uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy Analysis
K231482 · Shanghai United Imaging Healthcare Co., Ltd. · JAK · Nov 29, 2023 · Radiology
Device Facts
Record ID
K231482
Device Name
uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy Analysis
Applicant
Shanghai United Imaging Healthcare Co., Ltd.
Product Code
JAK · Radiology
Decision Date
Nov 29, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
uCT ATLAS Astound is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS Astound is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography. uCT ATLAS Astound is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. * Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information. uCT ATLAS is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography. uCT ATLAS has the capability to image a whole organ in a single rotation. Organs include, but not limited to head, heart, liver, kidney, pancreas, joints, etc. uCT ATLAS is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of protocols that have been approved and published by either a governmental body or professional medical society. * Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.
Device Story
Computed tomography (CT) system producing cross-sectional images via reconstruction of X-ray transmission data; includes uWS-CT-Dual Energy Analysis software for material composition visualization. New functionality: CT-guided intervention providing real-time CT fluoroscopy (12 IPS) with in-room view and X-ray control. Hardware additions include scanning room monitor, exposure foot pedal, hand cart, suspension gear, and couch-side controller. Operated by clinicians in clinical settings; allows real-time scan parameter adjustment and 2D/3D entry path planning. Benefits include reduced radiation dose, fast scanning for cardiac/dynamic whole-organ imaging, and improved procedural guidance for interventions.
Clinical Evidence
Bench testing only. Performance evaluation for CT-guided intervention, dosimetry, and image quality metrics conducted to verify design specifications. Conformance to IEC 60601-1, 60601-1-2, 60601-2-44, and NEMA XR-25/28/29 standards demonstrated.
Technological Characteristics
Computed tomography X-ray system; X-ray tube/detector assembly; software for image reconstruction and dual-energy analysis. Hardware for intervention: monitor, foot pedal, suspension gear. Standards: IEC 60601-1 (safety), IEC 60601-2-44 (CT performance), ISO 10993 (biocompatibility). Connectivity: DICOM. Software: post-processing package for dual-energy material decomposition.
Indications for Use
Indicated for head, whole body, cardiac, and vascular CT imaging; low-dose lung cancer screening in patients meeting established governmental or professional society criteria.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
uCT ATLAS Astound with uWS-CT-Dual Energy Analysis (K223028)
uCT ATLAS with uWS-CT-Dual Energy Analysis (K203448)
{0}------------------------------------------------
November 29, 2023
Shanghai United Imaging Healthcare Co., Ltd. % Xin Gao Regulatory Affairs Manager No. 2258 Chengbei Rd., Jiading Industrial District SHANGHAI, SHANGHAI 201807 CHINA
### Re: K231482
Trade/Device Name: uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy Analysis Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: November 1, 2023 Received: November 1, 2023
## Dear Xin Gao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231482
### Device Name
uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy Analysis
### Indications for Use (Describe)
uuCT ATLAS Astound is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS Astound is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography.
uCT ATLAS Astound is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. * Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.
uCT ATLAS is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography.
uCT ATLAS has the capability to image a whole organ in a single rotation. Organs include, but not limited to head, heart, liver, kidney, pancreas, joints, etc.
uCT ATLAS is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of protocols that have been approved and published by either a governmental body or professional medical society.
* Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is dark blue. The logo is simple and modern.
## K231482
# 510(k) SUMMARY
- Date of Preparation 1. November 1, 2023
- 2. Sponsor Identification
Shanghai United Imaging Healthcare Co.,Ltd.
No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China Contact Person: Xin GAO Position: Regulatory Affairs Manager Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com
#### 3. Identification of Subject Device
Device Name: uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy Analysis Common Name: Computed Tomography X-ray System Model(s): uCT ATLAS Astound; uCT ATLAS Regulation Name: Computed Tomography X-ray System Regulation Number: 21 CFR 892.1750 Regulatory Class: II Product Code: JAK Review Panel: Radiology
#### 4. Identification of Predicate/Reference Device(s)
Predicate Device(s) Device Name: uCT ATLAS Astound with uWS-CT-Dual Energy Analysis 510(k) Number: K223028 Regulation Name: Computed Tomography X-ray System Regulation Number: 21 CFR 892.1750 Regulatory Class: II Product Code: JAK Review Panel: Radiology
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is in bold, dark teal font. To the right of the text is a stylized "U" shape, also in dark teal. The logo is simple and modern.
Device Name: uCT ATLAS with uWS-CT-Dual Energy Analysis 510(k) Number: K203448 Regulation Name: Computed Tomography X-ray System Regulation Number: 21 CFR 892.1750 Regulatory Class: II Product Code: JAK Review Panel: Radiology
Reference device(s): Device Name: uCT 760, uCT780 510(k) Number: K230162 Regulation Name: Computed Tomography X-ray System Regulation Number: 21 CFR 892.1750 Regulatory Class: II Product Code: JAK Review Panel: Radiology
# 5. Device Description:
The proposed device CT system with uWS-CT-Dual Energy Analysis includes image acquisition hardware, image acquisition, reconstruction and dual energy analysis software, and associated accessories.
The proposed CT system is designed to use less radiation dose. Further, the fast scanning capability benefits the clinical applications, especially for cardiac imaging, dynamic whole organ imaging and fast body and vascular imaging.
The computer system delivered with the CT scanner is able to run post processing applications optionally.
The Computed Tomography System family scanners referenced in this submission are comparable in indications for use, and are substantially equivalent in design, material, functionality, technology, energy source and are substantially equivalent to the predicate devices. The reason for this submission is to support the following additional function:
CT intervention provides real-time CT fluoroscopy at 12 IPS with in-room view and in-room X-ray control. It allows the user to adjust the scan parameters during operation, and scan modes can be switched according to technician's operation requirements. Entry path planning based on 2D and 3D images.
The CT guided intervention will be applicable for the UIH qualified CT systems. This indication will also be applicable for future qualified UIH CT systems.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, which is dark gray and has a white line running vertically through the center. The logo is simple and modern, and the use of bold font and dark gray color gives it a strong and professional look.
#### Indications for Use 6.
uCT ATLAS Astound is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS Astound is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography.
uCT ATLAS Astound is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
* Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.
uCT ATLAS is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography.
uCT ATLAS has the capability to image a whole organ in a single rotation. Organs include, but not limited to head, heart, liver, kidney, pancreas, joints, etc.
uCT ATLAS is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
* Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011: 365:395-409) and subsequent literature, for further information.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is divided vertically into two halves by a white line. The logo is simple and modern in design.
uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.
#### 7. Comparison of Technological Characteristics with the Predicate Device
There is no change between the proposed device and the predicate device except CT-Guided Intervention function. Below is the comparison table between the proposed device, predicate device, and the reference device related to CT Guided Intervention function and the Hardware modifications due to the CT Guided Intervention:
| Software functions | Subject<br>devices | Predicate devices<br>(K223028 and<br>K203448) | Reference device<br>#1 (K230162) | Discussion |
|--------------------------|--------------------|-----------------------------------------------|----------------------------------|------------|
| Single Axial Scan | Yes | -- | Yes | Same |
| Single Helical Scan | Yes | -- | Yes | Same |
| Fluoro Mode | Yes | -- | Yes | Same |
| Continuous Mode | Yes | -- | Yes | Same |
| 2D & 3D path<br>planning | Yes | -- | Yes | Same |
| 3D image Guidance | Yes | -- | Yes | Same |
| Hardware addition | Subject<br>devices | Predicate devices<br>(K223028 and<br>K203448) | Reference device<br>#1 (K230162) | Discussion |
| Scanning room<br>monitor | Yes | -- | Yes | Same |
| Exposure Foot Pedal | Yes | -- | Yes | Same |
| Hand Cart | Yes | -- | Yes | Same |
| Suspension Gear | Yes | -- | Yes | Same |
| Couch Side Controller | Yes | -- | Yes | Same |
#### 8. Performance Data
Non-Clinical Testing
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other and are in a bold, sans-serif font. To the right of the words is a stylized "U" symbol. The symbol is made up of two vertical bars with a horizontal bar connecting them at the top.
Non-clinical testing including dosimetry and image performance tests were conducted for the uCT ATLAS Astound with uWS-CT Dual Energy Analysis and uCT ATLAS with uWS-CT Dual Energy Analysis to verify that the proposed device met all design specifications as it is Substantially Equivalent (SE) to the predicate device.
UNITED IMAGING HEALTHCARE claims conformance to the following standards and guidance:
# Electrical Safety and Electromagnetic Compatibility (EMC)
- > ANSI AAMI ES60601-1:2005+A1:2012+A2:2021, Medical electric for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)].
- A ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests [Including Amendment 1 (2021)]
- A IEC 60601-1-3: 2008+AMD1:2013+A2:2021, Edition 2.2, Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment.
- A IEC 60601-2-44 Edition 3.2: 2016 Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography
- A IEC 60825-1: 2014, Edition 3.0, Safety of laser products - Part 1: Equipment classification and requirements.
- > IEC 60601-1-6:2010+A1:2013+A2:2020. Edition 3.2. Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- > NEMA XR 25-2019, Computed Tomography Dose Check
- > NEMA XR 28-2018, Supplemental Requirements For User Information And System Function Related To Dose In CT
- A NEMA XR 29-2013, Standard Attributes on CT Equipment Related to Dose Optimization and Management
- A IEC 61223-3-5:2019 Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance and constancy tests - Imaging performance of computed tomography X-ray equipment
# Software
- A NEMA PS 3.1-3.20(2016): Digital Imaging and Communications in Medicine (DICOM)
- > Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- A Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in bold. The color of the logo is a dark teal.
# Biocompatibility
- ISO 10993-5: 2009, Edition 3.0, Biological evaluation of medical devices Part A 5: Tests for in vitro cytotoxicity.
- A ISO 10993-10: 2010, Edition 3.0, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
## Other Standards and Guidance
- A ISO 14971: 2019, Edition 3.0, Medical Devices - Application of risk management to medical devices
- Code of Federal Regulations, Title 21, Part 820 Quality System Regulation A
- A Code of Federal Regulations, Title 21, Subchapter J - Radiological Health
- A Provision for Alternate Measure of the Computed Tomography Dose Index (CTDI) to Assure Compliance with the Dose Information Requirements of the Federal Performance Standard for Computed Tomography
In this submission, following reports related to image quality metrics and CT Guided Intervention function are renewed and submitted:
- A Biocompatibility Report
- A IEC 61223-3-5 Test Report
- > Performance Evaluation for CT Guided Intervention
### Summarv
The features described in this premarket submission are supported with the results of the testing mentioned above, the proposed device was found to have a safety and effectiveness profile that is similar to the predicate device.
#### 9. Conclusions
Based on the comparison and analysis above, the proposed device has same intended use, similar performance, safety equivalence, and effectiveness as the predicate device. The differences above between the proposed device and predicate device do not affect the intended use, technology characteristics, safety, and effectiveness. The proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.