Nexiva™ Closed IV Catheter System with NearPort™ IV Access
K231239 · Becton Dickinson Infusion Therapy Systems, Inc. · FOZ · Sep 22, 2023 · General Hospital
Device Facts
Record ID
K231239
Device Name
Nexiva™ Closed IV Catheter System with NearPort™ IV Access
Applicant
Becton Dickinson Infusion Therapy Systems, Inc.
Product Code
FOZ · General Hospital
Decision Date
Sep 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nexiva™ Closed IV Catheter System with NearPort™ IV Access is intended to be inserted into a patient's peripheral venous system for short term use to administer fluids and/or medications and to sample blood. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 24-18 GA (0.7-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when connected to the NearPort™ IV Access.
Device Story
Over-the-needle peripheral IV catheter system; includes radiopaque Vialon™ tubing, needle, needle shield, septum, stabilization platform, extension tubing, clamp, Luer adapter, vent plug, and MaxZero™ needle-free connector. Designed as a closed system to contain blood during insertion; septum wipes needle upon withdrawal to reduce exposure risk; passive needle tip protection prevents needlestick injuries. Features Instaflash™ technology for immediate blood return visualization. NearPort™ IV Access allows compatibility with needle-free blood collection devices (e.g., PIVO™ Pro) and supports power injection. Used in clinical settings by healthcare providers for fluid/medication administration and blood sampling. Output is venous access; enables therapeutic delivery and diagnostic sampling. Benefits include reduced blood exposure, improved safety, and expanded compatibility with blood collection and power injection systems.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated through bench testing (ISO 10555-1/5, ISO 23908, ISO 594-1/2), biocompatibility (ISO 10993-1/7), packaging (ASTM D4169), sterilization validation (ISO 11135-1), MR compatibility (ASTM F2182-19), microbial ingress evaluation, and human factors usability testing.
Technological Characteristics
Closed peripheral IV catheter system. Materials: Polyurethane (tubing), Stainless Steel (needle), Polycarbonate (grip/shield/connector), Copolyester (adapter), Thermoplastic Elastomer (wings), Acetal Copolymer (clamp), Silicone (septum). Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: None. Energy: User-operated. Power injection: Up to 300 psi (24-18 GA).
Indications for Use
Indicated for peripheral venous access in any patient population for short-term administration of fluids/medications and blood sampling. 24-18 GA devices are compatible with power injectors up to 300 psi when using the NearPort™ IV Access.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 22, 2023
Becton Dickinson Infusion Therapy Systems Inc. Paul Holman Regulatory Affairs Manager 9450 South State Street Sandy, Utah 84070
Re: K231239
Trade/Device Name: Nexiva™ Closed IV Catheter System with NearPort™ IV Access Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ, FPA Dated: August 25, 2023 Received: August 25, 2023
Dear Paul Holman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231239
#### Device Name
Nexiva™ Closed IV Catheter System with NearPort™ IV Access
### Indications for Use (Describe)
The Nexiva™ Closed IV Catheter System with NearPort™ IV Access is intended to be inserted into a patient's peripheral venous system for short term use to administer fluids and/or medications and to sample blood. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 24-18 GA (0.7-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when connected to the NearPort™ IV Access.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
bd.com
Image /page/3/Picture/2 description: The image shows the BD logo. The logo consists of two parts: a stylized orange sun-like symbol on the left and the letters "BD" in blue on the right. The sun-like symbol is a circle with rays emanating from it, and the letters "BD" are in a bold, sans-serif font.
# K231239 510(k) Summary (21 CFR §807.92) Nexiva™ Closed IV Catheter System with NearPort™ IV Access
| Submitter | Submitter Name: | Becton Dickinson Infusion Therapy Systems Inc. | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|--------------|
| Information | Submitter Address: | 9450 South State Street, Sandy, Utah 84070 | |
| | Contact Person: | Paul Holman, Regulatory Affairs Manager | |
| | Email Address: | paul.holman@bd.com | |
| | Phone Number: | (801) 522-5000 | |
| | Date of Preparation: | September 22, 2023 | |
| Subject Device | Trade Name: | Nexiva™ Closed IV Catheter System with NearPort™ IV Access | |
| | Common Name: | Peripheral Intravascular or IV Catheter | |
| | 510(k) Reference: | K231239 | |
| | Regulation Number: | 21 CFR §880.5200 | |
| | Regulation Name: | Catheter, intravascular, therapeutic, short-term less than 30 days | |
| | Regulatory Class: | II | |
| | Product Code: | FOZ, FPA | |
| | Classification Panel: | General Hospital | |
| Predicate | Trade Name: | BD Nexiva™ Closed IV Catheter System | |
| Device | Common Name: | Peripheral Intravascular or IV Catheter | |
| | 510(k) Reference: | K183399 | |
| | Regulation Number: | 21 CFR §880.5200 | |
| | Regulation Name: | Catheter, intravascular, therapeutic, short-term less than 30 days | |
| | Regulatory Class: | II | |
| | Product Code: | FOZ, FPA | |
| | Classification Panel: | General Hospital | |
| Reason for | | The purpose of this submission is to notify the FDA of the introduction of the Nexiva™ | |
| Submission | Closed IV Catheter System with NearPort™ IV Access, allowing for compatibility with a<br>needle-free blood draw device (e.g., PIVO™ Pro Needle-free Blood Collection Device). | | |
| Device | The Nexiva™ Closed IV Catheter System with NearPort™ IV Access is an over-the-<br>needle, intravenous (IV) catheter. These devices have a radiopaque Vialon™ catheter<br>tubing, needle, needle shield, septum, stabilization platform, long extension tubing, clamp,<br>Luer adapter, and vent plug. The needle and catheter tubing are protected by a needle<br>cover. A MaxZero™ needle-free connector with protective cover is provided in the<br>package. | | |
| bd.com | | | |
| The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury. | | | |
| These devices have Instaflash™ Needle Technology, allowing for immediate visualization of blood return along the catheter. Continuous blood return is seen in the long extension tubing. The vent plug and the NearPort™ IV Access prevent blood leakage from the long extension tubing during insertion. | | | |
| Both the long extension tubing and NearPort™ IV Access are suitable to administer fluids and/or medications and to sample blood. | | | |
| In addition, the NearPort™ IV Access is compatible with the PIVO™ Pro Needle-free Blood Collection Device for needle-free blood draws and is suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa). | | | |
| These devices are available with wings. The catheter hub and wings are color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green). | | | |
| Indications for Use<br>(21 CFR § 807.92(a)(5)) | The Nexiva™ Closed IV Catheter System with NearPort™ IV Access is intended to be inserted into a patient's peripheral venous system for short term use to administer fluids and/or medications and to sample blood. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 24-18 GA (0.7-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when connected to the NearPort™ IV Access. | | |
| Technological<br>Characteristics | Technological characteristics of the subject device are substantially equivalent to the predicate device. The subject device, Nexiva™ Closed IV Catheter System with NearPort™ IV Access, achieves its intended use based on the same technology and principles of operation as the predicate device. A comparison of the subject and predicate device technological characteristics is provided in the table below | | |
| | SUBJECT (K231239) | PREDICATE (K183399) | Comparison & |
{4}------------------------------------------------
| Attribute | SUBJECT (K231239)<br>Nexiva™ Closed IV Catheter<br>System with NearPort™ IV<br>Access | PREDICATE (K183399)<br>BD Nexiva™ Closed IV Catheter<br>System | Comparison &<br>Discussion |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 21 CFR 880.5200<br>Class II | 21 CFR 880.5200<br>Class II | |
| Classification | FOZ - Intravascular Catheter<br>FPA - Intravascular Administration<br>Set | FOZ - Intravascular Catheter<br>FPA - Intravascular Administration<br>Set | Same |
| Attribute | SUBJECT (K231239)<br>Nexiva™ Closed IV Catheter<br>System with NearPort™ IV<br>Access | PREDICATE (K183399)<br>BD Nexiva™ Closed IV Catheter<br>System | Comparison &<br>Discussion |
| Indications for<br>Use | The Nexiva™ Closed IV Catheter<br>System with NearPort™ IV Access is<br>intended to be inserted into a<br>patient's peripheral venous system<br>for short term use to administer<br>fluids and/or medications and to<br>sample blood. These devices may be<br>used for any patient population with<br>consideration given to adequacy of<br>vascular anatomy, procedure being<br>performed, fluids being infused, and<br>duration of therapy. The 24-18 GA<br>(0.7-1.3 mm) devices are suitable<br>for use with power injectors set to a<br>maximum pressure of 300 psi (2068<br>kPa) when connected to the<br>NearPort™ IV Access. | Nexiva™ closed IV catheter<br>systems are intended to be inserted<br>into a patient's vascular system for<br>short term use to sample blood,<br>monitor blood pressure, or<br>administer fluids. These catheters<br>may be used for any patient<br>population with consideration<br>given to adequacy of vascular<br>anatomy, procedure being<br>performed, fluids being infused,<br>and duration of therapy. The 22-18<br>GA (0.9-1.3 mm) devices are<br>suitable for use with power<br>injectors set to a maximum<br>pressure of 300 psi (2068 kPa)<br>when access ports not suitable for<br>use with power injectors are<br>removed. | Modifications have<br>been made to re-<br>phrase and align with<br>product use<br>specifications.<br>The Indications for<br>Use is being<br>narrowed to limit use<br>to the peripheral<br>venous system only<br>to administer fluids<br>and/or medications<br>and to sample blood.<br>Narrowing the<br>Indications for Use<br>does not raise any<br>new or different<br>questions of safety or<br>effectiveness.<br>Additionally, power<br>injection capability<br>has been expanded to<br>include the 24 GA<br>configuration, using<br>the same test method<br>as the predicate<br>device as supported<br>by performance<br>verification testing,<br>which did not raise<br>any new or different<br>questions of safety<br>and effectiveness. |
| Intended Use | Intravascular access | Intravascular access | Same |
| Fundamental<br>Scientific<br>Technology | Closed peripheral intravenous<br>catheter system designed with an<br>integrated extension tubing<br>incorporating a Y (dual)-port<br>injection site. Incorporates<br>Instaflash™ technology to assist<br>with flashback visualization.<br>Nexiva™ Closed IV Catheter<br>System with NearPort™ IV Access<br>is compatible with the PIVOTM Pro<br>Needle-free Blood Collection | Closed peripheral intravascular<br>catheter system designed with an<br>integrated extension set<br>incorporating a single port or Y<br>(dual)-port injection site.<br>Incorporates Instaflash™<br>technology to assist with flashback<br>visualization | The subject device<br>uses the same<br>fundamental<br>technology compared<br>to the predicate device<br>(K183399).<br>The NearPort™ IV<br>Access component has<br>the same fundamental<br>scientific technology |
| Attribute | SUBJECT (K231239)<br>Nexiva™ Closed IV Catheter<br>System with NearPort™ IV<br>Access | PREDICATE (K183399)<br>BD Nexiva™ Closed IV Catheter<br>System | Comparison &<br>Discussion |
| | Device for needle-free blood draws. | | however, the split<br>septum design enables<br>compatibility with a<br>needle-free blood<br>draw device as<br>supported by human<br>factors validation<br>testing, which did not<br>raise any new or<br>different questions of<br>safety and<br>effectiveness. |
| | Primary Components Material Composition | | |
| ISO 10993-1<br>Biocompatibility<br>Contact Type<br>and Duration | Body Contact: Externally<br>communicating device<br>Contact: Circulating blood<br>Contact Duration: Limited (A)<br>to Prolonged (B) (≤ 24 hrs to<br>30 days) | Body Contact: Externally<br>communicating device<br>Contact: Circulating blood<br>Contact Duration: Limited (A) to<br>Prolonged (B) (≤ 24 hrs to 30<br>days) | Same |
| Needle | Stainless Steel | Stainless Steel | Same |
| Catheter Tubing | Polyurethane with radiopaque<br>barium sulfate | Polyurethane with radiopaque<br>barium sulfate | Same |
| Grip | Polycarbonate + White Colorant | Polycarbonate + White Colorant | Same |
| Tip Shield | Polycarbonate + Gray Colorant | Polycarbonate + Gray Colorant | Same |
| Catheter<br>Adapter | Copolyester | Copolyester | The material<br>formulation has been<br>modified; however,<br>the material resin type<br>remains the same.<br>Testing in accordance<br>with ISO 10993-1<br>requirements did not<br>raise any new or<br>different questions<br>related to biological<br>safety |
| Catheter<br>Adapter Wings | Thermoplastic Elastomer + Gauge<br>Specific Colorant<br>Green (18 GA)<br>Pink (20 GA)<br>Blue (22 GA)<br>Yellow (24 GA)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.