K231255 · Novo Nordisk, Inc. · FMI · Aug 25, 2023 · General Hospital
Device Facts
Record ID
K231255
Device Name
NovoFine® Plus
Applicant
Novo Nordisk, Inc.
Product Code
FMI · General Hospital
Decision Date
Aug 25, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NovoFine® Plus needles are intended for use with pen injector devices for the subcutaneous injection of drugs.
Device Story
NovoFine® Plus is a sterile, single-use hypodermic needle designed for attachment to pen injector devices. Operation involves removing the protective tab, screwing the needle onto the injector, removing caps, and performing the injection. After use, the needle is detached and disposed of. The device facilitates subcutaneous drug delivery in clinical or home settings by patients or healthcare professionals. It serves as a mechanical interface between the drug-filled pen injector and the patient's subcutaneous tissue. The 29G needle variant expands the existing product family, maintaining the same principle of operation as the predicate. It provides a consistent, sterile pathway for medication administration, benefiting patients by enabling reliable, self-administered subcutaneous injections.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by design verification testing, including material and dimensional requirements per ISO 9626 and performance testing per ISO 11608-2, alongside biocompatibility evaluation per ISO 10993-1:2018.
Technological Characteristics
Sterile, single-use, stainless steel hypodermic needle. Complies with ISO 9626 (stainless steel needle tubing) and ISO 11608-2 (needle-based injection systems). Manual operation; no energy source or software components.
Indications for Use
Indicated for patients requiring subcutaneous drug delivery via pen injector devices. No specific age, gender, or disease state restrictions provided.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 25, 2023
Novo Nordisk Inc. Hiral Shah Manager, Regulatory Affairs P.O Box 846 Plainsboro, New Jersey 08536
Re: K231255
Trade/Device Name: NovoFine® Plus Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: August 9, 2023 Received: August 11, 2023
Dear Hiral Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan Sthe
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery, General Hospital, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231255
Device Name NovoFine® Plus
Indications for Use (Describe)
Indications for Use: NovoFine® Plus needles are intended for use with pen injector devices for the subcutaneous injection of drugs.
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5 510(k) Summary
## 21 CFR 807.87(h)
As required by 21 CFR 807.92(a)
(1) Date Prepared: August 25, 2023
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The subject of this Special 510(k) submission is the NovoFine® Plus with gauge 29G, hereafter named 'NovoFine® Plus (subject needle)'. The NovoFine® Plus (subject needle) meets all applicable product and quality standards for hypodermic single lumen needle products. NovoFine® Plus product family will be expanded to include NovoFine® Plus (subject needle).
#### Submitter's Name and Address
Novo Nordisk Inc. 800 Scudders Mill Road Plainsboro, New Jersey 08536
Contact Person: Hiral Palkhiwala Shah Senior Manager, Regulatory Affairs Tel: 609-787-7603 Email: hpaw@novonrdisk.com
### (2) Name of Device
Proprietary Name: Common or usual name: Classification:
Product Code:
NovoFine® Plus Pen Needle Class II device; 21 CFR 880.5570, (hypodermic single lumen needle) FMI (hypodermic single lumen needle)
### (3) Substantial Equivalence
Predicate Device: NovoFine® Plus 32G 4 mm ETW cleared under K202005 on December 19, 2020.
{4}------------------------------------------------
### (4) Device Description
NovoFine® Plus is a sterile single use needle for subcutaneous injection of drugs with a pen injector device.
#### Mode of operation
Prior to giving an injection, the protective tab is removed from the outer needle cap of the single-use disposable needle. With the disposable needle remaining in the outer needle cap, it is then carefully screwed onto the injection delivery device until tight and then the needle outer and inner caps are removed. Use the needles as described in the instructions for use that comes with the pen-injector device and as instructed by the healthcare professional.
After the injection, the needle is removed from the skin. The needle is detached from the injection device and disposed of in accordance with national/local regulations. For each subsequent injection, a new disposable needle must be used.
## (5) Intended Use
Intended for use with pen injector devices for the subcutaneous injection of drugs.
### Comparison with Predicate Devices
The NovoFine® Plus (subject needle) has the same intended use, indications for use, intended users, basic design, principle of operation, component materials, biocompatibility, shelf life and application of design controls for its development as the NovoFine® Plus cleared under K202005.
Specifically for biocompatibility, an evaluation according to ISO 10993-1:2018 'Biological evaluation of medical devices Part 1 : Evaluation and testing within a risk management process' has been made, which is valid for both the NovoFine® Plus (subject needle) and the NovoFine® Plus cleared under K202005. The evaluation, performed considering all parts of the device having direct or indirect contact with the patient, concludes that NovoFine® Plus does not pose a risk of cytotoxicity, skin irritation, skin sensitisation, material-mediated pyrogenicity, acute or subacute toxicity, genetic toxicity, implantation reactions or any other biological hazard as defined in ISO 10993-1:2018 when used as intended.
In addition, the manufacturing, quality control and clinical evaluation process for the NovoFine® Plus (subject needle) are the same as for the NovoFine® Plus cleared under K202005.
The purpose of this submission is to introduce the 29G gauge needle into the NovoFine® Plus product family, where the predicate device has a 32G gauge needle (see Table 1).
{5}------------------------------------------------
Table 1 Comparison of the characteristics of the NovoFine® Plus (subject needle) and the NovoFine® Plus cleared under K202005 (predicate needle)
| Feature | NovoFine® Plus (subject needle)* | NovoFine® Plus cleared under<br>K202005 (predicate needle) |
|-------------------------|----------------------------------|-------------------------------------------------------------------------------------------|
| 510(k)Number | K231255 | K202005 |
| Intended Use | Same as predicate | Intended for use with pen injector<br>devices for the subcutaneous injection<br>of drugs. |
| Needle Gauge Size (s) | 29G | 32 G |
| Needle Length Size (s) | Same as predicate | 4 mm |
| Needle insertion method | Same as predicate | Manual |
| Provided Sterile | Same as predicate | Sterile - steam |
* The NovoFine® Plus product family will include the 32G and 29G gauge needle. Only the 29G gauge needle information is presented in this column
### (6) Testing
The subject device NovoFine® Plus (subject needle) has the substantially equivalent technological characteristics as the predicate device NovoFine® Plus cleared under K202005.
Novo Nordisk has verified the design of the subject device as part of its design control process in accordance with the Quality System Regulation. This testing included confirmation of compliance to material and dimension requirements, as per ISO 9626 Stainless steel needle tubing for the manufacture of medical devices, and to performance, as per ISO 11608-2: Needle-based injection systems for medical use – Requirements and test methods– Part 2: Needles.
The 29G gauge needle does not introduce critical differences or new risks to the intended use of the device.
### (7) Conclusion
Based on the design equivalency and the functional testing. Novo Nordisk has determined that the NovoFine® Plus (subject needle) is substantially equivalent to the NovoFine® Plus cleared under K202005, which is currently marketed in the United States. Differences between the devices do not raise any significant issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.