K231341 · Zhejiang Tianqing Manufacturing Technology Group Co., Ltd. · HEB · Aug 14, 2023 · Obstetrics/Gynecology
Device Facts
Record ID
K231341
Device Name
Organic cotton tampon, Viscose tampon
Applicant
Zhejiang Tianqing Manufacturing Technology Group Co., Ltd.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Aug 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
The tampon is intended for insertion into the vagina for the absorption of menstrual or other vaginal discharge.
Device Story
Organic cotton and viscose tampons; traditional cylindrical, bullet-like shape; available as digital or applicator-based devices. Applicators feature smooth, rounded tips for insertion. Components include absorbent pledget (100% organic cotton or 100% viscose), security veil (polyethylene/polyethylene terephthalate), and removal string (100% organic cotton or polyester/cotton). Applicators made of polyethylene and polypropylene. Provided non-sterile for single-use. Used by patients for menstrual hygiene management. Device functions via passive absorption of vaginal fluids. No clinical decision-making or healthcare provider intervention required; patient self-use.
Clinical Evidence
No clinical data. Bench testing only, including performance testing (dimensions, absorbency, removal string strength, fiber shedding, integrity, chemical residues), biocompatibility (ISO 10993-5, -10, -23, -11), and microbiology testing (Staphylococcus aureus growth, TSST-1 production, and vaginal microflora impact).
Technological Characteristics
Materials: 100% organic cotton or 100% viscose pledget; polyethylene/polyethylene terephthalate overwrap; organic cotton or polyester/cotton string; polyethylene/polypropylene applicator. Passive absorption principle. Non-sterile, single-use. Dimensions vary by absorbency (light, regular, super, super plus). Complies with ISO 10993-1 and FDA guidance for menstrual tampons.
Indications for Use
Indicated for use by individuals requiring absorption of menstrual or other vaginal discharge.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 14, 2023
Zhejiang Tianqing Manufacturing Technology Group Co., Ltd. % Grace Liu Consultant Shenzhen Joyantech Consulting Co. Ltd 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518000 China
K231341 Re:
Trade/Device Name: Organic cotton tampon. Viscose tampon Regulation Number: 21 CFR§ 884.5470 Regulation Name: Unscented Menstrual Tampon Regulatory Class: II Product Code: HEB Dated: August 3, 2023 Received: August 3, 2023
Dear Grace Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Reginald K. Avery -S
for
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231341
#### Device Name
Organic cotton tampon, Viscose tampon
Indications for Use (Describe)
The tampon is intended for insertion into the absorption of menstrual or other vaginal discharge.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K231341
# 1. Contact Details
#### 1.1 Applicant information
| Applicant Name | Zhejiang Tianqing Manufacturing Technology Group Co., Ltd. |
|----------------|-----------------------------------------------------------------|
| Address | Lianshi Industrial Park, Nanxun District, Huzhou City, Zhejiang |
| | Province, China |
| Website | www.tianqingglobal.com |
| Contact person | Roy Du |
| Phone No. | +86-13817862379 |
| E-mail | roy.d@tianqingglobal.com |
| Date Prepared | August 8, 2023 |
| | |
#### 1.2 Submission Correspondent
| | Shenzhen Joyantech Consulting Co., Ltd | |
|-------------------------|-------------------------------------------------------------|--|
| | 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan | |
| 早远大成 | District, Shenzhen, Guangdong Province, China | |
| Phone No. | +86-755-86069197 | |
| Contact person | Grace Liu, Field Fu | |
| Contact person's e-mail | grace@cefda.com, field@cefda.com | |
| Website | http://www.cefda.com | |
# 2. Device Information
| Trade name | Organic cotton tampon, Viscose tampon |
|---------------------|---------------------------------------|
| Common name | Unscented Menstrual Tampon |
| Classification | II |
| Classification name | Tampon, Menstrual, Unscented |
| Product code | HEB |
| Regulation No. | 21 CFR 884.5470 |
## 3. Legally Marketed Predicate Device
| Trade Name | Tosama 100% Organic Cotton Menstrual Tampon |
|-----------------------|---------------------------------------------|
| 510(k) Number | K151170 |
| Product Code | HEB |
| Manufacturer | TOSAMA, d.o.o. |
| Design Related Recall | None |
# 4. Device Description
The organic cotton tampon and viscose tampon are traditional unscented menstrual tampons. Except for the raw materials, they share the same design. The organic cotton tampon and viscose tampon, both have two types i.e., digital tampon and applicator tampon. The tampon applicator is available in two colors - milky and pink. Each device consists of a tampon, including an absorbent 100% organic cotton or 100% viscose pledget ("absorbent core") surrounded by polyethylene and polyethylene terephthalate
{4}------------------------------------------------
overwrap ("security veil"), a 100% organic cotton or polyester and cotton removal string ("withdrawal cord"), and an applicator (only for the applicator tampon). The applicator is made of polyethylene and polypropylene. The tampon is of the traditional cylindrical, bullet-like shape. The applicator has a smooth, rounded tip to ease insertion.
Both, organic cotton tampon and viscose tampon, are provided in 4 absorbencies: light (≤6g), regular (6~9g), super (9~12g) and super plus (12~15g). Each device is individually wrapped in a polyethylene/nonwoven laminated film or cellophane film and packaged in sealed multi-unit containers for retail sale. It is provided non-sterile and for single use only.
# 5. Intended Use/Indication for Use
The tampon is intended for insertion into the vagina for the absorption of menstrual or other vaginal discharge.
## 6. Technological Characteristics Comparison
| Comparison item | Subject Device (K231341) | | Predicate Device<br>(K151170) | Comment | |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------|
| Manufacturer | Zhejiang Tianqing Manufacturing<br>Technology Group Co., Ltd. | | TOSAMA, d.o.o. | None | |
| Product Name | Organic cotton<br>tampon | Viscose<br>tampon | Tosama 100%<br>Organic Cotton<br>Menstrual<br>Tampon | None | |
| Product Code | HEB | | HEB | Same | |
| Regulation Number | 21 CFR § 884.5470 | | 21 CFR § 884.5470 | Same | |
| Classification | Class II | | Class II | Same | |
| Intended Use / Indications for<br>Use | The tampon is intended for insertion<br>into the vagina for the absorption of<br>menstrual or other vaginal discharge. | | The Tosoma 100%<br>Organic Cotton<br>Menstrual Tampon is<br>intended for insertion<br>into the vagina for<br>the absorption of<br>menstrual or other<br>vaginal discharge. | Same | |
| Single Use | Yes | | Yes | Same | |
| Sterility | Non-sterile | | Non-sterile | Same | |
| Design | Tampon with cylindrical shape and<br>bullet-like tip.<br>Applicator with smooth, rounded tip. | | Tampon with<br>cylindrical shape and<br>bullet-like tip.<br>Applicator with<br>smooth, rounded tip. | Same | |
| Absorbency | | | | | |
| Light | ≤6g | | N/A | | |
| Regular | 6~9g | | 6~9g | Different | |
| Super | 9~12g | | 9~12g | | |
| Super plus | 12~15g | | 12~15g | | |
| Product dimensions | | | | | |
| Digital tampon | | | | | |
| Light | Pledget length | (43~49) mm | | N/A | |
| | Pledget diameter | (9.5~12.5) mm | N/A | | |
| | Removal string<br>length | (120~160) mm | | | |
| Regular | Pledget length | (44~50) mm | (43.0~46.0) mm | | |
| | Pledget diameter | (10.5~13.5) mm | (11.0~12.0) mm | | |
| | Removal string<br>length | (120~160) mm | (130~160) mm | | |
| Super | Pledget length | (45~51) mm | (48.0~51.0) mm | | |
| | Pledget diameter | (11.5~14.5) mm | (12.0~13.0) mm | | |
| | Removal string<br>length | (120~160) mm | (130~160) mm | | |
| Super plus | Pledget length | (45~51) mm | (48.0~51.0) mm | | |
| | Pledget diameter | (12.5~15.5) mm | (14.0~15.0) mm | | |
| | Removal string<br>length | (120~160) mm | (130~160) mm | Different | |
| Applicator tampon | | | | | |
| | Pledget length | (43~49) mm | | | |
| | Pledget diameter | (9.5~12.5) mm | | | |
| | Removal string<br>length | (120~160) mm | | | |
| Light | Applicator length | 130.1 mm | N/A | | |
| | Applicator<br>diameter | 15.2 mm | | | |
| | Pledget length | (44~50) mm | (43.0~46.0) mm | | |
| | Pledget diameter | (10.5~13.5) mm | (11.0~12.0) mm | | |
| | Removal string<br>length | (120~160) mm | (130~160) mm | | |
| Regular | Applicator length | 130.1 mm | 120 mm | | |
| | Applicator<br>diameter | 15.2 mm | 13 mm | | |
| | Pledget length | (45~51) mm | (43.0~46.0) mm | | |
| | Pledget diameter | (11.5~14.5) mm | (12.0~13.0) mm | | |
| | Removal string<br>length | (120~160) mm | (130~160) mm | | |
| Super | Applicator length | 131.0 mm | 120 mm | | |
| | Applicator<br>diameter | 16.7 mm | 16 mm | | |
| Super plus | Pledget length | (45~51) mm | (43.0~46.0) mm | | |
| | Pledget diameter | (12.5~15.5) mm | (14.0~15.0) mm | | |
| | Removal string<br>length | (120~160) mm | (130~160) mm | | |
| | Applicator length | 131.0 mm | 120 mm | | |
| | Applicator<br>diameter | 16.7 mm | 18 mm | | |
| Component<br>Materials | Pledget | 100% Organic<br>Cotton | 100% Viscose | 100% Organic<br>Cotton | Different |
| | Overwrap | Polyethylene and<br>Polyethylene<br>terephthalate | Polyethylene and<br>Polyethylene<br>terephthalate | 100% Organic<br>Cotton | |
| | Removal string | 100% Organic<br>Cotton | PolyesterAnd<br>Cotton | 100% Organic Cotton | |
| | Applicator | Polyethylene<br>and<br>Polypropylene | Polyethylene<br>and<br>Polypropylene | TPO | |
| | Additives<br>and<br>Finishing<br>Agents | Anti-wicking agent<br>of removal string | Repellan T | Repellan T | |
| | | Finishing agent of<br>pledget | Fiber finishes | Fiber finishes | Not available |
| | Complies with ISO 10993-1 | | Yes | | Yes |
| | Compliance with microbiology<br>requirements of FDA Guidance<br>for Tampons | | Yes | | Yes |
| Labeling | | Complied with 21 CFR part 801 and<br>FDA Menstrual Tampons and Pad<br>guidance | | Complied with 21<br>CFR part 801 and<br>FDA Menstrual<br>Tampons and Pads<br>guidance | Same |
Table 1 Technological Characteristics Comparison Table
{5}------------------------------------------------
{6}------------------------------------------------
The subject devices have the same indication for use as the predicate device; therefore, the subject device has the same intended use as the predicate device. The differences in technological characteristics between the subject and predicate devices, outlined in Table 1, do not raise different questions for safety and effectiveness.
## 7. Summary of Non-clinical Testing
Non-clinical testing was conducted to verify that the subject devices meet all design and performance specifications similarly to the predicate device. The following tests were conducted.
#### Performance Testing
The following performance characteristics were assessed on the organic cotton tampon and the viscose tampon in accordance with the FDA guidance document "Guidance for Industry and FDA Staff - Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s)" issued on July 27, 2005.
- > Dimensions
- > Absorbency
- > Removal string strength
- > Fiber shedding
- > Tampon integrity
{7}------------------------------------------------
- > Chemical residues
#### Biocompatibility Testing
Biocompatibility studies were performed on the organic cotton tampon and the viscose tampon in accordance with the FDA guidance document "Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process" issued on September 4, 2020 as follows:
- A In vitro cytotoxicity test per ISO 10993-5:2009
- > Skin sensitization test per ISO 10993-10:2021
- > Vaginal irritation test per ISO 10993-23:2021
- A Acute systemic toxicity test per ISO 10993-11:2017 (tampon only)
All the above tests, except acute systemic toxicity, were performed on the applicator. The tampon and applicator were tested separately. The results demonstrated that the subject device is non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic.
#### Microbiology Testing
Per the FDA guidance document "Guidance for Industry and FDA Staff - Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s)" issued on July 27, 2005, the following microbiology testing was conducted on the highest absorbency of the final, finished form of the subject devices (i.e., the organic cotton tampon and the viscose tampon), and the test results showed that the subject devices do not:
- > enhance the growth of Staphylococcus aureus;
- > increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1); and do not
- > alter the growth of normal vaginal microflora.
#### 8. Clinical Testing
No clinical study is included in this submission.
#### 9. Conclusions
The nonclinical tests demonstrate that the proposed device is as safe and effective as the legally marketed device (K151170). Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.