K231256 · Carecubes, Inc. · LGM · Jan 3, 2024 · General Hospital
Device Facts
Record ID
K231256
Device Name
Carecube Negative Pressure Isolation Chamber
Applicant
Carecubes, Inc.
Product Code
LGM · General Hospital
Decision Date
Jan 3, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5450
Device Class
Class 2
Indications for Use
The Carecube Negative Pressure Isolation Chamber is a patient isolation unit (PIU) intended to be use in hospitals as a negative pressure isolation chamber. The Carecube is intended to be used by a single patient for up to 24 hours of continuous use. The Carecube Negative Pressure Isolation Chamber Canopy is intended for multi-patient use with a maximum of ten (10) patients. The Arm Access, Auxiliary Arm Access and Trunk ports of the Carecube Negative Pressure Isolation Chamber are intended for singlepatient use only. The unit is designed to be used with the required personal protective equipment (PPE).
Device Story
Carecube is a modular, negative pressure patient isolation unit (PIU) for hospital use. It encloses a patient to prevent airborne particulate cross-contamination between the patient and healthcare practitioners. The device features a thermoplastic polyurethane (TPU) canopy and a powder-coated steel frame. It includes glove arms, trunk ports, conduit ports, and lean-in windows to allow clinicians to perform low-to-moderate complexity interventions (e.g., blood draws, medication administration, cardiac auscultation) while the patient remains isolated. The system uses a fan and H13 HEPA filter to maintain negative pressure and airflow (35 air changes per hour). It is designed for stationary use in hospitals for up to 24 hours. The output is a contained environment that protects staff from infectious disease particulates. It benefits patients by allowing necessary medical care while maintaining strict isolation, reducing the risk of hospital-acquired infections or environmental contamination.
Clinical Evidence
Prospective, single-center, non-blinded study with 10 healthy volunteers (ages 21-47). Evaluated patient safety and comfort during 3 hours of containment. Endpoints included vital signs (BP, temp, pulse, O2 saturation), anxiety levels (STAID), and environmental monitoring (temp, humidity, CO2, O2). Results showed the device is safe for patient and provider use. Bench testing included biocompatibility (ISO 10993), electrical safety (IEC 60601), and functional performance (HEPA filtration, viral penetration per ASTM F1671).
Technological Characteristics
Modular negative pressure isolation chamber. Materials: Thermoplastic Polyurethane (TPU) envelope, powder-coated steel frame. Filtration: H13 HEPA (99.97% at 0.3μ). Airflow: 35 air changes per hour (ACH). Dimensions: 118"L x 54"W x 82"H. Power: Electrical fan/blower. Connectivity: None. Sterilization: Reusable components validated via cleaning/disinfection protocols (AAMI TIR30/ST98).
Indications for Use
Indicated for temporary isolation of patients in hospital settings to prevent particulate (biological) cross-contamination between user and patient, while enclosing the patient from the external environment. Intended for temporary housing prior to transfer to a permanent hospital setting. Enables low-moderate complexity medical interventions including blood draw, medication administration, abdominal palpation, cardiac auscultation, and connection to IV/monitoring cables. Contraindicated for morbidly obese patients and patients weighing less than 45 lbs.
Regulatory Classification
Identification
A patient care reverse isolation chamber is a device consisting of a roomlike enclosure designed to prevent the entry of harmful airborne material. This device protects a patient who is undergoing treatment for burns or is lacking a normal immunosuppressive defense due to therapy or congenital abnormality. The device includes fans and air filters which maintain an atmosphere of clean air at a pressure greater than the air pressure outside the enclosure.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 3, 2024
Carecubes, Inc. Steve Bollinger Operational Lead 300 Oxford Drive Suite 330 Monroeville, Pennsylvania 15146
Re: K231256
Trade/Device Name: Carecube Negative Pressure Isolation Chamber Regulation Number: 21 CFR 880.5450 Regulation Name: Patient Care Reverse Isolation Chamber Regulatory Class: Class II Product Code: LGM Dated: April 27, 2023 Received: May 1, 2023
Dear Steve Bollinger:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231256
Device Name Carecube Negative Pressure Isolation Chamber
### Indications for Use (Describe)
The Carecube is a patient isolation unit (PIU) designed for the temporary isolation of patients within a hospital setting to prevent particulate (biological) cross-contamination between user and patient, while enclosing the contaminated patient from the external environment. Device should only be used in a hospital setting. This is for temporary housing of a patient prior to transfer to an appropriate hospital destination. Transfer to a more permanent hospital setting should occur as soon as possible. The Carecube is designed with features that enable low-moderate complexity medical interventions. This includes the following procedures: blood draw, medication, palpating abdomen, cardiac auscultation, and connection to IV line/monitoring cables.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY K231256 Carecube Negative Pressure Isolation Chamber
# I. Submitter Information
Carecubes, Inc. 3101 20th Street San Francisco, CA 94110
| Contact Person: | Steve Bollinger |
|-------------------|---------------------|
| Telephone Number: | (508) 942-6396 |
| Email: | steve@carecubes.com |
| Date Prepared: | January 03, 2023 |
### II. Device
| Trade Name: | Carecube Negative Pressure Isolation Chamber |
|-------------------------|----------------------------------------------|
| Common Name: | Patient Isolation Unit (PIU) |
| Classification Name: | Chamber, Patient Isolation |
| Produce Classification: | Class II |
| Regulation Number: | 21 CFR 880.5450 |
| Product Code: | LGM |
{5}------------------------------------------------
### III. Predicate Device:
| Predicate Device Name | 510(k) | Company Name |
|---------------------------------------------------|---------|----------------------|
| CAPSULS™ Patient Isolation Unit | K052798 | ISOVAC Products, LLC |
| Reference Device | | |
| ORCA™ Operational Rescue<br>Containment Apparatus | K163223 | ISOVAC Products, LLC |
#### IV. Description of Device:
The Carecube Negative Pressure Isolation Chamber is a modular, negative pressure, isolation patient care system with both reusable and disposable components that can rapidly augment capacity to care for airborne infectious diseases and other biological threats for common patient care protocols with features built into the unit to allow ease of execution. The device is intended to be used indoors, in a hospital setting, as a temporary isolation of patients with suspected or confirmed diagnosis of infectious disease to prevent Healthcare Practitioners (HCP) exposure to pathogenic biological airborne particulates and is designed with features that enable low-moderate complexity medical interventions.
#### V. Intended Use:
The Carecube Negative Pressure Isolation Chamber is a patient isolation unit (PIU) intended to be use in hospitals as a negative pressure isolation chamber. The Carecube is intended to be used by a single patient for up to 24 hours of continuous use. The Carecube Negative Pressure Isolation Chamber Canopy is intended for multi-patient use with a maximum of ten (10) patients. The Arm Access, Auxiliary Arm Access and Trunk ports of the Carecube Negative Pressure Isolation Chamber are intended for singlepatient use only. The unit is designed to be used with the required personal protective equipment (PPE).
#### VI. Indications for Use:
The Carecube is a patient isolation unit (PIU) designed for the temporary isolation of patients within a hospital setting to prevent particulate (biological) cross-contamination between user and patient, while enclosing the contaminated patient from the external environment. Device should only be used in a hospital setting. This is for temporary housing of a patient prior to transfer to an appropriate hospital destination. Transfer to a more permanent hospital setting should occur as soon as possible. The Carecube is designed with features that enable low-moderate complexity medical interventions. This includes the following procedures: blood draw, medication administration, palpating abdomen, cardiac auscultation, and connection to IV line/monitoring cables.
{6}------------------------------------------------
### VII. Comparison of Technological Characteristics with Predicate Devices
The indications for use and principle of operation of the Carecube Negative Pressure Isolation Chamber are equivalent to those of the predicate device, ISOVAC CAPSULS™ Patient Isolation Unit (K052798) and reference device, ISOVAC ORCA™ Operational Rescue Containment Apparatus (K163223). Differences between the devices result primarily from technological advances which have occurred since the ISOVAC CAPSULS™ device was introduced in 2005.
| PARAMETER | PROPOSED<br>DEVICE | PREDICATE<br>DEVICE | COMPARISON |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Carecube Negative<br>Pressure Isolation<br>Chamber | CAPSULS™ Patient<br>Isolation Unit | |
| Manufacturer | Carecubes, Inc. | ISOVAC | |
| 510(k) | K231256 | K052798 | |
| Product<br>Classification | II | II | Same |
| Product Code | LGM | LGN | Equivalent Carecube<br>(LGM) is not used<br>for transport (LGN) |
| Regulation<br>Number | 880.5450 | 880.5450 | Same |
| Regulation<br>Name | Chamber, Patient<br>Isolation | Chamber, Patient<br>Isolation | Same |
| Intended Use | Negative Pressure<br>Isolation Chamber | Negative Pressure<br>Isolation Chamber | Same |
| Indications for<br>Use | The Carecube is a patient<br>isolation unit (PIU)<br>designed for the<br>temporary isolation of<br>patients within a hospital<br>setting to prevent<br>particulate (biological)<br>cross-contamination<br>between user and patient,<br>while enclosing the<br>contaminated patient from<br>the external environment.<br>Device should only be<br>used in a hospital setting.<br>This is for temporary<br>housing of a patient prior<br>to transfer to an<br>appropriate hospital<br>destination. Transfer to a<br>more permanent hospital<br>setting should occur as<br>soon as possible. The<br>Carecube is designed with<br>features that enable low-<br>moderate complexity<br>medical interventions.<br>This includes the<br>following procedures:<br>blood draw, medication<br>administration, palpating<br>abdomen, cardiac<br>auscultation, and<br>connection to IV<br>line/monitoring cables. | The CAPSULS<br>(Containment and<br>Protection System Utilizing<br>Life Support) is a portable<br>Patient Isolation Unit<br>(PIU), which prevents<br>particulate (biological and<br>radiological) cross-<br>contamination between the<br>patient and the external<br>environment; and with<br>features that enable medical<br>intervention to the patient<br>via end-user supplied<br>medical equipment. The<br>CAPSULS is intended to<br>be used for:<br>• The transport and<br>isolation of<br>patients on<br>aircraft,<br>ambulances, ships<br>and any vehicle<br>capable of safely<br>transporting a<br>patient on a<br>standard litter.<br>• The temporary<br>isolation, with or<br>without transport,<br>of patients within<br>hospitals or<br>medical facilities. | Same except the<br>Carecube<br>indications are a<br>subset of the<br>predicate device<br>because the former<br>is not used for<br>transport. |
| Provided<br>Sterile | No | No | Same |
| Reusable | Yes | No | Different<br>Reprocessing<br>validation was<br>performed to ensure<br>safe reuse |
| Contra-<br>indications | • Morbidly<br>obese patients<br>• Patients<br>weighing less<br>than 45 lbs. or<br>less<br>• Oxygen-rich | Unknown | |
| OTC Use | Yes | Yes | Same |
| Materials of<br>Construction | Thermoplastic Polyurethane (TPU) envelope Powder-coated steel frame | Clear polyurethane (PUR) or polyvinylchlori de (PVC) envelope Plastic resin stanchions and integral ribs | Different<br>Materials used in both devices represent plastic materials commonly utilized in medical devices and do not raise new issues of safety or effectiveness |
| Dimensions | 118" overall assembled length<br>54" width<br>82" height | 81" overall length<br>26" width<br>20" | Different<br>The Carecube is longer to accommodate stationary use and a medical bed for patient isolation up to 24 hours.<br>Clinical testing demonstrates no new questions of safety or effectiveness. |
| Weight | 120 lbs. | 21 lbs. | Different<br>Carecube is heavier based on design for stationary use and does not raise new questions of safety or effectiveness. |
| Air Filtration | H13 HEPA (99.97% particulates @ 0.3μ) | M95, P100 (99.97% particulates @ 0.3μ) HEPA | Different<br>Both devices have an inline outbound filter, with differences in model type and size.<br>Performance testing demonstrates that, despite the differences, the Carecube is effective in air filtration. |
| Air Flow | Airflow for patient<br>head to foot is<br>controlled by the fan<br>and HEPA filter<br>blower manufacturer<br>at 150 CFM. | Airflow rate is set by<br>the blower<br>manufacturer at 4<br>CFM. | Different<br>Air flow capabilities<br>are larger due to the<br>larger room and<br>containment area.<br>Safety and<br>effectiveness is<br>demonstrated in<br>patient comfort and<br>in performance<br>testing. |
| Pressure | Negative Pressure<br>(NP) | Negative Pressure<br>(NP)<br>Positive Pressure (PP) | Different<br>The CAPSULS™<br>PIU allows the use<br>of PP or NP which<br>is not necessary for<br>the Carecube's<br>intended use. Both<br>devices utilize NP to<br>isolate a patient and<br>prevent the patient<br>from infecting<br>others. The<br>Carecube negative<br>pressure<br>performance was<br>validated. |
| Air Changes | 35 air changes per<br>hour (ACH) | 20-38 air changes per<br>hour (ACH) | Different<br>(The Carecube ACH<br>is within the ACH<br>range for the<br>predicate device.) |
| Glove Arms | Multiple Glove Arms | Multiple Glove Arms | Same |
| Other Access<br>Ports | Trunk Ports<br>Conduit Ports<br>Lean-In Windows | Access Ports<br>Conduit Port | Different<br>The Carecube offers<br>additional access<br>ports to facilitate<br>stationary use for<br>patient isolation.<br>The additional ports<br>do not raise different<br>questions of safety or<br>effectiveness and<br>performance was<br>demonstrated. |
### Table 5-1. General Technological Characteristics Comparison
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
!The Carecube Negative Pressure Isolation Chamber is to be used by a single patient for up to 24 hours of
{10}------------------------------------------------
continuous use. The Carecube Negative Pressure Isolation Chamber canopy is intended for multi-patient re-use with a labeled maximum of ten (10) uses. The glove and trunk ports of the Carecube Negative Pressure Isolation Chamber are intended for single patient use only.
#### VIII. Performance Data
The following performance data was considered in support of the substantial equivalence determination.
### Performance Testing - Bench
The following tests were performed to demonstrate that the proposed Carecube Negative Pressure Isolation Chamber met the applicable design and performance requirements and support a determination of substantial equivalence. Where applicable, testing was done per applicable ISO and other international standards:
- Biocompatibility ●
- EMC and Electrical Safety
- Usability Testing
- Packaging Testing
- Shelf-Life Testing
- Canopy Robustness
- Functional Performance Testing ●
{11}------------------------------------------------
| Performance | Standard/Test<br>Method | Proposed Device<br>Result | Predicate Result |
|---------------------------------------------------|-----------------------------------------------|---------------------------|--------------------------|
| Biocompatibility | | | |
| Cytotoxicity | ISO 10993-5 | Non-cytotoxic | Not available |
| Sensitization | ISO 10993-10 | Non-sensitizing | Not available |
| Intracutaneous<br>Reactivity | ISO 10993-23 | Non-irritant | Not available |
| Acute Systemic<br>Toxicity | ISO 10993-11 | Non-toxic | Not available |
| Volatile<br>Organic<br>Compounds;<br>Particulates | Toxicological Risk<br>Assessment | Non-toxic | Not available |
| EMC and Electrical<br>Safety | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6 | Pass | Not available |
| Usability /Human<br>Factors | IEC 62366-1 | Pass | Not available |
| Packaging Testing | ASTM D4169<br>ASTM D4332 | Pass | Pressure/Leak<br>testing |
| Shelf-life Testing | ASTM F1980 | 12 months | Not available |
| Functional Performance Testing | | | |
| Canopy Robustness | Internal | Pass | Not available |
| Airflow and Smoke<br>Testing | Internal | Pass | Not available |
| Differential Pressure | Internal | Pass | Not available |
| Tagged particle<br>Clearance | Internal | Pass | Not available |
| Tagged particle<br>Containment | Internal | Pass | Not available |
| Noise | Internal | Pass | Not available |
| Viral Penetration | ASTM F1671 | Pass | Not available |
| Cleaning Resiliency | Internal | Pass | Not available |
| Cleaning Validation | AAMI TIR30<br>AAMI ST98 | Pass | Not available |
| Disinfection Validation | AAMI TIR30<br>AAMI ST98 | Pass | Not available |
| Flammability | 16 CFR Part<br>1610.7 | Pass | Not available |
| Repeat Use Testing | Internal | Pass | Not available |
| Usage Life Testing | | | |
| System Usage Life | Internal | Pass | Not available |
| Filter usage Life | Internal | Pass | Not available |
# Table 5-2. Summary of Performance Testing – Bench and Biocompatibility
Carecubes, Inc.
{12}------------------------------------------------
{13}------------------------------------------------
### Biocompatibility
Biocompatibility testing was conducted in accordance with the FDA Guidance Document "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process," September 4, 2020, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
Based on the results of the biocompatibility testing performed on the Carecube Negative Pressure Isolation Chamber, the proposed device meets the requirements outlined in ISO 10993-1:2018.
### Electrical Safety and Electromagnetic Compatibility (EMC)
EMC and Electrical Safety evaluation was conducted on the proposed Carecube Negative Pressure Isolation Chamber. Based on the results of the EMC testing, the Carecube Negative Pressure Isolation Chamber meets the requirements of IEC 60601-1.
### Software Verification and Validation Testing
The proposed Carecube Negative Pressure Isolation Chamber does not contain software; therefore, the proposed device does not require software verification and validation testing.
### Performance Testing - Animal
This submission does not include any animal performance testing. It was determined that no such testing was required to demonstrate substantial equivalence.
### Performance Testing - Clinical
A non-significant risk (NSR) clinical study was conducted to evaluate patient comfort and environmental safety of the Carecube Negative Pressure Isolation Chamber. This study was intended to evaluate general patient safety and comfort within the Carecube (not associated with a specific disease state or condition), therefore allowing for the use of normal, healthy volunteers to achieve the study endpoints. This study was conducted as a prospective, single center, non-blinded study. The purpose of this study was to ensure patient safety and comfort during containment within the Carecube. Patient safety and comfort was assessed through routine monitoring of subject's vital signs (blood pressure, body temperature, pulse and oxvgen saturation), assessment of level of anxiety using the STAID evaluation, as well as monitoring of the Carecube environment (temperature, relative humidity, carbon dioxide (CO2) and atmospheric oxygen (O2) on an hourly basis. Subject data was recorded at baseline + 5 minutes and hourly, for a minimum of three (3) consecutive hours. This study consisted of a minimum of ten (10) Subjects and came from a cross-section of males and females, ages 21-47 years old.
| Table 5-3. Summary of Clinical Testing | | |
|----------------------------------------|--|--|
|----------------------------------------|--|--|
| Performance | Standard/Test<br>Method | Result | Comparison to<br>Predicate |
|----------------------|-------------------------|--------|----------------------------|
| Human Use<br>Comfort | Internal | Pass | Not available |
{14}------------------------------------------------
Based on the results of the study, it was determined that the Carecube Negative Pressure Isolation Chamber is safe for healthcare provider and patient use.
#### IX. CONCLUSION
The conclusions drawn from the nonclinical and clinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified as the CAPSUL Patient Isolation Unit, cleared under the 510(k) submission number K052798.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.