K231351 · Siemens Medical Soultions USA, Inc. · LLZ · Jul 13, 2023 · Radiology
Device Facts
Record ID
K231351
Device Name
Chondral Quant
Applicant
Siemens Medical Soultions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Jul 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Knee cartilage segmentation
—
—
—
Training data: more than 31 cases; Testing data: more than 100 cases (3T MRI), more than 11 cases (7T MRI), more than 9 cases (1.5T MRI)
—
—
—
Knee cartilage volumetry
—
—
—
Training data: more than 31 cases; Testing data: more than 100 cases (3T MRI), more than 11 cases (7T MRI), more than 9 cases (1.5T MRI)
—
—
—
Knee cartilage thickness
—
—
—
Training data: more than 31 cases; Testing data: more than 100 cases (3T MRI), more than 11 cases (7T MRI), more than 9 cases (1.5T MRI)
—
—
—
Indications for Use
Chondral Quant is a musculoskeletal post-processing software application that allows assessment of knee cartilage condition based on Magnetic Resonance Imaging (MRI).
Device Story
Chondral Quant is a musculoskeletal post-processing software application for knee cartilage assessment. It takes morphological 3D MRI series of the knee as input; optionally accepts parametric maps (e.g., T2/T2*) as secondary input. The device performs automated cartilage segmentation, sub-segmentation, volumetry, and thickness calculations. It registers morphological images with parametric maps to transfer segmentations. Used in clinical settings by healthcare professionals via the syngo.via platform. Operates in interactive mode (manual user adjustments) or 'PACS-ready' fully-automated mode. Outputs include statistical tables (volume, thickness, mapping results), segmentation maps, cartilage thickness maps, 3D models, and PDF reports. Assists clinicians in evaluating cartilage status, potentially aiding in the diagnosis or monitoring of joint conditions.
Clinical Evidence
No clinical data. Bench testing only. Performance validated using >31 training cases and >120 testing cases across 1.5T, 3T, and 7T MRI systems. Verification and validation performed per 21 CFR 820.30, including unit, integration, and system testing.
Technological Characteristics
Software-only musculoskeletal post-processing application. Operates within syngo.via framework. Conforms to IEC 62304 (software life cycle). Features automated segmentation, volumetry, thickness calculation, and image registration. Bias field correction included. Supports MR modalities.
Indications for Use
Indicated for assessment of knee cartilage condition in patients undergoing MRI. Intended for use by healthcare professionals.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 13, 2023
Siemens Medical Soultions USA, Inc. % Milind Dhamankar Clinical Affairs Professional 40 Liberty Boulevard Malvern, Pennsylvania 19335
Re: K231351
Trade/Device Name: Chondral Quant Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ, LNH Dated: April 28, 2023 Received: May 9, 2023
Dear Milind Dhamankar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231351
Device Name
Chondral Quant
Indications for Use (Describe)
Chondral Quant is a musculoskeletal post-processing software application that allows assessment of knee cartilage condition based on Magnetic Resonance Imaging (MRI).
| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) |
|----------------------------------------------------------------|
|----------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the text "K231351" at the top. Below that is the Siemens Healthineers logo. The Siemens part of the logo is in teal, and the Healthineers part is in orange. There are also some orange dots to the right of the Healthineers text.
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR § 807.92.
| 1. General Information | |
|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment: | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Malvern, PA 19355, USA<br>Registration Number: 2240869 |
| Date Prepared: | July 12, 2023 |
| Manufacturer: | Siemens Healthcare GmbH<br>Henkestr. 127<br>91052 Erlangen<br>Germany<br>Registration Number: 3002808157 |
| 2. Contact Information | |
| | Milind Dhamankar<br>Clinical Affairs Professional<br>Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Malvern, PA 19355, USA<br>Phone: +1(610) 517-9484<br>Fax: (610) 448-6547<br>E-mail: milind.dhamankar@siemens-<br>healthineers.com |
| 3. Device Name and Classification | |
| Device/ Trade name:<br>Classification Name: | Chondral Quant<br>System, Image Processing, Radiological<br>Picture Archiving and Communications<br>System |
| Classification Panel:<br>CFR Code:<br>Classification:<br>Product Code: | Radiology<br>21 CFR § 892.2050<br>II<br>Primary: LLZ |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
#### 4. Legally Marketed Predicate Device1
Trade name:
510(k) Number: Classification Name:
Classification Panel: CFR Code: Classification: Product Code:
syngo.MR Applications - syngo.MR Brain Morphometry K182904 Picture Archiving and Communication System (PACS) Radiology 21 CFR § 892.2050 11 Primary: LLZ Secondary: LNH
#### 5. Intended Use / Indications for Use
Chondral Quant is a musculoskeletal post-processing software application that allows assessment of knee cartilage condition based on Magnetic Resonance Imaging (MRI).
#### 6. Device Description
The medical device Chondral Quant, software version VA10A, is a musculoskeletal postprocessing application that allows evaluating the status of knee cartilage based on Magnetic Resonance Imaging (MRI).
The software is part of the syngo OpenApps framework and can be used from within syngo.via like any other syngo.via workflow.
Version VA10A is the initial version of this medical device.
Chondral Quant processes a morphological 3D series of the knee joint and performs an automated segmentation of the knee's cartilage. The segmentation may be modified by the user. Chondral Quant will also perform a sub-segmentation of the knee cartilage. Chondral Quant additionally performs volumetry and thickness calculation on the segmented areas. Optionally, it is possible to provide a parametric map as secondary input to Chondral Quant. Commonly, this will be a T2 or T2* map. Chondral Quant will perform a registration of the morphological image and the parametric map and transfer the segmentation to the parametric map.
A statistical evaluation of the mapping results for the 21 sub zones will be additionally provided in this case.
All output will be provided in the form of a table showing statistical evaluation of the assessment (volume and mean, median and standard deviation of thickness and mapping results for every region). Additionally, Chondral Quant will generate output maps (segmentation map, cartilage thickness map) and 3D models of the segmented cartilage
<sup>1</sup> The predicate device has not been subject to a design-related recall.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
(zone model and thickness model). Finally, a PDF report containing the table values and the 3D models will be generated as output.
Alternatively, the application Chondral Quant may also be executed in a "PACS-ready" mode, i.e. fully-automated and without user interaction. The results will be sent to PACS automatically. In this case the series names will be marked with a prefix "AUTO GENERATED" for a clear indication of the automatic mode.
The subject device, Chondral Quant with software VA10A, consists of new and modified features that are similar to what is currently offered on the predicate device. The subject device includes the following modifications in comparison to the predicate device: new body region compared to predicate device:
- -Automatic segmentation and volumetry of the cartilage
- -Thickness calculation
#### 7. Substantial Equivalence
Chondral Quant with software version VA10A is substantially equivalent to the following predicate device:
| Predicate Device | FDA Clearance Number<br>and Date | Product<br>Code | Manufacturer |
|----------------------------------------------------------|----------------------------------|-----------------|----------------------------|
| syngo.MR Applications –<br>syngo.MR Brain<br>Morphometry | K182904 - July 5, 2019. | LLZ, LNH | Siemens Healthcare<br>GmbH |
The following summary comparison provides an overview of changes Chondral Quant with software VA10A compared to the predicate device. Hardware comparison is not applicable as Chondral Quant is a SW-only device.
| | Subject Device<br>Chondral Quant VA10A | Predicate Device<br>syngo.MR Applications - syngo.MR<br>Brain Morphometry (K182904) |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Purpose | Chondral Quant is a musculoskeletal<br>post-processing software application<br>that allows assessment of knee cartilage<br>condition based on Magnetic<br>Resonance Imaging (MRI).<br>Chondral Quant processes a<br>morphological 3D series of the knee<br>joint and performs an automated<br>segmentation of the knee's cartilage.<br>Chondral Quant additionally performs<br>volumetry and thickness calculation on<br>the segmented areas. | syngo.MR Applications is a syngo based<br>postacquisition image processing<br>software for viewing, manipulating,<br>evaluating, and analyzing MR, MR-PET,<br>CT, PET, CT-PET images and MR spectra.<br>syngo.MR Brain Morphometry offers a<br>comprehensive package for the<br>automatic calculation of the volume<br>properties of different brain structures<br>using MPRAGE datasets, which are<br>typically acquired for a typical MR<br>examination of the head. |
| Features / Functionalities | | |
| Integration into<br>general syngo.via<br>workflow concept | Yes | Yes |
| Support Images<br>Modalities | MR | MR |
| Data<br>Acquisition Protocol | MRI Image Series | MRI Image Series |
| Body Region | Knee | Brain |
| Clinical Use Case | Automatic segmentation and<br>volumetry of the cartilage | Automatic segmentation and<br>volumetry of different brain structures |
| | Segmentation<br>algorithm | |
| Template registration | | Yes |
| | Atlas-based initialization for the bone<br>segmentation | Free tissue classification |
| Bias field correction | Yes | Yes |
| | Preprocessing bias field correction | - |
| Skull-stripping | No | Yes |
| Tissue classification | Yes | Yes |
| | Cartilage segmentation | Brain tissue classification |
| Segmentation quality<br>check | No | Yes |
| | Volumetry | Yes |
| Volume calculation | | Lobe-wise GM volumes |
| Thickness calculation | Yes | No |
| | 3D Segmentation | Yes |
| Clinical sub-regions | 21 clinical sub-regions | 57 clinical sub-regions |
| | Output | - DICOM<br>- Maps and statistical evaluations |
| Workflow | - PACS-ready without user interaction<br>- Interactive mode (manual changes by user possible) | |
| Normative References<br>/ Deviation Maps | No | Yes |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
### 8. Technological Characteristics
The subject device Chondral Quant with software version VA10A is substantially equivalent to the predicate device with regard to the operational environment, programming language, operating system and performance.
The subject device conforms to the standard for medical device software (IEC 62304).
#### 9. Nonclinical Tests
The following performance testing was conducted on the subject devices.
| Performance Test | Tested Software | Source/Rationale for test |
|---------------------------------------|----------------------|------------------------------------------------------------------------------------------------------|
| Subsystem<br>Verification Report | Chondral Quant VA10A | Guidance for the Content of<br>Premarket Submissions for<br>Software Contained in Medical<br>Devices |
| Solution Validation<br>Summary Report | | |
| Algorithm | Cases | Equipment: |
|----------------|---------------------|------------------|
| Training data: | more than 31 | 3T MRI systems |
| Testing data: | more than 100 cases | 3T MRI systems |
| | more than 11 cases | 7T MRI systems |
| | more than 9 cases | 1.5T MRI systems |
Non-clinical tests such as unit test, integration testing, and system test are passed.
The system test results indicate that open defects were identified which had no impact on safety and effectiveness of Chondral Quant with software version VA10A.
## 10. Clinical Tests / Publications
No clinical tests were conducted for the subject device.
Clinical publications and other support documents were referenced to provide information on the use, testing and validation of the Subject Device.
No animal testing has been performed.
#### 11. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971, to identify and provide mitigation of potential hazards early in the design cycle and continuously throughout the development of the product. Siemens Healthcare GmbH adheres to recognized and established industry standards, to minimize hazards. Furthermore, the device is intended for healthcare professionals.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
| Recognition<br>Number | Product<br>Area | Title of Standard | Reference Number<br>and date | Standards<br>Development<br>Organization |
|-----------------------|--------------------------|-----------------------------------------------------------------------------------------|---------------------------------------------------------|------------------------------------------|
| 5-125 | General I<br>(QS/<br>RM) | Medical devices - Application of<br>risk management to medical<br>devices | 14971 Third Edition<br>2019-12 | ISO |
| 5-114 | General I<br>(QS/<br>RM) | Medical devices - Part 1:<br>Application of usability engineering<br>to medical devices | 62366-1:2015,<br>Edition 1.0 2015-02 | ANSI AAMI IEC |
| 13-79 | Software/<br>Informatics | Medical device software - Software<br>life cycle processes | 62304 Edition 1.1<br>2015-06<br>CONSOLIDATED<br>VERSION | IEC |
Chondral Quant with software version VA10A conforms to the following standards:
#### 12. Conclusion as to Substantial Equivalence
The extensive testing of Chondral Quant with software version VA10A has been successfully completed. All risk mitigations, as identified in the Risk Analysis and all relevant SSRS/FS requirements for Chondral Quant with software version VA10A have been tested and verified successfully.
Verification and validation of the product within the meaning of the Quality System Regulation (21 CFR § 820.30) have been performed by trained personnel. Chondral Quant with software version VA10A has been found to be validated for its intended use.
Indications for Use for subject device is different compared to the predicate device, however the intended use is the same. syngo.via is the hosting platform for both the subject device and the predicate device. Both the devices are integrated into the already cleared and marketed general syngo.via workflow concept. Chondral Quant processes a morphological 3D series of the knee joint and performs an automated segmentation of the knee's cartilage. Chondral Quant additionally performs volumetry and thickness calculation on the segmented areas and optionally performs registration of a parametric map while syngo.MR Brain Morphometry processes a morphological 3D data series and offers a comprehensive package for the automatic calculation of the volume properties of different brain structures.
The difference between the predicate device and the subject device doesn't impact the safety and effectiveness of the subject device.
Therefore, it is Siemens' opinion that the safety and effectiveness of the subject device have been fully verified by objective evidence, and that the subject device performs as safely and effectively as the predicate device (K182904) and the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.