The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.
Device Story
AI-based computer-assisted detection (CADe) device; processes live colonoscopy video input; identifies regions consistent with colorectal lesions (polyps/adenomas). Device connects between video processor and endoscopic display; overlays green square markers on detected lesions; provides audible alert. Used in clinical settings by endoscopists. System 200 model adds 4K UHD support via internal 'wrapper' library converting 4K to HD for processing, then back to 4K for display. Optional footswitch/USB K-switch allows manual toggle of visual overlays. Output aids endoscopist in lesion detection; does not replace clinical decision-making. Benefits include real-time visual assistance during screening/surveillance.
Clinical Evidence
Bench testing only. Non-inferiority of performance with Olympus CV-1500 EVIS X1 UHD processor established via per-frame assessment on 42 pre-recorded procedures. Verification/validation of software, EMC/electrical safety (IEC 60601-1/1-2), and usability studies for new accessories performed. Performance metrics (sensitivity 86.5%, AUC 0.796) identical to predicate.
Technological Characteristics
Hardware-based CADe system; interfaces with video processors via SDI/HD-SDI. System 200 includes 4K UHD videoboard and internal wrapper library. Connectivity: wired video input/output, USB for footswitch. Software: AI-based detection algorithm (v3.0.2). Standards: IEC 60601-1, IEC 60601-1-2, SMPTE 259M/292M.
Indications for Use
Indicated for patients undergoing screening and surveillance endoscopic mucosal evaluations to aid endoscopists in real-time detection of colonic mucosal lesions (polyps, adenomas) during standard white-light endoscopy.
Regulatory Classification
Identification
A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope.
Special Controls
In combination with the general controls of the FD&C Act, the gastrointestinal lesion software detection system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including detection of gastrointestinal lesions and evaluation of all adverse events.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Standalone algorithm performance testing;
(ii) Pixel-level comparison of degradation of image quality due to the device;
(iii) Assessment of video delay due to marker annotation; and
(iv) Assessment of real-time endoscopic video delay due to the device.
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Performance data must demonstrate electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety testing for any hardware components of the device.
(5) Software verification, validation, and hazard analysis must be provided. Software description must include a detailed, technical description including the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, the associated inputs and outputs, mapping of the software architecture, and a description of the video signal pipeline.
(6) Labeling must include:
(i) Instructions for use, including a detailed description of the device and compatibility information;
(ii) Warnings to avoid overreliance on the device, that the device is not intended to be used for diagnosis or characterization of lesions, and that the device does not replace clinical decision making;
(iii) A summary of the clinical performance testing conducted with the device, including detailed definitions of the study endpoints and statistical confidence intervals; and
(iv) A summary of the standalone performance testing and associated statistical analysis.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 19, 2023
Cosmo Artificial Intelligence - AI Ltd % Roger Gray VP Quality and Regulatory Donawa Lifescience Consluting Srl Piazza Albania 10 Rome, 00153 ITALY
Re: K231143
Trade/Device Name: GI Genius System 100 and GI Genius System 200 Regulation Number: 21 CFR 876.1520 Regulation Name: Gastrointestinal lesion software detection system Regulatory Class: Class II Product Code: ONP Dated: April 20, 2023 Received: April 21, 2023
Dear Roger Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K231143
Device Name
GI Genius System 100 and GI Genius System 200
#### Indications for Use (Describe)
The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------|-----------------------------------------------|
|------------------------------------------------|-----------------------------------------------|
N Prescription Use (Part 21 CFR
unter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cosmo Pharmaceuticals. The logo features a stylized figure in light blue with an orange circle above it, followed by the word "COSMO" in large, teal letters. Below "COSMO" is the word "Pharmaceuticals" in a smaller, sans-serif font.
# 510(k) Summary
| 510(k) Reference: | K231143 | | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------|--|--|
| Device Name: | GI Genius System 100 and GI Genius System 200 | | |
| Type of 510(k) submission: | Special | | |
| Date of submission: | 15 May 2023 | | |
| 510(k) Owner and Submitter: | Cosmo Artificial Intelligence - Al Ltd<br>Riverside II, Sir John Rogerson's Quay<br>Dublin D02 KV60<br>Ireland | | |
| FDA Establishment Reg. Number: | 3018899987 | | |
| Specification Developer: | Linkverse Srl<br>via Ostiense 131/L<br>00154 Rome, Italy | | |
| Owner/Operator Reg. Number: | 3018901422 | | |
| 510(k) Application Correspondent: | Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome, Italy | | |
| Phone:<br>Email: | +39 06 578 2665<br>rgray@donawa.com | | |
| FDA Product Code: | QNP | | |
| FDA Regulation Number: | 21 CFR 876.1520 | | |
| FDA Classification Name: | Gastrointestinal lesion software detection system | | |
| Classification Panel: | Gastroenterology and Urology | | |
| FDA Classification: | Class II | | |
### Indications for Use:
The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.
The indications for use statement is identical to that for the original unmodified (predicate) device, as cleared under K211951.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Cosmo Pharmaceuticals. The logo features a stylized human figure in light blue with a yellow circle above its head, followed by the word "COSMO" in teal. Below "COSMO" is the word "Pharmaceuticals" in a smaller, lighter font.
#### Device Description:
GI Genius is an artificial intelligence-based device that has been trained to process colonoscopy images containing regions consistent with colorectal lesions like polyps, including those with flat (non-polypoid) morphology.
GI Genius is compatible with Video Processors featuring SDI (SMPTE 259M) or HD-SDI (SMPTE 292M) output ports and endoscopic display monitor featuring SDI (SMPTE 259M) or HD-SDI (SMPTE 292M) input ports.
GI Genius is connected between the video processor and the endoscopic display monitor. When first switched on, the endoscopic field of view is clearly identified by four corner markers, and a blinking green square indicator appears on the connected endoscopic display monitor to state that the system is ready to function.
During live video streaming of the endoscopic video image, GI Genius generates a video output on the endoscopic display monitor that contains the original live video together with superimposed green square markers that will appear when a polyp or other lesion of interest is detected, accompanied by a short sound. These markers will not be visible when no lesion detection occurs.
The operating principle of the subject device is identical to that of the predicate device, this being a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device includes hardware to support interfacing with video endoscopy systems.
The baseline clinical validation for the subject device was conducted and reviewed in DEN200055 and is still applicable to the versions of the device that are the subject of this submission.
#### Design changes:
This Special 510(k) submission describes the design changes incorporated into GI Genius following FDA clearance under K211951. The device software version number as cleared under K211951 was 2.0.0; the device software version that is the subject of this Special 510(k) is 3.0.2.
Two types of design change are detailed in this submission. The first covers planned design changes that fall within the FDA guideline for submittal of a new 510(k), and the second covers minor design changes incorporated into the design that have been the subject of internal documentation. In addition, a labeling change is planned, to include a 4K UHD video processor in the list of compatible units in the User Manual.
As a result of user requests, optional accessories are planned to be offered for use with Gl Genius, these being a footswitch/USB K-switch to operate as follows:
- Footswitch: the GI Genius software overlays on the main display image the markers around the target objects. Such overlaid markers can be switched ON/OFF by the user by means of a button on the hardware unit front panel. The footswitch allows the user to operate the same switching function by foot as an alternative to the button on the front panel.
- . USB K-switch: Allows communication between the footswitch and the GI Genius hardware unit, by converting the footswitch signal to a command already recognized by the GI Genius software, which, as a result, does not need to be modified.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Cosmo Pharmaceuticals. The logo features a stylized human figure in light blue with a yellow circle above its head, followed by the word "COSMO" in blue, and the word "Pharmaceuticals" in a smaller font below it. The logo is clean and modern, with a focus on the company's name and its connection to the pharmaceutical industry.
A new endoscopy video processor is to be added to the list of video processors that are compatible with the GI Genius, namely the Olympus CV-1500 EVIS X1, which operates in accordance with the 4K UHD standard. The current GI Genius videoboard does not support the 4K UHD standard, so a new videoboard that does support 4K UHD functionality will be sourced and incorporated into the GI Genius that is the subject of this Special 510(k).
Changing the videoboard would normally necessitate a change to the polyp detection Al-based software for the device, so to allow maintenance of the current unchanged software version, the videoboard software communication will be interfaced by a new internal library, working as a 'wrapper'. The wrapper will convert the 4K UHD video stream from the video processor to HD as an input to the GI Genius software, and then convert the GI Genius software screen overlay from HD to 4K in output. Such design allows the original video frames of the endoscopy video processor not to be altered.
With the understanding that many GI Genius users will not require 4K UHD capability, at least for the time being, this design change will result in a range extension for the GI Genius, with two versions being available, these being:
- GI Genius System 100 (the current model, as cleared under K211951, but with new device model name, and ability to use footswitch/K-switch), and
- . GI Genius System 200 (the new model with 4K UHD capability, including full 4K UHD compatibility with the Olympus CV-1500 EVIS X1 video processor, and ability to use footswitch/K-switch).
## Non-clinical testing:
The following verification / validation activities have been carried out:
- . Verification of the revised software at the system level has been carried out, incorporating tests sufficient for the verification of the SRS.
- . Validation of the revised software at the user level has been carried out, incorporating tests sufficient for the validation of the SRS.
- Verification of the three new SRS related to the internal library wrapper has been carried out at the system level and found to meet requirements.
- . Validation of the new SRS related to the internal library wrapper at the user level has been carried out and a Validation Test Report was generated.
- Risk mitigation measures identified during Risk Management have been successfully verified or validated, as applicable.
- . Electromagnetic Compatibility (EMC) and Electrical Safety compliance tests have been successfully completed on the hardware variant with the 4K video board according to IEC 60601-1 and IEC 60601-1-2 requirements, operating with software version 3.0.2.
- Tests according to the Standalone Performance Testing Protocol v2.0, submitted as part of the K211951 predicate device submission, have been repeated for the applicable parts of the subject device.
- As part of the above protocol, non-inferiority of performance of GI Genius with the Olympus CV-1500 EVIS X1 UHD video processor has been established by means of a per-frame assessment on 42 pre-recorded procedures. Since the Olympus CV-1500 simultaneously outputs both HD and UHD video signals, the non-inferiority test has been conducted on the UHD video signal only and on GI Genius System 200 only, as this represents the worst-case condition for compatibility testing.
- . The footswitch and the USB K-switch send a single switching command to the GI Genius software to turn ON/OFF the overlay on target objects. The Electromagnetic Compatibility (EMC) and
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Cosmo Pharmaceuticals. The logo features a stylized human figure in light blue with a yellow circle above its head, followed by the word "COSMO" in large, teal letters. Below "COSMO" is the word "Pharmaceuticals" in a smaller, light gray font.
Electrical Safety compliance tests have been successfully completed on the footswitch and USB Kswitch, according to IEC 60601-1 and IEC 60601-1-2 requirements.
- . Given that the user interface of the GI Genius system will not change, a usability study has been performed with respect to use of the footswitch and USB K-switch, to verify the acceptability of the risk control measures.
The results of the above testing aid demonstration of substantial equivalence of the subject device with the predicate device, as the same test protocols have been used where applicable.
#### Substantial equivalence
The predicate device for the subject device is the pre-modification version of the same device, GI Genius, FDA-cleared under DEN200055 on 9 April 2021:
| Predicate Device: | GI Genius |
|----------------------|---------------------------------------------------|
| Sponsor: | Cosmo Artificial Intelligence - Al Ltd |
| De Novo Number: | K211951 |
| Clearance Date: | 23 July 2021 |
| FDA Product Code: | QNP |
| Classification Name: | Gastrointestinal lesion software detection system |
| Regulation No: | 21 CFR 876.1520 |
| Class: | II |
#### Predicate device comparison table:
Table 1 provides evidence of substantial equivalence of the subject device with the predicate device.
| Table 1: Predicate device comparison table | | | | | | | |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--|--|--|
| Characteristic | Subject device | | Predicate device | Comparison | | | |
| Device name | GI Genius<br>System 200 | GI Genius<br>System 100 | GI Genius | N/A | | | |
| Manufacturer | Linkverse Srl, Italy | | Linkverse Srl, Italy | Same | | | |
| FDA clearance | This submission | | K211951 | N/A | | | |
| FDA Reg name | Gastrointestinal lesion software<br>detection system | | Gastrointestinal lesion software<br>detection system | Same | | | |
| FDA Reg # | 21 CFR 876.1520 | | 21 CFR 876.1520 | Same | | | |
| FDA Product Code | QNP | | QNP | Same | | | |
| Indications for Use | The GI Genius System is a computer-<br>assisted reading tool designed to aid<br>endoscopists in detecting colonic<br>mucosal lesions (such as polyps and<br>adenomas) in real time during<br>standard white-light endoscopy<br>examinations of patients undergoing<br>screening and surveillance<br>endoscopic mucosal evaluations.<br>The GI Genius computer-assisted<br>detection device is limited for use<br>with standard white-light endoscopy<br>imaging only. This device is not<br>intended to replace clinical decision<br>making. | | The GI Genius System is a computer-<br>assisted reading tool designed to aid<br>endoscopists in detecting colonic<br>mucosal lesions (such as polyps and<br>adenomas) in real time during<br>standard white-light endoscopy<br>examinations of patients undergoing<br>screening and surveillance<br>endoscopic mucosal evaluations.<br>The GI Genius computer-assisted<br>detection device is limited for use<br>with standard white-light endoscopy<br>imaging only. This device is not<br>intended to replace clinical decision<br>making. | Same | | | |
| Video delay, signal in to<br>signal out | 0.74 µs | 1.52 µs | 1.52 µs | Improved for<br>System 200 * | | | |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Cosmo Pharmaceuticals. The logo features a stylized human figure in blue with an orange circle above its head. To the right of the figure is the company name "COSMO" in blue, with the word "Pharmaceuticals" in a smaller font size underneath.
| Table 1: Predicate device comparison table | | | |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-------------------|
| Characteristic | Subject device | Predicate device | Comparison |
| Lesion-based sensitivity | 86.5 % | 86.5 % | Same |
| Frame level performance<br>(150 videos / 338<br>polyps)* | True positive: 269,223<br>True negative: 5,239,128<br>False positive: 104,669<br>False negative: 192,567 | True positive: 269,223<br>True negative: 5,239,128<br>False positive: 104,669<br>False negative: 192,567 | Same |
| True positive rate per<br>frame | Mean: 58.30 %<br>% of polyps: 100 % | Mean: 58.30 %<br>% of polyps: 100 % | Same |
| False positive rate per<br>frame | Mean: 1.96 % | Mean: 1.96 % | Same |
| Frame-based TPr/FPr<br>ROC curve, AOC | 0.796 | 0.796 | Same |
| False positive clusters<br>per patient | < 500 ms: 40 less than baseline<br>> 500 ms: 1 more than baseline | < 500 ms: 40 less than baseline<br>> 500 ms: 1 more than baseline | Same |
| Additional video<br>processor | Yes: Olympus<br>CV-1500 EVIS X1<br>with UHD output | Yes: Olympus<br>CV-1500 EVIS X1<br>with HD output<br>N/A | Same |
| Optional footswitch | Yes, connected via K-switch to USB<br>port | No | New functionality |
| Electrical safety | IEC/EN 60601-1 (including<br>footswitch/K-switch) | IEC/EN 60601-1 | Same |
| Electromagnetic<br>compatibility | IEC/EN 60601-1-2 (including<br>footswitch/K-switch) | IEC/EN 60601-1-2 | Same |
| LAN port | Yes, non-functional to user | Yes, non-functional to user | Same |
Note *: UHD pixels are 8 times faster that HD pixels. To cope with this increase, 4 parallel processors are used, resulting in a twofold lower video delay value.
The subject device and the predicate device have many identical or similar characteristics. None of the identified differences introduce new aspects of safety or effectiveness.
#### Conclusion
The subject and predicate devices have identical indications for use and fundamental technological characteristics. Any differences in performance between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device, which is already in interstate commerce within the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.