The Orthodontic Screw is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only. The orthodontic screw is intended in patients aged 12 years and older.
Device Story
Mini Screws provide fixed anchorage in maxillary/mandibular bone for orthodontic appliances. Device consists of head (hexagonal socket for rotational force/attachment of chains/springs), neck (smooth surface), and body (threaded, cylindrical). Applied by orthodontists/dentists via hand tool or low-torque rotary instrument; typically self-drilling. SLA surface treatment (sandblasted, large-grid, acid-etched) on apical half increases stability. Used temporarily during treatment; removed upon completion. Benefits include firm anchorage to facilitate tooth movement. Single-use, sterile, prescription-only.
Clinical Evidence
Bench testing only. Mechanical strength testing performed per TS EN ISO 19023 and ASTM F543. Biocompatibility evaluated per EN 10993-1. Sterilization validated per EN 11137-1/2.
Technological Characteristics
Material: Pure Titanium Grade 4 (ISO 5832-2, ASTM F67-13). Surface: SLA (sandblasted, large-grid, acid-etched) on apical 1/2. Dimensions: 1.4-2.0mm diameter, 5-16mm length. Connection: Hexagon socket. Sterilization: Gamma radiation. Packaging: Class 100,000 clean room (ISO 11607).
Indications for Use
Indicated for patients aged 12 years and older requiring fixed anchorage for orthodontic tooth movement. Contraindications: Not evaluated for MRI safety.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 04, 2024
Proimtech Saglik Urunleri Anonim Sirketi Hakan Cevik General Manager Imes Sitesi, No:3 Dudullu Osb Mahallesi Imes 305.Sokak Istanbul. Istanbul 34773 TURKEY
Re: K231502
Trade/Device Name: Proimtech Mini Screw Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: OAT Dated: May 12, 2024 Received: August 8, 2024
Dear Hakan Cevik:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231502
Device Name Mini Screws
Indications for Use (Describe)
The Orthodontic Screw is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only. The orthodontic screw is intended in patients aged 12 years and older.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary 510k number: K231502 Mini Screws Proimtech Saglik Urunleri Anonim Sirketi
| Company Name | : | Proimtech Saglik Urunleri A.S. |
|---------------------------|---|-----------------------------------------------------------------------------------|
| Authorised Person / Title | : | Hakan Çevik / General Manager |
| Address-1 | : | Dudullu OSB Mahallesi Imes 305 Sokak C Blok No:3 PK:34775<br>Umraniye / İstanbul |
| Address-2 | : | Dudullu OSB Mahallesi Imes 501 Sokak E Blok No:25 PK:3477.<br>Umraniye / İstanbul |
| Phone | : | +90 (216) 999 5041-44 |
| Web/Mail | : | www.bilimplant.com / info@bilimplant.com |
| Official Contact | : Hakan Cevik, General Manager of Proimtech Saglik Urunleri Anonim Sirket |
|------------------|---------------------------------------------------------------------------|
| Phone Number | : +90 535 355 51 00 |
| E-Mail | : hakancevik@bilimplant.com |
### DEVICE INFORMATION
| Preparation Date | : September 04, 2024 |
|---------------------|-------------------------------------------|
| Device Trade Name | : Mini Screws |
| Common Name | : Mini Screw |
| Classification Name | : screw, fixation, bone (21 CFR 872.3640) |
| Device Class | : II |
| Product Code | : OAT |
| Regulation Number | : 872.3640 |
| Review Panel | : Dental |
## INDICATION FOR USE
The Orthodontic Screw is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only. The orthodontic screw is intended in patients aged 12 years and older.
{5}------------------------------------------------
### DEVICE DESCRIPTION
The Mini Screws system has been designed to provide a firm anchorage point in the oral cavity for orthodontic treatment. Mini screws are manufactured with CP Ti Gr 4 and are SLA treated. The Mini Screw system is designed to be applied with different orthodontic solutions and is versatile enough to satisfy different clinical cases. Titanium orthodontic mini screws are intended to provide fixed anchorage for orthodontic treatment used by implanting in the maxillary and mandibular bone.
These mini screws used for anchoring purposes can be divided into 3 parts as head, neck and body.
## ► Head Part:
Head part, it is the part where both the rotational force necessary for the screw to move in the bone is applied and where it is used as a fulcrum for the force to occur. The heads of the mini screws are designed in the shape of a hexagon in which the screwdriver bit is fully inserted. The hole in the head is designed for the attachment of chains, elastic wires or springs to initiate the necessary tooth movements.
## ► Neck Part:
Neck part It is the part between the gingiva and the cortical bone, has no grooves and a shiny surface. The maximum compression of the screw occurs in this area and during insertion into the cortical bone.
## ► Body Part:
Body part is the threaded part of the screw that remains in the bone. The inner part of the body (core) provides the resistance of the screw against fracture and the groove depth provides the resistance of the screw against pulling.
The head part of the screws is cross-shaped or button-shaped and this part is the part where the necessary rotational force is applied to ensure that the screw advances in the bone. It is also the part where mechanics such as wire, spring or rubber are connected to the teeth to be moved. The neck part is designed to be smooth to make the use more comfortable, especially in patients with thin gingival biotype. The body part is designed as cylindrical and grooved in order to keep the screw fixed in the bone. The tip is angled conical for ease of application.
### > Length of Screws:
Screw lengths vary between 5 mm and 16 mm. The length of the screw is determined by some factors such as the angle of application, thickness of the mucosa, bone quality and proximity of neighboring structures.
### > Screw Diameters:
Screw diameters vary between 1.4 and 2 mm. The stability of the screws is affected by the screw surface area in contact with the cortical bone. As the diameter increases, the contact surface area will increase and the stability will also increase.
{6}------------------------------------------------
## > Surface Characteristics of Screws:
The tips of the mini screws with 0 1.4 mm, Ø 1.65 mm and Ø 2 mm diameters used for orthodontic anchorage have a sandblasted and acid etched (SLA) surface structure and have a surface roughness of 1.5-2 Ra. Since similar screws on the market are not expected to be osseointegrated, the surfaces of most of them are not treated. However, in order to apply high anchorage force and increase screw stability, the ends of these products are treated with "Sandblasted, large-grid, acid etched" (SLA) process, which is sandblasting followed etching.
## > Groove Structure of Screws:
Screws can be classified as self-tapping (ST) and self-drilling (SD) according to their thread design. Our mini screws, which are used for tooth mobilization are placed directly as SD without predrilling. However, in areas with dense cortical bone, the need for pilot slot preparation with drills may rarely be encountered. Our mini screws are placed with a hand tool or a low-torque rotary instrument (contraangle handpiece) at the recipient site.
## > Connection Type of Products:
The connection of the mini screws features a hexagon socket.
Image /page/6/Picture/6 description: The image contains three different views of a medical device. The first view shows a screw with a hexagonal head and a threaded shaft. The second view shows a spherical component with a hexagonal indentation. The third view shows a tool with a blade and a mechanism for gripping the screw.
## Hexagon Connection Structure
Image /page/6/Picture/8 description: The image shows three different views of a screw. The first view is a side view of the screw, showing the threads and the head. The second view is a top-down view of the screw head, showing a cross-shaped indentation. The third view shows the screw being inserted into a piece of material using a screwdriver.
## Length of SLA Surface Treatment Area:
All our mini screws have 1/2 apical part with SLA surface.
{7}------------------------------------------------
#### MATERIALS:
Only materials approved for medical purposes are used in the manufacture of products: Pure Titanium Grade 4 according to standard ISO 5832-2 and ASTM F67-13.
## PERFORMANCE AND SAFETY SPECIFICATION
### Surface Characterization:
SLA is used to roughen the surface by applying a strong acid to the surface sandblasted with coarse grainy particles. The SLA technique has been used as an alternative to TPS coatings to achieve better surface chemistry and topography. This process is applied in combination after osseointegration to achieve macroroughness and micropits, as well as grit-sandblasting and acid etching to increase surface roughness.
### MRI Compatibility:
The Mini Screw has not been evaluated for safety in the MR environment. It has not been tested for heating or unwanted movement in the MR environment. The safety of Mini Screws in the MR environment is unknown. Performing an MR exam on a person who has this medical device may result in injury or device malfunction.
The product design has been made to include all kinds of safety and performance requirements considering the intended use and location of the product. In this direction, especially;
- √ The product is manufactured from pure titanium (Grade 4) raw material in accordance with EN 1642, TS EN ISO 19023, ISO 5832-2 and ASTM F67-13 Standard.
- > Our products are packaged in a Class 100,000 Clean Room according to ISO 11607 standard to ensure microbiological safety. They are then sterilized with Gamma radiation in accordance with the parameters specified in our user manual.
- > The gamma sterilization method declared by us has been validated in accordance with EN 11137-1/2 Standard.
- > The biocompatibility of our finished products has been carried out in accordance with EN 10993-1 Standard.
- > Tests for the mechanical strength of our products have been carried out in accordance with TS EN ISO 19023 and ASTM F543 Standard.
### Primary Predicate Device and Reference Device
For primary device Mini Screws: K161197; Orthodontic screw, OSSTEM Implant Co., Ltd.
For reference device Mini Screws: K152297; OBS Anchorage Screw, BOMEI
For reference device Mini Screws: K191041; Orthodontic Fixation Screw (Smart Anchor Miniscrew), GNI Co., LTD
For reference device Mini Screw: K122069; Storm Mini Screw, Lancer Orthodontics
{8}------------------------------------------------
For reference device Mini Screw: K161335; Dual Top Screw System, Jeil Medical Corporation
# SUMMARIES OF TECHNOLOGICAL CHARACTERISTICS & SUBSTANTIAL EQUIVALENCE DISCUSSION
{9}------------------------------------------------
| Technological<br>Characteristics | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | | | | | | | removed after<br>orthodontic<br>treatment has<br>been<br>completed. The<br>Storm Mini<br>Screw is<br>provided sterile<br>and is intended<br>for single use<br>only. | |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|---------------------------------|------------------------------------------|---------------------------|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| | Mini Screws | Orthodontic screw | OBS Anchorage Screw | Orthodontic Fixation<br>Screw (Smart Anchor<br>Miniscrew) | Storm Mini<br>Screw | Dual Top<br>Screw<br>System | Design | Image: screw | Image: screw | Image: screw | Image: screw | Image: screw | Image: screw | Image: screw |
| 510(k)<br>Number | K231502 | K161197 | K152297 | K191041 | K122069 | K161335 | Materi<br>al | Titanium (Ti Grade 4, ASTM F67) | | Ti-6Al-4V ELI, ASTM F 136 | Stainless Steel (ISO 5832-<br>1:2007/ ASTM F138-13)<br>Titanium Alloy (ISO 5832-<br>3:1996/ ASTM F136-13) | Ti-6Al-4V ELI, ASTM F 136 | Ti-6Al-4V ELI,<br>ASTM F 136 | Titanium<br>Alloy (ASTM<br>F136) |
| Manufacturer | Proimtech Saglik Urunleri A.S | OSSTEM Implant Co., Ltd. | BOMEI | GNI Co., LTD | Lancer<br>Orthodontics | Jeil Medical<br>Corporation | Body<br>Diamet<br>er | Ø1.4 mm<br>Ø1.65 mm<br>Ø2 mm | Ø1.2 mm<br>Ø1.4 mm<br>Ø1.6 mm<br>Ø1.8 mm | | Ø1.5<br>Ø2 | Ø1.2 mm<br>Ø1.4 mm<br>Ø1.6 mm<br>Ø1.8 mm<br>Ø2 mm | Ø1.5 mm<br>Ø2 mm | Ø1.3 – Ø2.5 |
| Product Code | OAT | OAT | OAT | OAT | OAT | OAT | | | | | | | | |
| Indications for<br>Use | The Orthodontic Screw is indicated<br>for use as a fixed anchorage point for<br>attachment of orthodontic<br>appliances to facilitate the<br>orthodontic movement of teeth. It is<br>used temporarily and is removed<br>after orthodontic treatment has been<br>completed. Screws are intended for<br>single use only. The orthodontic<br>screw is intended in patients aged 12<br>years and older. | The Orthodontic Screw is<br>indicated for use as a fixed<br>anchorage point for<br>attachment of orthodontic<br>appliances to facilitate the<br>orthodontic movement of<br>teeth. It is used temporarily<br>and is removed after<br>orthodontic treatment has<br>been completed. Screws are<br>intended for single use only | Temporary anchorage<br>screws are intended to<br>provide a fixed anchorage<br>point for attachment of<br>orthodontic appliances to<br>facilitate the orthodontic<br>movement of the teeth. It<br>is used temporarily and is<br>removed after orthodontic<br>treatment has been<br>complete. | The Orthodontic Fixation<br>Screw[Smart Anchor<br>Miniscrew] is intended<br>for use as a temporary<br>anchor for orthodontic<br>treatment for use in<br>patients aged 12 and<br>older. | The Storm Mini<br>Screw is a<br>threaded<br>titanium dental<br>implant screw<br>intended to<br>provide a fixed<br>anchorage<br>point for !he<br>attachment .of<br>orthodontic<br>appliances and<br>facilltate the<br>orthodontic<br>movement of,<br>teeth. it is used<br>temporarily<br>and must be | The Dual<br>Top Screw<br>System is<br>intended<br>for use as a<br>temporary<br>anchor for<br>orthodontic<br>treatment<br>for use in<br>patients<br>aged 12<br>and older. | | | | | | | | |
{10}------------------------------------------------
{11}------------------------------------------------
| Length | Ø1.4 x 6 mm<br>Ø1.4 x 7 mm<br>Ø1.4 x 8 mm<br>Ø1.4 x 9 mm<br>Ø1.4 x 10 mm<br>Ø1.65 x 6 mm<br>Ø1.65 x 7 mm<br>Ø1.65 x 8 mm<br>Ø1.65 x 9 mm<br>Ø1.65 x 10 mm<br>Ø1.65 x 11 mm<br>Ø1.65 x 12 mm<br>Ø2 x 6 mm<br>Ø2 x 7 mm<br>Ø2 x 8 mm<br>Ø2 x 9 mm<br>Ø2 x 10 mm<br>Ø2 x 11 mm<br>Ø2 x 12 mm<br>Ø2 x 14 mm | 6 mm<br>8 mm<br>10 mm | Ø1.5 x 8 mm<br>Ø1.5 x 10 mm<br>Ø1.5 x 12 mm<br>Ø2 x 8 mm<br>Ø2 x 10 mm<br>Ø2 x 12 mm<br>Ø2 x 14 mm | 6 mm<br>8 mm<br>10 mm | 8 mm<br>10 mm | 5 mm – 16<br>mm |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Single<br>Use | Yes | Yes | Yes | Yes | Yes | Yes |
| Surface | Sandblasted Large grit Acid etched (SLA) | Acid etching | --- | Machined (not acid<br>etched) | --- | No surface<br>treatment |
| Steriliz<br>ation | Radiation Sterile | Radiation Sterile | Non-Sterile (Steam<br>sterilized by user) | Non-Sterile (Steam<br>sterilized by user) | Sterile via beta<br>irradiation | Non-Sterile<br>(Steam<br>sterilized by<br>user) or<br>Gamma-<br>Sterilized |
| Shelf<br>life | 5 years | 8 years | 5 years | 5 years | 5 years | - |
| Target<br>population | Professional use only – qualified<br>dentists, orthodontist or oral<br>surgeon. Strictly reserved to<br>specialized and trained users. | Professional use only –<br>qualified dentists. Strictly<br>reserved to specialized and<br>trained users. | Professional use only -<br>qualified dentists. Strictly<br>reserved to specialized and<br>trained users. | Professional use only –<br>qualified dentists. Strictly<br>reserved to specialized<br>and trained users. | People having<br>orthodontia<br>work | Professional use only –<br>qualified<br>dentists.<br>Strictly<br>reserved to<br>specialized<br>and trained<br>users. |
| Principl<br>e of<br>Operati<br>on | Anchorage mini screws are used to<br>move the teeth. | Orthodontic screw is inserted<br>into either jaw to help the<br>orthodontist move the correct<br>teeth and stop the wrong<br>teeth from moving in the<br>wrong direction. | Anchorage mini screws are<br>used to move the teeth. | Anchorage mini screws<br>are used to move the<br>teeth. | Dental and<br>orthodontic<br>specialists<br>offices | Anchorage<br>mini screws<br>are used to<br>move the<br>teeth. |
| Substantial Equivalence Discussion | | | | | | |
{12}------------------------------------------------
#### Similarities
The device has similar characteristics compared to the following respects. Indication for use, product code, design, diameter, disposable, intended population and operating principle.
#### Difference
The differences between the subject device is the surface treatment. However, testing data such as mechanical testing provided in the submission show that these differences do not raise issues in performance. Additionally, the subject device are available in slightly different however the such dimensional changes do not raise any performance concerns.
#### Conclusion
On the basis of the above discussion, it is concluded that the device in question is substantially equivale device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.