Phoenix Contact Lens Case is indicated for storage of soft (hydrophilic), rigid gas permeable and hard contact lenses during chemical disinfection.
Device Story
Phoenix Contact Lens Case consists of three models (CL-01, CL-02, CL-03) designed for storage of contact lenses during chemical disinfection. Models feature hinged, self-sealing caps and varying well capacities (1.5ml to 2.0ml). CL-01 and CL-02 are manufactured from LDPE; CL-03 is manufactured from polypropylene. Device is intended for patient self-use at home. Users place lenses into wells with chemical disinfectant solution for storage. Device provides a physical container to facilitate the disinfection process; it does not perform disinfection itself. Benefits include secure storage and protection of lenses during the chemical disinfection cycle.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 standards: Cytotoxicity (ISO 10993-5), Intracutaneous irritation (ISO 10993-23), Skin sensitization (ISO 10993-10), Acute systemic toxicity (ISO 10993-11), Material-mediated pyrogenicity (USP <151>), and Acute ocular irritation (ISO 10993-23). All tests concluded the device is non-cytotoxic, non-sensitizing, non-irritant, non-pyrogenic, and shows no systemic toxicity.
Technological Characteristics
Materials: LDPE (CL-01, CL-02) and Polypropylene (CL-03). Design: Hinged, self-sealing caps; 1.5ml or 2.0ml well capacities. Biocompatibility: Tested per ISO 10993-5, 10, 11, 23 and USP <151>. No electronic components or software.
Indications for Use
Indicated for storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses during chemical disinfection. Not for use with heat disinfection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
August 30, 2023
Phoenix Innovative Healthcare Manufacturers Pvt. Ltd. Michael Stuart CFO EL-209, Shil Mahape Road Electronic Zone; MIDC TTC Industry Area, Mahape Navi Mumbai, Maharashtra 400710 India
Re: K231123
Trade/Device Name: Phoenix Contact Lens Case - dome top flat pack (CL-01); Phoenix Contact Lens Case - classic flat pack (CL-02); Phoenix Contact Lens Case - sunglass shape flat pack (CL-03) Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LRX Dated: April 20, 2023 Received: July 24, 2023
Dear Michael Stuart:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed Bennett N. by Bennett N. Walker -S Walker -S Date: 2023.08.30 09:57:47 -04'00'
for Angelo Green Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K231123
Device Name
Phoenix Contact Lens Case - dome top flat pack (CL-01);
Phoenix Contact Lens Case - classic flat pack (CL-02);
Phoenix Contact Lens Case - sunglass shape flat pack (CL-03)
Indications for Use (Describe)
Phoenix Contact Lens Case is indicated for storage of soft (hydrophilic), rigid gas permeable and hard contact lenses during chemical disinfection.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) #: | K231123 | | 510(k) Summary | Prepared on: | 2023-08-30 | Page | 1 |
|-----------|---------|--|----------------|--------------|------------|------|---|
|-----------|---------|--|----------------|--------------|------------|------|---|
Contact Details 21 CFR 807.92(a)(1)
| Applicant Name | Phoenix Innovative Healthcare Manufacturers Private Limited |
|-----------------------------|---------------------------------------------------------------------------------------------------------------|
| Applicant Address | EL-209, Shil Mahape Road Electronic Zone; MIDC TTC Industry Area, Mahape Navi Mumbai Maharashtra 400710 India |
| Applicant Contact Telephone | 9548804274 |
| Applicant Contact | Mr. Michael Stuart |
| Applicant Contact Email | mikes@phoenix-hs.com |
Device Name 21 CFR 807.92(a)(2)
| Device Trade Name | Phoenix Contact Lens Case - dome top flat pack (CL-01);<br>Phoenix Contact Lens Case - classic flat pack (CL-02);<br>Phoenix Contact Lens Case - sunglass shape flat pack (CL-03) |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Contact Lens Case |
| Classification Name | Opthalmic |
| Regulation Number | 886.5928 |
| Product Code | LRX |
Legally Marketed Predicate Devices 21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K112832 | Optego/Eye Care Cure | LRX |
| K013232 | Bausch & Lomb | LRX |
Device Description Summary 21 CFR 807.92(a)(4)
There are three models of the Phoenix Contact Lens Case:
CL-01 "Dome Top Flat Pack" - made with LDPE and has 1.5ml wells on each side
CL-02 "Classic Flat Pack" - made with LDPE and has 1.5 ml wells on each side
CL-03 "Sunglass Shape Flat Pack" made with Polypropylene and has 2.0 ml wells on each side
All three models have hinged self sealing caps and are available in white, black, blue, orange, green, and natural.
The Phoenix contact lens cases are intended for storage during chemical disinfection of soft, rigid gas permeable or hard contact lenses. It is not to be used with heat disinfection.
Intended Use/Indications for Use 21 CFR 807.92(a)(5)
Phoenix Contact Lens Case is indicated for storage of soft (hydrophilic), rigid gas permeable and hard contact lenses during chemical
{4}------------------------------------------------
# Indications for Use Comparison
The indications for use are the same.
# Technological Comparison
The proposed device has the same classification and prodicate devices. The proposed device has the same intended use as the predicate devices. The proposed device has a similar design and is materials to the predicate devices.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The nonclinical tests submitted were:
Cytotoxicity per ISO 10993-5 Intracutaneous skin irritation per ISO 10993-23 Guinea pig maximization (GPMT) skin sensitization per ISO 10993-10 Acute systemic injection per ISO 10993-11 Material mediated pyrogenicity per USP <151> Acute ocular irritation testing per ISO 10993-23
The tests concluded the following:
Based on the results obtained under laboratory testing conditions, the extract of test item, Contact Lens Case was found to be "non-cytotoxic" to the subconfluent monolayer of L-929 mouse fibroblast cells.
Based on the above results of the experiment, it is concluded that the polar extracts of Contact Lens Case was found to be "Non-sensitizer" to the skin of the Guinea pigs under the experimental conditions employed.
Based on the results of the experiment, it is concluded that the polar extracts of test item, Contact Lens Case was "Nonirritant" to the skin of New Zealand White Rabbits under the experimental conditions and the dose employed as per the ISO 1093 Part 23:2021 (E) Specification.
Based on the results of the experiment. it is concluded that the polar extracts of test item. Contact Lens Case when administered to Swiss Albino Mice through and intraperitoneal routes respectively at a dose volume of 50 mL/kg body weight did not reveal any systemic toxicity under the experimental conditions employed.
Based on the results of the experiment, it is conduded that the extract Lens Case evaluated for progen test in New Zealand White Rabbits is Non-pyrogenic as it meets the requirements of progen test as per U.S. Pharmacopoeia, and General Chapters: <151> Pyrogen Test.
Under the experimental conditions employed and based on the observed results of the experiment. it is colar and non-polar extract of test tiem, Contact Lens Case did not produce any irritant effects to the eyes of New Zealand White Rabbits as per ISO 10993 "Biological Evaluation of Medical Devices" Part 23:2021(E) "Test for Irritation".
All testing demonstrates substantial equivalence.
### Page 2
## 21 CFR 807.92(a)(5)
## 21 CFR 807.92(a)(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.