The Multi-parameter detector is intended to be used for measuring and displaying of non-invasive blood pressure (NIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), forehead temperature (TEMP), weight and height in adults. The Multi-parameter detector is also intended to estimate body mass index (BMI), Body fat percentage, visceral fat grade, basal metabolism and free fat mass (muscle mass, bone mass and body water percentage) using the BIA (Bio-electrical Impedance Analysis) method.
Device Story
Multi-parameter detector measures physiological parameters in adults; inputs include oscillometric NIBP, photoelectric SpO2/PR, infrared forehead temperature, and BIA-based body composition (8-electrode tactile method). Device integrates cleared SpO2/PR technology (K170047). Used in hospital outpatient clinics and experience rooms by clinicians. Outputs displayed on device; data used for clinical assessment of patient health status. Benefits include comprehensive, non-invasive monitoring of vital signs and body composition metrics.
Clinical Evidence
Single-center, randomized, self-controlled clinical trial. NIBP accuracy evaluated in 119 subjects; mean difference ±5.0mmHg, SD ≤8.0mmHg. Temperature accuracy evaluated in 105 subjects; error ≤±0.2°C, repeatability ≤±0.3°C. SpO2/PR accuracy (Arms 1.49) verified in 10 healthy subjects (Fitzpatrick 2-6) against cleared predicate data. No adverse events reported.
Technological Characteristics
Materials: Biocompatible (ISO 10993-5, 10993-10). Sensing: Oscillometric (NIBP), Photoelectric (SpO2/PR), Infrared (TEMP), BIA (50kHz/250kHz, 8-electrode). Power: 100V-240V AC. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 80601-2-61, ISO 80601-2-56, ASTM E 1965-98.
Indications for Use
Indicated for adults for non-invasive measurement/monitoring of blood pressure, SpO2, pulse rate, forehead temperature, weight, height, and body composition (BMI, body fat, visceral fat, basal metabolism, muscle mass, bone mass, body water) via BIA.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Microlife Upper Arm Automatic Digital Blood Pressure Monitor (K222979)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
June 15, 2024
Zhongshan Jinli Electronic Weighing Equipment Co., Ltd. % Liz Li, Counselor Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District Shenzhen, Guangdong 518000 China
Re: K231116
Trade/Device Name: Multi-parameter detector Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, DQA, FLL, PUH Dated: September 11, 2023 Received: September 20, 2023
Dear Liz Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K231116
Device Name Multi-parameter detector
Indications for Use (Describe)
The Multi-parameter detector is intended to be used for measuring and displaying of non-invasive blood pressure (NBP). non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), forehead temperature (TEMP), weight and height in adults.
The Multi-parameter detector is also intended to estimate body mass index (BMI), Body fat percentage, basal metabolism and free fat mass, bone mass and body water percentage) using the BIA (Bio-electrical Impedance Analysis) method.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Zhongshan Jinli Electronic Weighing Equipment Co., Ltd. Multi-parameter detector
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
## 510(k) Summary
### 1. Contact Details
1.1 Applicant information
| Applicant Name | Zhongshan Jinli Electronic Weighing Equipment Co.,<br>Ltd. |
|-------------------------|----------------------------------------------------------------------------|
| Address | 283rd South Min'an Road, Xiaolan Town, Zhongshan<br>City, Guangdong, China |
| Phone No. | +86-760-28133793 |
| Contact person | Jerry Liang |
| Contact person's e-mail | jerry@kinleehealth.com |
| Date Prepared | Dec.30.2023 |
### 1.2 Submission Correspondent
| Image: Logo | Shenzhen Joyantech Consulting Co., Ltd. |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| | 1713A, 17th Floor, Block A, Zhongguan Times Square,<br>Liuxian Avenue, Xili Town, Nanshan District, Shenzhen,<br>Guangdong Province, China |
| Phone No. | +86 755-86069197 |
| Contact person | Liz Li |
| Contact person's e-mail | liz@cefda.com |
| Website | http://www.cefda.com |
### 2. Device information
| Device Trade name | Multi-parameter detector |
|-----------------------------|---------------------------------------------------|
| Device Common name | Multi-parameter detector |
| Classification | II |
| Classification name | System, Measurement, Blood-Pressure, Non-Invasive |
| Product code | DXN |
| Subsequent Product<br>Codes | DQA, FLL, PUH |
| Regulation No. | 21 CFR 870.1130 |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{5}------------------------------------------------
Zhongshan Jinli Electronic Weighing Equipment Co., Ltd. Multi-parameter detector
### 3. Primary Predicate Device
| Trade Name | Microlife Upper Arm Automatic Digital Blood Pressure<br>Monitor |
|----------------|-----------------------------------------------------------------|
| 510(k) Number | K222979 |
| Product Code | DXN |
| Regulation No. | 21 CFR 870.1130 |
| Manufacturer | Microlife Intellectual Property GmbH |
### 4. Secondary Predicate Device
| Trade Name | Vitals360® Multi-Vitals Mobile Monitor |
|-----------------------------|----------------------------------------|
| 510(k) Number | K210086 |
| Product Code | MWI |
| Subsequent Product<br>Codes | DQA, DSH DXN, FLL |
| Regulation No. | 21 CFR 870.2300 |
| Manufacturer | VoCare, Inc. |
### 5. Device Description
The Multi-parameter detector is a device designed for measuring of the patient's physiological parameters. The Multi-parameter detector can monitor the patient's blood oxygen saturation (SpO2) and pulse rate (PR) non-invasively by the photoelectric method. It can also measure non-invasive blood pressure (NIBP, the pressures of systolic and diastolic) by the oscillating method and body temperature (TEMP) by the infrared radiation energy technology.
The Multi-parameter detector are non-invasive body composition analyzers. The devices employ BIA (Bio-electrical Impedance Analysis) method and then measure body composition using an experimentally derived algorithm.
The Multi-parameter detector integrates an FDA market cleared devices, All-in-One Health Monitor (K170047) for measurement of oxygen saturation of arterial hemoqlobin (SpO2).
### 6. Intended use
The Multi-parameter detector is intended to be used for measuring and displaying of non-invasive blood pressure (NIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), forehead temperature (TEMP), weight and height in adults.
The Multi-parameter detector is also intended to estimate body mass index (BMI), Body fat percentage, visceral fat grade, basal metabolism and free fat mass (muscle mass, bone mass and body water percentage) using the BIA (Bio-electrical Impedance Analysis) method.
{6}------------------------------------------------
### 7. Substantial Equivalence Comparison
| Item | Proposed Device | Primary Predicate Device (K222979) | Secondary Predicate Device (K210086) | Comments |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Trade name | Multi-parameter detector | Microlife Upper Arm Automatic Digital<br>Blood Pressure Monitor | Vitals360® Multi-Vitals Mobile Monitor | N/A |
| Classification | II | II | II | Same |
| Product Code | DXN | DXN | MWI | |
| Subsequent<br>Product<br>Codes | DQA, FLL, PUH | N/A | DQA, DSH DXN, FLL | Similar |
| Intended<br>use/Indications<br>for use | The Multi-parameter detector is<br>intended to be used for measuring<br>and displaying of non-invasive blood<br>pressure (NIBP), non-invasive<br>monitoring of functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2), pulse rate (PR), forehead<br>temperature (TEMP), weight and<br>height in adults.<br>The Multi-parameter detector is<br>intended to estimate body mass<br>index (BMI), Body fat percentage,<br>visceral fat grade, basal metabolism<br>and free fat mass (muscle mass,<br>bone mass and body water<br>percentage) using the BIA<br>(Bio-electrical Impedance Analysis)<br>method. | The Upper Arm Blood Pressure Monitor,<br>Model BP3KV1-5W is a device intended<br>to measure the systolic and diastolic<br>blood pressure and pulse rate of an adult<br>individual by using a non-invasive<br>oscillometric technique in which an<br>inflatable cuff is wrapped around the<br>upper arm for a circumference range<br>from 22 to 52cm. The device is suitable<br>for use by adults, including adults with<br>conditions of diabetes, pregnancy, or<br>pre-eclampsia. The device detects the<br>appearance of irregular heartbeat during<br>measurement and gives a warning signal<br>with the reading once the irregular<br>heartbeat is detected. The device can be<br>used in connection with a smart phone<br>via Bluetooth or with a personal<br>computer (PC) via USB. The<br>measurement data can be transferred to<br>a smart phone running the Microlife<br>Connected Health+ mobile software<br>(App) or a PC running the Microlife BP<br>Analyzer+ (BPA+) software. | Vitals360® device is intended to be used for<br>measuring, displaying, reviewing and storing<br>of non-invasive blood pressure (NIBP),<br>non-invasive monitoring of functional oxygen<br>saturation of arterial hemoglobin (SpO2),<br>pulse rate (PR), forehead temperature<br>(TEMP), ECG, weight and height in adults no<br>less than 18 years of age.<br>This VITALS360® device is intended for use<br>by trained adults only who can use<br>smartphones proficiently.<br>This VITALS360® device is intended for use<br>in a clinical or home environment.<br>This VITALS360® device is a reusable device<br>following thorough cleaning between uses. | Different |
| Environment of<br>Use | Hospital outpatient clinic and<br>experience room. | Home | Clinical or home environment | Same |
| OTC or Rx | Rx | OTC | Rx | Identical to the<br>secondary |
| Patient<br>Population | Adult | Adult | No less than 18 years of age. | Similar |
| Power Source | 100V-240V~50Hz/60Hz 1.0A | 4 AA batteries, Or<br>AC adapter 6 V DC 600 mA | (100-240) VAC, 50/60Hz, 0.5A | Identical to the<br>secondary |
{7}------------------------------------------------
#### VOL_006:001_510(k) Summary
#### Zhongshan Jinli Electronic Weighing Equipment Co., Ltd.
| NIBP | | | | |
|-------------------------------------------------|----------------------|-------------------------------------------------------|-----------------------------------------------------------------|-------------------------------|
| Code: DXN | | | | |
| Features | Proposed Device | Primary Predicate Device (K072325) | Secondary Predicate Device (K210086) | Comments |
| Method | Oscillometric method | Oscillometric method | Oscillometric method | Same |
| Pressure<br>measurement<br>range -<br>Systolic | 60mmHg-260mmHg | 60 mmHg-255 mmHg | 60 mmHg-230mmHg | |
| Pressure<br>measurement<br>range -<br>Diastolic | 30mmHg-200mmHg | 40mmHg-200mmHg | 40 mmHg-130mmHg | Similar |
| Accuracy or<br>pressure<br>indicator | ±3mmHg | Pressure within ± 3 mmHg or<br>2% of reading >200mmHg | Mean deviation values: ±5 mmHg.<br>Standard deviation ≤ 8 mmHg. | |
| Cuff pressure<br>range | 0~300mmHg | 0~299mmHg | 0~300mmHg | Identical to the<br>secondary |
| SpO2 | | | | |
|--------------------|-----------------|------------------------------------|--------------------------------------------------------|-------------------------------|
| Code: DQA | | | | |
| Features | Proposed Device | Primary Predicate Device (K072325) | Secondary Predicate Device (K210086) | Comments |
| Measuring<br>range | 70%~100% | N/A | 70%~100% | Identical to the<br>secondary |
| Accuracy | $\pm$ 3% | N/A | $\pm$ 2% (during 90~100%),<br>$\pm$ 4% (during 70~89%) | Different |
{8}------------------------------------------------
Version: A/0
| TEMP | | | | |
|-----------------------------------------|--------------------------------|------------------------------------|--------------------------------------|-------------------------------|
| Code: FLL | | | | |
| Features | Proposed Device | Primary Predicate Device (K072325) | Secondary Predicate Device (K210086) | Comments |
| Fundamental<br>scientific<br>technology | Infrared technology | N/A | Infrared technology | Identical to the<br>secondary |
| Unit of<br>measure | °C/°F | N/A | °C/°F | Identical to the<br>secondary |
| Measurement<br>site | Forehead | N/A | Forehead | Identical to the<br>secondary |
| Temperature<br>measurement<br>range | 22.0°C~42.9°C (71.6°F~109.2°F) | N/A | 34.0°C~43.0°C (93.2°F~109.4°F) | Different |
| Temperature<br>measurement<br>accuracy | 22°C~42.9°C ±0.3°C | N/A | ±0.3°C | Identical to the<br>secondary |
| Body composition | | | | |
|---------------------------------------|-------------------------------------------------------------------------|------------------------------------|--------------------------------------|-----------|
| Code: PUH | | | | |
| Features | Proposed Device | Primary Predicate Device (K072325) | Secondary Predicate Device (K210086) | Comments |
| Technology | BIA (bioelectrical impedance<br>analysis) | | | |
| Operating frequency range | 50kHz/250kHz | | | |
| Electrode type | Tactile | | | |
| Number/Place<br>ment of<br>Electrodes | 8 electrodes Placed on Hands and<br>feet (Tetra-polar Electrode method) | N/A | N/A | Different |
| Patient<br>Position | Upright | | | |
| Impedance<br>measurement<br>range | 200Ω~800Ω | | | |
{9}------------------------------------------------
#### K231116
#### VOL_006:001_510(k) Summary
#### Zhongshan Jinli Electronic Weighing Equipment Co., Ltd.
Version: A/0
The subject device, Multi-parameter detector, is substantially equivalent to the predicate device K222979 and K210086. This condusion is based upon comparison on intended use and technological characteristics. Any difference between devices or concerns of safety or effectiveness.
{10}------------------------------------------------
Zhongshan Jinli Electronic Weighing Equipment Co., Ltd.
### 8. Non-clinical Testing
The following data were provided in support of the substantial equivalence determination:
#### 1) Electrical Safety, Electromagnetic Compatibility
IEC 60601-1:2005, AMDI:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
IEC 60601-1-2:2014 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests
#### 2) Performance testing
IEC 80601-2-30: 2018 Medical electrical equipment - Part 2: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
ISO 80601-2-61: 2017 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
ISO 80601-2-56: 2017 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
ASTM E 1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
In addition, we tested the following items according to our internal standards.
- Height
- Weight
- Impedance
#### 3) Biocompatibility test
Biocompatibility testing was conducted in accordance with the 2020 FDA quidance "Use of International Standard ISO 10993, Biological Evaluation of Medical Device Part 1: Evaluation and Testing."
The subject device has passed biocompatibility testing in according to following standards.
ISO 10993-5:2019 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
#### 4) Software validation
The subject device contains an embedded software system.
{11}------------------------------------------------
Zhongshan Jinli Electronic Weighing Equipment Co., Ltd.
Software verification and validation testing were conducted as recommended by FDA's Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
#### 5) Human Factor Engineering and Usability Engineering Evaluation
The usability document of the subiect device was determined according to IEC 62366-1:2015 & FDA guidance of Applying Human Factors and Usability Engineering to Medical Devices
### 9. Clinical testing
Clinical studies were conducted to verify the accuracy of subject device. The clinical studies were conducted per following standards:
- 1) ISO 81060-2: 2018 Non-invasive sphygmomanometers Part 2: Clinical investigation of intermittent automated measurement type
- 2) ISO 80601-2-56: 2017+AMD1: 2018 Medical electrical equipment Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
The Multi-parameter detector integrates an FDA market cleared devices, All-in-One Health Monitor (K170047) for measurement of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The SpO2 and PR cleared under K170047 were originally cleared under K063641, which is in compliance with ISO 9919: 2005.
The pulse oximeter accuracy was tested in ten healthy subjects, aged 22-40, with skin tones varying from Fitzpatrick 2-6. 3 subjects had dark skin with Fitzpatrick 5 or 6. The Arms of 70%-100% is 1.49, it meets the accuracy claims of ±3%.
Therefore, there is no test of these two parts (SpO2 and PR) in this clinical test.
Clinical testing has been performed under an approved protocol with subject informed consent. Clinical test results support device accuracy claims for the specified measurement range.
The Clinical studies was a single-center, randomized, self-controlled clinical trial for accuracy validation. Compared with registered mercury thermometer and desktop sphygmomanometer, the accuracy of body temperature and blood pressure measurement was used as the main effectiveness observation indications, and easy for use and device stability were used as the main safety observation indications to evaluate the safety and effectiveness of the Multi-parameter detector (body temperature and blood pressure) produced by Zhongshan Jinli Electronic Weighing Equipment Co., Ltd.
In the evaluation of the accuracy of temperature measurement, 105 subjects did not exceed the maximum allowable error of ±0.2℃ in the temperature display range of 35.0 ℃~ 42.0 ℃, and the clinical repeatability did not exceed the range of ±0.3 ℃, which met the temperature measurement standards.
In the evaluation of the accuracy of Blood Pressure Measurement, a total of 119 subjects completed the experiment. The average value of the difference between the blood pressure value (systolic or diastolic blood pressure) measured by the test device and the reference blood pressure measurement value is within the range of ±5.0mmHg, and the standard deviation is not greater than 8.0mmHg, which meets the standard requirements.
{12}------------------------------------------------
Zhongshan Jinli Electronic Weighing Equipment Co., Ltd.
The interface friendliness of the device and the satisfaction rate of the operation response time were both 100.00%. The satisfaction rate of the whole device function and stability evaluation index was 100.00%. There were no adverse events and serious adverse events in the clinical trial, and the incidence rate was 0.00%, and the device has good safety.
### 10. Other information (such as required by FDA guidance/Test)
N/A
### 11.Conclusions
Performance testing and compliance with voluntary standards demonstrate that the proposed subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.