K231446 · Fitone Latex Products Co., Ltd. Guangdong · KGO · Sep 22, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231446
Device Name
Single-Use Latex Sterile Surgical Gloves (SG100)
Applicant
Fitone Latex Products Co., Ltd. Guangdong
Product Code
KGO · General, Plastic Surgery
Decision Date
Sep 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The Single-Use Latex Sterile Surgical Gloves are device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Device Story
Single-use sterile surgical gloves made of natural rubber latex; intended for use by operating room personnel to protect surgical wounds from contamination. Device is provided radiation-sterilized (SAL 10^-6) and is powder-free. Available in multiple sizes (6 through 9). Used in clinical/surgical settings to provide a barrier between the wearer and the patient. Benefits include reduction of contamination risk during surgical procedures. Device is disposable.
Clinical Evidence
No clinical data. Bench testing only. Testing included physical dimensions, tensile strength, elongation, hole detection (ASTM D5151), residual powder content (ASTM D6124), protein content (ASTM D5712), and biocompatibility (ISO 10993-5, 10, 11). All results met acceptance criteria.
Technological Characteristics
Natural rubber latex; powder-free; radiation sterilized (SAL 10^-6); shelf life 3 years. Complies with ASTM D3577-19 (Type I), ASTM D5151-19, ASTM D6124-06, and ASTM D5712-15. Biocompatibility per ISO 10993-1, 10, and 11.
Indications for Use
Indicated for operating room personnel to protect surgical wounds from contamination. Intended for use by healthcare professionals.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
September 22, 2023
Fitone Latex Products Co., Ltd. Guangdong % Kvra Kang Director Landlink Healthcare Technology (Shanghai) Co., Ltd Room 1308, Baohua International Plaza, West Guangzhong Road 555. Jingan District Shanghai, 200072 China
Re: K231446
Trade/Device Name: Single-Use Latex Sterile Surgical Gloves (SG100) Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: September 11, 2023 Received: September 11, 2023
Dear Kyra Kang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B2: Personal Protective Equipment, Reprocessing & Disinfection Devices Team OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K231446
Device Name
Single-Use Latex Sterile Surgical Gloves (SG100)
Indications for Use (Describe)
The Single-Use Latex Sterile Surgical Gloves are device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
### K231446
#### l. Submitter
Fitone Latex Products Co., Ltd. Guangdong No.5 Huitong road, Lingbei Industrial Zone, Suixi, 524338 Zhanjiang, Guangdong, China
Contact person: Christine Ou Position: Manager Tel.: 0759-7905808 E-mail: market-intl@fitonelatex.com
Preparation date: Sept. 19, 2023
#### Submission Correspondent
Ms. Kyra Kang Landlink Healthcare Technology (Shanghai) Co., Ltd. E-mail: kyra.kang@landlink-health.com
#### US Agent
Qihui Zhang ZYPPEL LLC 1337 Massachusetts Avenue #158 Arlington MA, MA US 02476
#### II. Proposed Device
| Device | Trade Name: Single-Use Latex Sterile Surgical Gloves |
|--------------------|------------------------------------------------------|
| Model: | SG100 |
| Common name: | Surgeon's Gloves |
| Regulation Number: | 21 CFR 878.4460 |
| Regulatory Class: | Class I |
| Product code: | KGO |
| Review Panel: | General Hospital |
{4}------------------------------------------------
#### III. Predicate Device
| 510(k) Number: K212848 | |
|------------------------|-------------------------------------------|
| Trade name: | Sterile Latex Surgical Gloves Powder Free |
| Common name: | Surgeon's Gloves Classification: Class I |
| Product Code: | KGO |
| Manufacturer | Pentavest Holdings Sdn Bhd |
#### IV. Device Description
- The proposed device, Single-Use Latex Sterile Surgical Gloves is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination.
- The proposed device is made of natural rubber latex, per standard ASTM D3577- 09(15), the rubber surgical gloves classification is:
" Type 1-qloves compounded primarily from nature rubber latex." The qloves are powderfree and available in white in sizes 6, 6.5, 7, 7.5, 8, 8.5 and 9.
The proposed device is provided Radiation sterilized to achieve the sterility Assurance Level (SAL) of 10-6. The latex surgical glove shelf life is 3 years.
#### V. Indication for use
The Single-Use Latex Sterile Surgical Glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
{5}------------------------------------------------
| Characteristic | Proposed device<br>K231446 | Predicate device<br>K212848 | Discussion |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product name | Single-Use Latex<br>Sterile Surgical<br>Gloves | Sterile Latex Surgical<br>Gloves Powder Free | Similar |
| Indications For<br>Use | The Single-Use Latex<br>Sterile Surgical Glove<br>is a device made of<br>natural rubber<br>intended to be worn<br>by operating room<br>personnel to protect a<br>surgical wound from<br>contamination. | A Sterile Latex Surgical<br>Gloves Powder Free is a<br>device made of natural<br>rubber intended to be worn<br>by operating room personnel<br>to protect a surgical wound<br>from contamination. | Similar |
| Product Code | KGO | KGO | Same |
| Regulation<br>Number | 21 CFR 878.4460 | 21 CFR 878.4460 | Same |
| Powdered or<br>Powder free | Powdered free | Powdered free | Same |
| Classification as<br>per ASTM<br>D3577- 2019 | Type I - gloves<br>compounded primarily<br>from natural rubber<br>latex | Type I - gloves<br>compounded primarily from<br>natural rubber latex | Same |
| Type of use | Over the counter use | Over the counter use | Same |
| Shelf Life | 3 Years | 3 Years | Same |
| Sterilization | Radiation, SAL- 10-6 | Radiation, SAL- 10-6 | Same |
| Single use | Yes | Yes | Same |
| Label and<br>Labeling | Meet FDA's<br>Requirement | Meet FDA's Requirement | Same |
| Protein Content | 141µg/dm² | 50µg/dm² | Similar |
# VI. Comparison of Technological Characteristics with the Predicate Device
Table
{6}------------------------------------------------
| Characteristics | Acceptance criteria of the<br>standard | | Remarks |
|-----------------------------------------------------|------------------------------------------------------|------------------------------------------------------|----------------------------------------------------------------------------------------------|
| | Subject device<br>K231446 | Predicate device<br>K212848 | |
| Dimensions Length:-<br>Min 265 mm | 281 mm | 380 mm | Similar<br>Meets ASTM D3577-2019,<br>Standard Specification for<br>Rubber Surgical Gloves |
| Size | | | |
| 6.0 (76+/-6mm) | 76mm | 74mm | Similar |
| 6.5 (83+/-6mm) | 84mm | 86mm | Meets ASTM D3577-2019,<br>Standard Specification for<br>Rubber<br>Surgical Gloves |
| 7.0 (89+/-6mm) | 90mm | 92mm | |
| 7.5 (95+/-6mm) | 96mm | 98mm | |
| 8.0 (102+/6mm) | 102mm | 105mm | |
| 8.5 (108+/6mm) | 109mm | 110mm | |
| 9.0 (114+/6mm) | 115mm | 116mm | |
| Cuff, Palm, Finger Tip<br>Min<br>0.10 mm | Cuff- 0.16mm<br>Palm-0.18mm<br>Finger Tip-<br>0.22mm | Cuff- 0.12mm<br>Palm-0.16mm<br>Finger Tip-<br>0.21mm | |
| Tensile Strength<br>24Mpa minimum<br>Before Aging | 37Mpa | 28.55Mpa | Similar<br>Meets ASTM D3577-2019,<br>Standard Specification for |
| Ultimate Elongation<br>750% minimum<br>Before Aging | 814% | 870% | Rubber Surgical Gloves |
| Stress at 500%<br>5.5 MPa Max<br>Before Aging | 3.5Mpa | 5.1Mpa | |
| Tensile Strength 18Mpa<br>minimum<br>After Aging | 32.49Mpa | 23.48Mpa | Similar<br>Meets ASTM D3577-2019,<br>Standard Specification for<br>Rubber<br>Surgical Gloves |
| Ultimate Elongation<br>560% minimum After<br>Aging | 760% | 731% | |
{7}------------------------------------------------
| Characteristics | Acceptance criteria of the<br>standard | | Remarks |
|---------------------------------------------------------------------------|----------------------------------------|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject device<br>K231446 | Predicate<br>device<br>K212848 | |
| Freedom from<br>Holes<br>AQL 1.5 | AQL 1.5 | AQL 1.0 | Similar<br>Meets ASTM D3577-2019 and<br>ASTM D5151-2019,<br>Standard Test Method for<br>Detection of Holes in Medical<br>Gloves |
| Powder residue for<br>powder free glove<br>Powder content <<br>2 mg/Glove | 0.21 mg/Glove | 0.40 mg/Glove | Similar<br>Meets ASTM D3577-2019 and<br>ASTM<br>D6124-06,(Reapproved 2017),<br>Standard Test Method for<br>Residual Powder on Medical<br>Gloves |
{8}------------------------------------------------
| Characteristics | Biocompatibility | Acceptance criteria of the standard | | Remarks |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|---------|
| | | Subject device<br>K231446 | Predicate device<br>K212848 | |
| Skin Irritation &<br>Skin Sensitization | ISO 10993-<br>10,Biological<br>evaluation of medical<br>devices - Part 10:<br>Tests for irritation and<br>skin<br>sensitization | Non- irritant and<br>Non-Sensitizer | Non- irritant and<br>Non-Sensitizer | Same |
| In vitro cytotoxicity | ISO 10993-<br>5:2009(E),Biological<br>Evaluation of Medical<br>Devices - Part 5-Tests<br>for<br>in vitro Cytotoxicity | Cytotoxic | Cytotoxic | Same |
| Material Mediated<br>pyrogenicity | ISO 10993-11:2017(E)<br>,Biological Evaluation<br>of Medical Devices -<br>Part 11, Tests for<br>Systemic Toxicity and<br>USP 41<br><151>Pyrogen Test | Non pyrogenic | No data available | ----- |
| Acute Systemic<br>Toxicity | ISO 10993-11:2017(E)<br>,Biological Evaluation<br>of Medical Devices -<br>Part 11, Tests for<br>Systemic Toxicity | Under the<br>conditions of<br>study the<br>device extracts<br>do not pose an<br>acute systemic<br>toxicity<br>concern | Under the<br>conditions of study<br>the<br>device extracts do<br>not pose an acute<br>systemic toxicity<br>concern | Same |
| Bacterial<br>Endotoxin | USP 42 <85> | <20EU/pair of<br>gloves | <20EU/pair of<br>gloves | Same |
{9}------------------------------------------------
#### VII. Non-Clinical Testing
Non clinical tests were conducted in accordance with following standards to verify that the proposed device met all design specifications.
- · ASTM D3577-19: Standard Specification for Rubber Surgical Gloves.
- · ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
- ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
- ASTM D5712-15 Standard Test Method for Analysis of Aqueous Extractable Protein in Natural Rubber in Latex, Natural Rubber, and Elastomeric Products Using the Modified Lowry Method
- ASTM D7160-05 Standard Practice for Determination of Expiration Dating for medical Gloves
- ISO 10993-1: 2009 Biological evaluation of medical devices Part1: Evaluation and testing within a risk management process
- · ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
- ISO 10993-11:2017 Biological Evaluation of Medical Devices Part 5: Tests For systemic toxicity
- · ISO 11137-1-2006/ (R) 2010 validation of sterilization process
- · ISO 11137-2:2013, sterilization of health care products radiation part 2: Establishing the sterilization dose
{10}------------------------------------------------
| Test<br>Methodology | Purpose | Acceptance Criteria | Results |
|-------------------------|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| ASTM D3577-19 | Physical Dimension | The actual measured<br>dimension of the<br>gloves shall meet<br>the stated<br>tolerance specified<br>in Table 2 of the<br>ASTM D3577-19 | Meets the<br>requirement |
| ASTM D3577-19 | Determination of<br>Physical Properties | Before and after<br>Accelerated aging,<br>the gloves shall<br>conform to the<br>physical<br>requirements in<br>Table 3 of ASTM<br>D3578-19 | Meets the<br>requirement |
| ASTM D5151-19 | Water Leak Test for<br>Detection of Holes | The gloves shall be<br>free from hole<br>when tested in<br>accordance with<br>the method given<br>in ASTM D5151 | Meets the<br>requirement AQL<br>1.5 |
| ASTM D6124-<br>06(2017) | Residual Powder<br>Content Test | The powder residue<br>content shall be<br>not more than<br>2mg per gloves. | Meets the<br>requirement |
| ASTM D5712-15 | Protein Content | <200µg/dm² | Meets the<br>requirement |
| ISO 10993-10:<br>2010 | Skin Sensitization | Non-sensitizing | Meets the<br>requirement |
| ISO 10993-10:<br>2010 | Skin irritation | Non-irritating | Meets the<br>requirement |
| ISO 10993-11:<br>2017 | Systemic toxicity | Non-systemic toxicity | Meets the<br>requirement |
{11}------------------------------------------------
#### VIII.Clinical Testing
No clinical study is included in this submission.
#### IX. Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in this 510(K) submission, the Single-Use Latex Sterile Surgical Gloves is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K212848.
{12}------------------------------------------------
K231446 Page 10 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.