps of the brain. The package allows<br>the user to generate information<br>regarding relative changes in the PET-<br>FDG glucose metabolism. |
| | cNeuro cPET additionally allows the user<br>to generate information regarding<br>relative changes in PET brain amyloid<br>load between a subject's images and a<br>normal database, which may be the<br>result of brain neurodegeneration.<br>PET co-registration and fusion display<br>capabilities with MRI allow PET findings<br>to be related to brain anatomy.<br>cNeuro cPET aids physicians in the image<br>interpretation of PET studies conducted<br>on patients being evaluated for | CortexID Suite additionally allows the<br>user to generate information in PET<br>brain amyloid load between a subject's<br>images and a normal database, which<br>may be the result of brain<br>neurodegeneration.<br>PET co-registration and fusion display<br>capabilities with CT and MR allow PET<br>findings to be related to brain anatomy<br>and offers visualization of structural<br>abnormalities, which may result from<br>brain injury, trauma, disorder, disease or |
| | cognitive impairment, or other causes of<br>cognitive decline. | dysfunction, such as subdural<br>hematoma, tumor, stroke, or<br>cerebrovascular disease, etc. |
| | | CortexID Suite may aid physicians in the<br>image interpretation process of PET<br>studies conducted on patients being<br>evaluated for cognitive impairment, or<br>other causes of cognitive decline, and is<br>an adjunct to other diagnostic<br>evaluations. |
| Product Code(s) | LLZ - System, Image Processing,<br>Radiological<br>KPS - System, Tomography, Computed,<br>Emission | KPS - System, Tomography, Computed,<br>Emission |
| Regulation Number/<br>Description | 21 CFR 892.2050 / Medical image<br>management and processing system<br>21 CFR 892.1200 / Emission computed<br>tomography system | 21 CFR 892.1200 / Emission computed<br>tomography system |
| Import of PET and<br>MRI Images | Upload DICOM files from folder or<br>through connection to PACS. PET images<br>are mandatory, but MRI images are<br>optional. | Upload DICOM files from folder or<br>through connection to PACS. PET images<br>are mandatory, but MRI images are<br>optional. |
| Supported tracers | Supports FDG, Flutemetamol,<br>Florbetaben and Florbetapir. | Supports FDG, Flutemetamol,<br>Florbetaben and Florbetapir, and the<br>research tracer C11 PIB. |
| Volumes of interest<br>(VOIs) | A VOI-template is defined outlining VOIs<br>corresponding to the different brain<br>lobes, anterior and posterior cingulate,<br>as well as reference VOIs corresponding<br>to cerebellar gray matter, whole<br>cerebellum and pons. There is also a<br>cortical composite VOI defined by<br>merging all the cortical VOIs. | A VOI-template is defined outlining VOIs<br>corresponding to the different brain<br>lobes, anterior and posterior cingulate,<br>as well as reference VOIs corresponding<br>to cerebellar gray matter, whole<br>cerebellum and pons. There is also a<br>cortical composite VOI defined by<br>merging all the cortical VOIs. |
| Method for<br>quantification | Fully automated registration to establish<br>the transformation between the PET<br>image and a standard template space. A<br>VOI-template is then transformed to PET<br>native space and is used to quantify<br>tracer uptake. | Fully automated registration to establish<br>the transformation between the PET<br>image and a standard template space.<br>The PET image is then transformed<br>standard space and a VOI-template is<br>used to quantify tracer uptake. |
| Comparison to tracer<br>specific reference<br>data | Provides reference data for all<br>supported tracers. The reference data is<br>used to compute regional and voxel-<br>based Z-scores. | Provides reference data for FDG and<br>Flutemetamol. The reference data is<br>used to compute regional and voxel-<br>based Z-scores. |
| Output from<br>quantification | Results are presented as SUVR images,<br>Z-score images, regional SUVR and Z-<br>scores and 3D SSP Z-score images. For<br>amyloid tracers results include | Results are presented as SUVR images,<br>Z-score images, regional SUVR and Z-<br>scores and 3D SSP SUVR and Z-score<br>images |
| | Centiloids for a cortical composite region. | |
| Co-registration with MRI | If the patient's MRI is available, this is co-registered with the PET for display purposes, and it is also used during image quantification. | If the patient's MRI is available, this is co-registered with the PET for display purposes. |
| Report | Provides a PDF report that can be sent to PACS. The report summarizes region-based SUVR and Z-scores and shows 3D stereotactic surface projections of voxel-based Z-scores as well as Centiloids for a cortical composite region for amyloid tracers. | Provides a PDF report that can be sent to PACS. The report summarizes region-based SUVR and Z-scores and shows 3D stereotactic surface projections of voxel-based Z-scores. |
A comparison of the subject device and predicate device is provided below.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped design made up of small squares in blue and red, followed by the word "Combinostics" in a dark gray sans-serif font. The logo is simple and modern, with a focus on the company's name.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a brain-shaped icon made up of small squares in red and blue, followed by the word "Combinostics" in a dark gray sans-serif font. The logo is simple and modern, and the brain icon suggests that the company is involved in some kind of cognitive or analytical work.
## SUMMARY OF PERFORMANCE TESTING
Summary of Non-Clinical Testing
The cNeuro cPET software complies with NEMA PS 3.1 - 3.20 (2021) Digital Imaging and Communications in Medicine (DICOM) Set (Radiology) standard.
cNeuro cPET employs the same fundamental scientific technology as its predicate device, CortexID Suite. cNeuro cPET uses the equivalent DICOM image data input requirements. It has equivalent display, formatting, archiving and visualization technologies compared to the predicate device. cNeuro cPET utilizes essentially the same methodology to quantify and assess uptake of FDG and amyloid tracers. The information is presented using volumes of interest or voxel-based maps of the brain. cNeuro cPET utilizes existing PET co-registration and fusion technologies and provides for PET & MR co-registration and fusion capabilities. Thorough testing of these capabilities has not raised any safety or effectiveness issues.
The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- Requirements Reviews
- Design Reviews
- Integration testing (System verification)
- Performance testing (Bench testing, verification)
- o Safety testing (Verification)
The substantial equivalence determination is based on the software documentation for a MODERATE level of concern device.
Combinostics performed testing of cNeuro cPET using PET images from 2,275 subjects. Data were from all tracers supported by cNeuro cPET, i.e., Flutemetamol, Florbetaben and FDG. The performance evaluation included a comparison with the predicate device,
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Combinostics. The logo consists of a stylized brain made of blue and red squares on the left, followed by the word "Combinostics" in a dark gray sans-serif font. The brain graphic is composed of multiple squares of varying sizes, arranged to resemble the shape of a brain.
assessment of robustness using a large dataset from the Alzheimer's Disease Neuroimaging Initiative (ADNI), evaluation of test-retest variability, and comparison with Standard of Truth (SoT) based on majority/consensus visual read and histopathology using Florbetaben Phase III data. Centiloids for amyloid tracers were validated according the methods described by W. E. Klunk et al., The Centiloid Project: standardizing quantitative amyloid plaque estimation by PET. Alzheimers. Dement. 11, 1-4 (2015).
Results showed agreement with the predicate device with correlations (R) ranging from 0.98 – 0.99 depending on cohort. Furthermore, a similar strong correlation (R=0.98 – R=0.99) was obtained for the comparison with a FreeSurfer based PET-MR quantification method used in ADNI. Evaluation of the consistency of cNeuro cPET quantification showed low test-retest variability (TRT) in the quantification of repeat scans. For Flutemetamol, the best results were obtained using Pons as the reference region (TRT = 1.2% for PET-MR, TRT = 1.7% for PET-only), and for Florbetapir, the lowest variability was obtained by using the whole cerebellum as the reference region (TRT = 1.3% for PET-MR, TRT = 1.8% for PET-only).
Categorization of Florbetaben scans into positive and negative based on cNeuro cPET quantification showed agreement with SoT based on histopathology (94.4% for PET-only and 96.3% for PET-MR). The agreement in the comparison of cNeuro cPET categorization with SoT based on majority visual read (5 readers) gave similar agreement (94.8% for PET-only and 92.8% for PET-MR). When excluding all scans where readers had different opinions, i.e., there was unanimous categorization by the readers of the scans into negative or positive, the agreement with cNeuro cPET categorization was 98.8% for PET-only and 98.2% for PET-MR.
The correlation of cNeuro cPET Centiloids for Flutemetamol, Florbetaben and Florbetapir with the corresponding values produced by the Centiloid Project ranged from R2 = 0.85 - 0.99 depending on tracer and reference region.
The results show that cNeuro cPET is accurate and robust, as well as effective and is suitable for its intended use. cNeuro cPET is substantially equivalent to the predicate device.
## CONCLUSION
Based on performance testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed cNeuro cPET are assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.