K231514 · Daeshin Enterprise Co., Ltd. · GEX · Feb 14, 2024 · General, Plastic Surgery
Device Facts
Record ID
K231514
Device Name
Azure
Applicant
Daeshin Enterprise Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 14, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Azure(DS-22CO) is intended for use in non-fractionated mode for Incision, Excision, Ablation, Vaporization, and Coagulation of Human body soft Tissues in Dermatology, Plastic surgery, General Surgery, Gynecology, Neurosurgery, and in Podiatry.
Device Story
Azure is a CO2 laser surgical instrument (10.6µm wavelength) used for soft tissue procedures. System components include a main body with touch LCD, articulated arm, handpiece, foot switch, and power cord. The operator uses the touch interface to configure parameters and the foot switch to trigger laser output. The device delivers laser energy to the target tissue to perform surgical tasks. It is intended for use by clinicians in clinical settings. The device provides visual feedback via the monitor and allows for precise tissue interaction, aiding in surgical precision and patient outcomes.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), usability (IEC 60601-1-6), and performance testing (IEC 60601-2-22).
Technological Characteristics
CO2 laser surgical instrument; 10,600nm wavelength; 30W (CW) / 15J (Ultra) max energy; 1Hz-500Hz pulse frequency; 60µs-900µs pulse duration; articulated arm delivery; 5mW 650nm InGaAIP aiming laser; LCD touch screen interface. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of human soft tissues in dermatology, plastic surgery, general surgery, gynecology, neurosurgery, and podiatry.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 14, 2024
Daeshin Enterprise Co., Ltd. % Seung Ahn CEO Chemron FDA Korea S-303, S-304, 338, Hakdong-ro Gangnam-gu. Seoul 06099 Korea, South
Re: K231514
Trade/Device Name: Azure Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 12, 2024 Received: January 12, 2024
Dear Seung Ahn:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Shlomit Shlomit Halachmi -S Date: 2024.02.14 Halachmi -S 15:38:39 -05'00'
Shlomit Halachmi
for
Tanisha Hithe Assistant Director
{2}------------------------------------------------
DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231514
Device Name Azure
Indications for Use (Describe)
Azure(DS-22CO) is intended for use in non-fractionated mode for Incision, Excision, Ablation, Vaporization, and Coagulation of Human body soft Tissues in Dermatology, Plastic surgery, General Surgery, Gynecology, Neurosurgery, and in Podiatry.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1H2zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#401, 402 Woolim e-Biz Center, 28, Digital-ro 33-gil, Guro-gu, Seoul, Republic of Korea Tel: +82-2-2108-2198 / Fax: +82-2-2108-2181
### 510(k) Summary
510(k) number: K231514
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date Prepared: Feb 13, 2024
I. Submitter
Submitter DAESHIN ENTERPRISE CO., Ltd. Kyu Kim #401, 402 Woolim e-Biz Center, 28, Digital-ro 33-gil, Guro-gu Seoul, Republic of Korea Email: dse khw@dselaser.com Tel: +82-2-2108-2198 Fax: +82-2-2108-2181
Official Correspondent Chemron FDA Korea Seung Hyun Ahn S-303, S-304, 338, Hakdong-ro, Gangnam-gu, Seoul, Republic of Korea Email: chemronsayou@naver.com / ahn 3060@naver.com Tel: +82-2-568-7744 / Fax: +82-2-543-4746
#### II. Subject Device
- Trade (Proprietary) Name: Azure
- Manufacturer: DAESHIN ENTERPRISE CO., Ltd.
- Common Name: Powered Laser Surgical Instrument
- Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
- Product Code: GEX
- Review Panel: General & Plastic Surgery
- Regulation Number: 21 CFR 878.4810
- Device Classification: Class II
- Submission Type: 510(k) Traditional
{5}------------------------------------------------
## DAESHIN ENTERPRISE CO., Ltd. #401, 402 Woolim e-Biz Center, 28, Digital-ro 33-gil, Guro-gu, Seoul, Republic of Korea Tel: +82-2-2108-2198 / Fax: +82-2-2108-2181
### III. Predicate Device
The subject device is substantially equivalent to the following predicate device, which is legally marketed.
| Predicate Device | 510(k) Number | Company Name |
|------------------|---------------|--------------|
| Finexel | K213557 | SNJ Co., Ltd |
### IV. General Description
Azure consists of main body, articulated arm, handpiece, foot switch, power cord, protective glasses. In a main body, there are a monitor displaying all information needed by an operator at a glance and a touch LCD manipulated by simple hand touch, which facilitates operation of the equipment at user's convenience. The laser output is released by the foot switch. This laser operation device uses CO2 medium based wave of 10.6µm.
### V. Indications for Use
Azure(DS-22CO) is intended for use in non-fractionated mode for Incision, Ablation, Vaporization, and Coagulation of Human body soft Tissues in Dermatology, Plastic surgery, General Surgery, Gynecology, Neurosurgery, and in Podiatry.
### VI. Comparison of Technological Characteristics with the Predicate Device
Azure has a substantially equivalent to the predicated device with respect to the intended use and basic technological characteristics. Please refer to the next page for the comparison table.
{6}------------------------------------------------
## #401, 402 Woolim e-Biz Center, 28, Digital-ro 33-gil, Guro-gu, Seoul, Republic of Korea Tel: +82-2-2108-2198 / Fax: +82-2-2108-2181
| Category | Predicate Device<br>(SNJ Co., Ltd) | Subject Device<br>(DAESHIN ENTERPRISE<br>CO., Ltd.) | Decision | Explanation |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Finexel | Azure | - | - |
| Manufacturer | SNJ Co., Ltd | DAESHIN ENTERPRISE CO.,<br>Ltd. | - | - |
| 510K Number | K213557 | K231514 | - | - |
| Classification<br>Product Code | GEX | GEX | Same | - |
| Regulation<br>Number | 21 CFR 878.4810 | 21 CFR 878.4810 | Same | - |
| Class | Class II | Class II | Same | - |
| Intended Use | Finexel CO2 laser system is<br>intended for use isn non-<br>fractionated mode for<br>Incision, Excision, Ablation,<br>Vaporization, and<br>Coagulation of Human body<br>soft Tissues in Dermatology,<br>Plastic surgery, General<br>Surgery, Gynecology,<br>Neurosurgery, and in<br>Podiatry. | Azure(DS-22CO) is intended<br>for use in non-fractionated<br>mode for Incision, Excision,<br>Ablation, Vaporization, and<br>Coagulation of Human body<br>soft Tissues in Dermatology,<br>Plastic surgery, General<br>Surgery, Gynecology,<br>Neurosurgery, and in<br>Podiatry. | Same | - |
| Dimension and<br>Weight<br>of Device | 350(W) x 400(L) x 1000(H)<br>mm<br>Approx. 50kg | 330(W) x 330(L) x 1060(H)<br>mm<br>Approx. 45kg | Different | The dimensions and weight of<br>the subject device are different<br>from those of the predicate<br>device. However, the dimension<br>and weight difference is only in<br>physical specification, which will<br>not raise any issues in safety<br>and effectiveness. |
| Wavelength | 10,600nm | 10,600nm | Same | - |
| Laser Type<br>(Source) | CO2 Laser | CO2 Laser | Same | |
| Pulse<br>Frequency | N/A | 1Hz ~ 500Hz | Different | The pulse frequency of the subject device is different from that of the predicate device, but it falls within the range of values for the predicate device. Therefore, it will not raise any issues in safety and effectiveness. |
| Pulse Duration | N/A | 60 $\mu$ s ~ 900 $\mu$ s | Different | The pulse duration of the subject device is different from that of the predicate device, but it falls within the range of values for the predicate device. It will not raise any issues in safety and effectiveness. |
| Energy (max) | N/A | 30W (CW mode)<br>15J (Ultra mode) | Different | It is different from the Predicate Device, but it is included within the range of the value the Predicate Device and it will not raise any issues in safety and effectiveness. |
| Spot Size | N/A | 0.4mm(100mm Handpiece)<br>0.2mm(50mm Handpiece)<br>0.2~1.0mm(Zoom<br>Handpiece) | Slightly<br>Different | The difference in spot size is very small, approximately 0.2mm, so it will not affect safety or effectiveness. |
| Fluency<br>(Energy<br>Density) | N/A | 72243J/cm²<br>(cw, 50mm) | Different | Fluency of the subject device is different from that of the predicate device, but it falls within the range of values for the predicate device. It will not raise any issues in safety and effectiveness. |
| Aiming laser | 5mW, 650nm, InGaAIP | 5mW, 650nm, InGaAIP | Same | |
| Laser Delivery | Articulated Arm | Articulated Arm | Same | |
| User Interface | LCD Touch Screen | LCD Touch Screen | Same | |
| Standards met | IEC 60601-1-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-2-22<br>IEC 60825-1 | IEC 60601-1-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-2-22<br>IEC 60825-1 | Same | |
{7}------------------------------------------------
# #401, 402 Woolim e-Biz Center, 28, Digital-ro 33-gil, Guro-gu, Seoul, Republic of Korea Tel: +82-2-2108-2198 / Fax: +82-2-2108-2181
{8}------------------------------------------------
## #401, 402 Woolim e-Biz Center, 28, Digital-ro 33-gil, Guro-gu, Seoul, Republic of Korea Tel: +82-2-2108-2198 / Fax: +82-2-2108-2181
#### VII. Testing Data
#### i. Non-Clinical Testing
Non-Clinical Testing was performed for this 510(k) Notification. The following performance data were provided in support of the substantial equivalence determination:
- Biocompatibility Testing – The biocompatibility evaluation for the subject device was conducted in accordance with the quidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" as recognized by FDA.
- Shelf Life Testing – Expected service life rationale report has been attached in this submission.
- 트 Software - Verification and validation testing was conducted on the software interface and firmware and the documentation provided is as recommended in the FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
- . Electromagnetic Compatibility and Electrical Safety - Electromagnetic Compatibility and Electrical Safety testings were conducted on Azure. The following standards were complied with the device:
- Electromagnetic Compatibility: IEC 60601-1-2
- Electrical Safety Testing: IEC 60601-1
- Laser Safety: IEC 60825-1
- Usability: IEC 60601-1-6
- . Performance Testing - Performance testing was conducted on Azure according to the following standard:
- IEC 60601-2-22
The subject device passed all tests and successfully met all acceptance criteria and test requirements.
{9}------------------------------------------------
#401, 402 Woolim e-Biz Center, 28, Digital-ro 33-gil, Guro-gu, Seoul, Republic of Korea Tel: +82-2-2108-2198 / Fax: +82-2-2108-2181
#### ii. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate devices. These types of devices, including the predicated devices, have been on the market for many years with a proven safety and efficacy for the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
#### VIII. Conclusion
Based on the similarities of the subject and predicate device, it is concluded that Azure is substantially equivalent to the predicate device. The summary submitted includes only information that is also covered in the body of the 510(k). It does not include any puffery or unsubstantiated labeling claims, any raw data, any trade secret or confidential commercial information and any patient identification information.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.