Multi-Unit DAS System abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
Device Story
Multi-Unit DAS System consists of titanium alloy (ASTM F136) multi-unit abutments and Ti-Base copings designed to support single or multi-unit dental prostheses. The system functions as a two-piece abutment: a straight multi-unit abutment threads directly into compatible OEM dental implants; a Ti-Base coping attaches to the abutment; a zirconia superstructure (ISO 13356) is bonded to the Ti-Base using G-CEM LinkAce™ cement. The device is used in dental clinics by clinicians. Compatibility with specific OEM implant systems (Astra Tech, Biomet 3i, MegaGen, NobelActive, Zimmer) is achieved via reverse engineering. Digitally designed custom abutments are manufactured at a Talladium-validated milling center. The system provides functional and esthetic rehabilitation for edentulous patients. Mechanical performance is supported by ISO 14801 testing; biocompatibility is supported by ISO 10993-5 testing.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence is supported by bench testing, including reverse engineering analysis of implant compatibility, mechanical testing per ISO 14801, biocompatibility testing (ISO 10993-5), and MR environment safety analysis.
Technological Characteristics
Materials: Titanium alloy (ASTM F136/ISO 5832-3) for abutments/screws; Zirconia (ISO 13356) for superstructures. Design: Two-piece system (multi-unit abutment + Ti-Base). Sterilization: Non-sterile, validated moist heat cycle (ANSI/AAMI/ISO 17665-1). Connectivity: None (mechanical).
Indications for Use
Indicated for partially or fully edentulous patients requiring dental implant-supported single-unit or multi-unit prostheses in the maxillary or mandibular arch.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
November 17, 2023
Talladium España, SL % Kevin Thomas Vice President & Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real. Suite 400 San Diego, California 92130
Re: K231559
Trade/Device Name: Multi-Unit DAS System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous dental implant abutment Regulatory Class: Class II Product Code: NHA Dated: October 20, 2023 Received: October 20, 2023
Dear Kevin Thomas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices
{2}------------------------------------------------
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231559 Device Name
#### Multi-Unit DAS System
Indications for Use (Describe)
Multi-Unit DAS System abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
Compatible Implant Systems
| Implant Compatibility | Implant Body Diameter, mm | Implant Platform, mm |
|-------------------------------|---------------------------|----------------------|
| Astra Tech OsseoSpeed TX | 3.5 | 3.5/4.0 |
| Astra Tech OsseoSpeed TX | 4.0 | 3.5/4.0 |
| Biomet 3i OSSEOTITE® Certain® | 3.25 | 3.4 |
| Biomet 3i OSSEOTITE® Certain® | 4.0 | 4.1 |
| MegaGen AnyRidge | 3.5, 4.0, 4.4, 4.9, 5.4 | 3.5 |
| NobelActive® | 3.5 | 3.5 (NP) |
| NobelActive® | 4.3, 5.0 | 3.9 (RP) |
| Zimmer Tapered Screw-Vent® | 3.7, 4.1 | 3.5 |
All digitally designed custom abutments for use with Multi-Unit DAS System abutments are to be sent to a Talladium validated milling center for manufacture.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K231559 Talladium España, SL Multi-Unit DAS System
November 16, 2023
## ADMINISTRATIVE INFORMATION
| Manufacturer Name | Talladium España, SL<br>Virginia Woolf, 17<br>Lleida, Lleida, ES 25005<br>Telephone +34 973-289-580 | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Official Contact | Xavier Soca Filella, General Manager | |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone +1 858-792-1235<br>Fax +1 858-792-1236<br>Email kthomas@paxmed.com<br>flarson@paxmed.com | |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Multi-Unit DAS System |
|------------------------|---------------------------------------------------------------|
| Common Names | Endosseous dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1 |
| | (Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices) |
| Reviewing Division | Division of Dental and ENT Devices |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K221966, Dynamic TiBase, Talladium España, SL
Reference Devices K230143, DESS® Dental Smart Solutions, Terrats Medical SL K220200, Paltop Conical Implant System, Paltop Advanced Dental Solutions, Ltd K101732, Astra Tech Implant System, Astra Tech AB K063341, 3i OSSEOTITE® Certain® Dental Implants, Implant Innovations, Inc. K140091, Xpeed AnyRidge Internal Implant System, MegaGen Implant Company, Ltd. K110955, AnyRidge Internal Implant System, MegaGen Implant Company, Ltd.
{5}------------------------------------------------
K142260, NobelActive®, Nobel Biocare AB K013227, Screw Vent Implant; Tapered Screw Vent Implant, Sulzer Dental, Inc. K072589, Tapered Screw-Vent Implant, 4.1mmD, Zimmer Dental, Inc.
#### INDICATIONS FOR USE STATEMENT
Multi-Unit DAS System abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
| Implant Compatibility | Implant Body Diameter, mm | Implant Platform, mm |
|-------------------------------|---------------------------|----------------------|
| Astra Tech OsseoSpeed TX | 3.5 | 3.5/4.0 |
| Astra Tech OsseoSpeed TX | 4.0 | 3.5/4.0 |
| Biomet 3i OSSEOTITE® Certain® | 3.25 | 3.4 |
| Biomet 3i OSSEOTITE® Certain® | 4.0 | 4.1 |
| MegaGen AnyRidge | 3.5, 4.0, 4.4, 4.9, 5.4 | 3.5 |
| NobelActive® | 3.5 | 3.5 (NP) |
| NobelActive® | 4.3, 5.0 | 3.9 (RP) |
| Zimmer Tapered Screw-Vent® | 3.7, 4.1 | 3.5 |
#### Compatible Implant Systems
All digitally designed custom abutments for use with Multi-Unit DAS System abutments are to be sent to a Talladium validated milling center for manufacture.
### SUBJECT DEVICE DESCRIPTION
Multi-Unit DAS System abutments are designed for retention of single-unit and multi-unit restorations. The multi-unit abutments are provided in a straight design (no angulation in the base portion) that threads directly to the OEM implant. For each of the compatible OEM implant lines, the multi-unit abutments are provided with gingival heights ranging from 1 mm to 5 mm, a prosthetic platform diameter of 4 mm, and a prosthetic post height of 1.4 mm. The multi-unit abutments are the base of a two-piece abutment. The second piece is a metal coping, called a Ti-Base in this submission, that attaches to the multi-unit abutment (not directly to an implant). For permanent restorations a zirconia superstructure is attached to the Ti-Base, and additional gingival height and angulation may be provided in the zirconia superstructure.
The subject device metal copings (Ti-Bases) include: a straight, prepable design with an additional gingival height of 1.5 mm and a prepable 9 mm prosthetic post; and Dynamic Ti-Bases in three (3) designs, each with an additional gingival height of 0.5 mm and a cut-out in the prosthetic post to accommodate a restoration with an angled screw channel when clinically necessary. The Dynamic Ti-Base prosthetic post heights are 4.5 mm (maximum height) / 3.0 mm (cut-out height), and 9.0 mm/3.5 mm. Multi-Unit DAS System Ti-Bases with a 9.0 mm post height may be shortened to no less than 4 mm for a single-unit restoration. The prepable Ti-Base has a platform diameter of 4 mm (platform to the multi-unit abutment) and a prosthetic platform diameter of 4 mm. The Dynamic Ti-Bases have a platform diameter of 4 mm (platform to the multi-unit abutment) and a prosthetic platform diameter of 4.15 mm.
The compatibility between the subject device abutments and the OEM implants listed in the Indication for Use Statement was established by reverse engineering analysis of the OEM implants, OEM abutments, and OEM abutment screws.
All subject device abutments and abutment screws are made of titanium alloy (Ti-6Al-4V) conforming to ASTM F136 and ISO 5832-3.
{6}------------------------------------------------
510(k) Summary Page 3 of 8
All zirconia copings (superstructures) for use with the subject device Dynamic Ti-Base will be made at a Talladium España, SL validated milling center under FDA quality system regulations, and the material will conform to ISO 13356.
The design parameters for the CAD-CAM zirconia superstructure for the Multi-Unit DAS System are:
Minimum wall thickness - 0.25 mm
Minimum post height for single-unit restorations - 4.0 mm
Maximum gingival height in the zirconia superstructure - 5.24 mm for compatible Biomet 31 OSSEOTITE® Certain®, MegaGen AnyRidge, NobelActive®, and Zimmer Tapered Screw-Vent® implants; 5.76 mm for compatible Astra Tech OsseoSpeed TX implants
Minimum gingival height - 0.5 mm (in the Ti-Base)
Maximum angulation - 30° for compatible Biomet 3i OSSEOTITE® Certain®, MegaGen AnyRidge, NobelActive®, and Zimmer Tapered Screw-Vent® implants; 25° for compatible Astra Tech OsseoSpeed TX implants
The recommended cement for bonding the zirconia superstructure to the Dynamic TiBases to create the final two-piece abutment is G-CEM LinkAce™, cleared as GAM-200 in K120243.
### PERFORMANCE DATA
Non-clinical data submitted or referenced to demonstrate substantial equivalence included:
- . provided in this submission was moist heat sterilization for subject devices provided non-sterile to the end user, validated to a sterility assurance level of 105 by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2;
- provided in this submission was reverse engineering analysis (of OEM implants, OEM abutments, and OEM abutment screws) demonstrating compatibility between the subject device abutments and the OEM implants listed in the Indication for Use Statement;
- . provided in this submission was mechanical testing conducted according to ISO 14801 to support the performance of the subject device abutments in conjunction with the compatible OEM implants;
- referenced from K221966 (provided in K212108) was biocompatibility testing according to ISO 10993-5 ● (cytotoxicity) covering the finished subject devices manufactured from titanium alloy (ASTM F136) with cemented zirconia superstructures (ISO 13356); and
- . non-clinical analysis also was performed to evaluate the subject devices (including all abutments, abutment screw, and materials) in the MR environment using scientific rationale and published literature (TO Woods, JG Delfino, and S Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Allovs Used in Medical Devices, " Journal of Testing and Evaluation, Volume 49, No. 2, 2021, pp. 783-795); the analysis addressed parameters per the FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (issued May 2021) including magnetically induced displacement force and torque.
No clinical data were included in this submission.
## EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device K221966, and the reference device K230143.
{7}------------------------------------------------
The primary predicate device K221966 is for support of substantial equivalence in terms of abutment designs, materials, manufacturing, biocompatibility, and sterilization. The reference device K230143 is for support of substantial equivalence in terms of abutment designs. The reference devices K101732, K063341, K140091, K110955, K142260, K013227, and K072589 are for the OEM implant sizes.
The subject device abutments are substantially equivalent in intended use to the abutments cleared in primary predicate device K221966 and the reference device K230143. All are intended for use with endosseous dental implants in the maxilla and mandible to provide functional and esthetic rehabilitation. The Indications for Use Statement (IFUS) for the subject device is identical to that of the primary predicate device K221966 except for the marketing names and the list of compatible OEM implants. The IFUS for the subject device also is similar to that of the reference device K230143 except for the marketing names and the list of compatible OEM implants.
All subject device multi-unit abutments are similar in design, materials, and technological characteristics to the multi-unit abutments cleared in the primary predicate device K221966 and the reference device K230143. The subject device and the reference device K220200 include multi-part abutments (finished device comprising a non-engaging multi-unit abutment, a metal coping, and a zircoma superstructure) for single-unit restorations.
The subject device abutments and prosthetic components are manufactured from identical materials, in the identical facilities using the identical manufacturing processes as used for Talladium España, SL abutments and abutment screws previously cleared in the primary predicate device K221966.
The subject device multi-unit abutments and the multi-unit abutments in the reference device K230143 have similar or identical ranges of implant-abutment platform diameter, prosthetic platform diameter, and gingival height.
The subject device Ti-Base components are to be used with a zirconia superstructure with similar design parameters as the primary predicate K221966. The zirconia material and recommended cement to bond the superstructure to the prosthetic component is the same as the primary predicate K221966.
All abutment screws are similar in design, materials, and technological characteristics to those cleared in the primary predicate device K221966, except for the threads and lengths.
Subject device components that are provided non-sterilized by the same moist heat cycle as in the primary predicate K221966. The subject devices are provided in pouches manufactured from a polyethylene terephthalate (PET) and cast polypropylene (CPP) laminate, identical to the primary predicate K221966.
The subject device Multi-Unit Abutments are to be used with the subject device Ti-Bases to create the final two-piece abutment. The risks associated with the use of an angled zirconia superstructure bonded to the subject device Ti-Bases attached to the subject device Multi-Unit Abutments were mitigated by mechanical testing performed according to ISO 14801.
Minor differences in the designs, dimensions, sizes, or compatible implant lines among the subject device, the primary predicate device K221966, and the reference device K230143 do not affect substantial equivalence. These minor differences do not impact safety or effectiveness because these differences are related to the compatible implant designs and are mitigated by the mechanical performance testing.
{8}------------------------------------------------
#### CONCLUSION
The subject device, the primary predicate device, and the reference device have the same intended use, have similar technological characteristics, and are made of identical or similar materials. The subject device and the primary predicate device encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate device listed above.
{9}------------------------------------------------
# Substantial Equivalence – Indications for Use Statement
| | Indications for Use Statement | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Subject Device | | |
| K231559<br>Multi-Unit DAS System<br>Talladium España, SL | Multi-Unit DAS System abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or | |
| Compatible Implant Systems | | |
| Implant Compatibility | Implant Body Diameter, mm | Implant Platform, mm |
| Astra Tech OsseoSpeed TX | 3.5<br>4.0 | 3.5/4.0<br>3.5/4.0 |
| Biomet 3i OSSEOTITE® Certain® | 3.25<br>4.0 | 3.4<br>4.1 |
| MegaGen AnyRidge | 3.5, 4.0, 4.4, 4.9, 5.4 | 3.5 |
| NobelActive® | 3.5<br>4.3, 5.0 | 3.5 (NP)<br>3.9 (RP) |
| Zimmer Tapered Screw-Vent® | 3.7, 4.1 | 3.5 |
| All digitally designed custom abutments for use with Multi-Unit DAS System abutments are to be sent to a Talladium validated milling center for manufacture. | | |
| Primary Predicate Device | | |
| K221966<br>Dynamic TiBase<br>Talladium España, SL | Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edem | |
| | | |
| Implant Compatibility | Implant Body Diameter, mm | Implant Platform, mm |
| Kontact™ Dental Implant System | 3.6<br>4.2<br>4.8<br>5.4 | 3.6<br>4.2<br>4.8<br>5.4 |
| All digitally designed custom abutments for use with Dynamic TiBase abutments are to be sent to a Talladium validated milling center for manufacture. | | |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
ully edentulous patient.
ulous patient.
{10}------------------------------------------------
# Substantial Equivalence – Indications for Use Statement
| Indications for Use Statement | | |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Reference Device | | |
| K231043<br>Terrats Medical SL<br>DESS® Dental Smart Solutions | DESS Multi-Unit Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. | |
| Compatible Implant Systems | | |
| Compatible Implant Systems | Implant Body Ø, mm | Implant Platform Ø, mm |
| Internal Hex Connection | | |
| Legacy1 | 3.7 | 3.5 |
| | 4.2 | 3.5 |
| | 4.7 | 4.5 |
| Legacy2, simplyLegacy2,<br>Legacy3, simplyLegacy3,<br>Legacy4 | 3.7 | 3.5 |
| | 4.2 | 3.5 |
| | 4.7 | 4.5 |
| | 5.2 | 4.5 |
| Internal Conical Connection | | |
| InterActive | 3.2 | 3.0 |
| | 3.7 | 3.0 |
| | 4.3 | 3.4 |
| | 5.0 | 3.4 |
| Simply Iconic™ | 3.2 | 3.0 |
| | 3.7 | 3.0 |
| | 4.2 | 3.0 |
| | 4.7 | 3.0 |
| | 4.7 | 3.4 |
| | 5.2 | 3.4 |
| | 5.7 | 3.4 |
{11}------------------------------------------------
# Table of Substantial Equivalence – Technological Characteristics
| Comparison | Subject Device | Primary Predicate Device | Reference Device |
|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| | K231559<br>Multi-Unit DAS System<br>Talladium España, SL | K221966<br>Dynamic TiBase<br>Talladium España, SL | K231043<br>Terrats Medical SL<br>DESS® Dental Smart Solutions |
| Reason for Predicate Device | Not applicable | Abutment designs, materials, manufacturing, biocompatibility, and<br>sterilization | Abutment designs |
| Product Codes | NHA | NHA | NHA |
| Intended Use | Functional and esthetic rehabilitation of the edentulous mandible or maxilla | Functional and esthetic rehabilitation of the edentulous mandible or maxilla | Functional and esthetic rehabilitation of the edentulous mandible or maxilla |
| Abutment Designs | | | |
| Abutment Types | Multi-unit, straight (0°) | CAD-CAM Titanium Base Abutments | Multi-unit, straight (0°), 17°, 30° |
| Prosthesis Attachment | Screw retained | Cement-retained, screw-retained | Screw retained |
| Restoration | Multi-unit | Single-unit and multi-unit | Multi-unit |
| Prosthetic Interface Connections | Internal | Internal | Internal |
| Implant-Abutment Platform Diameter | 3.4 mm - 4.1 mm | 3.6 mm - 5.4 mm | 3.0 mm - 4.5 mm |
| Prosthetic Platform Diameter | 4.0 mm, Multi-unit Abutments<br>4.15 mm Ti-Base for Multi-unit Abutments | 4.0 mm, 4.5 mm | 4.8 mm |
| Gingival Height | 1 mm - 5 mm | 1 mm (minimum) | 1 mm - 5 mm |
| Abutment Angulation, degrees | Straight (0°) | Straight (0°) | Straight (0°), 17°, 30° |
| Zirconia Superstructure Design Parameters | | | |
| Minimum wall thickness | 0.25 mm | 0.55 mm | Not applicable |
| Minimum post height for single-unit restorations | 4.0 mm | 4.0 mm | Not applicable |
| Minimum gingival height (GH) of the<br>superstructure | 0 mm; all Ti-Bases have minimum GH 0.5 mm | 0 mm; all bases have minimum GH of 1 mm | Not applicable |
| Maximum gingival height (of the superstructure) | 5.24 mm for Biomet 3i OSSEOTITE® Certain®, MegaGen AnyRidge,<br>NobelActive®, and Zimmer Tapered Screw-Vent® implants<br>5.76 for Astra Tech OsseoSpeed TX implants | 5.37 mm | Not applicable |
| Angulation of Finished Abutment | Up to 30° for Biomet 3i OSSEOTITE® Certain®, MegaGen AnyRidge,<br>NobelActive®, and Zimmer Tapered Screw-Vent® implants<br>Up to 25° for Astra Tech OsseoSpeed TX implants | Up to 30° | Not applicable |
| Cement to bond zirconia superstructure to the<br>abutment base | G-CEM LinkAce™ (cleared as GAM-200 in K120243) | G-CEM LinkAce™ (cleared as GAM-200 in K120243) | Not applicable |
| Abutment Material | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136 | Ti-6Al-4V ELI |
| Abutment Screw Material | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136 | Ti-6Al-4V ELI |
| Superstructure Material | Zirconia, ISO 13356 | Zirconia, ISO 13356 | Not applicable |
| How Provided | | | |
| Sterilization | Non-sterile | Non-sterile | Non-sterile |
| Usage - All Components | Single patient, single use | Single patient, single use | Single patient, single use |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.