Flume Catheter

K231101 · Flume Catheter Company, Ltd. · EZL · Jun 29, 2023 · Gastroenterology, Urology

Device Facts

Record IDK231101
Device NameFlume Catheter
ApplicantFlume Catheter Company, Ltd.
Product CodeEZL · Gastroenterology, Urology
Decision DateJun 29, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.5130
Device ClassClass 2

Indications for Use

The FLUME catheter is intended for bladder drainage via urethral or suprapubic route.

Device Story

The FLUME catheter is a sterile, single-use, indwelling urinary catheter; constructed from polyurethane-based polymers. It features a dual-lumen shaft and an inflatable retention balloon. The larger lumen facilitates urine drainage, while the smaller lumen manages balloon inflation/deflation with sterile water. The distal end includes two opposing eye holes for drainage. The device is placed by clinicians using standard institutional practices for either urethral or percutaneous suprapubic bladder access. It functions as a passive drainage conduit to manage urine output, benefiting patients requiring temporary or long-term bladder catheterization.

Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing was conducted per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute/subacute systemic toxicity, material-mediated pyrogenicity, subcutaneous implantation, and genotoxicity). Bacterial endotoxin testing (BET) was performed to confirm compliance with pyrogen limit specifications.

Technological Characteristics

Materials: Polyurethane-based polymers. Design: Dual-lumen tube with inflatable retention balloon and distal drainage eye holes. Sterilization: Sterile, single-use. Biocompatibility: ISO 10993-1 compliant. Connectivity: None (mechanical device).

Indications for Use

Indicated for patients requiring bladder drainage via urethral or suprapubic route.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The Department of Health & Human Services seal is a stylized representation of a human figure, while the FDA acronym and name are written in blue, with the acronym in a square. June 29, 2023 Flume Catheter Company, Ltd. % Elaine Duncan President Paladin Medical, Inc. P.O. Box 560 Stillwater, MN 55082 Re: K231101 > Trade/Device Name: FLUME catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZL, KOB Dated: Mav 26, 2023 Received: May 30, 2023 Dear Elaine Duncan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Jessica K. Nguyen -S Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K231101 Device Name FLUME catheter Indications for Use (Describe) The FLUME catheter is intended for bladder drainage via urethral or suprapubic route. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## SPECIAL 510(k) SUMMARY | Submitted on behalf of: | The Flume Catheter Company Ltd.<br>Company Contact Name: Roger Holmes | |----------------------------|------------------------------------------------------------------------| | Address: | The Granary, Mill House Frensham,<br>Farnham, GU10 3EE, United Kingdom | | Telephone: | 44(0) 207 808 9125 | | by: | Paladin Medical, Inc.<br>PO Box 560<br>Stillwater, MN 55082 | | Telephone: | 715-549-6035 | | SUBMISSION CONTACT PERSON: | Elaine Duncan, MSME, RAC, FAIMBE<br>President, Paladin Medical, Inc. | | DATE PREPARED: | June 28, 2023 | | TRADE NAME: | FLUME catheter | | COMMON NAME: | Catheter, Retention Type, Balloon & Suprapubic | | REGULATION NAME: | Urological Catheter and Accessories | | REGULATION NUMBER: | 876.5130 | | REGULATORY CLASS: | II | | PRODUCT CODE: | EZL, KOB | #### PREDICATE DEVICE: FLUME catheter K212151. The predicate device has not been subject to a design-related recall. #### DESCRIPTION of the DEVICE: The FLUME catheter is intended for the drainage of urine from the bladder. The FLUME catheter is a sterile indwelling urinary catheter intended for single use. The FLUME catheter is made using polyurethanebased polymers. The catheter has an inflatable retention balloon attached to the catheter shaft. The catheter has a dual lumen tube. The larger lumen is for urine drainage. The smaller lumen is used to inflate and deflate the balloon with sterile water. The distal end has two opposite eye holes, which are used for drainage. The product is available in standard sizes. The catheter may be placed following the institutions' standard practice for either urethral bladder drainage or via percutaneous suprapubic access to the bladder. #### INDICATIONS FOR USE: The FLUME catheter is intended for bladder drainage via urethral or suprapubic route. #### SUBSTANTIALLY EQUIVALENT TO: The FLUME catheter is substantially equivalent to FLUME catheter, K212151. There are no changes or differences made to the FLUME catheter to extend the indication for use to include suprapubic drainage. The only change in this Special 510(k) is to the current indication for use and instructions for use. According to the risk assessment, there is no difference to the patient when the FLUME catheter is placed suprapubic. #### SUMMARY of NON_CLINICAL TESTING: - . Biocompatibility testing from the predicate device was leveraged for the subject device since the materials and method of manufacture are identical. The biocompatibility testing endpoints completed for the predicate device align with the biocompatibility evaluation expectations to support suprapubic use. Biocompatibility was conducted per International Standard ISO 10993-1. "Biological Evaluation of medical devices – Part 1: Evaluation and Testing within a risk management process: - Cytotoxicity, o - Sensitization, O - Irritation, O - Acute Systemic Toxicity, o - Subacute Systemic Toxicity, O - Material Mediated Pyrogenicity O {4}------------------------------------------------ - Subcutaneous Implantation, о - O Genotoxicity. - Bacterial endotoxin testing (BET) was conducted to ensure the device meets pyrogen limit • specifications. CONCLUSION: The FLUME catheter is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...