EConsole1 is indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures (excluding fluoroscopic, and mammographic applications). The main features of this software are controlling and interfacing the detector, controlling the x-ray generator acquisition settings, storing acquired images, data management and image processing.
Device Story
EConsole1 is a digital image processing console software for digital X-ray detectors. It interfaces with X-ray detectors and generators to acquire, process, store, and manage radiographic images. Operated by qualified medical personnel, it allows input of patient and shot information; it does not control X-ray exposure or generator power, but interfaces with generator protocols to select/modify exposure parameters (kVp, mA, mAs) as defined by the generator manufacturer. Output includes processed digital images for viewing, annotation, measurement, and stitching. The device supports DICOM 3.0 for transmission to PACS and printing. It functions as a workstation-based system to replace traditional film/screen systems, facilitating clinical decision-making through enhanced image management and display.
Clinical Evidence
No clinical data. Bench testing only. Software verification and validation (code, module, integration, and dynamic tests) confirmed that all software specifications met acceptance criteria and risk controls were effective.
Technological Characteristics
Medical software for digital X-ray image processing. Interfaces with digital detectors and X-ray generators via manufacturer-provided protocols. Supports DICOM 3.0 for network connectivity (PACS/printers). Implements image processing functions including viewing, annotation, measurement, and stitching. Complies with IEC 62304 (software life cycle), ISO 14971 (risk management), and IEC 62366-1 (usability).
Indications for Use
Indicated for general radiographic imaging of human anatomy in general-purpose diagnostic procedures. Excludes fluoroscopic, angiographic, and mammographic applications.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below.
October 18, 2023
DRTECH Corporation % Kim Hanbyul Assistant Manager Suite No. 1, 2 Floor/Suite No. 2, 3 Floor, 29, Dunchon-Daero 541 Beon-Gil, Jungwon-Gu Seongnam-si. Gyeonggi-do 13216 SOUTH KOREA
### Re: K231225
Trade/Device Name: EConsole1 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: September 19, 2023 Received: September 19, 2023
#### Dear Kim Hanbyul:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231225
Device Name EConsole1
Indications for Use (Describe)
EConsole1 is indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures (excluding fluoroscopic, and mammographic applications).
The main features of this software are controlling and interfacing the detector, controlling the x-ray generator acquisition settings, storing acquired images, data management and image processing.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92
### 1. Date Prepared [21 CFR 807.92(a) (1)]
October 17, 2023
#### 2. Submitter's Information [21 CFR 807.92(a)
(1)]
- Name of Sponsor: DRTECH Corporation .
- Address: Suite No.1, 2 Floor / Suite No. 2, 3 Floor, 29, Dunchon-daero 541 beon-gil, ● Jungwon-gu, Seongnam-si, Gyeonggi-do, 13216, Republic of Korea ● Contact Name: Hanbyul Kim
- Telephone No.: + 82-31-779-7720 ●
- Fax No.: + 82-31-779-7790 ●
- Email Address : hbkim95(@drtech.com ●
- Registration Number: 3005172103
- Name of Manufacturer: Same as Sponsor ●
#### 3. Identification of the Subject Device [21 CFR 807.92(a) (2)]
- Trade Name: EConsole 1
- . Common Name: Radiological Image Processing System
- . Classification Name: Medical image management and processing system
- Classification Panel: ● Radiology
- Classification Regulation: 21 CFR 892.2050 ●
- Product Code: ● LLZ
- . Device Class:
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
II
II
#### 4.1 Primary predicate device
- 510(k) Number: K10033 ●
- Applicant: IMFOU CO., LTD ●
- Trade Name: FEEL-DRCS ●
- Classification Name: Medical image management and processing system ●
- Classification Panel: Radiology ●
- Classification Regulation: 21 CFR 892.2050
- Product Code: LLZ ●
- Device Class: ●
{4}------------------------------------------------
#### 4.2 Reference predicate device
- 510(k) Number: K152172 ●
- Applicant: DRTECH Corporation ●
- Trade Name: EConsole 1 ●
- Classification Name: Medical image management and processing system .
- Classification Panel: Radiology ●
- Classification Regulation: 21 CFR 892.2050 ●
- Product Code: LLZ ●
- Device Class: . II
#### 5. Description of the Device [21 CFR 807.92(a) (4)]
Radiological Image Processing Software, EConsolel is complete digital image processing console software specialized for the digital X-ray detector series developed by DRTECH Corporation.
EConsole1 not only processes the acquired images but also complies with DICOM standards which allow the user to transmit and receive data with the PACS system and print images through the DICOM printer.
#### 6. Indication for Use [21 CFR 807.92(a)(5)]
EConsole1 is indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures (excluding fluoroscopic, angiographic, and mammographic applications).
The main features of this software are controlling and interfacing the detector, controlling the x-ray generator acquisition settings, storing acquired images, data management and image processing.
#### 7. Technological Characteristics [21 CFR 807.92(a)(6)]
EConsole1 is medical software. It digitalizes the signal sent from the detector and displays the x-ray image. Also, it can enter patient information, shot information, and other necessary references for convenience.
Compared with the predicate device, the technological characteristics of the proposed device, Econsole1, are substantially equivalent to those of the proposed device is functionally similar to the predicate device.
Additionally, our device does not exert control over generator power that produces ionizing radiation. The control function of the X-ray generator is dependent on the specific X-ray Generator company, as EConsole1 can only interface and be controlled using the protocol provided by that particular company. Moreover, EConsole1 can only select or modify values of X-ray exposure parameters (such as kVp, mA, second, or kVp; mAs) according to the defined values set by each X-ray company. However, EConsole1 does not have control over exposure, electrical changes, or the calibration of the X-ray system. Therefore, it is necessary for the radiological technician to check the X-ray exposure conditions on the X-ray Generator before conducting any X-ray procedures. If the X-ray generator does not allow interface with external software, such as EConsole1, the software will not be able to interface with the X-ray Generator.
{5}------------------------------------------------
| Parameter | Subject Device | Predicate Device | Reference Device | Comparison<br>Analysis |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| 510(k) Number | Unknown | K110033 | K152172 | N/A |
| Model Name | EConsole1 | FEEL-DRCS | EConsole1 | N/A |
| Manufacturer | DRTECH Corporation | IMFOU CO., LTD | DRTECH Corporation | N/A |
| Common Name | Radiological<br>Image<br>Processing System | Radiological<br>Image<br>Processing System | Radiological<br>Image<br>Processing System | Identical |
| Classification<br>Name | Medical<br>image<br>management<br>and<br>processing system | Medical<br>image<br>management<br>and<br>processing system | Medical<br>image<br>management<br>and<br>processing system | Identical |
| Classification<br>Panel | Radiology | Radiology | Radiology | Identical |
| Classification<br>Regulation | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 | Identical |
| Product Code | LLZ | LLZ | LLZ | Identical |
| Device Class | Class II | Class II | Class II | Identical |
| Intended Use | EConsole1 is<br>indicated for use in<br>general radiographic<br>images of human<br>anatomy. It is<br>intended to replace<br>radiographic<br>film/screen systems in<br>all general-purpose<br>diagnostic procedures<br>(excluding<br>fluoroscopic,<br>angiographic, and<br>mammographic<br>applications).<br>The main features of<br>this software are<br>controlling and<br>interfacing the<br>detector, controlling<br>the x-ray generator<br>acquisition settings,<br>storing acquired<br>images, data<br>management and<br>image processing | feel-DRCS software<br>using a digital X-ray<br>detector is the digital X-<br>ray image processing<br>system designed to<br>acquire images and<br>process acquired images<br>efficiently.<br>The main features of this<br>software are controlling<br>and interfacing the<br>detector, acquiring<br>images after X-ray,<br>storing acquired images,<br>managing data,<br>optimizing window level<br>and width of acquired<br>images, rotating images,<br>zooming images,<br>measuring images and so<br>on.<br>feel-DRCS is compatible<br>with DICOM 3.0<br>standard. It can transfer<br>images processed in<br>PACS and print images<br>with a film printer. | The Econsole1<br>software is indicated<br>for use in general<br>radiographic images<br>of human anatomy<br>(excluding<br>fluoroscopic,<br>angiographic, and<br>mammographic<br>applications). | Identical |
| Acquisition devices | Digital X-ray Detector | Digital X-ray Detector | compatible with DICOM 3.0 by using DICOM and network systems.<br>feel-DRCS is not approved for the acquisition of mammographic image data and is meant to be used by qualified medical personnel only. All users must be qualified to create and diagnose radiological image data.<br>feel-DRCS is not approved for the acquisition of mammographic image data. | Identical |
| Software Function | Image viewing<br>Image search<br>Image storage<br>Image annotation<br>Image measurement<br>Image processing<br>Image stitch<br>Generator Control | Image viewing<br>Image search<br>Image storage<br>Image annotation<br>Image measurement<br>Image processing<br>Image stitch<br>Generator Control | Image viewing<br>Image search<br>Image storage<br>Image annotation<br>Image measurement<br>Image processing<br>Image stitch | Identical |
| DICOM Compliance | Yes | Yes | Yes | Identical |
# 8. Substantial Equivalence Comparison [21 CFR 807.92(b)]
{6}------------------------------------------------
When compared to the predicate devices (K110033), the device presented in this submission, EConsole1, has the same of the followings as the predicate device:
- Intended Use
- Performance properties
- Technological characteristics
- Software function
There are no significant differences between the EConsole1 and the predicate device that would have a negative impact on the product's use. Therefore, the subject device is considered substantially equivalent to the predicate device.
{7}------------------------------------------------
#### 9. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]
The EConsole1 complies with the following international and FDA-recognized consensus standards:
| Recognition No. | Standard No. | Title of Standard | Remark |
|-----------------|--------------------------------------------------------------|-----------------------------------------------------------------------------|--------|
| 13-79 | IEC 62304 Edition 1.1 2015-06<br>CONSOLIDATED VERSION | Medical device software - Software<br>life cycle processes | |
| 12-349 | NEMA PS 3.1 - 3.20 2022d | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set | |
| 5-135 | ISO 20417 First edition 2021-04<br>Corrected version 2021-12 | Medical devices - Information to be<br>supplied by the manufacturer | |
| 5-129 | IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED<br>VERSION | Medical devices - Part 1:<br>Application of usability<br>engineering to med | |
| 5-125 | ISO 14971 Third Edition 2019-12 | Medical devices - Application of<br>risk management to medical device | |
#### Software Verification and Validation
As part of this submission, the Software Documentation for a Medical Device with Moderate Level of Concern, as outlined in the FDA's Guidance Document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005, is also provided.
The data indicates that medical devices containing software adhere to special controls consistently. During the product development of EConsole1, various non-clinical tests such as code, module, integration, and dynamic tests were conducted. A risk analysis was carried out, and risk controls were put in place to mitigate identified hazards. The test results suggest that all software specifications meet the acceptance criteria. Furthermore, the verification and validation testing were deemed acceptable, thereby supporting the claim of substantial equivalence.
#### Cybersecurity
EConsole1 conforms to Section 6 of the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff (October 2, 2014), to maintain confidentiality, integrity, and availability. Additionally, EConsolel is linked through DICOM standards to medical devices and a PACS system, which stores data generated by the medical devices. It retrieves image data via network communication based on DICOM standards. Therefore EConsole I assures an adequate degree of protection for cybersecurity.
{8}------------------------------------------------
#### DRTECH Corporation Suite No.1, 2 Floor / Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, 13216, Republic of Korea
# DRTECH
#### 10. Summary of Clinical Data [21 CFR 807.92(b)(2)]
Clinical studies are unnecessary to validate the safety and effectiveness of the software in EConsole1, the subject of this 510(k) notification.
#### 11. Conclusion [21 CFR 807.92(b)(3)]
EConsole1 is substantially equivalent to the predicate device. It has the same intended use, performance properties and technological characteristics as its predicate device.
EConsole1 passed all Verification and Validation testing, which means that the features, functions and technology were all demonstrated to perform well. Performance data also demonstrates that EConsole1 is both safe and effective in its intended performance.
Therefore, EConsole1 does not introduce any new potential safety risks and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.