K231333 · Globus Medical, Inc. · JDQ · Aug 4, 2023 · Orthopedic
Device Facts
Record ID
K231333
Device Name
STRETTO™ Cable System
Applicant
Globus Medical, Inc.
Product Code
JDQ · Orthopedic
Decision Date
Aug 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The STRETTO™ Cable System is indicated for use in: general orthopedic trauma surgery (e.g., fractures of the olecranon, patella, femur - including periprosthetic, pelvis, acetabulum, humerus and ankle, and acromioclavicular dislocations); prophylactic banding during total joint procedures; and, temporary reduction techniques for ORIF (Open Reduction Internal Fixation) procedures. STRETTO™ Screw Anchors are indicated for fractures that may not be securely held by either a screw or a cerclage device alone, and where cerclage is used in combination with bone screws and/or plates to provide internal fixation of fractured bone. STRETTO™ Press-In Anchors are indicated for use with a cerclage cable and plate to augment long bone fracture fixation, particularly when the use of screws would be inhibited, as in the presence of intramedullary implants.
Device Story
STRETTO™ Cable System consists of cables, crimps, and anchors used for bone fixation. Anchors include press-in styles for ANTHEM fracture plates and screw-in styles for ANTHEM screw hex recesses. System provides internal fixation for fractures or prophylactic banding. Used in orthopedic trauma surgery by surgeons. Device functions by securing bone fragments via cable tensioning and crimping, often in conjunction with plates or screws. Benefits include stabilization of fractures where traditional screw fixation is insufficient or inhibited by existing implants.
Clinical Evidence
Bench testing only. Mechanical testing (static and dynamic tension) performed using ASTM E8 as a guide. Bacterial endotoxin testing (BET) conducted per ANSI/AAMI ST72:2011. No clinical data provided.
Technological Characteristics
Materials: titanium alloy, commercially pure titanium, cobalt chromium alloy, cobalt chromium molybdenum alloy, or stainless steel. Components: cables, crimps, and anchors. Anchors include press-in and screw-in variants. Mechanical testing guided by ASTM E8. Sterilization/biocompatibility testing per ANSI/AAMI ST72:2011.
Indications for Use
Indicated for patients requiring orthopedic trauma surgery for fractures (olecranon, patella, femur, pelvis, acetabulum, humerus, ankle), acromioclavicular dislocations, prophylactic banding during total joint procedures, or temporary reduction for ORIF. Screw anchors indicated for fractures requiring combined cerclage and screw/plate fixation. Press-in anchors indicated for long bone fracture augmentation when screw use is inhibited by intramedullary implants.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
Predicate Devices
Pioneer Laboratories SDB Cerclage System (K992616)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
August 4, 2023
Globus Medical Inc. Jennifer Antonacci Group Manager, Regulatory Affairs 2560 General Armistead Ave. Audubon, Pennsylvania 19403
Re: K231333
Trade/Device Name: STRETTO™ Cable System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ, HWC, HRS Dated: May 8, 2023 Received: May 8, 2023
Dear Jennifer Antonacci:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Shumaya Ali -S" in a large, clear font. The text is horizontally oriented and appears to be the primary focus of the image. The background is plain, with a faint design that does not distract from the text.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K231333
Device Name STRETTO™ Cable System
#### Indications for Use (Describe)
The STRETTO™ Cable System is indicated for use in: general orthopedic trauma surgery (e.g., fractures of the olecranon, patella, femur - including periprosthetic, pelvis, acetabulum, humerus and ankle, and acromioclavicular dislocations); prophylactic banding during total joint procedures; and, temporary reduction techniques for ORIF (Open Reduction Internal Fixation) procedures.
STRETTO™ Screw Anchors are indicated for fractures that may not be securely held by either a screw or a cerclage device alone, and where cerclage is used in combination with bone screws and/or plates to provide internal fixation of fractured bone.
STRETTO™ Press-In Anchors are indicated for use with a cerclage cable and plate to augment long bone fracture fixation, particularly when the use of screws would be inhibited, as in the presence of intramedullary implants.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary: STRETTO™ Cable System
Company: Globus Medical Inc. 2560 General Armistead Ave. Audubon, PA 19403 610-930-1800
Contact: Jennifer Antonacci, Ph.D. Group Manager, Requlatory Affairs
- August 3, 2023 Date Prepared:
Device Name: STRETTO™ Cable System
Common Name: Bone fixation cerclage cable
Classification: Per 21 CFR as follows: &88.3010 Bone Fixation Cerclage (primary) §888.3030 Single/multiple component metallic bone fixation appliances and accessories §888.3040 Smooth or threaded metallic bone fixation fastener
> Product Code: JDQ (primary), HWC, HRS Regulatory Class: II
Primary Predicate: Pioneer Laboratories SDB Cerclage System (K992616)
# Additional
| Predicates: |
|-------------------------------------------------------------------------------------------------------|
| Pioneer Laboratories Hex Button Device (K992617)<br>Synthes Variable Angle Positioning Pins (K162124) |
# Purpose:
The purpose of this submission is to request clearance for the STRETTO™ Cable System.
### Device Description:
The STRETTO™ Cable System implants are comprised of cables, crimps and anchors. Anchors include a thru hole for the cables to pass through, and are available in multiple styles. Press-in anchors mate with ANTHEM fracture plates and screw anchors mate with the hex recess of ANTHEM screws. The cable is secured with the crimp and may be used with the anchors. STRETTO™ Implants are manufactured from titanium allov. commercially pure titanium. cobalt chromium alloy, cobalt chromium molybdenum alloy, or stainless steel.
{4}------------------------------------------------
# Indications for Use:
The STRETTO™ Cable System is indicated for use in general orthopedic trauma surgery, including olecranon, patella, femur (including periprosthetic fractures), pelvic, acetabular, humeral and ankle fractures, acromioclavicular dislocations, prophylactic banding during total joint procedures, and temporary reduction techniques for ORIF (Open Reduction Internal Fixation) procedures.
STRETTO™ Screw Anchors are indicated for fractures that may not be securely held by either a screw or a cerclage device alone, and where cerclage is used in combination with bone screws and/or plates to provide internal fixation of fractured bone.
STRETTO™ Press-In Anchors are indicated for use with a cerclage cable to augment fracture stabilization with plates used in long bone fixation, and when the use of screws would be inhibited, as in the presence of intramedullary implants.
# Performance Data:
Mechanical testing (static and dynamic tension) was conducted using ASTM E8 as a guide. Performance data demonstrate substantial equivalence to the predicate devices. An engineering analysis was conducted for the anchors to demonstrate substantial equivalence to the predicate devices. Bacterial endotoxin testing (BET) was conducted in accordance with ANSI/AAMI ST72:2011.
### Technological Characteristics:
The STRETTO™ Cable System has similar technological characteristics as the predicate devices including design, intended use, material composition, function, and range of sizes.
### Basis of Substantial Equivalence:
The STRETTO™ Cable System has been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided supports substantial equivalence of the subject devices to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.