ABPMpro Ambulatory Blood Pressure Recorder is a Non-Invasive oscillometric and ECG device intended to acquire Ambulatory non-invasive Blood Pressure signals and 3 Channel ECG signals from the upper body surfaces. Cardiac rhythm is acquired via 3 Channel ECG signals. Additional physiological signals like plethysmogram, body position and activity are also measured. The Recorder is intended for adults and children who are over the age of 12 years. ABPMpro Software allows transfer of Blood Pressure-, ECG- and data from additional signals from the Recorder to a Windows-based PC-based computer program via USB cable for the purpose of creating reports and printouts. The Software does not perform diagnostics. Physicians carry out diagnostics evaluations of this data. The system is only for measurement, recording and display. It makes no diagnosis.
Device Story
Portable device worn on upper arm via cuff; records oscillometric blood pressure, 3-channel ECG, plethysmogram, body position, and activity. Used in hospital, clinic, or home settings by patients; data recorded over 24 hours. Post-recording, device connects to Windows-based PC via USB; ABPMpro Analyzer Software downloads data for physician review. Software provides reports/printouts; performs no diagnostics. Physician evaluates data to inform clinical decisions. Benefits include long-term physiological monitoring for diagnostic evaluation.
Clinical Evidence
Prospective clinical validation study (N=90 general, N=36 ambulatory) evaluated BP accuracy against auscultatory reference (aneroid sphygmomanometer) per ISO 81060-2:2018. Results confirmed device meets accuracy criteria for adolescents and adults. Usability validation confirmed safety and effectiveness with no use-related errors.
Technological Characteristics
Portable recorder with integrated accelerometer/position sensor, brachial cuff, optional 3-channel ECG, and LED plethysmogram sensor. Powered by rechargeable Li-ion battery. Connectivity via USB to PC. Biocompatible materials per ISO 10993-1, -5, -10. Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-47, and 80601-2-30.
Indications for Use
Indicated for adults and children over 12 years old for ambulatory non-invasive blood pressure and 3-channel ECG monitoring, including plethysmogram, body position, and activity tracking.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 27, 2023
SOMNOmedics GmbH % Cherita James Regulatory Consultant M Squared Associates/ a PPG Company 127 West 30th St Floor 9 New York, New York 10001
Re: K231104
Trade/Device Name: ABPMpro Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: DSH, DXN Dated: October 19, 2023 Received: October 19, 2023
Dear Cherita James:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K231104
Device Name ABPMpro
#### Indications for Use (Describe)
ABPMoro Ambulatory Blood Pressure Recorder is a Non-Invasive oscillometric and ECG device intended to acquire Ambulatory non-invasive Blood Pressure signals and 3 Channel ECG signals from the upper body surfaces. Cardiac rhythm is acquired via 3 Channel ECG signals. Additional physiological signals like plethysmogram, body position and activity are also measured. The Recorder is intended for adults and children who are over the age of 12 years. ABPMpro Software allows transfer of Blood Pressure-, ECG- and data from additional signals from the Recorder to a Windows-based PC-based computer program via USB cable for the purpose of creating reports and printouts. The Software does not perform diagnostics. Physicians carry out diagnostics evaluations of this data. The system is only for measurement, recording and display. It makes no diagnosis."
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(K) SUMMARY
The following information is provided as required by 21 CFR § 807.87 for low impedance cream lic2® 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
Sponsor: SOMNOmedics GmbH Am Sonnenstuhl 63 Randersacker, Germany D-97236
- Contact: Cherita James M Squared Associates/a PPG Comapny 127 West 30th Street 9th Floor New York, New York 10001 Ph: 347-954-0624 CJames(@MSquaredAssociates.com
Submitted Date:, October 19, 2023
Proprietary Name: ABPMpro
Common Name: ambulatory ECG and non-invasive blood pressure monitor and recorder
Regulatory Class: II
870.2800 Recorder, Magnetic Tape, Medical; 870.1130 System, Measurement, Regulation:
Blood-Pressure, Non-Invasive
Panel: Cardiovascular
Product Codes: DSH, DXN
Predicate Device: Primary- Vasomedical-Biox, K092785
Reference- SOMNOscreen plus , K201054
Device Description: The ABPMpro is a portable device for recording physiological signals. The ABPMpro is used as a long-term blood pressure as well as a long-term ECG device.
The ABPMpro is a portable device worn on the upper arm by means of a cuff and used as an oscillometric 24h ambulatory blood pressure recorder.
The ABPMpro consists of the following hardware:
- . ABPMpro recorder with integrated accelerometer and position sensor;
- . Brachial blood pressure cuff (3 sizes available)
- . Optional: 3-channel ECG sensor
- . Optional: Plethysmogram-sensor (LED)
Following the completion of the patient recording session, the device, cuff and electrodes are removed from the patient. The ABPMPro is connected to the PC using the USB cable. With the
{5}------------------------------------------------
downloaded ABPMpro Analyzer Software open on the PC, the user can download the recorded data. The user can select the patient data and begin analysis.
The device is intended for prescription use.
Indications for Use: ABPMpro Ambulatory Blood Pressure Recorder is a Non-Invasive oscillometric and ECG device intended to acquire Ambulatory non-invasive Blood Pressure signals and 3 Channel ECG signals from the upper body surfaces. Cardiac rhythm is acquired via 3 Channel ECG signals. Additional physiological signals like plethysmogram, body position and activity are also measured. The Recorder is intended for adults and children who are over the age of 12 years.
ABPMpro Software allows transfer of Blood Pressure-, ECG- and data from additional signals from the Recorder to a Windows-based PC-based computer program via USB cable for the purpose of creating reports and printouts. The Software does not perform diagnostics. Physicians carry out diagnostics evaluations of this data. The system is only for measurement, recording and display. It makes no diagnosis.
Performance: Performance testing was conducted to confirm compliance to device specifications. All functions were verified to operate as designed. Signal recorded by the ABPMpro underwent performance testing demonstrating that all signal types being recorded comply with the performance criteria set forth by SOMNOmedics that includes the minimum performance specification recommended by the applicable standards for IEC 80601-2-30, which sets forth the definitive standard for devices intended for use in evaluating automated noninvasive sphygmomanometer, and IEC 60601-2-47 for ambulatory ECG.
Usability validation was performed under actual conditions within a clinical study to assess the effectiveness of implemented risk mitigations and confirm safety of the device when used as intended. No incidents of use-related errors were reported. The device was found to be safe and effective for the intended users, use, and use environments. The ABPMpro is not validated for use in exercise stress testing or strenuous exercise. Under certain circumstances, for instance transport in vehicles can have an impact on the measurement results.
#### Standards
SOMNOmedics GmbH has demonstrated compliance with the following standards for the
{6}------------------------------------------------
#### ABPMpro.
| Standard |
|----------------------------------------------------------------------------------------------------------------------|
| ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) |
| Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601- |
| 1:2005, MOD) |
| IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential |
| performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential |
| performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used |
| in the home healthcare environment |
| IEC 60601-2-47 Edition 2.0 2012-02 |
| Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performance of |
| ambulatory electrocardiographic systems |
| IEC 80601-2-30: Edition 2.0 2018-03 |
| Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of |
| automated non-invasive sphygmomanometers |
| IEC 62366-1:2015 + COR1:2016 Medical devices - Part 1: Application of usability engineering to medical devices |
| [Including CORRIGENDUM 1 (2016)] |
| ISO 14971:2007 Medical devices - Application of risk management to medical devices |
| IEC 62304:2006 + A1:2015 Medical device software - Software life cycle processes [Including Amendment 1 |
| (2016)] |
| ISO 81060-2 Third edition 2018-11 |
| Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type |
| ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a |
| risk management process |
| ISO 10993-5 Third edition 2009-06-01 |
| Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity |
| ISO 10993-10 Fourth edition 2021-11 |
| Biological evaluation of medical devices - Part 10: Tests for skin sensitization |
Software validation and cybersecurity assessment for Tier 2 risk level are provided to support the substantial equivalence of the subject device to the predicates.
Clinical Testing: SOMNOmedics GmbH sponsored an open, prospective, clinical validation study to evaluate the accuracy of the APBMpro oscillometric, upper arm device for ambulatory blood pressure (BP) measurement in general population according to the AAMI/ESH/ISO and FDA recognized Universal Standard (ISO 81060-2:2018) both at rest (primary objective) and under dynamic exercise (secondary objective). The clinical investigation received Ethics Committee approval and was conducted in accordance with good clinical practice (GCP) as described in 21 CFR 812.28(a)(1).
Patient population
General validation study:
N=90 age between 12-76, Mean age 39.7 ± 15.1, male 47 (52%), female 43 (48%) Ambulatory validation study:
N=36 age between 22-65, mean age 40.9 ± 11.0, male 20 (56%), female 16 (44%) From Publication:
{7}------------------------------------------------
#### Reference blood pressure
A calibrated (accuracy according to the test protocol for Boso Nova S, Jungingen, Germany is ± 1 mmHg), standard aneroid sphygmomanometer, was used for simultaneous auscultatory reference BP measurements by two observers, using a dual-head teaching stethoscope (Prestige Medical, Northridge, USA). Four different cuff sizes (14-21, 22-32, 33-41, >41 cm) were selected to meet the requirements of the AAMI/ESH/ISO universal standard.
#### Procedure
The 'same arm sequential method' described in the universal standard was used.
This study showed that the ABPMpro fulfils the accuracy criteria of the ISO 81060-2:2018, both in the general validation as well as in the ambulatory validation study. Thus, the device is recommended for clinical use in adolescents and adults.
{8}------------------------------------------------
## Substantial Equivalence Comparison
| Features /<br>Technical<br>Information | ABPMpro | Vasomedical-Biox<br>Primary predicate | SOMNOscreen plus<br>Reference predicate | Substantial<br>Equivalence |
|----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | DXN, DSH | DSH, DXN | MNR, OLV | |
| K-Number | Not yet assigned. | K092785 | K201054 | |
| Indications for Use: | ABPMpro Ambulatory Blood Pressure<br>Recorder is a Non-Invasive oscillometric<br>and ECG device intended to acquire<br>Ambulatory non-invasive Blood Pressure<br>signals and 3 Channel ECG signals from the<br>upper body surfaces. Cardiac rhythm is<br>acquired via 3 Channel ECG signals.<br>Additional physiological signals like<br>plethysmogram, body position and activity<br>are also measured. The Recorder is intended<br>for adults and children who are over the age<br>of 12 years.<br>ABPMpro Software allows transfer of<br>Blood Pressure-, ECG- and data from<br>additional signals from the Recorder to a<br>Windows based PC-based computer<br>program via USB cable for the purpose of<br>creating reports and printouts. The Software<br>does not perform diagnostics. Physicians<br>carry out diagnostics evaluations of this<br>data.<br>The system is only for measurement,<br>recording and display. It makes no<br>diagnosis. | Vasomedical-Biox Combined<br>Ambulatory ECG and Blood<br>Pressure Recorder is a Non-<br>Invasive oscillometric device<br>intended to acquire Ambulatory<br>3 Channel ECG signals and<br>non-invasive Blood Pressure<br>signals from the upper body<br>surfaces. Cardiac rhythm is<br>acquired via 3 Channel ECG<br>signals. The Recorders are<br>intended for adults and children<br>who are over the age of six<br>years.<br>Vasomedical-Biox Ambulatory<br>ECG CB Series Analysis<br>System Software allows<br>transfer of ECG and Blood<br>Pressure data from the<br>Recorder to a Windows based<br>PC-based computer program<br>via a removable and large<br>capacity storage card (SD) for<br>the purpose of creating reports<br>and printouts. The Software<br>does not perform diagnostics.<br>Physicians carry out<br>diagnostics evaluations of this<br>data.<br>The system is only for<br>measurement, recording and<br>display. It makes no diagnosis. | The SOMNOscreen®<br>plus is a non-life-<br>supporting portable<br>physiological signal<br>recording device<br>intended to be used for<br>testing adults and<br>children/adolescents<br>suspected of having<br>sleep-related breathing<br>disorders. | The Vaosimedical-Biox<br>Combined is for blood<br>pressure and ECG<br>measurements like the<br>ABPMpro. The<br>additional signals are<br>covered by the reference<br>predicate. The intended<br>patient population is<br>different. For<br>Vaosimedical-Biox<br>Combined the patient<br>population is adults and<br>children who are over<br>the age of six years. For<br>ABPMpro it is from 12<br>years old. The<br>Vaosimedical-Biox<br>Combined has just little<br>bit more range than the<br>ABPMpro. The signals<br>will be transferred to<br>software for display. No<br>diagnosis will be done<br>by the software. The<br>indications are similar. |
| Prescription Use | Yes | Yes | Yes | Same |
| Patient population | >12 yo | > 6 yo | >2 yo | Subject device has a |
| Features /<br>Technical<br>Information | ABPMpro | Vasomedical-Biox<br>Primary predicate | SOMNOscreen plus<br>Reference predicate | Substantial<br>Equivalence |
| | | | | more limited age range<br>that has been tested<br>clinically. |
| Use Environment | Hospital, clinic and home use | Hospital, clinic and home use | Hospital or clinic | Same as primary<br>predicate |
| Measurement<br>Range<br>B/P<br>Pulse rate | Heart Rate:<br>Measurement range: 28-260bpm<br>Accuracy: 5%<br>BP:<br>Measurement range: 0-300mmHg<br>Accuracy: 1.5% | BP:<br>Measurement range: Systolic:<br>50 - 260 mmHg<br>Diastolic: 30 - 180 mmHg | Heart rate:<br>18 to 300 bpm | The measurement range<br>for BP of the primary<br>predicate is covered by<br>the ABPMpro. The pulse<br>rate of the reference<br>device is more than the<br>ABPMpro. However,<br>from a medical<br>perspective the edges<br>will not be reached by a<br>human in this<br>application. |
| Channels | See table sensor comparison | See table sensor comparison | See table sensor<br>Comparison | See Table Sensor<br>Comparison below |
| Portable Design | Yes | Yes | Yes | Same |
| Number of Patients<br>can monitor<br>simultaneously | 1 | 1 | 1 | Same |
| Data Collection | Yes | Yes | Yes | Same |
| Report Generation | Optional | Optional | Optional | Same |
| Capable of Data<br>Transfer for Analysis<br>and Report<br>Generation | Yes | Yes | Yes | Same |
| Data Analysis<br>(Computer or<br>Computer Assisted) | Optional | Optional | Optional | Same |
| Data transfer to PC | Via USB cable. | Via flash card | Via flash card | Different but both are<br>common ways to<br>exchange data from a<br>device to the PC. |
| Power Supply | Rechargeable Li ION battery | AA Alkaline battery | Rechargeable Li ION<br>battery | Same as reference |
| Features /<br>Technical<br>Information | ABPMpro | Vasomedical-Biox<br>Primary predicate | SOMNOscreen plus<br>Reference predicate | Substantial<br>Equivalence |
| Operation time per<br>charge | 24 hours | 24 hours | - | Same as primary<br>predicate |
{9}------------------------------------------------
{10}------------------------------------------------
## Sensor Comparison
| Features/Technical<br>Information | ABPMpro | Vasomedical-Biox<br>Primary predicate | SOMNOs…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.