K231453 · Nvision Biomedical Technologies, Inc. · HWC · Feb 16, 2024 · Orthopedic
Device Facts
Record ID
K231453
Device Name
SnapHammer Hammertoe Correction System
Applicant
Nvision Biomedical Technologies, Inc.
Product Code
HWC · Orthopedic
Decision Date
Feb 16, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SnapHammer Hammertoe Correction System is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe, and mallet toe. Cannulated implants in the SnapHammer Hammertoe Correction System can be used with K-wires for the delivery of implants or the temporary stabilization of outlying joints (e.g. MTP Joint).
Device Story
SnapHammer Hammertoe Correction System consists of additively manufactured titanium fixation implants with snap-off posts; system includes stainless steel drills, inserters, and sizers. Used by orthopedic surgeons in clinical/surgical settings for hammertoe, claw toe, and mallet toe reconstruction. Surgeon inserts implant into bone; snap-off post facilitates quick insertion and is removed after threading. Provides mechanical fixation of osteotomies; stabilizes lesser toe joints. Benefits patient through stable bone fixation during healing process.
Clinical Evidence
Bench testing only. Mechanical performance testing included torsion, driving torque, and axial pullout per ASTM F543, plus static and dynamic bending. Additive manufacturing validation included tensile testing per ASTM F3001 and residual powder removal analysis per ASTM F1877.
Technological Characteristics
Titanium fixation device (additively manufactured); stainless steel instrumentation (drills, inserters, sizers). Diameters: 3.0mm and 3.50mm; lengths from 12mm. Mechanical fixation via threaded implant with snap-off post. Standards: ASTM F543 (mechanical), ASTM F3001 (tensile), ASTM F1877 (particle analysis).
Indications for Use
Indicated for fixation of osteotomies and reconstruction of lesser toes (hammertoe, claw toe, mallet toe) in patients requiring surgical correction. Cannulated implants allow use with K-wires for implant delivery or temporary stabilization of outlying joints (e.g., MTP joint).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
4Webb Medical Inc. Hammertoe Truss System (HTS) (K221283)
Nvision Vector Hammertoe Correction System (K213421)
{0}------------------------------------------------
February 16, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.
Nvision Biomedical Technologies. Inc % Analaura Villarreal-Berain Extremity Project Engineer Nvision Biomedical Technologies 4590 Lockhill Selma San Antonio, Texas 78249
Re: K231453
Trade/Device Name: SnapHammer Hammertoe Correction System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC, HTY Dated: January 17, 2024 Received: January 17, 2024
Dear Analaura Villarreal-Berain:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Eileen
Cadel -S
Digitally signed by
Eileen Cadel -S
Date: 2024.02.16
08:32:09 -05'00'
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K231453
Device Name SnapHammer Hammertoe Correction System
Indications for Use (Describe)
The SnapHammer Hammertoe Correction System is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe, Cannulated implants in the SnapHammer Hammertoe Correction System can be used with K-wires for the delivery of implants or the temporary stabilization of outlying joints (e.g. MTP Joint).
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
# DATE PREPARED
02/15/2024
#### MANUFACTURER AND 510(k) OWNER
Nvision Biomedical Technologies, Inc. 4590 Lockhill Selma San Antonio, TX 78249, USA Telephone: (210) 545-3713 Fax: (866) 764-1139 Official Contact: Diana Langham, Director of Regulatory and Corporate Compliance
#### REPRESENTATIVE/CONSULTANT
Analaura Villarreal-Berain, Extremity Project Engineer Nvision Biomedical Technologies Telephone: (210) 545-3713 ext. 109 Email: analauravillarreal(@nvisionbiomed.com
#### PROPRIETARY NAME OF SUBJECT DEVICE
SnapHammer Hammertoe Correction System
COMMON NAME Screw, Fixation, Bone
## DEVICE CLASSIFICATION
Smooth or Threaded Metallic Bone Fixation Fastener (Classification Regulations: 21 CFR 888.3040, Product Codes: HWC, HTY, Class: II)
#### PREMARKET REVIEW
Orthopedic Device Panel
#### INDICATIONS FOR USE
The SnapHammer Hammertoe Correction System is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe, and mallet toe. Cannulated implants in the SnapHammer Hammertoe Correction System can be used with K-wires for the delivery of implants or the temporary stabilization of outlying joints (e.g. MTP Joint).
#### DEVICE DESCRIPTION
The SnapHammer Hammertoe Correction System is comprised of an additively manufactured titanium fixation device attached to a snap-off post. The snap-off post aids in a quick insertion and snaps off after implant has been threaded in. The implants are offered in Ø3.0mm and Ø3.50mm diameters and in lengths ranging from 12mm. The system includes drills, implant inserters, and sizers manufactured from medical grade stainless steel.
{4}------------------------------------------------
## PREDICATE DEVICE IDENTIFICATION
The SnapHammer Hammertoe Correction System is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Predicate |
|---------------|--------------------------------------------------|------------|
| K221283 | 4Webb Medical Inc. Hammertoe Truss System (HTS)* | Primary |
| K213421 | Nvision Vector Hammertoe Correction System** | Additional |
*Additional clearances include K190926
** Additional clearances include K183055
The following reference devices are also cited in this submission:
- Nvision Biomedical's Quantum Cervical Plate (K210424)
## SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the SnapHammer Hammertoe Correction System. The following tests were performed to demonstrate safety based on recognized consensus standards and current industry practice:
- . Mechanical Performance: Torsion, driving torque, and axial pullout (per ASTM F543), as well as static and dynamic bending.
- . Additive Manufacturing: Tensile testing per ASTM F3001 and residual powder removal (particle analysis per ASTM F1877).
The results of these tests indicate that the SnapHammer Hammertoe Correction System is substantially equivalent to the predicate devices.
## EQUIVALENCE TO PREDICATE DEVICES
Nvision believes that the SnapHammer Hammertoe Correction System is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has a similar design, similar dimensions, and uses similar or identical materials as the devices cleared in K221283, K213421 and K210424. The subject device also has the same intended use, as well as similar technological characteristics (application and technique) as these predicates. The Indications for Use are equivalent and any minor differences in wording choices are insignificant. These technological characteristics have undergone testing and engineering analysis to ensure the device is as safe and effective as the predicates.
## CONCLUSION
Based on the testing performed, including torque, static and dynamic bending of device characteristics, it can be concluded that the subject device does not raise new issues of safety or efficacy compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed SnapHammer Hammertoe Correction System are assessed to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.