K231230 · Optina Diagnostics, Inc. · HKI · Dec 5, 2023 · Ophthalmic
Device Facts
Record ID
K231230
Device Name
Optina-4C (MHRC-C1N)
Applicant
Optina Diagnostics, Inc.
Product Code
HKI · Ophthalmic
Decision Date
Dec 5, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
Optina-4C™ is intended to capture images of the retina at multiple wavelengths (colors) under mydriatic conditions.
Device Story
Optina-4C (MHRC-C1N) is a non-contact mydriatic fundus camera used by optometrists and ophthalmologists in clinical settings. It captures a series of 92 monochromatic retinal images across a spectral range of 450 nm to 905 nm (5 nm steps) using a Tunable Laser Source (TLS). The device features a 31.5° field-of-view, a focus wheel for refractive error correction (-15 to +15 diopters), and an external fixation target. Patients are positioned via chin and forehead rests. The system aligns using 700 nm light before sequential image acquisition (10 ms per frame). Images are displayed on a monitor and saved for clinical consultation. The device aids practitioners in visualizing retinal structures at specific wavelengths to support diagnostic decision-making.
Clinical Evidence
Bench testing only. Compliance verified with ANSI AAMI ES60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 15004-1 (ophthalmic instruments). Spectral accuracy and reliability were validated using an eye model and reference materials. Biocompatibility of patient-contacting components (chin/forehead rests) was assessed per ISO 10993-1. No clinical data provided.
Technological Characteristics
Mydriatic fundus camera; Tunable Laser Source (TLS) for monochromatic illumination; 31.5° field-of-view; -15 to +15 diopter focus range. Complies with ISO 15004-1, ISO 10940:2009, and ANSI Z80.36:2021. Connectivity: Network enabled. Software: Embedded Off-The-Shelf and Software-of-Unknown-Provenance (SOUP).
Indications for Use
Indicated for capturing retinal images at multiple wavelengths under mydriatic conditions in patients requiring ophthalmic examination by eye care practitioners.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Optina Diagnostics Inc. Sarah Lemaire QA/RA Manager 8200 Decarie Boulevard Montreal, QC H4P 2P5 Canada
Re: K231230
Trade/Device Name: Optina-4C (MHRC-C1N) Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: November 6, 2023 Received: November 7, 2023
Dear Sarah Lemaire:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Elvin Y. Ng-S
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231230
Device Name Optina-4C (MHRC-C1N)
Indications for Use (Describe)
Optina-4C™ is intended to capture images of the retina at multiple wavelengths (colors) under mydriatic conditions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K231230
Image /page/3/Picture/1 description: The image contains the logo for Optina Diagnostics. The logo consists of a blue circular graphic on the left and the company name on the right. The word "Optina" is in gray, and the word "DIAGNOSTICS" is in a smaller, teal font below it.
Special 510(k): Device Modification
Optina-4C™ Model: MHRC-C1N
## 510(k) Summary
| APPLICANT: | Optina Diagnostics, Inc | | |
|-----------------------------------------|----------------------------------------|--|--|
| | 8200 Decarie Blvd, Suite 220 | | |
| | Montreal (QC) | | |
| | H4P 2P5 | | |
| | Phone: 514-394-0797 | | |
| ESTABLISHMENT<br>REGISTRATION<br>NUMBER | Registration Number: 3014606890 | | |
| | Owner Operator Number: 10081796 | | |
| DATE PREPARED: | August 5, 2022 | | |
| CONTACT PERSON: | Sarah Lemaire | | |
| | QA/RA Manager | | |
| | Email: slemaire@optinadx.com | | |
| | Phone: 579-421-0126 | | |
| TRADE NAME: | Optina-4C™ | | |
| MODEL: | MHRC-C1N | | |
| COMMON NAME: | Fundus camera | | |
| CLASSIFICATION<br>NAME: | Camera, Ophthalmic, Ac-Powered (21 CFR | | |
| | 886.1120, Product Code HKI) | | |
| DEVICE<br>CLASSIFICATION: | Class II | | |
| PREDICATE DEVICES | MHRC-C1 (K200254) | | |
### 1. Description of the Modified Device
Optina-4C™ (model: MHRC-C1N) is a mydriatic fundus camera (also called retinal camera) that presents eye care practitioners (optometrists and ophthalmologists) with a series of the retina obtained sequentially at specific wavelengths (colors) in the spectral range 905 nm to 450 nm in steps of 5 nm.
Pictures of the retina are obtained on a field-of-view of 31.5° without contact with the eye. The patient is positioned in front of the device with the chin on the chinrest and forehead on the forehead rest and a
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Optina Diagnostics. On the left side of the logo is a blue circular design that is made up of several overlapping circles. To the right of the circular design is the company name, "Optina", in gray text, with the word "DIAGNOSTICS" in smaller, teal text underneath.
positioning system is used to align the camera relative to the patient's eye. An external fixation target is available to guide the patient's eye. A focus wheel allows for the accommodation for eye refractive error in the range of -15 to +15 diopters. The images are displayed on a monitor and can be saved on the computer for future consultation.
The illumination light of Optina-4C™ is provided by a Tunable Laser Source (TLS). The TLS selects a narrow band of light from a broadband white illumination source. Only a single monochromatic band can output the TLS at a time. The alignment of the retinal camera relative to the patient's eye phase is performed with the illumination light set at a wavelength of 700 nm. The image acquisition phase consists of the sequential acquisition of a series of 92 monochromatic images at wavelengths from 905 nm in steps of 5 nm. Each frame is captured with an exposure time of 10 ms, resulting in a total acquisition time of 920 ms.
The 92 images can be visualized one by one in Optina-4C™ acquisition software. The retinal images captured by Optina-4C™ are monochromatic images having a spectral bandwidth of "10 nm centered on the wavelength indicated in the upper left corner of the visualization pane, with a spectral accuracy of 7.5 nm.
## 2. Indications for Use
Optina-4C™ is intended to capture images of the retina at multiple wavelengths (colors) under mydriatic conditions.
### 3. Technological Characteristics
Optina-4C™ is similar in fit, form, and function to the predicate mydriatic fundus cameras. The main elements of the hardware including contact areas, the illumination of the retina, and the monochromatic images obtained with Optina-4C™ are the same as for the predicate device.
In comparison with the predicate device, Optina-4C™ has updated software and labeling and features network connectivity. Performance testing was performed to verify that this technological difference raised no new concerns regarding the safety and efficacy of the device.
## 4. Performance Testing
The following performance data were conducted to support the substantial equivalence determination.
#### ● Software evaluation
Optina-4C™ uses embedded Off-The-Shelf software and Software-of-Unknown-Provenance. This software was evaluated for use in accordance with the FDA's Guidance for Industry "Off-The-Shelf Software Use in Medical Devices" (September 2019) based on a "Moderate Level of Concern" and
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Optina Diagnostics. On the left side of the logo is a blue circular graphic that is made up of many overlapping circles. To the right of the graphic is the company name "Optina" in gray, with the word "DIAGNOSTICS" in a smaller font size and blue color underneath.
IEC 62304:2006 + A1:2015 based on a "Safety class B".
#### ● Electrical safety and electromagnetic compatibility
lt was verified that the Optina-4C™ complies with the standard ANSI AAMI ES60601-1:2005/(R)2012+A1:2012 for electrical safety and with the standard IEC 60601-1-2:2014 for electromagnetic compatibility.
#### ● Evaluation of recognized consensus standards for ophthalmic cameras
Optina-4C™ was found to comply with the recognized consensus standard ISO 15004-1:2020 specifying general requirements for ophthalmic instruments.
The following performance data were conducted on the marketed device (MHRC-C1). No changes that can affect the evaluation or results of these recognized consensus standards were implemented on the Optina-4C™.
#### ● Spectral accuracy and reliability of the retinal images
The spectral accuracy of the illumination light of the MHRC-C1 was verified using a spectrometer. The spectral accuracy and reliability of the retinal images were evaluated in an eye model using a reference material with tabulated spectral bands.
#### ● Biocompatibility
The intact skin from the chin and forehead of the patient is intended to contact the MHRC-C1 for a short period of time. The biocompatibility of the MHRC-C1 was assessed in accordance with ISO 10993-1:2018.
#### ● Evaluation of recognized consensus standards for ophthalmic cameras
The MHRC-C1 was found to comply with the recognized consensus standard ISO 10940:2009 specifying product requirements for fundus camera.
The MHRC-C1 was found to comply with the recognized consensus standard ANSI Z80.36: 2021 specifying fundamental requirements for optical radiation safety for ophthalmic instruments.
### 5. Substantial Equivalence
In conclusion, the performance testing support that the Optina-4C™ meets the recognized consensus standards for a fundus camera and that retinal imaging at multiple wavelengths (colors) may be accurately and reliably achieved. The proposed device did not raise new concerns regarding its safety and efficacy for its intended use and was therefore deemed substantially equivalent to the predicate device.
3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.