TraumaCad Neo is indicated for assisting healthcare professionals to analyze orthopedic conditions and to plan orthopedic procedures by overlaying on relevant radiological images visual information such as measurements and prosthesis templates. Clinical judgment and experience are required to properly use the software is not intended for primary radiological image interpretation or radiological appraisal. Device is not intended for use on mobile phones.
Device Story
TraumaCad Neo is a software application for orthopedic surgical planning and post-operative image review. It receives digital radiological images via the Quentry® cloud service. Surgeons use the software on PCs, Macs, iPads, or Android tablets to perform measurements, position digital prosthesis templates from a library, and plan procedures. The device supports hip, knee, foot/ankle, and upper limb modules. It allows simultaneous viewing of up to three images. The software facilitates a film-less workflow by enabling pre-operative planning and post-operative review. It does not perform primary image interpretation; clinical judgment is required. The output assists surgeons in visualizing surgical outcomes, potentially improving procedural accuracy and planning efficiency.
Clinical Evidence
Bench testing only. Performance evaluation was based on established test methods used for the predicate device to demonstrate conformity to intended use and substantial equivalence.
Technological Characteristics
Radiological image processing system; software-based. Supports DICOM. Cloud-integrated via Quentry®. Operates on Windows 10+, iOS 16+, macOS 11+, and Android 11+. Supports PC, Mac, iPad, and Android tablets. Features include automatic scaling, interactive template positioning, and template rotation. No patient contact. Non-life sustaining.
Indications for Use
Indicated for healthcare professionals to analyze orthopedic conditions and plan orthopedic procedures via overlaying measurements and prosthesis templates on radiological images. Not for primary radiological image interpretation or appraisal. Not for use on mobile phones.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Brainlab Ltd. % Yael Guttentag QM & RA Senior Manager 35 Efal Street Petach-Tikva, 4951132 Israel
Re: K231498
November 20, 2023
Trade/Device Name: TraumaCad Neo Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: October 18, 2023 Received: October 18, 2023
Dear Yael Guttentag:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K231498
Device Name TraumaCad Neo
Indications for Use (Describe)
TraumaCad Neo is indicated for assisting healthcare professionals to analyze orthopedic conditions and to plan orthopedic procedures by overlaying on relevant radiological images visual information such as measurements and prosthesis templates. Clinical judgment and experience are required to properly use the software is not intended for primary radiological image interpretation or radiological appraisal. Device is not intended for use on mobile phones.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> </span> </span> </span> | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> </span> </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Æ BRAINLAB 510(k) Summary
# K231498
Pursuant to CFR 807.92, the following 510(k) Summary is provided:
| 1. (a) | Submitter<br>Address: | Brainlab Ltd<br>35 Efal Street<br>Petach-Tikva, Israel 4951132 |
|--------|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. (b) | Manufacturer<br>Address: | Brainlab Ltd<br>35 Efal Street<br>Petach-Tikva, Israel 4951132 |
| | Mfg. Phone: | Tel.: +972-3-929-0929 |
| | Contact Person: | Mrs. Yael Guttentag |
| | Date: | May 22, 2023 |
| 2. | Device &<br>Classification<br>Name: | Radiological Image Processing System -<br>classified as Class 2 LLZ, Regulation Number 21 CFR 892.2050<br>TraumaCad Neo |
| 3. | Predicate Device: | TraumaCad Mobile 2.0 (K160001) |
| 4. | Description: | TraumaCad Neo allows surgeons to evaluate digital images while<br>performing various pre-operative surgical planning and evaluation of<br>images. This software application enables surgeons to plan operations on<br>screen, execute measurements, and facilitates the film-less orthopedic<br>practice. TraumaCad Neo allows post-operative review of images<br>obtained after the surgical procedure. The program features an extensive<br>regularly updated library of digital templates from leading manufacturers.<br>TraumaCad Neo supports DICOM and is integrated to communicate with<br>Quentry®, a proprietary web-based cloud service from Brainlab. It is through<br>Quentry®, where the medical staff can upload images to plan their expected<br>results prior to the procedure to create a smooth surgical workflow from start<br>to finish. |
| 5. | Indications for<br>Use: | TraumaCad Neo is indicated for assisting healthcare professionals to<br>analyze orthopedic conditions and to plan orthopedic procedures by<br>overlaying on relevant radiological images visual information such as<br>measurements and prosthesis templates. Clinical judgment and experience<br>are required to properly use the software. The software is not intended for<br>primary radiological image interpretation or radiological appraisal. Device is<br>not intended for use on mobile phones. |
{4}------------------------------------------------
- 6. Comparison of With respect to technology and intended use, TraumaCad Neo is Technological substantially equivalent to its predicate device. Based upon the outcomes Characteristics: from the Risk Analysis and Performance Testing Evaluation, Brainlab Ltd believes that the modification of TraumaCad Mobile 2.0 (predicate device) which allows it to become TraumaCad Neo does not raise additional safety or efficacy concerns. The following comparison table depicts the changes.
| | Submitted Device | Predicate Device |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Features/ Characteristics | TraumaCad Neo | TraumaCad Mobile<br>Release 2.0 |
| Product Code | LLZ | LLZ |
| Indication for Use | TraumaCad Neo is indicated<br>for assisting healthcare<br>professionals to analyze<br>orthopedic conditions and to<br>plan orthopedic procedures<br>by overlaying on relevant<br>radiological images visual<br>information such as<br>measurements and<br>prosthesis templates. Clinical<br>judgment and experience<br>are required to properly use<br>the software. The software is<br>not intended for primary<br>radiological image<br>interpretation or radiological<br>appraisal. Device is not<br>intended for use on mobile<br>phones. | The TraumaCad Mobile<br>Release 2.0 program is<br>indicated for assisting<br>healthcare professionals in<br>preoperative planning of<br>orthopedic surgery. The<br>device allows for<br>overlaying of prosthesis<br>templates on radiological<br>images, and includes tools<br>for performing<br>measurements on the<br>image and for positioning<br>the template. Clinical<br>judgment and experience<br>are required to properly<br>use the software. The<br>software is not for primary<br>image interpretation. The<br>software is not for use on<br>mobile phones. |
| Operating System | Microsoft Windows 10 and<br>above<br>iOS 16.x and above<br>MAC OS 11 and above<br>Android 11 and above | MS Windows 7,<br>8 or 10<br>iOS 7.x, 8.x or<br>9.x MAC OS X |
| Devices Supported | PC<br>MAC<br>iPads<br>Android tablets | PC/MAC<br>iPad 3, 4 and 5 (Air) |
| Browsers Supported | Microsoft Edge<br>Firefox<br>Chrome<br>Safari (MAC/iOS) | Minimum<br>requirements:<br>iOS based<br>browsers:<br>- Safari 7<br>- Chrome 30<br>Mac based<br>browsers:<br>- Firefox 26<br>- Safari 5<br>Windows based<br>browsers:<br>- Chrome 30<br>- Firefox 26<br>- Safari 5<br>- IE 11 |
| Image Input | Can receive digital images<br>from Quentry® | Can receive digital images<br>from various sources |
| Means of Collecting Data | Obtained from pre-obtained<br>digital images via Quentry® | Obtained from pre-<br>obtained digital images via<br>PACS or via Quentry® |
| Number of Images that can<br>simultaneously viewed on<br>the screen | Up to 3 | Up to 2 |
| Runs on Server | yes | yes |
| Hip Module | yes | yes |
| Knee Module | yes | yes |
| Foot and Ankle Module | yes | yes |
| Upper Limb Module | yes | yes |
| Digital Prosthetic<br>Templates | yes | yes |
| Interactive template<br>positioning | yes | yes |
| Automatic Scaling | yes | yes |
| Template support from<br>manufacturers | yes | yes |
| Permits Template<br>Rotation | yes | yes |
| Pre-Operative Planning | yes | yes |
| Post Operative Review | yes | no |
| Patient Contacting | no | no |
| Control of Life Sustaining<br>Devices | no | no |
| Healthcare professional<br>intervention for<br>interpretation of images | yes | yes |
| 510(k) # | Pending | K160001 |
{5}------------------------------------------------
510k Notification: TraumaCad Neo
{6}------------------------------------------------
- The Performance Evaluation of TraumaCad Neo was based upon well-7. Performance Evaluation: established test methods which demonstrated conformity to the intended use. These test methods were the same which were used to demonstrate the substantial equivalence of the predicate device TraumaCad Mobile 2.0.
- 8. Conclusion: The intended use and the fundamental technological characteristics of TraumaCad Neo are the same as those in the TraumaCad Mobile 2.0, which is the predicate device. Any additions or differences do not affect the safety and effectiveness of the device. The performance tests have been completed and successfully confirm the performance of the device. Based upon this data, Brainlab Ltd believes that TraumaCad Neo is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.