Thermogard HQ™ Temperature Management System, Thermogard XP® Temperature Management System
K231182 · Zoll Circulation, Inc. · NZE · Nov 3, 2023 · Cardiovascular
Device Facts
Record ID
K231182
Device Name
Thermogard HQ™ Temperature Management System, Thermogard XP® Temperature Management System
Applicant
Zoll Circulation, Inc.
Product Code
NZE · Cardiovascular
Decision Date
Nov 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Thermogard HQ™ Temperature Management System Temperature reduction in adult patients where clinically indicated, e.g., in hyperthermic patients. Thermogard XP® Temperature Management System Temperature reduction in adult patients where clinically indicated, e.g., in hyperthermic patients.
Device Story
System consists of Thermogard console (HQ or XP) connected to STx™ Vest via modified Start-Up Kit (SUK). Console circulates thermally conditioned saline through vest; heat exchange between patient and fluid modulates core temperature. Modification involves longer tubing and 0.25-inch hose barb valved connectors to interface with vest. Used in clinical settings for therapeutic temperature management. Healthcare providers operate console to set desired cooling parameters; system automatically adjusts coolant flow to achieve/maintain target temperature. Benefits include controlled, consistent patient cooling.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including dimensional verification, functional testing, temperature control performance, shipping/handling, electrical safety, electromagnetic compatibility, and usability testing.
Technological Characteristics
Thermal regulating system using circulating saline as heat exchange medium. Consists of console, modified SUK, and STx™ Vest. Materials contacting saline are identical to previously cleared predecessors. Connectivity via hose barb connectors. System uses automated flow control to maintain target temperature.
Indications for Use
Indicated for temperature reduction in adult patients where clinically indicated, such as in hyperthermic patients.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 3, 2023
ZOLL Circulation, Inc Sean Manning Senior Manager, Regulatory Affairs 2000 Ringwood Avenue San Jose, California 91531
Re: K231182
Trade/Device Name: Thermogard HQTM Temperature Management System, Thermogard XP® Temperature Management System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal regulating system Regulatory Class: Class II Product Code: NZE Dated: October 12, 2023 Received: October 12, 2023
Dear Sean Manning:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231182
Device Name
Thermogard HQ Temperature Management System Thermogard XP Temperature Management System
Indications for Use (Describe)
Thermogard HO™ Temperature Management System Temperature reduction in adult patients where clinically indicated, e.g., in hyperthermic patients.
Thermogard XP® Temperature Management System Temperature reduction in adult patients where clinically indicated, e.g., in hyperthermic patients.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <span style="font-size: 10px;"> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10px;"> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
#### I. GENERAL INFORMATION
| Date Prepared: | October 31, 2023 |
|-----------------------------|----------------------------------------------------------------------------------------------|
| Submitter: | ZOLL Circulation, Inc. |
| Address: | 2000 Ringwood Avenue<br>San Jose, CA 95131 |
| Phone: | 707-696-6666 |
| Fax: | 408-541-1030 |
| Contact Person: | Sean Manning, Senior Manager Regulatory Affairs |
| Secondary Contact: | Nazma Chaudhry, Director Regulatory Affairs |
| Trade Name: | Thermogard HQ™ Temperature Management System<br>Thermogard XP® Temperature Management System |
| Common Name: | Thermal Regulating System |
| Classification/Name: | Class II; System, Thermal Regulating |
| Regulation: | 21 CFR 870.5900, Thermal Regulating System |
| Product Code: | NZE |
| Primary Predicate Device: | K100071 EMCOOLSpad |
| Secondary Predicate Device: | K162523 The IQool™ System |
#### II. DEVICE DESCRIPTION
The subject of this 510(k) is the new surface feature which makes minor modifications to the existing IVTM Start-Up Kits (SUKs) to enable connection of a Thermogard console to the STx™ Vest provides for circulation of thermally conditioned fluid supplied by the Thermogard console (either Thermogard HQ™ or Thermogard XP®). Transfer of heat between the patient and the circulating fluid results in modification of patient core temperature for the purpose of therapeutic patient temperature modulation.
The existing, commercialized Thermogard Systems (Thermogard HQTM and Thermogard XP®) are used in conjunction with catheters placed in the vasculature to support intravascular cooling. The TGHQ-SURF accessory is a modification of the model TGHQ-500D accessory which was verified, validated, and commercialized. The CG-SURF accessory is a modification of the model CG-500D accessory which was verified, validated, and commercialized. These Surface SUK models are identical to their predecessors with the exception of longer inflow and outflow tubing lengths and replacing the Luers for catheters with 0.25inch hose barb valved connectors for the STx™ Vest. The Surface SUK also has a label on the fluid spike to indicate the size of bag of fluid (3000ml) to use and to mark it as non-sterile. The Surface SUKs are otherwise identical to their predecessors in terms of materials that contact the heat exchange fluid (i.e., saline), manufacturing process, and packaging materials.
There are no changes to the commercially available Thermogard XP® (K213031) and Thermogard HO™ (K220008) consoles.
There are no changes to the commercially available STxTM Vest.
{4}------------------------------------------------
#### INDICATIONS FOR USE III.
Thermogard HQ™ Temperature Management System
Temperature reduction in adult patients where clinically indicated, e.g., in hyperthermic patients.
Thermogard XP® Temperature Management System
Temperature reduction in adult patients where clinically indicated, e.g., in hyperthermic patients.
#### SUBSTANTIAL EQUIVALENCE DISCUSSION IV.
| Thermogard Temperature Management System Non-Confidential 510(k) Summary | | | | | | |
|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Comparison of Subject to Predicate Device: | | | | | | |
| Description | Subject Device | Predicate Device | Secondary Predicate<br>Device | | | |
| Proprietary Name | Thermogard HQ and<br>XP Temperature<br>Management System | EMCOOLSpad | IQool™ System | | | |
| Common Name | Thermal Regulating<br>System | Thermal Regulating<br>System | Thermal Regulating<br>System | | | |
| Product<br>Classification Code | NZE | NZE | NZE | | | |
| Product Regulation<br>Number & Name | 21 CFR 870.5900,<br>Thermal Regulating<br>System | 21 CFR 870.5900,<br>Thermal Regulating<br>System | 21 CFR 870.5900,<br>Thermal Regulating<br>System | | | |
| Device Class | Class II | Class II | Class II | | | |
| Device Description | System, Thermal<br>Regulating | System, Thermal<br>Regulating | System, Thermal<br>Regulating | | | |
| Indications For Use | Temperature reduction<br>in adult patients where<br>clinically indicated,<br>e.g., in hyperthermic<br>patients. | Temperature reduction<br>in adult patients where<br>clinically indicated,<br>e.g., in hyperthermic<br>patients. | Temperature reduction<br>in adult patients where<br>clinically indicated,<br>e.g., in hyperthermic<br>patients. | | | |
| Technical<br>Specifications | Achieves cooling with<br>patient contacting<br>Cooling Pads through<br>which coolant (saline)<br>is circulated. Cooling is<br>controlled by the<br>settings on the device.<br>Consistent temperature<br>reduction is maintained<br>and controlled by the<br>system pushing more<br>or less coolant through<br>the pads as needed<br>achieve or maintain<br>desired cooling. | Achieves cooling<br>through patient<br>contacting Cooling<br>Pads which are filled<br>with a high thermal<br>conductivity material.<br>Pads must be pre-<br>cooled before use. To<br>control cooling, pads<br>are removed from the<br>patient's skin to slow<br>cooling or to stop<br>cooling. The pads are<br>re-applied to the<br>patient's skin to<br>recommence the<br>cooling process. | Achieves cooling with<br>patient contacting<br>Cooling Pads through<br>which coolant is<br>circulated. Cooling is<br>controlled by the<br>settings on the device.<br>Consistent temperature<br>reduction is maintained<br>and controlled by the<br>system pushing more or<br>less coolant through the<br>pads as needed achieve<br>or maintain desired<br>cooling. | | | |
{5}------------------------------------------------
The subject device is identical to the predicates in the indications for use. The mechanism of action to achieve the intended use is equivalent between the subject and predicate devices.
#### V. SUMMARY OF THE NONCLINICAL TESTS PERFORMED
Testing was completed for the Thermogard Temperature Management System used with the Surface Start-Up Kit and STx™ Vest which demonstrated intended, labeled device performance. Performance testing included dimensional verification, functional testing, temperature, shipping and handling, and standards testing included electrical safety, electromagnetic compatibility, and usability. All testing confirmed that the Thermogard Temperature Management System used with the Surface Start-Up Kit and STx™ Vest operates as described and is substantially equivalent to the predicate devices.
#### SUMMARY OF THE CLINICAL TESTS PERFORMED VI.
Clinical performance data to demonstrate substantial equivalence was not required.
#### VII. CONCLUSION
The Thermogard HQTM Temperature Management System and the Thermogard XP® Temperature Management System used with the Surface Start-Up Kit and STx™ Vest meet the design, performance, and safety specifications when used in accordance with the labeling. It was demonstrated through performance testing, design and features, and non-clinical testing that the proposed devices are substantially equivalent to those of the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.