K231138 · Oscimed SA · LRK · Jan 16, 2024 · Dental
Device Facts
Record ID
K231138
Device Name
Aniti-Snore Mouthguard+
Applicant
Oscimed SA
Product Code
LRK · Dental
Decision Date
Jan 16, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended to aid in the reduction of snoring for adults at least 18 years old.
Device Story
Anti-Snore Mouthguard+ is a thermoforming mandibular advancement device (MAD) used for snoring reduction. The device consists of upper and lower trays constructed from a rigid copolymer frame overmolded with ethylene vinyl acetate (EVA) thermal impression material. Users perform a 'boil-and-bite' procedure in a home environment to customize the fit to their dentition. The trays are connected via flexible thermoplastic elastomer straps of varying lengths, which pull the lower jaw forward to increase pharyngeal space. The device allows for mandibular advancement up to +6mm. It is an over-the-counter (OTC) device intended for patient self-use. The output is the physical repositioning of the mandible, which helps maintain airway patency during sleep, potentially reducing snoring.
Clinical Evidence
Bench testing only. Testing included strap tensile testing, validation of the temperature indicator for the molding process, and force measurement/mandibular advancement testing to confirm advancement up to +6mm and ensure forces are clinically tolerable compared to the predicate.
Technological Characteristics
Thermoforming mandibular advancement device. Materials: rigid copolymer frame, ethylene vinyl acetate (EVA) overmolding, thermoplastic elastomer straps. Biocompatibility per ISO 10993-5 and ISO 10993-10 for prolonged contact (>24h to 30 days). Mechanical adjustment via interchangeable flexible straps. OTC 'boil-and-bite' customization.
Indications for Use
Indicated for adults at least 18 years old for the reduction of snoring.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 16, 2024
Oscimed SA % Michael Chen Quality Engineer Rook Quality Systems 1155 Mount Vernon Hwy. Suite 800 Dunwoody, Georgia 30038
Re: K231138
Trade/Device Name: Anti-Snore Mouth Guard+ Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: April 20, 2023 Received: April 21, 2023
Dear Michael Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K231138
Device Name
### Anti-Snore Mouthguard+
Indications for Use (Describe)
The device is intended to aid in the reduction of snoring for adults at least 18 years old.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: | K231138 | | 510(k) Summary | | Prepared on: 2024-01-12 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|--|----------------|--------------|-------------------------|
| Contact Details | | | | | 21 CFR 807.92(a)(1) |
| Applicant Name | Oscimed SA | | | | |
| Applicant Address | Allée du Quartz 3 2300 La Chaux-de-Fonds Suisse Switzerland | | | | |
| Applicant Contact Telephone | 41 32 926 63 36 | | | | |
| Applicant Contact | Mr. Jacques Magnin | | | | |
| Applicant Contact Email | jm@oscimedsa.com | | | | |
| Device Name | | | | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Anti-Snore Mouthguard+ | | | | |
| Common Name | Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea | | | | |
| Classification Name | Device, Anti-Snoring | | | | |
| Regulation Number | 872.5570 | | | | |
| Product Code | LRK | | | | |
| Legally Marketed Predicate Devices | | | | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | Product Code | |
| K200657 | SmartGuard Anti-Snore Device | | | LRK | |
| Device Description Summary | | | | | 21 CFR 807.92(a)(4) |
| The Anti-Snore Mouthguard+ is a thermoforming mandibular advancement device to reduce snoring. The device works by pulling the lower jaw forward. As such, the entirety of the device is tissue-contacting. The device combines an upper and lower molding with thermal impression material to conform to user's teeth morphology with an adjustable mechanism to position the user's lower jaw according to the user's need. The Anti-Snore Mouthguard+ is a "Boil and Bite" device where both the upper and lower moldings are constructed out of moldable material when heated in a water bath with rigid frames that allow for the straps to be attached to adjust the fitting. The device additionally comes with a ventilated storage box and Instructions for Use. This ventilated storage box is only provided for storage when the device is not in use and does not impact intended use of the device. | | | | | |
| Intended Use/Indications for Use | | | | | 21 CFR 807.92(a)(5) |
| The device is intended to aid in the reduction of snoring for adults at least 18 years old. | | | | | |
| Indications for Use Comparison | | | | | 21 CFR 807.92(a)(5) |
Indications for use of the subject device and predicate device are the same.
# Technological Comparison
Equivalence to Predicate Device
21 CFR 807.92(a)(6)
{5}------------------------------------------------
The subject device has the same intended use, inciple of operation, similar technological and material characteristics as the predicate device, SmartGuard Anti-Snore Device (K200657), Comparisons of the devices can be found in the substantial equivalence comparison table.
Comparison of Indication for Use
Both the subject and predicate devices are indicated for Over-the-Counter (OTC) use in alleviating snoring for adults 18 and older by way of mandibular repositioning.
Comparison of Technological Features
Oscimed SA's Anti-Snore Mouthguard+ is designed similarly as the predicate device and the mechanisms of action. The subject device is comprised of an upper and a lower tray, that are copolyester frames with a soft ethylene vinyl acetate (EVA) overmolding. The trays are connected via a flexible strays via hooks. The EVA overmolding is a thermal impression material that can be heat molded to the users dentition in a home environment via common "Boil-and-Bite" method for retention. The predicate device is designed similarly as a "Boil-and-Bite" device with separate upper and lower thermoplastic trays overmolded with thermoforming resin. The trays in the predicate device are connected via advancement bars.
The subject and predicate devices position the lower mandible anteriorly to increase pharyngeal space. The subject device achieves this by supplying flexible bands of varying lengths that allow for the user to adjust the level of lower mandible/tray anterior positioning to their comfort level, up to +6mm. The predicate device increases the pharyngeal space, providing advancement bars that position the lower jaw forward relative to the upper jaw, up to +6mm.
The subject device connects the upper and lower molding anteriorly on the maxilla and posteriorly on the mandible whereas the predicate connects the upper and lower molding posteriorly on the maxilla and anteriorly on the mandible. Although the connection points differ, the subject device distributes forces across the mouth resulting in force that is not greater than that applied by the predicate device at the front and rear connection points. This resulting force is proven through performance testing conducted comparing the forces applied to the anterior connection points by both the subject and predicate devices. The technological features of the subject and predicate devices are similar and achieve the same principle of operation in increasing pharyngeal space by repositioning the lower mandible. Therefore, the technological features of the subject device does not raise any questions of safety or performance.
### Comparison of Material Characteristics
The subject and predicate devices require biocompatibility testing for a surfacts intraoral surfaces for prolonged contact (>24h to 30 days), per "Intraoral Devices for Snoring and/or Obstructive Sleep Apnea – Class II Special Controls Guidance Document for Industry and FDA," ISO 10993 Part 5 and Part 10, Tests for in vitro Toxicity and Tests for irritation, respectively. A summary of the biocompatibility testing conducted for the subject device can be found in the Biocompatibility Section.
The subject device is comprised of a stiff copolymer frame overmolded with ethylene vinyl acetate thermal moldable resin, and a flexible strap. The flexible straps are thermoplastic elastomers which are rigid enough to position the lower tray and to increase pharyngeal space. The predicate device is made from thermoplastic trays (polycarbonate) overmolded with thermoforning resin (EVA) and connected with advancement bars. Both the subject and predicate devices use materials with similar characteristics to achieve the same principle of operation and do not raise any questions of the subject device.
### Substantial Equivalence Conclusion
Based on the comparison of indication of use, material characteristics, technological features, and principle of operation, Occimed SA's Anti-Snore Mouthguard+ is substantially equivalent to the predicate device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-clinical validation testing was performed to demonstrate the safety and effectiveness of the following tests were performed:
· Strap Tensile Testing - ensures that the straps do not stretch under expected load to keep the device in the apropriate position while in use
• Validation of Temperature Indicator - ensures that the temperature indicator is accurate for when water bath is at the correct temperature to permit molding of the upper and lower moldings to the user's teeth
· Force Measurement and Mandibular Advancement Testing - evaluates the amount of advancement achieved for each strap up to +6mm and ensure that forces applied by the device are clinically tolerable through comparison of forces applied by the predicate
Oscimed SA believes that the aforementioned non-clinical testing demonstrates that the subject device is designed in such a way that, when used under the conditions and for the purposes intended, the safety and effectiveness, as well as the performance characteristic of the subject device is substantially equivalent to the predicate device and meet the testing requirements contained within the Intraoral Devices for Snoring and/or Obstructive Sleep Apnea - Class II Special Controls Guidance Document for Industry and FDA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.