K231179 · Anoxia Medical, Inc. · QJP · Dec 1, 2023 · Cardiovascular
Device Facts
Record ID
K231179
Device Name
Slinky Catheter
Applicant
Anoxia Medical, Inc.
Product Code
QJP · Cardiovascular
Decision Date
Dec 1, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Slinky Catheter is indicated for general intravascular use. including the neuro and peripheral vasculature. It can be used to facilitate introduction of diagnostic agents and therapeutic devices. It is not intended for use in coronary arteries.
Device Story
Slinky Catheter is a single-lumen, variable-stiffness catheter; features coil/braid reinforcement, radiopaque distal marker, and hydrophilic coating. Used in neuro and peripheral vasculature to facilitate delivery of diagnostic agents and therapeutic devices. Operated by physicians (e.g., interventional neuroradiologists) in clinical settings. Device provides a pathway for ancillary devices (guidewires, microcatheters, stent-retrievers). Benefits include enhanced vascular navigation and visualization under fluoroscopy. Single-use, sterile, EO-sterilized.
Clinical Evidence
Bench testing only. No clinical data. Performance validated via simulated use in tortuous anatomical models by interventional neuroradiologists, confirming compatibility with ancillary devices. Mechanical testing included kink resistance, torque strength, tensile force, burst pressure, and flow rate. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, complement activation, thrombogenicity) confirmed safety.
Technological Characteristics
Single-lumen catheter; 6F size; 132 cm length. Materials: Pebax, Grilamid, Polyblend, Pellethane, PTFE, 304 V Stainless Steel (braid/coil), Platinum/Iridium marker, Acrylic hub. Hydrophilic coating (Harland). EO sterilization. ISO 80369-7 compliant hub. ISO 10555-1 compliant integrity.
Indications for Use
Indicated for general intravascular use in neuro and peripheral vasculature to facilitate introduction of diagnostic agents and therapeutic devices. Not for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
December 1, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Anoxia Medical, Inc. % Bosmat Friedman, MSc Regulatory Consultant ProMedoss, Inc. 3521 Hatwynn Road Charlotte, North Carolina 28269
# Re: K231179
Trade/Device Name: Slinky Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY, DQO Dated: November 1, 2023 Received: November 1, 2023
# Dear Bosmat Friedman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231179
Device Name Slinky Catheter
Indications for Use (Describe)
The Slinky Catheter is indicated for general intravascular use. including the neuro and peripheral vasculature. It can be used to facilitate introduction of diagnostic agents and therapeutic devices. It is not intended for use in coronary arteries.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
### 510(k) Number: K231179
#### Date Prepared: November 30, 2023
| Submitter/Manufacturer | Anoxia Medical, Inc. |
|------------------------|----------------------------------------------|
| | 3475 Investment Boulevard, Suite #9 |
| | Hayward, CA 94545 |
| Contact | Henry Nita |
| | Telephone: 650-430-2045 |
| Trade Name | Slinky Catheter |
| Common/Usual Name | Distal Access Catheter |
| Regulation Description | Percutaneous Catheter |
| Regulation Number | 21 CFR 870.1250; 21 CFR 870.1200 |
| Product Code | QJP, DQY, DQO |
| Device Class | Class II |
| Classification Panel | Neurology, Cardiovascular |
| Predicate Devices | SOFIA PLUS/Distal Access Catheters (K150366) |
#### Device Description:
The Slinky® Catheter is a single lumen, variable stiffness catheter with a coil and braid reinforcement. It has a radiopaque marker at the distal tip for enhanced fluoroscopic visualization, a lubricious inner, and a hydrophilic coating on the distal 35 cm to enhance lubricity and vascular navigation. The Slinky® Catheter is provided sterile via Ethylene Oxide (EO) sterilization, non-pyrogenic, and is for single use only.
#### Indications for Use:
The Slinky Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. It can be used to facilitate introduction of diagnostic agents and therapeutic devices. It is not intended for use in coronary arteries.
#### Comparison of Technological Characteristics with the Predicate Device:
The Slinky® Catheter has similar indications for use and technological characteristics as the predicate device as presented in the following table. The differences do not raise new questions of safety or effectiveness.
{4}------------------------------------------------
| Comparison to the Predicate Device | | |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Predicate Device<br>SOFIA PLUS/Distal Access<br>Catheters<br>K150366 | Subject Device<br>Slinky® Catheter<br>K231179 |
| Classification,<br>Product Code | Class II, DQY, DQO | Class II, QJP, DQY, DQO |
| Generic Name | Percutaneous Catheter | Same |
| Regulation<br>Number | 21 CFR 870.1250; 21 CFR 870.1200 | Same |
| Classification<br>Panel | Neurology, Cardiovascular | Same |
| Indications<br>For Use | The SOFIA PLUS/Distal Access<br>Catheters are indicated for general<br>intravascular use, including the neuro<br>and peripheral vasculature. It can be<br>used to facilitate introduction of<br>diagnostic and therapeutic agents. It is<br>not intended for use in coronary<br>arteries. | The Slinky Catheter is indicated for<br>general intravascular use, including<br>the neuro and peripheral vasculature.<br>It can be used to facilitate introduction<br>of diagnostic agents and therapeutic<br>devices. It is not intended for use in<br>coronary arteries. |
| DIMENSIONS: | | |
| Catheter Size | 6F | 6F |
| Catheter<br>Effective Length | 115-135 cm | 132 cm |
| Proximal ID | 0.070" | 0.075" |
| Proximal OD | 0.0825" | 0.086" |
| Tip Configuration | Steam shapeable by user | Straight-cut or 30°-cut atraumatic soft<br>tip configurations |
| MATERIALS: | | |
| Catheter Body<br>Outer Layer | Polyurethane elastomer (Polyblend<br>and Pellethane), polyether block amide<br>(Pebax) and polyamide (Grilamid) | Pebax® 55D, Pebax® 63D, Grilamid™<br>L25, QFLEX 85A/Polyblend 60A,<br>Pellathane® 80A, Pebax® 35D |
| Catheter Body<br>Inner Layer | Stainless steel braid/coil, PTFE and<br>polyolefin elastomer | Braid and coil: 304 V Stainless Steel,<br>PTFE, QFLEX 75A/Polyblend 60A |
| Marker | Platinum/Iridium | Platinum/Iridium |
| Hub | Nylon | Acrylic |
| Strain Relief | Polyurethane | Polyolefin |
| Shaping Mandrel | Stainless steel | N/A |
| Coating | Hydrophilic coating: Hydak® | Hydrophilic coating: Harland |
| Guidewire<br>Compatibility | 0.035" | 0.014", 0.035", 0.038" |
| Method of<br>Supply | Sterile and Single Use | Same |
| Sterilization Method | Ethylene oxide | Same |
| Packaging Card | Polyethylene | Same |
| Packaging Hoop | HDPE | Polyethylene |
| Packaging Pouch | Tyvek® | Nylon and LDPE (top)<br>Tyvek® 1073B (bottom) |
| Packaging Carton Box | Pouch and IFU placed in carton box | Same |
| Shelf Life | Not specified | 12 Months |
| Accessories Included | Introducer sheath, shaping mandrel | None |
{5}------------------------------------------------
### Performance Testing:
To demonstrate the substantial equivalence of the subject Slinky® Catheter to the predicate device, the SOFIA PLUS/Distal Access Catheters, the performance, biocompatibility, and sterility of the Slinky® Catheter were evaluated via the following tests and assessments:
- Design Verification Bench Testing ●
- Biocompatibility .
- Shelf-Life .
- Sterilization ●
- Packaging Validation ●
# Bench Performance Testing
| Test | Test Method Summary | Conclusions |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Simulated Use | Test articles were used by three trained interventional<br>neuroradiologists in a tortuous anatomical bench test<br>model, demonstrating compatibility with ancillary<br>devices including introducers, microcatheters, and<br>guidewires. | Device performed as intended<br>under simulated use conditions. |
| Particulate Test | Test articles were tracked multiple times through the<br>tortuous anatomical model with ancillary devices.<br>Devices were inspected for coating anomalies pre- and<br>post-testing. | Particulate generation of the<br>subject device was similar to<br>the predicate device. |
| Product Compatibility | Simulated use testing by three physicians was<br>conducted, using a full-length silicone anatomical<br>tortuous path model. The procedure included using the<br>test catheter along with worst case ancillary devices<br>(i.e., smallest ID introducer, largest OD guidewire,<br>stiff micro guidewire, large stent-retriever). | The device was found to be<br>compatible with ancillary<br>devices. |
| Dimensional and Visual<br>Attributes | The physical and dimensional attributes were<br>evaluated and measured. | Device met the established<br>dimensional and visual<br>specifications. |
| Test | Test Method Summary | Conclusions |
| Kink Resistance | The distal, mid, and proximal sections of the catheter<br>were wrapped around varying size pin gauges. | Device met the established<br>kink resistance criteria. |
| Radio Detectability | The distal marker band and catheter body visibility<br>were evaluated under fluoroscopy. | Device radiopacity comparable<br>to the predicate device. |
| Catheter Hub | The hub functionality was evaluated during simulated<br>use, freedom from leakage, flow rate, dynamic burst,<br>particulate, and other bench testing completed. | Device hub meets the<br>requirements of ISO 80369-7. |
| Durability/Lubricity of<br>Hydrophilic Coating | Lubricious coating on the catheter surface was<br>evaluated after simulated use for defects and for<br>friction force. | Device met the established<br>friction and lubricity criteria.<br>The friction force was similar<br>to the predicate. |
| Tip Flexibility | Device tip stiffness was characterized on a cantilever<br>bend test. | The tip stiffness was similar to<br>the predicate. |
| Torque Strength | Devices were pre-conditioned by simulated use and<br>torqued to failure inside a tortuous anatomical model<br>with the distal tip held fixed. | Device torque strength is the<br>same as the predicate device. |
| Force at Break (Distal<br>and Hub) | Catheter force at break averaged 8 lbf for the hub and<br>5-6 lbf for the mid and distal segments and exceeds<br>acceptance criteria determined from withdrawal force<br>measurements. | Tensile strength meets the test<br>acceptance criteria. |
| Flow Rate | Flow rate was characterized at 100 psi with diagnostic<br>agents (e.g., saline, contrast media). | Device meets the specified<br>requirements for delivery of<br>diagnostic agents. |
| Static Burst Pressure | Device can withstand static pressure above labeled<br>100 psi maximum with margin of safety. | Device integrity is suitable for<br>intended clinical use and met<br>requirements of ISO 10555-1. |
| Freedom from Leakage | Testing followed the methods and criteria defined in<br>EN ISO 10555-1 after pre-conditioning by simulated<br>use. | Device integrity is suitable for<br>intended clinical use and met<br>requirements of ISO 10555-1. |
| Air Leakage | No air leakage at hub into syringe for 15 seconds. | Device integrity is suitable for<br>intended clinical use and met<br>requirements of ISO 10555-1. |
| Dynamic Burst | Devices were pre-conditioned by simulated use. Test<br>devices were then connected to a pressurized fluid<br>source, starting with 250 psi and further increased<br>every 10 seconds until a maximum of 600 psi was<br>applied. | Device met labeled maximum<br>infusion pressure of 100 psi. |
| Peak Tensile Force | Devices were pre-conditioned by simulated use.<br>Tensile testing was then conducted per the methods<br>defined in EN ISO 10555-1. | Device met acceptance criteria<br>determined from withdrawal<br>force testing. |
| Test | Test Method Summary | Conclusions |
| Cytotoxicity (ISO<br>10993-5) - MEM<br>Elution Assay | Cell culture observed for cytotoxic reactivity<br>following exposure to test extract with Grade<br><2 as acceptable. | Non-cytotoxic |
| Sensitization (ISO<br>10993-10) - Guinea Pig<br>Maximization<br>Sensitization | Observation of study animals exposed to test<br>extracts for delayed dermal sensitization. | No sensitization reaction |
| Irritation (ISO 10993-<br>10) - Intracutaneous<br>Reactivity | Observations for dermal reactivity were<br>conducted at 24, 48, and 72 hours after test<br>extract injection. Test article sites should not<br>show a significantly greater biological reaction<br>than sites injected with control article. | Non-irritant |
| Acute Systemic<br>Toxicity (ISO 10993-<br>11) | Study animals injected with the subject device<br>observed for abnormal clinical signs indicative<br>of toxicity during the 72-hour test period. | No signs of toxicity |
| Pyrogenicity (USP<br><151>, ISO 10993-11) | Study animals observed for individual<br>temperature rise following intravenous<br>injection of test extracts. | Non-pyrogenic |
| Hemolysis (ASTM<br>F756, ISO10993-4) –<br>Direct and Indirect<br>(extract) Hemolysis | The difference between the hemolytic indexes<br>of the subject device and the negative control<br>was evaluated after exposure to test extracts<br>and direct exposure to test article. | Direct contact - Non-hemolytic<br>Indirect contact - Non-hemolytic |
| SC5b-9 Complement<br>Activation Assay (ISO<br>10993-4) | Using an enzyme immunoassay, SC5b-9<br>concentration following exposure of test article<br>extracts to normal human serum (NHS)<br>compared to SC5b-9 concentration of the<br>control article exposed to NHS. | Not considered to be a potential<br>activator of the complement system |
| Thrombogenicity<br>(ASTM F2382) - Partial<br>Thromboplastin Time | Thrombogenicity potential of test article<br>compared to control articles. | Acceptable results |
| Platelet and Leukocyte<br>Count (ASTM F2888-19,<br>ISO 10993-4)<br>In vitro Blood Loop | | |
{6}------------------------------------------------
# Biocompatibility
Biocompatibility was assessed in accordance with ISO 10993-1 guidelines for a limited exposure (< 24 hours), externally communicating device with circulating blood contact. All studies were conducted according to 21 CFR Part 58 Good Laboratory Practices.
{7}------------------------------------------------
# Sterilization and Shelf-Life
The Slinky® Catheter is labeled as a single-use sterile device with shelf-life of 12 months. The sterilization process has been validated and process monitoring controls are in place to assure that the device is EO sterilized to achieve a minimum sterility assurance level (SAL) of 10 ° Shelf-life studies have been conducted and demonstrate that the product and packaging remain functional and sterile for the shelf-life period of 12 months.
{8}------------------------------------------------
# Conclusion:
The Slinky® Catheter has the same intended use and similar technological characteristics and principles of operation as the currently marketed predicate SOFIA PLUS/Distal Access Catheters. Any differences between the subject and predicate device do not raise new questions of safety and effectiveness. The subject device was evaluated through design verification and validation, biocompatibility, and sterility testing. Based on the information submitted in this 510(k) submission, the subject Slinky® Catheter is substantially equivalent to the currently marketed predicate SOFIA PLUS/Distal Access Catheters.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.