CoreLink Robotic Navigation Instruments

K231494 · Corelink, LLC · OLO · Oct 27, 2023 · Neurology

Device Facts

Record IDK231494
Device NameCoreLink Robotic Navigation Instruments
ApplicantCorelink, LLC
Product CodeOLO · Neurology
Decision DateOct 27, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

CoreLink Robotic Navigation Instruments are indicated to be used during the preparation screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures for skeletally mature patients, where reference to a rigid anatomical structure such as vertebra can be identified. These instruments are designed for use with the Globus ExcelsiusGPS system, which is indicated for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws.

Device Story

CoreLink Robotic Navigation Instruments consist of non-sterile, reusable stainless steel manual surgical instruments, including taps and drivers. Designed for use with the Globus ExcelsiusGPS Robotic Navigation Platform; instruments interface with the system to assist surgeons in precisely locating anatomical structures and guiding screw placement during spinal surgery. Used in open or minimally invasive procedures; surgeon operates instruments manually while the robotic system provides spatial positioning and orientation guidance based on pre-operative CT or intra-operative fluoroscopy. Output is physical guidance for instrument trajectory; aids surgeon in accurate pedicle screw implantation. Benefits include enhanced precision in spinal instrumentation and potential for improved surgical outcomes in spinal fixation procedures.

Clinical Evidence

No clinical data. Substantial equivalence established via engineering analysis and geometric comparison to predicate devices.

Technological Characteristics

Reusable stainless steel instruments (taps and drivers). Manual operation. Designed for integration with Globus ExcelsiusGPS robotic platform. No electronic components or software within the instruments themselves; connectivity is mechanical/physical interface with the robotic system.

Indications for Use

Indicated for skeletally mature patients undergoing spinal surgery requiring preparation for screw placement; used in open or minimally invasive procedures where rigid anatomical structures (vertebrae) can be identified via CT or fluoroscopy.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. October 27, 2023 CoreLink, LLC % Nathan Wright Engineer & Regulatory Specialist Empirical Technologies 4628 Northpark Drive Colorado Springs, Colorado 80918 Re: K231494 Trade/Device Name: CoreLink Robotic Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: October 27, 2023 Received: October 27, 2023 Dear Nathan Wright: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming {1}------------------------------------------------ #### K231494 - Nathan Wright product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali-S Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231494 Device Name CoreLink Robotic Navigation Instruments #### Indications for Use (Describe) CoreLink Robotic Navigation Instruments are indicated to be used during the preparation screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures for skeletally mature patients, where reference to a rigid anatomical structure such as vertebra can be identified. These instruments are designed for use with the Globus ExcelsiusGPS system, which is indicated for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K231494 - 510(K) SUMMARY | Submitter's Name: | CoreLink, LLC | |--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Address: | 2072 Fenton Logistics Park<br>St. Louis, Missouri 63026 | | Submitter's Telephone: | 888-349-7808 | | Contact Person: | Nathan Wright MS, RAC<br>Empirical Technologies<br>1-719-351-0248<br>nwright@empiricaltech.com<br>Image: Empirical Technologies Logo | | Date Summary was Prepared: | October 27, 2023 | | Trade or Proprietary Name: | CoreLink Robotic Navigation Instruments | | Device Classification Name: | Orthopedic Stereotaxic Instruments | | Classification & Regulation #: | Class II per 21 CFR §882.4560 | | Product Code: | OLO | | Classification Panel: | Orthopedic - Stereotaxic, Trauma, and Restorative Devices (DHT6C) | ## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: CoreLink Robotic Navigation Instruments are non-sterile, reusable instruments including taps and drivers that can be operated manually. These instruments are intended to be used with the Globus Medical Excelsius GPS® Robotic Navigation Platform to aid in implantation of CoreLink Pedicle Screw System implants (NextGen Pedicle Screw System, CoreLink Midline Fixation System (CentraFix), and Tiger Spine System). The instruments are manufactured from stainless steel. ### INDICATIONS FOR USE CoreLink Robotic Navigation Instruments are indicated to be used during the preparation screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures for skeletally mature patients, where reference to a rigid anatomical structure such as vertebra can be identified. These instruments are designed for use with the Globus ExcelsiusGPS system, which is indicated for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws. ## TECHNOLOGICAL CHARACTERISTICS The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of the safety and effectiveness. Specifically, the following characteristic are the similar between the subject and predicates: - Device design and dimensions ● - Indications for use ● - Materials of manufacture ● - Principles of operation {4}------------------------------------------------ #### Predicate Devices | 510k Number | Trade or Proprietary or Model Name | Manufacturer | Product Code | Predicate Type | |-------------|------------------------------------|---------------------|--------------------|----------------| | K171651 | EXCELSIUS GPS | Globus Medical Inc. | OLO | Primary | | K221776 | NextGen Pedicle Screw System | CoreLink, LLC | NKB | Additional | | K210539 | CoreLink Midline Fixation System | CoreLink, LLC | NKB | Additional | | K133369 | Tiger Spine System | CoreLink, LLC | NKB, KWP, MNH, MNI | Additional | ### PERFORMANCE DATA The CoreLink Robotic Navigation Instruments have been evaluated through an engineering analysis and geometric comparison to predicate devices to establish the safety and efficacy for accuracy performance. The results of this engineering analysis show that the subject is substantially equivalent to the cleared predicate. ## Conclusion The overall technology characteristics and engineering analysis lead to the conclusion that the CoreLink Robotic Navigation Instruments are substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...